The Flexi-Seal® SIGNAL™ Fecal Management System is an indwelling fecal management catheter intended for use to manage fecal incontinence through the collection of liquid to semi-liquid stool and to provide access to administer medications. The device is not intended for use on pediatric patients.
Device Story
Indwelling fecal management system; manages fecal incontinence in non-pediatric patients. Components: soft silicone catheter tube, retention balloon, collection bag, syringe, cinch clamp. Clinician inserts catheter into rectum; inflates low-pressure retention balloon with water/saline via dedicated port; balloon provides visual/tactile signal of optimal inflation. Catheter features odor-adsorbing material to reduce fecal odor. Sampling port allows stool collection; irrigation port allows medication administration. Device collects liquid/semi-liquid stool into external bag. Used in clinical settings by healthcare providers. Benefits: contains waste, reduces odor, protects skin, facilitates medication delivery.
Clinical Evidence
Bench testing only. No clinical data presented. Functional and structural parameters (material strength, catheter size, balloon size) were compared to the predicate device. Biocompatibility evaluated per ISO-10993.
Technological Characteristics
Collapsible silicone rubber catheter with odor-adsorbing additive. Features: low-pressure silicone retention balloon, Luer-lock inflation/irrigation ports, sampling port, 1L collection bag with gas filter. Non-sterile, single-use. Dimensions: 1,550mm-1,670mm length. No animal products, natural rubber latex, or DEHP.
Indications for Use
Indicated for management of fecal incontinence in non-pediatric patients via collection of liquid to semi-liquid stool and administration of medications.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
Flexi-Seal® SIGNAL™ Fecal Management System (K112342)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 21, 2015
ConvaTec, Ltd. Clare Williamson Regulatory Affairs Documentation Compliance Manager Unit 20, First Avenue, Deeside Industrial Park Deeside, Flintshire CH5 2NU UK
Re: K150350 Trade/Device Name: Flexi-Seal® Signal™ Fecal Management System Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: KNT Dated: June 18, 2015 Received: June 24, 2015
Dear Clare Williamson,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| Form Approved: OMB No. 0910-0120 | |
|-----------------------------------|--|
| Expiration Date: January 31, 2017 | |
| See PRA Statement below. | |
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
**Indications for Use**
| 510(k) Number (if known) | K150350 |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Flexi-Seal® Signal™ Fecal Management System |
| Indications for Use (Describe) | For use to manage fecal incontinence through the collection of liquid to semi-liquid stool and to provide access to administer medications. |
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
| FORM FDA 3881 (1/14) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 EF |
|----------------------|-------------|-------------------------------------------|
|----------------------|-------------|-------------------------------------------|
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for ConvaTec. The logo consists of the word "ConvaTec" in blue, followed by a blue circle with a stylized image of a building with three columns inside. The logo is simple and modern, and the blue color gives it a sense of trustworthiness and reliability.
Unit 20, First Avenue Deeside Industrial Park Deeside Flintshire CH5 2NU Telephone: 01244 584000 Facsimile: 01244 584011
## 510(k) Summary
Subject Device: Flexi-Seal® SIGNAL™ Fecal Management System
Date Prepared: February 06, 2015
Applicant: ConvaTec Ltd.
First Avenue Deeside Industrial Park Deeside Flintshire CH5 2NU UK
Contact: Clare Williamson
> Regulatory Affairs Documentation Compliance Manager Tel: +44 (0)1244 584155 Fax: +44 (0)1244 584311 Email: clare.williamson@convatec.com
Device Trade Name: Flexi-Seal® SIGNAL™ Fecal Management System
Common Name: Fecal Management System
Classification Name: Gastrointestinal Tube and Accessories (21 CFR §876.5980, Product Code: KNT)
Device Class: Class II
#### Predicate Device:
Device Trade Name: Flexi-Seal® SIGNAL™ Fecal Management System -ConvaTec Inc.
Common Name: Fecal Management System
Classification Name: Gastrointestinal Tube and Accessories (21 CFR §876.5980, Product Code: KNT)
Device Class: Class II
510(k) Substantial Equivalence: K112342 - determined substantially equivalent on April 26, 2012
Device Description:
{4}------------------------------------------------
The Flexi-Seal® SIGNAL™ Fecal Management System is comprised of a soft catheter tube assembly, a Luer-lock syringe, a collection bag with cap and a cinch clamp to pinch off flow in the catheter when required for medication retention or to stop waste flow. The components are contained in a rigid thermoformed plastic clamshell.
The catheter main drain tube is fabricated from collapsible silicone rubber with the addition of an odor adsorber, which is added in order to help contain the fecal odor which often permeates through the wall of the catheter.
The drain tube has a low-pressure silicone retention balloon at the distal end and a connector for attaching a collection bag (provided with the device and also available separately) at the proximal end. There is a recess (pocket) under the balloon for the clinician's finger, which allows the device to be positioned digitally.
Two ports are attached to the side of the catheter. One port is used to inflate the retention balloon with water or saline after the device has been inserted into the patient's rectum. This port also provides a visual and tactile signal of when the low pressure retention balloon is filled to its optimal volume. The other port is used to flush the device if needed and administer medication if prescribed.
An additional sampling port is located on the side of the catheter, which allows access to clinicians for stool sample collection.
A syringe (provided with the device) is used to fill and evacuate the retention balloon for insertion and removal.
The device, collection bag and syringe are intended for single us, are provided non-sterile and contain no components made with animal products, natural rubber latex or DEHP.
This Traditional 510(k) concerns modifications which do not alter the indications for use of the predicate device, the Flexi-Seal® SIGNAL™ Fecal Management System (ref. K112342).
### Intended use:
The Flexi-Seal® SIGNAL™ Fecal Management System is an indwelling fecal management catheter intended for use to manage fecal incontinence through the collection of liquid to semi-liquid stool and to provide access to administer medications. The device is not intended for use on pediatric patients.
## Summary of Technological Characteristics:
The Flexi-Seal® SIGNAL™ Fecal Management System has the same intended use and indications for use as the predicate device, however the tubing now incorporates odor adsorbing properties. The adsorber reduces the odor which often permeates through the wall of the catheter. These unpleasant odors during catheter use can cause major discomfort to the patients, caregivers and visitors. Therefore, the reduction of odor is advantageous to overall patient wellbeing.
The device specifications are essentially identical to those of the predicate device, although minor adjustments have been made to the catheter tubing to
{5}------------------------------------------------
ensure that the material softness remains. Catheter tube diameter, catheter tube hardness, catheter tube tensile strength, and balloon-catheter bonding were evaluated via bench testing and are substantially equivalent to the predicate device. There are no differences in operation between the Flexi-Seal® SIGNAL ™ Fecal Management System and the predicate device.
## Summary of Performance (Non-Clinical Testing) Data:
Non-clinical testing of the subject device for functional and structural parameters has been performed. In this testing, the device's performance has been found to be substantially equivalent to the aforementioned predicate device both functionally and structurally (material strength, catheter size, balloon size, etc.). The device has also been evaluated for biocompatibility in accordance with the US Food and Drug Administration's guidance entitled Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing', issued May 1, 1995, and has been found safe in such respect for its intended use.
In conclusion, the subject device has been demonstrated as safe and effective and substantially equivalent to the predicate device.
{6}------------------------------------------------
### Summary of Technological Characteristics
The following table summarises the technological characteristics of the subject and predicate devices (ref. K112342, determined substantially equivalent on April 26, 2012) and outlines the product characteristics and specifications which form the basis of the substantial equivalence discussion.
The intended use, technological characteristics and principles of operation of the Flexi-Seal® SIGNAL™ Fecal Management System remains the same as those of the predicate device.
| | Subject Device | Predicate | Comparison | |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|-------------|
| Parameter | Flexi-Seal® SIGNAL™ Fecal Management<br>System | Flexi-Seal® SIGNAL™ Fecal Management<br>System<br>(K112342) | Similarities | Differences |
| FDA Product<br>Code | KNT | KNT | Both devices are the same | None |
| FDA<br>Classification<br>Regulation | 21 CFR 876.5980 | 21 CFR 876.5980 | Both devices are the same | None |
| Regulatory<br>Class | Class II | Class II | Both devices are the same | None |
| Intended Use | To manage fecal incontinence through the<br>collection of liquid to semi-liquid stool and to<br>provide access to administer medications | To manage fecal incontinence through the<br>collection of liquid to semi-liquid stool and to<br>provide access to administer medications | Both devices are the same | None |
| Functional<br>configuration | Collapsible catheter with distal end secured in<br>the rectum and proximal end connected to a<br>collection bag | Collapsible catheter with distal end secured in<br>the rectum and proximal end connected to a<br>collection bag | Both devices are the same | None |
| Retention<br>feature | Soft annular balloon | Soft annular balloon | Both devices are the same | None |
| Balloon material | Silicone rubber | Silicone rubber | Both devices are the same | None |
| Inflation<br>management | Sealed water filled balloon | Sealed water filled balloon | Both devices are the same | None |
| Retention<br>balloon inflation<br>material | Water/saline (45ml) | Water/saline (45ml) | Both devices are the same | None |
| Catheter tube<br>length | 1,550mm - 1,670mm | 1,550mm - 1,670mm | Both devices are the same | None |
| Length of<br>sphincter section | >10cm | >10cm | Both devices are the same | None |
{7}------------------------------------------------
| Parameter | Subject Device | Predicate | Comparison | |
|-------------------------------------------------|------------------------------------------------|-------------------------------------------------------------|---------------------------|-------------|
| | Flexi-Seal® SIGNAL™ Fecal Management<br>System | Flexi-Seal® SIGNAL™ Fecal Management<br>System<br>(K112342) | Similarities | Differences |
| Drain channel | Large main drain tube | Large main drain tube | Both devices are the same | None |
| Inflation and<br>irrigation port<br>connections | Luer | Luer | Both devices are the same | None |
| Inflation lumen<br>diameter | Two lumens at 1.45 mm | Two lumens at 1.45 mm | Both devices are the same | None |
| Inflation lines | 2 | 2 | Both devices are the same | None |
| Irrigation lines | 1 | 1 | Both devices are the same | None |
| Sampling port | Closable snap seal for catheter syringe access | Closable snap seal for catheter syringe access | Both devices are the same | None |
| Collection bag<br>configuration | Disposable with gas filter | Disposable with gas filter | Both devices are the same | None |
| Collection bag<br>size | 1 litre disposable | 1 litre disposable | Both devices are the same | None |
| Method of bed<br>connection | Hanging strap on bag connector | Hanging strap on bag connector | Both devices are the same | None |
| Accessory<br>components | Syringe and cinch clamp | Syringe and cinch clamp | Both devices are the same | None |
| Flow stop<br>mechanism | External cinch clamp | External cinch clamp | Both devices are the same | None |
| Sterility status<br>(how provided) | Non-sterile | Non-sterile | Both devices are the same | None |
| Packaging | Thermoformed plastic clamshell | Thermoformed plastic clamshell | Both devices are the same | None |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.