K193486 · Biolase, Inc. · GEX · Mar 16, 2020 · General, Plastic Surgery
Device Facts
Record ID
K193486
Device Name
Epic 980
Applicant
Biolase, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 16, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Dental Soft Tissue Indications Incision, excision, vaporization, ablation and coagulation of oral soft-tissues including marginal and inter-dental gingival and epithelial lining of free gingiva and the following specific indications: - Excisional and incisional biopsies - Exposure of unerupted teeth - Fibroma removal - Frenectomy - Frenotomy - Gingival troughing for crown impressions - Gingivectomy - Gingivoplasty - Gingival incision and excision - Hemostasis and coagulation - Implant recovery - Incision and drainage of abscess - Leukoplakia - Operculectomy - Oral papillectomies - Pulpotomy - Pulpotomy as an adjunct to root canal therapy - Reduction of gingival hypertrophy - Soft-tissue crown lengthening - Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa - Vestibuloplasty - Tissue retraction for impression - Laser soft-tissue curettage - Laser removal of diseased, inflamed and necrosed soft-tissue within the periodontal pocket - Sulcular debridement (removal of diseased, inflamed and necrosed soft-tissue in the periodontal pocket to improve clinical indices including gingival bleeding index, probe depth, attachment loss and tooth mobility) - Reduction of bacterial level (decontamination) and inflammation - Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium - Lesion (tumor) removal - Removal of hyperplastic tissues - Laser assisted flap surgery - Removal of granulation tissue Whitening - Light activation for bleaching materials for teeth whitening - Laser-assisted whitening/bleaching of teeth Pain Therapy - Topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm, minor sprains and strains, and minor muscular back pain; the temporary increase in local blood circulation; the temporary relaxation of muscle.
Device Story
Epic 980 is a portable dental diode laser system; emits 980nm infrared energy via solid-state diode. Energy transmitted through flexible fiber optic cable to handpiece with single-use disposable tips. Operated by clinician via touch screen console and wireless footswitch. Used for oral soft-tissue surgery, teeth whitening, and pain therapy (topical heating). Provides precise tissue interaction for incision, excision, and coagulation; aids in periodontal decontamination and inflammation reduction. Benefits include minimally invasive tissue management, improved hemostasis, and temporary pain relief. Device is used in dental clinic settings.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing and comparison of technological characteristics to legally marketed predicate devices.
Technological Characteristics
Solid-state diode laser; 980nm wavelength; 10W max power; continuous/pulsed modes. Delivery via fiber optic cable, handpiece, and single-use quartz tips (200-400μm). Materials: medical-grade plastics, stainless steel, aluminum, brass. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, IEC 80601-2-60. Software developed per IEC 62304.
Indications for Use
Indicated for oral soft-tissue surgical procedures (incision, excision, ablation, coagulation), dental whitening, and temporary relief of minor muscle/joint pain/stiffness in dental patients.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Biolase, Inc Alicia Mszyca Director, Regulatory Affairs 4 Cromwell Irvine, California 92618
Re: K193486
Trade/Device Name: Epic 980 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: December 13, 2019 Received: December 17, 2019
Dear Alicia Mszyca:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Mavadia-Shukla, Ph.D. Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K193486
Device Name Epic 980
### Indications for Use (Describe)
Dental Soft Tissue Indications
Incision, excision, vaporization, ablation of oral soft-tissues including marginal and inter-dental gingival and epithelial lining of free gingiva and the following specific indications:
- · Excisional and incisional biopsies
- · Exposure of unerupted teeth
- Fibroma removal
- · Frenectomy
- · Frenotomy
- · Gingival troughing for crown impressions
- · Gingivectomy
- · Gingivoplasty
- · Gingival incision and excision
- Hemostasis and coagulation
- · Implant recovery
- · Incision and drainage of abscess
- · Leukoplakia
- · Operculectomy
- · Oral papillectomies
- · Pulpotomy
- · Pulpotomy as an adjunct to root canal therapy
- Reduction of gingival hypertrophy
- · Soft-tissue crown lengthening
- · Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa
- · Vestibuloplasty
- · Tissue retraction for impression
- · Laser soft-tissue curettage
- · Laser removal of diseased, inflamed and necrosed soft-tissue within the periodontal pocket
· Sulcular debridement (removal of diseased, inflamed and necrosed soft-tissue in the periodontal pocket to improve clinical indices including gingival bleeding index, probe depth, attachment loss and tooth mobility)
- · Reduction of bacterial level (decontamination) and inflammation
• Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium
- · Lesion (tumor) removal
- · Removal of hyperplastic tissues
- · Laser assisted flap surgery
- · Removal of granulation tissue
### Whitening
- · Light activation for bleaching materials for teeth whitening
- · Laser-assisted whitening/bleaching of teeth
{3}------------------------------------------------
### Pain Therapy
· Topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm, minor sprains and strains, and minor muscular back pain; the temporary increase in local blood circulation; the temporary relaxation of muscle.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">✖</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "BIOLASE" in a bold, sans-serif font. The word is all capitalized and is a dark blue color. The background is white, which makes the word stand out.
### 510(k) SUMMARY
#### I. SUBMITTER
Biolase, Inc. 4 Cromwell Irvine, CA 92618 USA Tel: (949) 226-8471 Fax: (949) 273-6688 Contact Person: Alicia Mszyca Date Prepared: December 13, 2019
#### II. DEVICE
| Name of Device: | Epic 980 |
|----------------------|----------------------------------------------------------------------------------------------------------|
| Common Name: | Dental Diode Laser |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery and<br>in dermatology (21 CFR 878.4810) |
| Device Class: | II |
| Product Code: | GEX |
### III. PREDICATE DEVICES
Epic 980 (K192430) QuickLase 980, 810 & Dual (K100474) SIROLase Advance (K103753) Elumi 810+980 (K152032) Curative980 Diode Laser (K082445)
#### IV. DEVICE DESCRIPTION
The Epic 980 diode laser is a surgical and therapeutic device designated for a wide variety of oral soft-tissue procedures and dental whitening as well as for use in providing a temporary relief of minor pain.
The device uses a solid-state laser diode to emit infrared laser energy which is transmitted via a flexible fiber optic cable to a handpiece that emits the energy to the treatment site.
The laser is comprised of a base console, a wireless footswitch which activates the laser and a detachable delivery system consisting of a fiber optic cable, surgical handpiece and single-use disposable tips designed and optimized for different applications.
{5}------------------------------------------------
## V. INDICATIONS FOR USE STATEMENT
Dental Soft Tissue Indications
Incision, excision, vaporization, ablation and coagulation of oral soft-tissues including marginal and inter-dental gingival and epithelial lining of free gingiva and the following specific indications:
- Excisional and incisional biopsies
- Exposure of unerupted teeth ●
- Fibroma removal ●
- Frenectomy
- Frenotomv
- Gingival troughing for crown impressions ●
- Gingivectomy
- Gingivoplasty
- Gingival incision and excision
- Hemostasis and coagulation
- Implant recovery
- Incision and drainage of abscess
- Leukoplakia
- Operculectomy
- Oral papillectomies
- Pulpotomy
- Pulpotomy as an adjunct to root canal therapy
- Reduction of gingival hypertrophy
- Soft-tissue crown lengthening
- . Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa
- Vestibuloplasty
- Tissue retraction for impression
- Laser soft-tissue curettage
- Laser removal of diseased, inflamed and necrosed soft-tissue within the periodontal pocket
- Sulcular debridement (removal of diseased, inflamed and necrosed soft-tissue ● in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)
- Reduction of bacterial level (decontamination) and inflammation ●
- Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium
- Lesion (tumor) removal
- Removal of hyperplastic tissues ●
- Laser assisted flap surgery
- Removal of granulation tissue
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "BIOLASE" in all capital letters. The word is written in a bold, sans-serif font. The color of the text is a dark blue.
Whitening
- · Light activation for bleaching materials for teeth whitening
- · Laser-assisted whitening/bleaching of teeth
Pain Therapy
- · Topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm, minor sprains and strains, and minor muscular back pain; the temporary
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS VI.
The Epic 980 subject device is the same as Epic 980 (K192430) except for the expanded indications for use already cleared for devices: Ouicklase 810, 980, Dual + (K100474) by QuickLase Limited, SIROLaser Advance (K103753) by Sirona, Elumi 810 + 980 (K152032) by Azena Medical, LLC and Curative980 (K082445) by OroScience, Inc.
The Epic 980 shares the same technological characteristic as the predicate devices including:
- the same laser source: solid state diode producing invisible infrared energy
- the same wavelength: 980nm
- the same intended use: incision, excision, vaporization, ablation and coagulation of ● oral soft tissue
- the same indications for use
- . the same patient-contacting components: glass fiber used in contact and non-contact mode the same use environment
- the same tissue type and application regimen
- the same principle of operation and emission mode: continuous wave, pulsed or . both the same control mechanism
- similar design consisting of software-operated portable laser unit, initiated by a ● footswitch similar delivery system comprising of an optical fiber, handpiece and single use disposable tips
- . the same human factors of user interface
Although some parameters such as maximum power output, power density, pulse rate differ among the devices, these differences do not result is a significantly different clinical performance since the settings and used for the expanded indications as well as the treatment regimen are essentially the same. Therefore, the consolidation of clinical applications creates no new risks or safety concerns.
{7}------------------------------------------------
Companson of the technological characteristics, intended use, indications for use of the Epic 980 subject and predicate devices:
| | Subject Device | Predicate Devices | | | | Reference Device | Pulse duration | 0.01 - 20 ms | 0.01 - 20 ms | 50, 30, 10ms | 0.01 – 0.99 s | variable | unknown |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|-------------------------------------------------------------------|-----------------------------------------------|-------------------------------------------------------|----------------------------------------------------------|--------------------|--------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Specification | Biolase Inc.<br>Epic 980 | Biolase, Inc.<br>Epic 980<br>(K192430) | QuickLase Limited<br>QuickLase 980, 810<br>and Dual+<br>(K100474) | Sirona<br>SIROLaser Advance<br>(K103753) | Azena Medical,<br>LLC<br>Elumi 810 + 980<br>(K152032) | OroScience, Inc.<br>Curative980 Diode<br>Laser (K082445) | Spot size tips | 200 – 400 μm | 200 – 400 μm | 400 μm | 200 - 320 μm | 400 μm | unknown |
| Laser medium | Solid state diode<br>laser | Solid state diode<br>laser | Solid state diode<br>laser | Solid state diode<br>laser | Solid state diode<br>laser | Solid state diode<br>laser | Power density | Up to<br>28294W/cm² | Up to<br>28294W/cm² | Up to<br>15915W/cm² | Up to<br>44563W/cm² | Up to<br>3138W/cm² | unknown |
| Wavelength | 980 ±10 nm | 980 ±10 nm | 980 ± 10 nm, or<br>810 ± 10nm, or<br>dual (810+980) | 970 ± 15nm | 980 ±10 nm, or<br>810 ± 10nm, or<br>dual (810+980) | 980 ± 10 nm | Aiming beam | diode<br>max 1mW<br>625 - 670 nm | diode<br>max 1mW<br>625 - 670 nm | diode<br>max 5mW<br>640 - 650nm | diode,<br>max 1mW<br>635 - 650nm | diode<br>max 5mW<br>640 - 660nm | diode<br>650nm |
| | The devices cleared under K100474 and K152032 operate in 3 different wavelengths: 980nm alone, 810nm<br>alone and dual (810+980nm). Biolase is claiming equivalence to the 980 nm version only.<br>Therefore, substantially equivalent. | | | | | | Control panel | Touch screen | Touch screen | Touch screen | Touch screen | Touch screen | Touch screen |
| Operating<br>modes | Continuous,<br>pulsed | Continuous,<br>pulsed | Continuous,<br>pulsed | Continuous,<br>chopped(pulsed),<br>peak pulse | Pulsed | Continuous,<br>pulsed | Activation | Footswitch | Footswitch | Footswitch | Footswitch | Footswitch | Footswitch |
| Max Power<br>(CW) | 10 W | 10 W | 10W | 7W | 2W | 10 W | Delivery<br>system | Fiber optic cable,<br>handpiece and<br>disposable tips | Fiber optic cable,<br>handpiece and<br>disposable tips | Quarts glass fiber | Fiber optic cable,<br>handpiece and<br>disposable tips | Quartz glass fiber<br>and tips, handpiece<br>and disposable tips | Optical fibers,<br>handpieces and<br>disposable tips |
| | Although Epic 980 is capable of reaching 10W max power, the power settings used for all expanded indications do<br>not exceed 1W, which is sufficient for effective performance and also falls under the maximum power of devices<br>cleared under K103753 and K152032. Therefore, substantially equivalent. | | | | | | Fiber Tips | Quartz single-use<br>tips varying in<br>length and core<br>diameter<br>(200 - 400μm) | Quartz single-use<br>tips varying in<br>length and core<br>diameter<br>(200 - 400μm) | Fiber<br>(400μm) | Single use tips<br>varying in core<br>diameter<br>(200 - 320 μm) | Quarts tips (400μm) | Single use tips |
| Max Peak<br>Power | 10 W | 10 W | 10W @980, or<br>10W @810, or<br>20W @ dual | 14W | 10W @980, or<br>10W @810, or<br>20W @ dual | unknown | Materials | Medical grade<br>plastics, stainless<br>steel, aluminum,<br>brass, and<br>electronic parts<br>and components | Medical grade<br>plastics, stainless<br>steel, aluminum,<br>brass, and<br>electronic parts<br>and components | Medical grade<br>plastics, stainless<br>steel, and electronic<br>parts and<br>components | Medical grade<br>plastics, stainless<br>steel, and electronic<br>parts and<br>components | Medical grade<br>plastics, stainless<br>steel, and electronic<br>parts and<br>components | Medical grade<br>plastics, stainless<br>steel, and electronic<br>parts and<br>components |
| Repetition rate<br>(Frequency) | Up to 20 kHz | Up to 20 kHz | Up to 20 kHz | Up to 20 kHz | 50 Hz | unknown | | | | | | | |
{8}------------------------------------------------
{9}------------------------------------------------
Intended use and Indications for Use
| Epic 980<br>(subject device) | Epic 980<br>K192430 | QuickLase 810,980 &<br>Dual+<br>K100474 | SiroLaser Advance<br>K103753 | Elumi 810 + 980<br>K152032 | Curative 980<br>K082445 |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Dental soft tissue<br>indications:<br>incision, excision,<br>vaporization, ablation<br>and coagulation of oral<br>soft-tissues including<br>marginal and inter-<br>dental gingival and<br>epithelial lining of free<br>gingiva and the<br>following specific<br>indications: | Dental soft tissue<br>indications:<br>Incision, excision,<br>vaporization, ablation<br>and coagulation of oral<br>soft-tissues including<br>marginal and inter-<br>dental gingival and<br>epithelial lining of free<br>gingiva and the<br>following specific<br>indications: | Intended for<br>incision, excision,<br>vaporization,<br>hemostasis and<br>treatment of oral soft-<br>tissue.<br>Examples: | Indicated for intra and<br>extraoral surgery<br>including incision.<br>excision, hemostasis,<br>coagulation and<br>vaporization of soft-<br>tissues including<br>marginal and inter-<br>dental gingival and<br>epithelial lining of free<br>gingiva and is indicated<br>for: | Soft tissue laser<br>intended for the<br>incision, excision,<br>vaporization,<br>hemostasis and<br>treatment of oral soft-<br>tissues. the following<br>are the oropharyngeal<br>indications for use: | Indicated for<br>incision, excision,<br>vaporization, ablation<br>and coagulation of oral<br>soft-tissues (intraoral<br>and extraoral) including<br>marginal and inter-<br>dental gingival and<br>epithelial lining of free<br>gingiva and the<br>following specific<br>indications: |
| excisional and<br>incisional biopsies | excisional and<br>incisional biopsies | excisional and<br>incisional biopsies | biopsies | excisional and<br>incisional biopsies | excisional and<br>incisional biopsies |
| exposure of unerupted<br>teeth | exposure of unerupted<br>teeth | exposure of unerupted<br>teeth | exposure of unerupted/<br>partially erupted teeth | exposure of unerupted<br>teeth | exposure of unerupted/<br>partially erupted teeth |
| fibroma removal | fibroma removal | fibroma removal | fibroma removal | fibroma removal | fibroma removal |
| frenectomy | frenectomy | frenectomy | frenectomy | frenectomy | frenectomy |
| frenotomy | frenotomy | frenotomy | frenotomy | frenotomy | frenotomy |
| gingival troughing for<br>crown impressions | gingival troughing for<br>crown impressions | gingival troughing for<br>crown impressions | gingival troughing | gingival troughing for<br>crown impressions | gingival troughing for<br>crown impressions |
| gingivectomy | gingivectomy | gingivectomy | gingivectomy | gingivectomy | gingivectomy |
| gingivoplasty | gingivoplasty | gingivoplasty | gingivoplasty | gingivoplasty | gingivoplasty |
| gingival incision and | gingival incision and | gingival incision and | gingival incision and | gingival incision and | gingival incision and |
| excision | excision | excision | excision | excision | excision |
| hemostasis and | hemostasis and | hemostasis and | hemostasis of donor | hemostasis and | hemostasis and |
| coagulation | coagulation | coagulation | site | coagulation | coagulation |
| implant recovery | implant recovery | implant recovery | implant recovery | implant recovery | implant recovery |
| incision and drainage of | incision and drainage of | incision and drainage of | incision and drainage of | incision and drainage of | incision and drainage of |
| abscess | abscess | abscess | abscess | abscess | abscess |
| leukoplakia | leukoplakia | leukoplakia | leukoplakia | leukoplakia | leukoplakia |
| operculectomy | operculectomy | operculectomy | operculectomy | operculectomy | operculectomy |
| oral papillectomies | oral papillectomies | oral papillectomies | papillectomy | oral papillectomies | oral papillectomies |
| pulpotomy | pulpotomy | pulpotomy | pulpotomy | pulpotomy | pulpotomy |
| pulpotomy as an | pulpotomy as an | pulpotomy as an | pulpotomy as an | pulpotomy as an | pulpotomy as an |
| adjunct to root canal | adjunct to root canal | adjunct to root canal | adjunct to root canal | adjunct to root canal | adjunct to root canal |
| therapy | therapy | therapy | therapy | therapy | therapy |
| reduction of gingival | reduction of gingival | reduction of gingival | reduction of gingival | reduction of gingival | reduction of gingival |
| hypertrophy | hypertrophy | hypertrophy | hypertrophy | hypertrophy | hypertrophy |
| soft-tissue crown | soft-tissue crown | soft-tissue crown | crown lengthening | soft-tissue crown | soft-tissue crown |
| lengthening | lengthening | lengthening | | lengthening | lengthening |
| treatment of canker | treatment of canker | treatment of canker | treatment of canker | treatment of canker | treatment of canker |
| sores, herpetic and | sores, herpetic and | sores, herpetic and | sores, herpetic and | sores, herpetic and | sores, herpetic and |
| aphthous ulcers of the | aphthous ulcers of the | aphthous ulcers of the | aphthous ulcers of the | aphthous ulcers of the | aphthous ulcers of the |
| oral mucosa | oral mucosa | oral mucosa | oral mucosa | oral mucosa | oral mucosa |
| vestibuloplasty | vestibuloplasty | vestibuloplasty | vestibuloplasty | vestibuloplasty | vestibuloplasty…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.