K082445 · Oroscience, Inc. · GEX · Oct 15, 2008 · General, Plastic Surgery
Device Facts
Record ID
K082445
Device Name
CURATIVE980 DIODE LASER
Applicant
Oroscience, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 15, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Curative980 Diode Lasers and the delivery accessories that are used with them to deliver laser energy are intended for use in the medical specialty of dental surgery. The Curative980 Diode Lasers (and the delivery accessories that are used with them to deliver laser energy) is indicated for use in a variety of applications requiring incision, excision, ablation, vaporization, coagulation and hemostasis of soft tissue in the medical specialty of dental surgery.
Device Story
Curative980 Diode Laser is a surgical instrument for dental soft tissue procedures. System includes a main console with a 980 nm AlGaAs solid-state laser diode for treatment and a 650 nm diode for aiming; a touch-screen control panel; footswitch; and fiber-optic delivery accessories (handpieces, tips). Operated by dental professionals in clinical settings. The device delivers laser energy to the target tissue to perform incision, excision, ablation, vaporization, coagulation, or hemostasis. The healthcare provider uses the handpiece to direct the laser beam, with visual feedback provided by the aiming beam. The device enables precise soft tissue management, potentially reducing bleeding and improving healing compared to traditional methods. Accessories include fiber strippers and cleavers for maintenance.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and intended use compared to legally marketed predicate devices.
Technological Characteristics
980 nm AlGaAs solid-state laser diode; 650 nm aiming beam; touch-screen interface; fiber-optic delivery system with reusable handpieces and disposable tips; footswitch-controlled; electrical power source; standard dental surgical laser architecture.
Indications for Use
Indicated for incision, excision, vaporization, ablation, coagulation, and hemostasis of oral soft tissue (intraoral and extraoral) in dental surgery. Specific applications include biopsies, frenectomy, gingivectomy, gingivoplasty, implant recovery, sulcular debridement, treatment of aphthous/herpetic ulcers, pulpotomy, crown lengthening, and light-activated tooth whitening.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Quanta System Diode Medical Laser 808, 940, or 980 nm (K072034)
B&W Tek, Inc. BWF-5 Medical Laser Series (K062363)
Submission Summary (Full Text)
{0}------------------------------------------------
# Section 5 - 510(k) Summary
#### I. General Information
| Submitter: | OroScience, Inc.<br>2275 East Bayshore Road, Suite 106<br>Palo Alto, CA 94303 |
|---------------------------|-------------------------------------------------------------------------------|
| Contact Person: | Phil Houle<br>Vice President R&D, Operations |
| Summary Preparation Date: | July 2, 2008 |
#### II. Names
| Device Names: | Curative980 Diode Lasers (and delivery device accessories) |
|---------------|------------------------------------------------------------|
|---------------|------------------------------------------------------------|
Primary Classification Names: Laser Powered Surgical Instruments (and Accessories)
#### III. Predicate Devices
- Quanta System Diode Medical Laser 808, 940, or 980 nm (K072034) .
- Hoya ConBio DioDent MICRO 980 nm (K063384) .
- Biolitec Ceralas 980 nm / SmilePro 980 (K050824) .
- Laser Dental Innovations LiteSaber™ 2000 (K993942) .
- Laser Dental Innovations StarLite 2006 (K052604) .
- B&W Tek, Inc. BWF-5 Medical Laser Series (K062363) .
### IV. Product Description
.
The Curative980 Diode Lasers are comprised of the following main components:
- Main console containing the major electrical components, including:
- > Control Touch-Screen Display Panel including touch controls;
- > 980 nm treatment laser (aluminum gallium arsenide (AlGaAs) solid state laser diode);
- 650 nm aiming beam diode laser; A
- Delivery device fiber-optic connector port; A
- A Emergency stop switch;
- Remote interlock connector (External door interlock connector); A
- A Connector ports for the footswitch and power cord;
- Handpiece holder (attaches to the top of the laser system console); A
- Fiber Spool (secures and organizes the optical fiber) attaches to the side of A the main console;
- Footswitch: .
- Medical grade power cord; .
- Delivery Devices: .
- Reusable, cleanable, sterilizable optical fibers; > Optical Fibers -
{1}------------------------------------------------
- A Handpieces -Reusable, cleanable, sterilizable handpieces; A Handpiece Tips - Disposable single-use tips;
- . Accessories:
- A Safety Glasses
- Tools:
- A Optical Fiber Striper;
- A Optical Fiber Cleaver.
### V. Indications for Use
The Curative980 Diode Lasers and the delivery accessories that are used with them to deliver laser energy are intended for use in the medical specialty of dental surgery.
The Curative980 Diode Lasers (and the delivery accessories that are used with them to deliver laser energy) is indicated for use in a variety of applications requiring incision, excision, ablation, vaporization, coagulation and hemostasis of soft tissue in the medical specialty of dental surgery.
### VI. Rationale for Substantial Equivalence
The Curative980 Diode Lasers share the same or similar indications for use, device operation, overall technical and functional capabilities, and therefore is substantially equivalent to the predicate devices.
### VII. Safety and Effectiveness Information
The review of the indications for use and technical characteristics provided demonstrates that the Curative980 Diode Lasers are substantially equivalent to the predicate devices.
## VIII. Conclusion
The Curative980 Diode Lasers were found to be substantially equivalent to the predicate devices.
The Curative980 Diode Lasers share identical indications for use, similar design features, and functional features with, and thus are substantially equivalent to, the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
OCT 1 5 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OroScience, Inc. % A Worden Consulting Ms. Anne Worden 3637 Bernal Avenue Pleasanton, California 94566
Re: K082445 Trade/Device Name: OroScience Curative980 Diode Lasers Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use ing eneral and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: October 9, 2008 Received: October 10, 2008
Dear Ms. Worden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Anne Worden
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 100-100.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH)'s) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Mark N. Millman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use Statement - Continued
510(k) Number (if known): K08
### Device Name: OroScience Curative980 Diode Lasers
Indications for Use:
The Curative980 Diode Lasers and the delivery accessories that are used with them to deliver laser energy are intended to deliver laser energy for incision, ablation, vaporization, coagulation and hemostasis of soft tissue in the medical specialty of dental surgery as follows:
### Dental Surgery
Indicated for incision, excision, vaporization, ablation and coagulation of oral soft tissue (intraoral and extraoral) including marginal and inter-dental gingival and epithelial lining of free gingival and the following specific indications:
- Biopsy ●
- Excisional and incisional biopsies .
- Excision of lesions ●
- Exposure of unerupted/partially erupted teeth .
- Fibroma removal .
- Frenectomy .
- Frenotomy 0
- Gingival troughing for crown impressions .
- . Gingival troughing
- . Gingivectomy
- . Gingivoplasty
- Gingival incision and excision .
- Hemostasis and coagulation .
- Hemostasis of donor site �
- Implant recovery .
- Removal of granulation tissue .
- Laser assisted flap surgery .
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use .(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
| 510(k) Number | 1082445 |
|---------------|---------|
|---------------|---------|
Page 1 of 2
017
{5}------------------------------------------------
# Indications for Use Statement - Continued
510(k) Number (if known): K08
### Device Name: OroScience Curative980 Diode Lasers
Indications for Use - Continued:
### Dental Surgery - Continued
- Debridement of diseased epithelial lining ●
- Treatment of aphthous ulcers .
- Sulcular debridement (removal of diseased or inflamed soft-tissue in the periodontal . pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)
- Laser soft tissue curettage .
- Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa .
- . Operculectomy
- Incision and drainage of abscess .
- Oral papillectomies; papillectomy .
- Removal of hyperplastic tissues .
- . Pulpotomy
- . Pulpotomy as an adjunct to root canal therapy
- . Reduction of gingival hypertrophy
- Reduction of bacterial level (decontamination) and inflammation .
- Soft tissue crown lengthening; crown lengthening .
- Tissue retraction for impressions .
- Leukoplakia .
- Vestibuloplasty .
- Light activation of bleaching materials for teeth whitening. .
- Laser-assisted bleaching/whitening for teeth .
V Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 2 of 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.