K193517 · Olympus Surgical Technologies America · KTI · Mar 18, 2020 · Ear, Nose, Throat
Device Facts
Record ID
K193517
Device Name
ViziShot 2 FLEX
Applicant
Olympus Surgical Technologies America
Product Code
KTI · Ear, Nose, Throat
Decision Date
Mar 18, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Indications for Use
The ViziShot 2 FLEX has been designed to be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) and fine needle biopsy (FNB) of submucosal and extramural lesions of the tracheobronchial tree. Do not use this device for any purpose other than its intended use.
Device Story
ViziShot 2 FLEX is a single-use, manual aspiration needle used with ultrasound endoscopes for tissue sampling. The device comprises a handle, sheath, needle, and removable stylet. During a procedure, the catheter is inserted into a bronchoscope's working channel and locked to the adapter biopsy valve. The clinician advances the needle to the target lesion under real-time ultrasound visualization. Suction is applied via a syringe at the handle's proximal end to obtain tissue samples. The device is used in surgical, endoscopy, or bronchoscopy suites by physicians. The output is a tissue sample for cytopathological or microbiological examination. The device benefits patients by enabling minimally invasive biopsy of tracheobronchial lesions.
Clinical Evidence
No clinical or animal studies were performed. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993 series), sterilization validation (ISO 11135-1), and mechanical performance testing (insertion/withdrawal forces, angulation, durability, echogenicity). A GLP bench study compared the ViziShot 2 FLEX 19G needle to the predicate, confirming similar tissue sampling metrics (weight, length, quality) for histologic and core analysis.
Technological Characteristics
Single-use aspiration needle (19G). Materials: Stainless Steel, PTFE, PEBAX, Nitinol. Manual operation. Dimensions: 70cm catheter length, 2.08mm working OD. Sterilization: Ethylene Oxide (EO) per ANSI/AAMI/ISO 11135-1. Packaging validated per ANSI/AAMI/ISO 11607-1. Shelf life: 3 years (accelerated aging per ASTM F1980-16).
Indications for Use
Indicated for patients requiring ultrasound-guided fine needle aspiration (FNA) or fine needle biopsy (FNB) of submucosal and extramural lesions of the tracheobronchial tree.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the acronym "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
March 18, 2020
Olympus Surgical Technologies America Mary Patella Senior Specialist, Regulatory Affairs 136 Turnpike Road Southborough, Massachusetts 01772-2104
# Re: K193517
Trade/Device Name: ViziShot 2 FLEX Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (flexible or rigid) and accessories Regulatory Class: Class II Product Code: KTI Dated: December 18, 2019 Received: December 19, 2019
### Dear Mary Patella:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James J. Lee Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K193517
Device Name ViziShot 2 FLEX
Indications for Use (Describe)
The ViziShot 2 FLEX has been designed to be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) and fine needle biopsy (FNB) of submucosal and extramural lesions of the tracheobronchial tree. Do not use this device for any purpose other than its intended use.
Type of Use (Select one or both, as applicable)
| <span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------------------------------|-----------------------------------------------|
|-------------------------------------------------------------------------------------|-----------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary Gyrus ACMI ViziShot 2 FLEX
# General Information
| Manufacturer: | Olympus Surgical Technologies<br>America<br>Gyrus ACMI, Inc.<br>136 Turnpike Rd.<br>Southborough, MA 01772-2104<br>Phone: 508-804-2600<br>Fax: 508-804-2624 |
|-----------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration Number: | 3003790304 |
| Contact Person: | Mary Anne Patella<br>Senior Specialist, Regulatory Affairs<br>508-804-2771<br>Maryanne.patella@olympus-osta.com) |
| Date Prepared: | December 18, 2019 |
| Device Description | |
| Classification Name:<br>CFR Citation Number:<br>Product Code:<br>Classification:<br>Review Panel<br>Trade Name:<br>Generic/Common Name: | Bronchoscope (flexible or rigid) and<br>accessories<br>21 CFR 874.4680<br>KTI<br>Class II<br>Ear Nose & Throat<br>Gyrus ACMI ViziShot 2 FLEX<br>Aspiration Needle |
| Predicate Device | |
| Cook Ireland, Ltd Echotip Procore<br>Endobronchial High Definition<br>Ultrasound Biopsy Needle | K160229 |
| Reference Device | |
| Gyrus ACMI ViziShot 2 FLEX | K163469 |
{4}------------------------------------------------
### Product Description
The subject device, ViziShot 2 FLEX is a single use aspiration needle to be used in conjunction with compatible ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) and fine needle biopsy (FNB) of submucosal and extramural lesions of the tracheobronchial tree.
The subject device consists of a handle, sheath, needle, and stylet. The sheath and needle are attached to the handle, and the removable stylet is located within the needle. Note that although, the device has a component called a needle, the device is often referred to as a needle as well. The device is a single-use sterile device.
Prior to a procedure, the flexible catheter portion is inserted into a bronchoscope's working channel (2.2mm) and advanced forward until fully inserted. The handle is then affixed to the channel port of the endoscope via a lever mechanism that locks onto the Adapter Biopsy Valve.
The needle is advanced through the bronchoscope to the sampling site while visualizing both the target and the needle in real time with ultrasound. The handle facilitates advancement of the needle during puncture of the targeted biopsy site. The sample is obtained by penetrating the lesion with the needle while applying suction at the proximal end of the handle. After completing the sampling, the vacuum from the syringe is released to atmosphere, the handle unlocked from the bronchoscope, and the catheter and needle pulled out from the working channel. The removed tissue can then be prepared for cytopathological or microbiological examination and testing.
The ViziShot 2 FLEX is available in one model only (NA-U403SX-4019), with a needle size of 19 gauge (19G). The two required accessories, the Adapter Biopsy Valve and the Merit Syringe with Stopcock, are packaged with the ViziShot 2 FLEX.
### Intended Use
The ViziShot 2 FLEX has been designed to be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) and fine needle biopsy (FNB) of submucosal and extramural lesions of the tracheobronchial tree. Do not use this device for any purpose other than its intended use.
Comparison of Proposed and Predicate DeviceThe proposed addition of "fine needle biopsy (FNB)" to the Indications for Use is the same as the primary predicate Cook Echotip Procore Endobronchial High Definition Ultrasound Biopsy Needle cleared via K160229.
The ViziShot 2 FLEX has the same technological characteristics and design as the as the reference predicate ViziShot 2 FLEX cleared under K163469. The subject and predicate device operate in the same manner to obtain a tissue biopsy using an ultrasound endoscope. Please refer to the following Substantial Equivalence Table.
{5}------------------------------------------------
-
| | Proposed Device | Predicate<br>Device<br>(K160229) | Reference Device<br>(K163469) | |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| Device Name/<br>Characteristics | Gyrus ACMI<br>ViziShot 2<br>FLEX | Cook Echotip<br>Procore<br>Endobronchial<br>High Definition<br>Ultrasound<br>Biopsy Needle | Gyrus ACMI<br>ViziShot<br>FLEX | Comment |
| Indications for<br>Use | The ViziShot 2<br>FLEX has been<br>designed to be<br>used with<br>ultrasound<br>endoscopes for<br>ultrasound<br>guided fine<br>needle<br>aspiration<br>(FNA) and fine<br>needle biopsy<br>(FNB) of<br>submucosal and<br>extramural<br>lesions of the<br>tracheobronchial<br>tree. Do not use<br>this device for<br>any purpose<br>other than its<br>intended use. | Echotip<br>Procore®<br>Endobronchial<br>High Definition<br>Ultrasound<br>Biopsy Needle<br>for use with<br>Olympus EBUS<br>scopes: This<br>device is used<br>with an<br>ultrasound<br>endoscope for<br>fine needle<br>biopsy, (FNB),<br>of submucosal<br>and extramural<br>lesions within or<br>adjacent to the<br>tracheobronchial<br>tree or<br>gastrointestinal<br>tract. | The ViziShot<br>FLEX has been<br>designed to be<br>used with<br>ultrasound<br>endoscopes for<br>ultrasound<br>guided fine<br>needle<br>aspiration<br>(FNA) of<br>submucosal and<br>extramural<br>lesions of the<br>tracheobronchial<br>tree. Do not use<br>this device for<br>any purpose<br>other than its<br>intended use. | Similar to<br>the Cook<br>Predicate. |
| Use Conditions | Surgical suite,<br>endoscopy or<br>bronchoscopy<br>suite, used with a<br>bronchoscope | Surgical suite,<br>endoscopy or<br>bronchoscopy<br>suite, used with<br>a bronchoscope | Surgical suite,<br>endoscopy or<br>bronchoscopy<br>suite, used with a<br>bronchoscope | Identical |
| Mechanics of<br>Action | Manual | Manual | Manual | Identical |
| Mode of Action | Single/multiple<br>puncture and<br>aspirate | Single/multiple<br>puncture and<br>aspirate | Single/multiple<br>puncture and<br>aspirate | Identical |
| General design | Handle, Sheath,<br>Needle, Stylet | Handle, Sheath,<br>Needle, Stylet | Handle, Sheath,<br>Needle, Stylet | Identical |
| Patient<br>Contacting<br>Materials | Stainless Steel,<br>PTFE, PEBAX,<br>Nitinol | Unknown | Stainless Steel,<br>PTFE, PEBAX,<br>Nitinol | Identical<br>to<br>reference<br>ViziShot 2<br>FLEX |
| Biocompatible | Yes | Yes | Yes | Identical |
| Single Use<br>Only | Yes | Yes | Yes | Identical |
| Working OD<br>(mm) | 2.08 | Unknown | 2.08 | Identical<br>to<br>reference<br>ViziShot 2<br>FLEX |
| Catheter Length<br>(cm) | 70 | Unknown | 70 | Identical<br>to<br>reference<br>ViziShot 2<br>FLEX |
| Needle Gauge | 19G | 22G, 25G | 19G | Identical<br>to |
| | Table 5.1 Comparison of Proposed and Predicate Device | |
|--|-------------------------------------------------------|--|
| | | |
{6}------------------------------------------------
| | Proposed Device | Predicate<br>Device<br>(K160229) | Reference Device<br>(K163469) | |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Device Name/<br>Characteristics | Gyrus ACMI<br>ViziShot 2<br>FLEX | Cook Echotip<br>Procore<br>Endobronchial<br>High Definition<br>Ultrasound<br>Biopsy Needle | Gyrus ACMI<br>ViziShot<br>FLEX | Comment |
| | | | | reference<br>ViziShot 2<br>FLEX |
| Typical Needle<br>Length (mm) | 20 | Unknown | 20 | Identical<br>to<br>reference<br>ViziShot 2<br>FLEX |
| Max Needle<br>Length (mm) | 40 | Unknown | 40 | Identical<br>to<br>reference<br>ViziShot 2<br>FLEX |
| Stylet OD (in) | 0.0204 | Unknown | 0.0204 | Identical<br>to<br>reference<br>ViziShot 2<br>FLEX |
| Stylet Surface<br>Finish | Polished | Unknown | Polished | Identical<br>to<br>reference<br>ViziShot 2<br>FLEX |
| Accessories | Syringe with<br>stopcock and<br>Adapter biopsy<br>valve are<br>provided with<br>device | Syringe with<br>stopcock and<br>Device adapter/<br>biopsy valve are<br>provided with<br>device | Syringe with<br>stopcock and<br>Adapter biopsy<br>valve are<br>provided with<br>device | Identical |
| Packaging | Needle<br>assembly,<br>syringe, adapter<br>placed in tray<br>with snap<br>downs and<br>Tyvek lid. Tray<br>placed in shelf<br>box prior to<br>sterilization. | Needle<br>assembly,<br>syringe, adapter<br>placed in tray<br>with snap<br>downs and lid. | Needle<br>assembly,<br>syringe, adapter<br>placed in tray<br>with snap<br>downs and<br>Tyvek lid. Tray<br>placed in shelf<br>box prior to<br>sterilization. | Identical |
| Sterilization | EO | EO | EO | Identical |
| Shelf Life | 3 Year at launch | Unknown | 1 Year at launch | Reference<br>device<br>subsequent<br>to launch<br>achieved 3<br>Year Shelf<br>Life |
# Summary of Perfomance Testing
The proposed ViziShot 2 FLEX in its final finished form is identical to the reference ViziShot 2 FLEX (previously marketed device) in formulation, processing, sterilization, and geometry and no other chemicals have been added (e.g., plasticizers, fillers, additives, cleaning agents, mold release agents).
{7}------------------------------------------------
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility testing:
Biocompatibility testing on all patient contacting surfaces has been performed in Accordance with the FDA Guidance Document "Use of International Standard ISO 10993-1 "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process". Biocompatibility testing included the following tests:
- ISO 10993-4: 2002 Biological evaluation of medical devices Part 4: ● Selection of tests for interactions with blood
- . ISO 10993-5: 2009 Biological evaluation of medical devices - Part5: Tests for in vitro cytotoxicity
- ISO 10993-10: 2010 Biological evaluation of medical devices. Tests for . irritation and sensitization
- ISO 10993-11:2006. Biological evaluation of medical devices. Tests for ● systemic toxicity
- United States Pharmacopeia 39, National Formulary 34, 2016 <151> Pyrogen ● Test
- United States Pharmacopeia 39, National Formulary 34, 2016 <85> Bacterial ● Endotoxins Test
ViziShot 2 FLEX is external communicating, blood path, indirect, and limited duration (<24 hour).
### Sterilization/Shelf Life testing:
The ViziShot 2 FLEX will be delivered in a sterile state and is intended for single patient use only. Sterilization (ethylene oxide) and packaging of the device was validated using the following standards:
- ANSI/AAMI/ISO 11607-1:2006 Packaging for terminally sterilized medical ● devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems
- ANSI/AAMI/ISO 11135-1:2014 Sterilization of health-care products ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices
Packaging integrity and performance testing on devices that had undergone accelerated aging support a labeled three year shelf life. Accelerated aging tests for the ViziShot 2 FLEX was conducted in accordance with ASTM F1980-16. Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.
{8}------------------------------------------------
#### Performance testing - Bench
During design verification, the output of the design process was evaluated against the physical and performance specifications. The following performance tests were conducted:
- · Sheath and Needle Insertion and Withdrawal Force
- · Stylet Insertion and Withdrawal Force
- Bronchoscope Angulation
- · Activation Force
- Plastic Deformation Angle
- · Penetration Force
- · Transmission Force
- Device Durability/Handle Durability
- Bronchoscope Adapter Sliding Force
- · Handle Durability
- · Sheath to Handle Joint Strength
- · Echogenicity
A GLP Comparison of the ViziShot 2 FLEX to the Echotip Procore Endobronchial High Definition Ultrasound Biopsy Needle was performed. The purpose of this study was to evaluate the test article. ViziShot 2 FLEX 19G Needle, and a predicate control article to determine if the needles obtain sufficient tissue samples for histologic and core sample analysis and that the test article needle provided similar quality of samples as the predicate. The performance testing in this study included observations on ease of needle passage through the tissue, and tissue sampling metrics to include weight, length, number of segments and quality of the tissues obtained for microscopic evaluation.
### Performance testing - Animal
No animal study was performed to demonstrate substantial equivalence.
#### Performance testing - Clinical
No clinical study was performed to demonstrate substantial equivalence.
#### Conclusion:
In summary, the Olympus ViziShot 2 FLEX is substantially equivalent to the predicate devices and presents no new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.