K131555 · Medsource International, LLC · FOZ · Jan 31, 2014 · General Hospital
Device Facts
Record ID
K131555
Device Name
MEDSOURCE IV SAFETY CATHETER
Applicant
Medsource International, LLC
Product Code
FOZ · General Hospital
Decision Date
Jan 31, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The MedSource IV Safety Catheter is indicated to sample blood or administer fluids intravenously. The MedSource IV Safety Catheter may be used for any patient population with consideration given to adequacy for the vascular anatomy appropriateness for the solution being administered and duration of the therapy.
Device Story
MedSource IV Safety Catheter is a single-use, sterile intravascular device for fluid delivery or drainage. It features a sharp needle attached to a hub, housed within a protective chamber. During insertion, the needle is locked; post-insertion, a user-activated lever or switch triggers a spring-loaded retraction mechanism, withdrawing the needle into the enclosed chamber to prevent accidental needle-stick injuries and exposure to patient body fluids. Used in clinical settings by healthcare professionals. The device is available in six gauges (14G-24G). By securing the contaminated needle immediately after use, it reduces the risk of bloodborne pathogen exposure for healthcare personnel.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included chemical analysis (volatile residue, heavy metals, buffering capacity), sterility (ETO), and physical testing (integrity per ISO 10555-1/5, flow-rate, and tensile strength). Results for the subject device were comparable to the predicate device.
Technological Characteristics
Intravascular catheter with spring-loaded needle retraction mechanism. Materials of construction are similar to predicate. Dimensions vary by gauge (14G-24G). Sterilization via ETO. Complies with ISO 10555-1, ISO 10555-5, ISO 10993-10, and ASTM standards for packaging and tensile testing.
Indications for Use
Indicated for blood sampling or intravenous fluid administration in any patient population, contingent upon vascular anatomy, solution appropriateness, and therapy duration.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
## KISISSE
JAN 3 1 2014
# 510(K) Summary
Section 5
Medsource International, LLC Submitter: 5346 Shoreline Drive Mound, MN 55364
Contact Person: Jennifer Ness, Quality and Regulatory Affairs Manager 5346 Shoreline Drive Mound, MN 55364 Phone: 952-241-8318
Date Prepared: Wednesday, November 06, 2013
General Information:
Common Name: Catheter, intravascular, therapeutic, short-term Regulatory Reference: 21 CFR §880.5200 Product Code: FOZ Classification: Class II Panel: General Hospital Proprietary Name: MedSource IV Safety Catheter Single Use: Yes Sterile: Yes
Indication for Use:
The MedSource IV Safety Catheter is indicated to sample blood or administer fluids intravenously. The MedSource IV Safety Catheter may be used for any patient population with consideration given to adequacy for the vascular anatomy appropriateness for the solution being administered and duration of the therapy.
#### Description of the Device:
The MedSource IV Safety Catheter is a safety medical device used for inserting a catheter into a patient's body for purpose of delivery of fluids or drainage of fluids from the patient's body. This device is engineered to protect healthcare workers using it against accidental needle stick injury and exposure to parts of the device that have come in contact with patient's blood or other body fluids, and have hence been potentially contaminated with infectious agents. In addition this device also secures the needle (used to insert the catheter) in an enclosed chamber, thus protecting healthcare and other personnel from exposure to patient body fluids after the device has been discarded postuse.
The device comprises a sharp needle attached to a hub which projects into a chamber. In the 'unactivated' state prior to insertion, the needle and hub are 'locked' in position. There exists a retraction mechanism for 'unlocking' the needle such that on activation of this mechanism it is withdrawn and completely enclosed within the chamber. This
{1}------------------------------------------------
mechanism for retracting the needle preferably includes a coiled spring or other elastic components. The retraction mechanism is activated by the user using only one hand by means of a lever or 'switch'.
| Model | Description | Gage | Color |
|----------|-------------------------------------------|------|--------|
| MS-84014 | IV SAFETY CATHETER: 14 GA X 1.75"(50/BX) | 14 | Orange |
| MS-84016 | IV SAFETY CATHETER: 16 GA X 1.16"(50/BX) | 16 | Gray |
| MS-84018 | IV SAFETY CATHETER: 18 GA X 1.16"(50/BX) | 18 | Green |
| MS-84020 | IV SAFETY CATHETER: 20 GA X 1.16" (50/BX) | 20 | Pink |
| MS-84022 | IV SAFETY CATHETER: 22 GA X 1"(50/BX) | 22 | Blue |
| MS-84024 | IV SAFETY CATHETER: 24 GA X .75" (50/BX) | 24 | Yellow |
The device is available in six gauges identified also by specific colors.
### Description of the Device Design:
The MedSource IV Safety Catheter is substantially equivalent to the Becton Dickinson Angiocath Autogaurd (K984059)
| Comparison Point | Predicate Devices | Result of Comparison |
|------------------------------------------------------------------------------------------------------|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| Intended Use | Becton Dickinson Angiocath<br>Autogaud (K984059) | Substantially equivalent |
| Technological<br>Characteristics<br>(Materials of<br>Construction,<br>Dimensions,<br>Performance in) | Becton Dickinson Angiocath<br>Autogaud (K984059) | Materials of construction-same<br>or similar<br>Performance-minor variations<br>Dimensions-minor variations |
| Instructions for Use | Becton Dickinson Angiocath<br>Autogaud (K984059) | Very similar |
Summary of Performance Testing Characteristics:
| | Submission Device ↓ | Predicate Device ↓ |
|---------------------|------------------------------|---------------------------------------------------|
| Comparison Point | MedSource IV Safety Catheter | Becton Dickinson Angiocath<br>Autogaurd (K984059) |
| Chemical Tests | | |
| Volatile Residue | <15.0mg | <15.0mg |
| Heavy Metals | <1ppm | <1ppm |
| Residue on Ignition | <5.0mg | <5.0mg |
| Buffering Capacity | Complies to USP-34 | Complies to USP-34 |
| Biological Tests | | |
{2}------------------------------------------------
| Sterility | Sterile ETO | Sterile ETO |
|-----------|----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| | Physical Tests | |
| Integrity | Complies to ISO 10555-1 and<br>ISO10555-5 | Complies to ISO 10555-1 and<br>ISO10555-5 |
| | | |
| Flow-rate | 14G: 300 ml/min<br>16G: 220 ml/min<br>18G: 105 ml/min<br>20G: 65ml/min<br>22G: 35ml/min<br>24G: 20ml/min | 14G: 300 ml/min<br>16G: 220 ml/min<br>18G: 105 ml/min<br>20G: 65ml/min<br>22G: 35ml/min<br>24G: 20ml/min |
| | | |
| | | |
| | | |
| | | |
| | | |
| Tensile | 18G: 10N<br>20G: 5N<br>22G: 5N<br>24G: 3N | 18G: 10N<br>20G: 5N<br>22G: 5N<br>24G: 3N |
| | | |
| | | |
| | | |
#### Conclusion:
As shown by data in the table above, there are no significant performance specification differences between the MedSource IV Safety Catheter and the substantially equivalent device. Therefore, we conclude that the performance specifications demonstrate that the Subject Device is as safe, as effective, and performs as well as or better than that of the legally marketed predicate device.
#### Reference Documents:
- 1. ISO 10555-5: Sterile, single use intravascular catheters
- 2. ISO 11607-1/2 Packaging for terminally sterilized medical devices
- 3. BS EN 868-5 Packaging materials and systems for medical devices which are to be sterilized. Heat and self-sealable pouches and reels of paper and plastic film construction-Requirements and test methods
- 4. DIN 58953-6 "Sterilization paper for bags and tubing pickings ".Testing for germ proof ness in moisture with passage of air
- 5. ASTM F 2054 Standard test method for Burst Testing of Flexible Package Seals Using Internal Air Pressurization within Restraining Plates.
- 6. ASTM F 1980-02 Standard Guide for Accelerated Aging of Sterile Medical Device Packages
- 7. ASTM D638 Plastic Tensile Testing
- 8. ISO 10993: Part 10 Biological evaluation of medical devices: Test of irritation and skin sensitization.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Image /page/3/Picture/3 description: The image is a low-resolution picture with a few dark spots and lines scattered across a white background. The dark spots and lines are concentrated in the center and right side of the image. The image is blurry and lacks detail.
#### January 31, 2014
MedSource International, LLC Ms. Jennifer Ness Manager, Quality Control and Regulatory Affairs 5346 Shoreline Drive Mound, MN 55318
Re: K131555
Trade/Device Name: MedSource IV Safety Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: May 23, 2013 Received: December 16, 2013
Dear Ms. Ness:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Ness
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
http://www.fda.gov/MedicalDevices/KResourcecenter/YourIndustry/default.htm.
Kwame O.
Ulmer-S
for
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation , Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use Statement SECTION 4
510(k) Number: K131555
Device Common Name: Catheter, intravascular, therapeutic, short-term Device Proprietary Name: MedSource IV Safety Catheter
Indications for Use: The MedSource IV Safety Catheter is indicated to sample blood, or administer fluids intravenously. The MedSource IV Safety Catheter may be used for any patient population with consideration given to adequacy for the vascular anatomy appropriateness for the solution being administered and duration of the therapy.
Prescription Usa (Part 21 CFR 601 Subpert D)
AND/OR
Over-The Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/5/Picture/10 description: The image shows a stylized logo of the letters FDA. The letters are outlined with a thick black line, and the letters are connected to each other. The letters are in a sans-serif font, and the letters are all uppercase. The logo is simple and modern.
Digitally signed by Richard C. Chapman Date: 2014.01.30 13:53:08 -05'00'
FDA 510k Submission: MedSource IV Safely Catheter
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.