K193266 · Medico USA, Inc. · GEX · Jan 22, 2021 · General, Plastic Surgery
Device Facts
Record ID
K193266
Device Name
MIIN
Applicant
Medico USA, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jan 22, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MIIN, Q-Switched Nd: YAG Laser Therapy System (1064nm or 532nm) is indicated for use in incision, excision, ablation, vaporization of soft tissue for general dermatologic and general surgical procedures for coagulation and hemostasis as follows: 1064 nm wavelength - Tattoo Removal (dark ink: blue and black) - Dermal Pigmented Lesions; including, but not limited to: Nevus of Ota, Lentigines, Nevi, Melasma and Cafe-au-lait - Removal or lightening of hair with or without adjuvant preparation. - Skin Resurfacing for Acne Scars and Wrinkles - Benign cutaneous lesions; including, but not limited to: striac and Scars (excludes the 650mm wavelength) - Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar (excludes the 650nm wavelength) 532 nm Wavelength - Tattoo removal (light ink: red, sky blue, green) - Vascular lesions including but not limited to: port wine birthmarks, telangiectasias, spider angioma, spider nevi - Epidermal Pigmented lesions; including, but not limited to: cafe-au-lait birthmarks, solar lentiginos, Becker's nevi, Freckles, Nevus spilus, seborrheic keratosis - Skin Resurfacing for Acne Scars and Wrinkles - Benign cutaneous lesions; including, but not limited to: striae and scars, (excludes the 6S0nm wavelength) - Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar (excludes the 650nm wavelength)
Device Story
MIIN is a Q-Switched Nd:YAG laser system (1064nm/532nm) used in clinical settings by physicians for dermatologic and surgical procedures. The device consists of a main body (laser module, power supply, control module, touch screen), articulated arm, and handpiece. It utilizes an optical cavity with an Nd:YAG crystal and a xenon lamp energy source. Operation relies on selective photothermal therapy and a blasting mechanism; laser energy targets specific chromophores (ink, carbon, melanin), causing pigment particles to fragment for macrophage phagocytosis and lymphatic clearance. Physicians select wavelength, energy, spot size, and fluency via the control panel. The device facilitates coagulation and hemostasis, benefiting patients through precise tissue ablation and pigment removal.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified via biocompatibility (ISO 10993), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), laser safety (IEC 60825-1), usability (EN 62366), and risk management (ISO 14971).
Technological Characteristics
Nd:YAG laser medium; Xenon lamp energy source; 1064nm/532nm wavelengths; Q-switched operation; articulated arm beam delivery; internal distilled water cooling; adjustable spot size (1-10mm); AC 110-230V power. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60825-1, and ISO 10993.
Indications for Use
Indicated for patients requiring soft tissue incision, excision, ablation, or vaporization for general dermatologic and surgical procedures, including tattoo removal, pigmented/vascular lesion treatment, hair removal, skin resurfacing, and scar reduction.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
RevLite Q-Switched Nd: YAG Laser Systems (K133254)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 22, 2021
Medico USA Inc. % Joyce Kwon CEO Provision Consulting Group Inc. 100 Barranca St. Suite 700 West Covina, California 91791
# Re: K193266
Trade/Device Name: Miin Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: November 22, 2019 Received: November 26, 2019
Dear Joyce Kwon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193266
Device Name MIIN
Indications for Use (Describe)
MIN, Q-Switched Nd: YAG Laser Therapy System (1064nm or 532nm) is indicated for use in incision, excision, ablation, vaporization of soft tissue for general dermatologic and general surgical procedures for coagulation and hemostasis as follows:
1064 nm wavelength
- · Tattoo Removal (dark ink: blue and black)
- · Dermal Pigmented Lesions; including, but not limited to: Nevus of Ota, Lentigines, Nevi, Melasma and Cafe-au-lait
- · Removal or lightening of hair with or without adjuvant preparation.
- · Skin Resurfacing for Acne Scars and Wrinkles
- · Benign cutaneous lesions; including, but not limited to: striac and Scars (excludes the 650mm wavelength)
· Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar (excludes the 650nm wavelength)
532 nm Wavelength
- · Tattoo removal (light ink: red, sky blue, green)
· Vascular lesions including but not limited to: port wine birthmarks, telangiectasias, spider angioma, spider nevi
· Epidermal Pigmented lesions; including, but not limited to: cafe-au-lait birthmarks, solar lentiginos, Becker's nevi, Freckles, Nevus spilus, seborrheic keratosis
· Skin Resurfacing for Acne Scars and Wrinkles
· Benign cutaneous lesions; including, but not limited to: striae and scars, (excludes the 6S0nm wavelength)
· Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar (excludes the 650nm wavelength)
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K193266 510(K) Summary
# Submitter
Peter Son MEDICO, Inc. 3407 W 6th st. Los Angeles, CA 90020 Tel : 213-703-1223
# Official Correspondent/ US Agent
Joyce Kwon Provision Consulting Group, Inc. 100 N Barranca St. Suite 700. West Covina, CA 91791 Phone: +1-909-493-3276
# Device Information
- · Device Name: MIIN
- · Classification Name: Powered Laser Surgical Instrument
- · Common Name: Q-Switched Nd:YAG Laser System
- · Classification: Class II
- Product Code: GEX
- · Regulation number: 21 CFR 878.4810
- Date Prepared: 01/16/2021
#### Predicate devices
- · Q-switched Nd:YAG Laser Therapy Systems (K133254)
#### Indication for use
MIIN, Q-Switched Nd: Y AG Laser Therapy System (1064nm or 532nm) is indicated for use in incision, excision, ablation, vaporization of soft tissue for general dermatologic and general surgical procedures for coagulation and hemostasis as follows:
1064 nm wavelength
- . Tattoo Removal (dark ink: blue and black)
- Dermal Pigmented Lesions; including, but not limited to: Nevus of Ota, Lentigines, Nevi, ● Melasma and Cafe-au-lait
- Removal or lightening of hair with or without adjuvant preparation. ●
- Skin Resurfacing for Acne Scars and Wrinkles ●
- Benign cutaneous lesions; including, but not limited to: striac and Scars (excludes the 650nm wavelength)
- Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral . part of the scar (excludes the 650nm wavelength)
{4}------------------------------------------------
532 nm Wavelength
- Tattoo removal (light ink: red, sky blue, green) ●
- Vascular lesions including but not limited to: port wine birthmarks, telangiectasias, spider . angioma, cherry angioma, spider nevi
- . Epidermal Pigmented lesions; including, but not limited to: cafe-au-lait birthmarks, solar lentiginos, senile lentiginos, Becker's nevi, Freckles, Nevus spilus, seborrheic keratosis
- Skin Resurfacing for Acne Scars and Wrinkles
- . Benign cutaneous lesions; including, but not limited to: striae and scars, (excludes the 6S0nm wavelength)
- . Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar (excludes the 650nm wavelength)
#### Device Description
MIN applies the Nd:YAG laser source and it is composed of main body part, hand piece part and arm part mainly. The main body part is consisted of laser module from which beam is irradiated, power module, control module and touch screen.
MIIN, The Q-Switched Nd:YAG Laser Systems is based on the Q-Switched Nd: YAG (1064nm) and frequency Nd: YAG (532nm) Laser Technology. There is one optical cavity containing the Nd: YAG Crystal
The MIIN Q-Switched Nd: YAG Laser Therapy Systems works based on laser selective photothermal therapy and blasting mechanism of Q-switched laser. Energy form particular wavelength with accurate dose will act on certain targeted color radicals: ink, carbon particles from derma and epidermis, exogenous pigment particles and endogenous melanophore from derma and epidermis. When suddenly being heated. pigment particles immediately blast into smaller pieces, which will be swallowed by macrophage phagocytosis and enters into lymph circulation system and finally be discharged out of body.
The physician is able to select the desired wavelength and the related output energy, spot size and fluency via control panel.
| Specification | Predicate device | Proposed Device | Discussion of<br>Differences |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|------------------------------|
| K Number | K133254 | K193266 | |
| Product Code | GEX | | Identical |
| Manufacturer | Cynosure (HOYA ConBio) | LTRA Global Co., Ltd. | |
| Device Name | RevLite Q-Switched Nd: YAG<br>Laser Systems | MIIN Q-Switched Nd: YAG<br>Laser Systems | Similar |
| Intended use | Incision, Excision, Ablation, Vaporization of Soft Tissue for<br>General Dermatology, Dermatologic and General Surgical<br>Procedures for Coagulation and Hemostasis. | | Identical |
| Indications for use | 1064nm wavelength<br>• Tattoo Removal (dark ink: blue and black) | | Identical |
#### Substantial Equivalence Discussion
{5}------------------------------------------------
| | Dermal Pigmented Lesions; including, but not limited to: Nevus of Ota, Lentigines, Nevi, Melasma and Cafe-au-lait Removal or lightening of hair with or without adjuvant preparation. Skin Resurfacing for Acne Scars and Wrinkles Benign cutaneous lesions; including, but not limited to: striae and Scars (excludes the 650nm wavelength) Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar (excludes the 650nm wavelength) 532nm Wavelength Tattoo removal (light ink: red, sky blue, green) Vascular lesions including but not limited to: port wine birthmarks, telangiectasias, spider angioma, cherry angioma, spider nevi Epidermal Pigmented lesions; including, but not limited to: cafe-au-lait birthmarks, solar lentigines, senile lentigines, Becker's nevi, Freckles, Nevus spilus, seborrheic keratosis Skin Resurfacing for Acne Scars and Wrinkles Benign cutaneous lesions; including, but not limited to: striae and scars, (excludes the 650nm wavelength) Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar (excludes the 650nm wavelength) | | |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|-----------|
| Laser Medium | Nd: YAG | Nd: YAG | Identical |
| Operating<br>Parameters | Q-Switched | Q-Switched | Identical |
| Energy Source | Xenon Lamp | Xenon Lamp | Identical |
| Wavelength (nm) | 1064nm and 532nm | | Identical |
| Specification | Predicate device | Proposed device | Discussion of<br>Differences |
|-----------------------------------|--------------------------------------------------------------|--------------------------------------------------------------|------------------------------------------------------------------------------------|
| Maximum Pulse<br>Energy | @1064nm wavelength:<br>1500mJ<br>@532nm wavelength:<br>450mJ | @1064nm wavelength:<br>3500mJ<br>@532nm wavelength:<br>500mJ | Similar |
| Pulse Duration | 7-20ns | 25ns | Similar |
| Repetition Rate | 1-10Hz | 0-10Hz | Identical |
| Nominal ocular<br>hazard distance | NOHD 3.3km | NOHD 93m | Less than<br>Predicate device,<br>safer |
| Spot Size | Adjustable Spot Size<br>2-8.5mm (Diameter) | Adjustable Spot Size<br>1-10mm (Diameter) | Wider adjustable<br>range of spot size<br>than predicate<br>device, more<br>useful |
{6}------------------------------------------------
| Beam delivery | Articulated Arm with<br>Handpiece | Articulated Arm with<br>Handpiece | Identical |
|----------------------------|-------------------------------------------------|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cooling | Internal distilled water<br>circulating cooling | Internal distilled water<br>circulating cooling | Identical |
| Anatomical Sites | Skin and subcutaneous<br>tissue | Skin and subcutaneous<br>tissue | Identical |
| Electrical<br>Requirements | AC 230V, 50/60 Hz | AC 110-230V, 50/60Hz | Identical |
| Sterilization | Sterile | Non-Sterile | K152856 and<br>K083203 share the<br>same indication<br>for use and<br>technology with<br>the subject<br>device(K193266),<br>the devices have<br>not done<br>sterilization for<br>their handpiece<br>tips. |
# Similarities
The subject device has same device characteristics with the Primary predicate devices (K133254) such as intended use, Indications for use, general shape (Design), structure, fundamental technologies, Operating Parameters, Laser medium, Wavelength, Pulse duration, repetition rate, spot size and Cooling method are similar. Technological characteristics are similar as predicate device. Testing including Biocompatibility tests (ISO 10993), Performance Testing (IEC 60825) and EMC Testing (IEC 60601-1, IEC 60601-1-2) has been finished to ensure the devices to comply with FDA Guidance on 'The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications'.
# Differences
The difference between the subject device and the primary predicate device is maximum Pulse Energy. The max energy of O-switch of MIIN (K193266) is 1400mJ(1064 S/25ns) and the long pulse energy is 3500mJ. The reference device K083889 has 1600mJ(5ns) (Q-switched) and 5000mJ (long pulsed). MIIN has lower pulse energy then its reference K083889, and does not raise any questions in safety and effectiveness in equivalence. For sterilization, K152856 and K083203 share the same indication for use and technology with the subject device(K193266), the devices have not done sterilization for their handpiece tips. It means that in all aspects, the subject device is equivalent in safety and effectiveness with predicate device. Therefore, this Maximum pulse energy difference doesn't affect device safety and effectiveness.
#### Conclusion
The MIIN Q-Switched Nd: YAG laser system constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This system has the same intended use and fundamental technology as its predicate devices. Therefore, MIIN Q-Switched Nd: Y AG laser system and the predicate device are substantially equivalent.
# Performance Data
{7}------------------------------------------------
# Biocompatibility testing
The biocompatibility evaluation for the Q-Switched Nd: YAG Laser Therapy Systems was conducted in accordance with the guidance "Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" as recognized by FDA. The biocompatible testing included the following tests:
- Cytotoxicity
- Sensitization ●
- . Irritation
The handpiece is considered skin and subcutaneous tissue contacting device.
# Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the O-Switched Nd: YAG Laser Therapy Systems. The device complies with the IEC 60601-1, standard for safety and the IEC 60601-1-2 standard for EMC.
#### Performance testing
Performance testing was conducted on the device according to IEC 60825-1.
#### Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw could indirectly result in minor injury to the patient or operator through incorrect or delayed information or through the action of a care provider.
The subject device was tested to evaluate its performance as below.
- MIIN Q-switched Nd: YAG Laser System is tested and evaluated according to EN60601-. 1:2006+A1:2013 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- Effect to the device by electromagnetic disturbances were tested and evaluated according to the ● EN60601-1-2:2015 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance. All the results presented here demonstrated the requirements and tests for electromagnetic disturbances.
- . MIIN Q-switched Nd: YAG Laser System is tested and evaluated according to EN60601-1-6:2010, AMD1:2015 Medical electrical equipment Part 1-6 General requirements for safety -Collateral Standard: Usability.
- MIIN O-switched Nd: YAG Laser System is tested and evaluated according to EN 60601-2-22:2013 "Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment"
- Safety of laser products is evaluated according to IEC 60825-1: 2014. All the results presented ● here demonstrated the equipment classification and requirements.
{8}------------------------------------------------
- Risk management was recorded according standard ISO 14971: 2012. All the results presented ● here demonstrated the application of risk management to medical devices.
- Usability was documented according to standard EN 62366: 2008. All the results presented here ● demonstrated the application of usability engineering to medical devices.
- Biocompatibility was tested and evaluated according to FDA-recognized consensus standard ISO . 10993-5: 2009 and ISO 10993-10: 2010.
### Conclusion:
The non-clinical data support the safety of the device and the performance testing report demonstrate that the MIN, the Q-Switched Nd: YAG Laser Therapy Systems should perform as intended in the specified use conditions. Medico considers the MIIN, the Q-Switched Nd: Y AG Laser Therapy Systems to be substantially equivalent to the predicate device and does not raise any new issues of safety or effective
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.