Avantgarde Faecal Management Systems, Avantgarde Faecal Management System Replacement Pouches
Applicant
Prosys International, Ltd.
Product Code
KNT · Gastroenterology, Urology
Decision Date
Mar 13, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Avantgarde Faecal Management System is intended for faecal management by the collection of semi-liquid or liquid faecal matter in bedridden adults.
Device Story
Avantgarde Faecal Management System is an invasive device for managing faecal matter in bedridden adults. It consists of a silicone catheter tube with an inflation cuff, irrigation port, 1,500ml collection bag, mounting plate, and 45ml syringe. The catheter is inserted into the rectum; the balloon is inflated to retain the device. Liquid or semi-liquid stool is diverted through the tube into the collection bag. The system includes a pressure release valve on the inflation port to prevent over-inflation. The device is used in clinical settings by healthcare providers to manage incontinence, reduce skin breakdown, and improve patient hygiene.
Clinical Evidence
Bench testing only. Testing included accelerated aging (ASTM F1980), seal strength (ASTM F88), leak detection (ASTM F1929), shipping performance (ASTM D4169), and urine collection bag requirements (ISO 8669-2). Biocompatibility testing was performed per ISO 10993, confirming the device is non-toxic.
Technological Characteristics
Silicone catheter with Parylene coating; 1,500ml collection bag; inflation cuff with pressure release valve; irrigation port (Luer). Non-sterile, single-use. 5-year shelf life. Standards: ASTM F1980, F88, F1929, D4169; ISO 8669-2, ISO 10993.
Indications for Use
Indicated for faecal management via collection of semi-liquid or liquid faecal matter in bedridden adults.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized depiction of a human figure. To the right of it is the FDA logo, with the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
March 13, 2020
ProSys International Ltd Graham Steer R&D Director Suite 303, Highland House 165 The Broadway SW19 1NE Wimbledon, Loudon UNITED KINDOM
Re: K193325
Trade/Device Name: Avantgarde Faecal Management Systems Avantgarde Faecal Management System Replacement Pouches Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: KNT Dated: February 13, 2020 Received: February 19, 2020
Dear Graham Steer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Shani P. Haugen, PhD Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K193325
Device Name
Avantgarde Faecal Management System Avantgarde Faecal Management System Replacement Pouches
Indications for Use (Describe)
The Avantgarde Faecal Management System (FMS) is intended for faecal management by the collection of semi liquid or liquid faecal matter in bedridden adults.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
*PRAStaff@fda.hhs.gov*
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for Pro Sys International Limited. The words "Pro" and "International Limited" are in blue, while "Sys" is in white. The logo is simple and professional, and it is likely used by a company that provides professional services.
K193325 Page 1 of 3
# 510(k) SUMMARY
In accordance with 21 CFR 807.92 and the Safe Medical Devices Act 1990, the following information is provided for the Avantgarde Faecal Management System 510(k) Premarket Notification. This submission was prepared in accordance with the FDA Guidance Document "Format for Traditional & Abbreviated 510(k)s'
| Subject Device(s): | Avantgarde Faecal Management System<br>Avantgarde Faecal Management System Replacement Pouches |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | 1st December 2019 |
| Applicant: | ProSys International Ltd<br>Suite 303, Highland House,<br>165 The Broadway<br>Wimbledon<br>London<br>SW19 1NE<br>United Kingdom |
| Contact: | Mr Graham Steer<br>R&D Director<br>ProSys International Ltd<br>Tel: + 44 (0) 208 944 7585<br>Fax: + 44 (0) 208 944 5434<br>Email: gsteer@prosysinternational.co.uk |
| Device Trade Name: | Avantgarde Faecal Management System |
| Classification Name: | Gastrointestinal tube and accessories |
| Product Code: | KNT |
| Regulation: | CFR 876-5980 |
| Device Class: | Class II |
# PREDICATE DEVICES:
Legally marketed Medical Devices to which Substantial Equivalence is claimed:
| Trade Name: | Flexi-Seal Signal |
|--------------------------------------|-----------------------------------------|
| Manufacturer: | ConvaTec Inc |
| 510(k) Substantial Equivalence: | K150350 - See FDA Approval in Section 3 |
| Determined Substantially Equivalent: | 21st July 2015 |
{4}------------------------------------------------
### DEVICE DESCRIPTION
The Avantgarde Faecal Management System (catheter) is an invasive medical device to be used inside the body. It consists of a silicone catheter tube with inflation cuff and irrigation port, a 1,500ml collection bag, mounting plate with an attachment strap and 45ml syringe. This tray of devices, is supplied as a Convenience Kit. In addition, other related medical devices to enable the Avantgarde work effectively, include the Avantgarde Faecal Management System Replacement Pouches which are to be supplied on a bundle basis.
This non-sterile device is for single use only.
The FMS catheter tube is constructed of a silicone material. The balloon once inflated retains the device in the rectum. The liquid or semi-liquid stool is diverted into the drainage tube, which provides a route for the liquid or semi-liquid faecal matter to pass into the collection bag.
Along the tube are two lumens with individual access ports. One port to inflate and deflate the balloon and the other to add water to flush and irrigate the catheter tube.
Its shelf life is 5 years from the date of production.
#### INTENDED USE
The Avantgarde Faecal Management System is intended for faecal management by the collection of semi-liquid or liquid faecal matter in bedridden adults.
## SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
The following table which forms part of this 510(k) submission contains a very brief summary of the technological characteristics of Avantgarde Faecal Management System compared to the predicate device. This table is supported by additional information in section 30 (Substantial Equivalence Discussion) and the Executive Summary section 28.
| Function | Avantgarde Faecal<br>Management System | ConvaTec Flexi Seal Signal<br>FMS | Similarities |
|------------------------------------|--------------------------------------------------|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Faecal Management | Faecal Management | Both devices are the same |
| Animal Derived<br>Material Content | None | None | Both devices are the same |
| Sterility | Non Sterile | Non Sterile | Both devices are the same |
| Medical Device Class | Class II | Class II | Both devices are the same |
| Inflation Port | Pressure Release Valve | Luer | Avantgarde is provided with a pressure<br>release valve to ensure the balloon/cuff<br>can not physically be over inflated. |
| Irrigation Port | Luer | Luer | Both devices are the same |
| Collection Bag | 1,500ml Silicone Disposable<br>Bags | 1,000ml Disposable<br>Collection Bags | Avantgarde has a larger collection<br>capacity but they are basically similar |
| Catheter Tube Length | 1,700mm | 1,550mm – 1,670mm | Avantgarde is slightly longer but they are<br>basically similar |
| Device Material | Constructed of Silicone with<br>Parylene coating | Silicone Rubber | Avantgarde has an enhanced waste<br>fecal flow rate but they are basically<br>the same |
{5}------------------------------------------------
# SUMMARY OF PERFORMANCE (NON CLINICAL TESTING) DATA
Non clinical testing of the subject device for functional and structural characteristics has been performed. In the testing, the device was found to be substantially equivalent to the predicate device. Typical testing included F1980 Standards guide for accelerated aging of sterile medical device package; F 88 Standard test method for seal strength of flexible barrier materials; F1929 Standard test method for detecting seal leaks in porous medical packaging by dye penetration; ASTM D 4169 Standard practice for performance testing of shipping containers and systems; W-ZK-08-65 Final Inspection criteria for faecal management system: ISO8669-2 Urine collection bags-part 2: Requirements and test methods.
In addition Avantgarde Faecal Management Systems had been tested against ISO 10993 by Soochow University and it confirms that Avantgarde is non-toxic.
In conclusion, the subject device (Avantgarde) has been shown to be as safe and effective and substantially equivalent to the predicate devices.
## END
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.