K193278 · Medsource Labs · FOZ · Aug 7, 2020 · General Hospital
Device Facts
Record ID
K193278
Device Name
MedSource CathMED IV Catheter
Applicant
Medsource Labs
Product Code
FOZ · General Hospital
Decision Date
Aug 7, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The MedSource CathMED IV Catheter is indicated to sample blood or administer fluids intravenously. The MedSource CathMED IV Catheter may be used for any patient population with consideration given to adequacy for the vascular anatomy appropriateness for the solution being administered and duration of the therapy.
Device Story
Intravascular catheter for blood sampling or fluid administration; used in clinical settings by healthcare providers. Device consists of a sharp needle with an over-the-needle catheter, protective cover, hub, and gauge chamber. Operation: clinician inserts needle/catheter into vein; catheter advanced off needle; needle removed; catheter remains in vein for short-term access. Provides venous access for therapy or sampling. No needle-stick prevention feature present; labeling includes warning regarding absence of safety mechanism. Available in 14G to 24G sizes.
Clinical Evidence
Bench testing only. Conformance to ISO 10555-1, ISO 10555-5, ISO 594-1, ISO 594-2, and USP <788> standards. Biocompatibility established via reference to predicate device data. Sterilization validated per ISO 11135:2014.
Technological Characteristics
Materials: Polyurethane/PTFE catheter, stainless steel needle, K-Resin hub/body, Polyacetal (POM) catheter holder. Design: Over-the-needle peripheral catheter with Luer taper fittings. Sterilization: Terminal Ethylene Oxide (EtO). No integrated software or electronics.
Indications for Use
Indicated for any patient population requiring intravenous blood sampling or fluid administration, contingent upon vascular anatomy, solution appropriateness, and therapy duration.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 7, 2020
MedSource Labs Emilie Andrews Quality Technician 8600 Shelby Court Chanhassen, MN 55317
Re: K193278
Trade/Device Name: MedSource CathMED IV Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: FOZ Dated: July 13, 2020 Received: July 13, 2020
Dear Emilie Andrews:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Payal Patel Assistant Director, General Hospital DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K193278
Device Name MedSource CathMED IV Catheter
#### Indications for Use (Describe)
The MedSource CathMed IV Catheter is indicated to sample blood or administer fluids intravenously. The MedSource CathMED IV Catheter may be used for any patient population with consideration given to adequacy for the vascular anatomy appropriateness for the solution being administered and duration of the therapy.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510 K-Summary- K193278
#### Date of Prepared: August 6, 2020
### A. Submitter:
MedSource International, LLC 8600 Shelby Court Chanhassen, MN 55317
#### B. Contact Person:
Emilie Andrews, Quality and Regulatory Affairs 8600 Shelby Court Chanhassen. MN 55317 Phone: 952-472-0131
#### C. Proposed Device:
Product Code: FOZ Trade Name: MedSource CathMED IV Catheter Common Name: Intravascular catheter Classification Name: Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days Regulatory Reference: 21 CFR §880.5200 Review Panel: General Hospital Classification: Class II
#### D. Predicate:
510(K): K131555 Product Code: FOZ Trade Name: MedSource IV Safety Catheter Common Name: Intravascular catheter Classification Name: Catheter, Intravascular, Therapeutic, Short-Term Less Than 30Days Regulatory Reference: 21 CFR §880.5200 Review Panel: General Hospital Classification: Class II
#### E. Indications for Use:
The MedSource CathMED IV Catheter is indicated to sample blood or administer fluids intravenously. The MedSource CathMED IV Catheter may be used for any patient population with consideration given to adequacy for the vascular anatomy appropriateness for the solution being administered and duration of the therapy.
{4}------------------------------------------------
## F. Device Description:
The MedSource CathMED IV Catheter is indicated to sample blood or administer fluids intravenously. The MedSource CathMED IV Catheter may be used for any patient population with consideration given to adequacy for the vascular anatomy appropriateness for the solution being administered and duration of the therapy.
Each MedSource CathMED IV Catheter is comprised of the following components: protective needle cover, color-coded catheter hub, radiopaque catheter tube, Teflon catheter holder, needle guide, medical grade stainless steel needle hub, porous plug and gauge chamber. The MedSource CathMED IV Catheter components combine for an ergonomic design.
The device inserts a catheter intravenously by inserting a sharp needle with an over-theneedle catheter. Once the needle and catheter have been passed into a vein the catheter can be advance off the needle is the needle is removed while the catheter remains in the vein for short term use to sample blood or administer fluids.
The MedSource CathMED IV Catheter gauges are color coded for positive identification and range from 14G to 24G
ORANGE = 14G GREY = 16G GREEN = 18G PINK = 20G BLUE = 22G YELLOW = 24G
## G. SUMMARY OF COMPARISON TABLE OF PROPOSED DEVICE and PREDICATE
The proposed device is substantially equivalent to the predicate device the MedSource Safety Catheter (K131555).
| Comparison<br>Point | Submission Device<br>MedSource CathMED IV<br>Catheter | Predicate Device<br>MedSource IV Safety<br>Catheter (K131555) | Discussion | Comments |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technology | | | | |
| Indications for<br>Use | The MedSource<br>CathMED IV Catheter is<br>indicated to sample blood<br>or administer fluids<br>intravenously. The<br>MedSource IV Catheter<br>may be used for any<br>patient population with<br>consideration given to<br>adequacy for the vascular<br>anatomy appropriateness<br>for the solution being<br>administered and | The MedSource IV Safety<br>Catheter is indicated to<br>sample blood or administer<br>fluids<br>intravenously. The<br>MedSource IV Safety<br>Catheter may be used for<br>any patient population<br>with consideration given to<br>adequacy for the vascular<br>anatomy appropriateness<br>for the<br>solution being | Same | |
| | duration of the therapy. | administered and duration<br>of the therapy. | | |
| Intended Use | The MedSource<br>CathMED IV Catheter is<br>indicated to sample blood<br>or administer fluids<br>intravenously. The<br>MedSource CathMED IV<br>Catheter may be used for<br>any patient population<br>with consideration given<br>to adequacy for the<br>vascular anatomy<br>appropriateness for the<br>solution being<br>administered and<br>duration of the therapy. | The MedSource IV Safety<br>Catheter is indicated to<br>sample blood or administer<br>fluids intravenously. The<br>MedSource IV Safety<br>Catheter may be used for<br>any patient population with<br>consideration given to<br>adequacy for the vascular<br>anatomy appropriateness<br>for the solution being<br>administered and duration<br>of the therapy. | Same | |
| Distal End<br>configuration | Beveled | Beveled | Same | |
| Proximal End<br>configuration | Cupper | Cupper | Same | |
| Needle Stick<br>Prevention<br>Feature | Not present | Push-button needle<br>shielding | Different | Absent in the<br>submission<br>device.<br>A warning was<br>added to the<br>labeling of this<br>product that a<br>Needle Stick<br>Prevention<br>Feature is not<br>present in this<br>device.<br>This does not<br>affect the safety<br>and<br>effectiveness of<br>the device. |
| Catheter<br>Length<br>Catheter O.D.<br>Needle Gauge | Gauge Length OD<br>mm mm | Gauge Length OD<br>mm mm | Different | one new Gauge<br>Length "20G<br>length 33 mm<br>OD 1.1 mm" |
| | 24G | 19 | 0.7 | |
| | 24G | 19 | 0.7 | |
| | 22G | 25 | 0.9 | |
| | 22G | 25 | 0.9 | |
| | 20G 33 1.1<br>20G 25 1.1<br>18G 45 1.3<br>16G 45 1.8<br>14G 45 2.1 | 20G 25 1.1<br>18G 25 1.3<br>16G 30 1.7<br>14G 45 2.1 | Same | added to the<br>submission<br>device.<br>This was added<br>to expand the<br>size availability<br>for patient care.<br>This does not<br>affect the safety<br>and<br>effectiveness of<br>the device. |
| Sterilization | Terminal Ethylene Oxide<br>(EtO gas) | Terminal Ethylene Oxide<br>(EtO gas) | Same | |
| | Materials | | | |
| Catheter | Polyurethane (PUR)<br>/ (PTFE)<br>Polytetrafluoroethylene | Polyurethane (PUR)<br>/ (PTFE)<br>Polytetrafluoroethylene | Same | |
| Needle | Stainless Steel | Stainless Steel | Same | |
| Catheter Body | K-Resin | K-Resin | Same | |
| Catheter<br>Holder | Polyacetal (POM) | Polyacetal (POM) | Same | |
| Needle Hub | K-Resin | K-Resin | Same | |
| Activation<br>Lever | Absent | Polyacetal (POM) | Different | Absent in the<br>submission<br>device.<br>A warning was<br>added to the<br>labeling of this<br>product that a<br>Needle Stick<br>Prevention<br>Feature is not<br>present in this<br>device.<br>This does not<br>affect the safety<br>and<br>effectiveness of |
| | | | | the device. |
| Flashback<br>Chamber | PUR/PTFE | PUR/PTFE | Same | |
| Hydrophobic<br>Filter | Stainless Steel | Stainless Steel | Same | |
| Materials | K-Resin | K-Resin | Same | |
| Sterilization<br>method | EtO | EtO | Same | |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
## H. Performance Data:
## I. Performance Testing:
- ISO 10555-5: Intravascular catheters Sterile and single-use catheters Part 5: ● Over-needle peripheral catheters
- ISO 10555-1 Intravascular catheters Sterile and single-use catheters Part 1: ● General requirements
- ISO 594-1Conical fittings with a 6% (Luer) taper for syringes, needles and certain ● other medical equipment. - Part 1: General requirements
- . ISO 594-2 Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings
## II. Biocompatibility
There were no changes to the material that impacted the biocompatibility of the device. A signed statement of conformance was included in place of testing as the testing from the predicate device was referenced.
- . USP <788> Particulate Matter in Injections
## III. Sterility
- ISO 11135:2014 Sterilization of health-care products Ethylene oxide -. Requirements for the development, validation and routine control of a sterilization process for medical devices
### I. Conclusion:
In summary, based on the indication for use, technological characteristics and performance testing, the differences between the MedSource CathMED IV Catheter and the predicate did not raise concern for safety and effectiveness. Therefore, we conclude that the subject device is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.