K152856 · Laseroptek Co., Ltd. · GEX · Jun 2, 2016 · General, Plastic Surgery
Device Facts
Record ID
K152856
Device Name
Helios III
Applicant
Laseroptek Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 2, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Helios III Q-Switched Nd:YAG Laser System delivers pulse wave laser light in the contact or noncontact mode for; Incision, excision, ablation, vaporization of soft tissue for general dermatology (1064 nm) Removal or lightening of unwanted hair with or without adjuvant preparation (1064 nm) Tattoo Removal (1064nm,532nm) · dark ink: blue and black (1064nm) · light ink: red, sky blue, green (532nm) Treatment of Vascular Lesions: (532nm) • port wine birthmarks · telangiectasias · spider angioma • cherry angioma · spider nevi Treatment of Pigmented Lesions: (1064nm,532nm) · café-au-lait birthmarks (532nm) · solar lentiginos (532nm) · senile lentiginos (532nm) · becker's nevi (532nm) · freckles (532nm) • nevus spilus (532nm ) • nevus of ota (1064nm)
Device Story
Helios III is a Q-Switched Nd:YAG laser system for dermatological procedures. Device uses Nd:YAG crystal as gain medium, Xe gas lamp as pumping source, and resonator for beam amplification. System delivers 1064 nm and 532 nm (frequency-doubled KTP) laser light. User operates device via GUI to set output and repetition rate; microprocessor controls power supply. Used in clinical settings by physicians. Output is laser beam for tissue interaction; clinical benefit includes removal of tattoos, pigmented/vascular lesions, and soft tissue surgery. Device is a floor-standing unit.
Clinical Evidence
Bench testing only. Included visual and mechanical inspection, electrical and mechanical safety testing, and functional performance testing (e.g., earthbond, software, transit, leakage current).
Technological Characteristics
Q-Switched Nd:YAG laser; 1064 nm and 532 nm wavelengths. Gain medium: Nd:YAG crystal. Pumping source: Xe gas lamp. Repetition rate: 1-10 Hz. Max pulse duration: 10 ns. Fluence: 1-8 J/cm2. Microprocessor-controlled via GUI. Electrical: 230 V, 50/60 Hz. Floor-standing unit.
Indications for Use
Indicated for soft tissue incision, excision, ablation, and vaporization in dermatology; hair removal/lightening; removal of dark (1064nm) and light (532nm) tattoos; treatment of vascular lesions (port wine, telangiectasias, spider/cherry angioma, spider nevi); and treatment of pigmented lesions (café-au-lait, solar/senile lentiginos, becker's nevi, freckles, nevus spilus, nevus of ota).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Cynosure, Inc.'s Revlite Q-Switched Nd:YAG Laser System (K133254)
Laseroptek Co. Ltd.'s Helios II Q-Switched Nd:YAG Laser System (K083203)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is a symbol of medicine and healing. The logo also includes the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" in a circular arrangement around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 2, 2016
Laseroptek Co. Ltd. Mr. Kevin J. Choi 28 Stafford Drive West Windsor, NJ 08550
Re: K152856 Trade/Device Name: Helios III Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: April 5, 2016 Received: April 12, 2016
Dear Mr. Choi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
#### 510(k) Number (if known) K152856
#### Device Name
Helios III O-Switched Nd:YAG Laser System
#### Indications for Use (Describe)
Incision, excision, ablation, vaporization of soft tissue for general dermatology (1064 nm) Removal or lightening of unwanted hair with or without adjuvant preparation (1064nm) Tattoo Removal: Dark ink (blue and black) (1064 nm) Tattoo Removal: light ink (red, sky blue, green) (532 nm) port wine birthmarks (532 nm) telangiectasias (532 nm) spider angioma (532 nm) cherry angioma (532 nm) spider nevi (532 nm) cafe-au-lait birthmarks (532 nm) solar lentiginos, senile lentiginos, becker's nevi, freckles, nevus spilus (532 mm) nevus of ota (1064 nm)
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
■ Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# This 510(k) Summary of safety and effectiveness is submitted in accordance with the requirements of 21 CFR §807.92
| Submitter's Information | |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Laseroptek Co. Ltd. |
| Address: | 204 Hyundai I Valley<br>223-12 Sangdaiwon, Jungwon<br>Sungnam-Si, Gyeonggido, 462-714<br>Rep. of Korea (South Korea) |
| Contact Person: | Hong Chu, Ph.D. (President & CEO) |
| Telephone / Fax / E-mail: | Tel) +82.31.737.9885<br>E-mail) hchu@laseroptek.com |
| Submitter: | Kevin Choi |
| Address: | 28 Stafford Drive, West Windsor, NJ 08850 |
| Contact Person: | Kevin J. Choi |
| Telephone / Fax / E-Mail: | Tel) 310-634-4480<br>E-mail) kjchoi1128@hotmail.com |
| Preparation Date: | 07/17/2015 |
| Device Trade Name: | Helios III |
| Common Name: | Q-Switched Nd:YAG Laser |
| Classification Name: | Instruments, Surgical, Powered, Laser 79-GEX<br>(21 CFR §878-4810) |
| Legally Marketed Predicate Devices: | • Cynosure, Inc.'s Revlite Q-Switched Nd:YAG<br>Laser System (K133254)<br>• Laseroptek Co. Ltd.'s Helios II Q-Switched |
Nd:YAG Laser System (K083203)
{4}------------------------------------------------
Intended Use of Helios III: The Helios III Q-Switched Nd:YAG Laser System delivers pulse wave laser light in the contact or noncontact mode for; Incision, excision, ablation, vaporization of soft tissue for general dermatology (1064 nm) Removal or lightening of unwanted hair with or without adjuvant preparation (1064 nm) Tattoo Removal (1064nm,532nm) · dark ink: blue and black (1064nm) · light ink: red, sky blue, green (532nm) Treatment of Vascular Lesions: (532nm) • port wine birthmarks · telangiectasias · spider angioma • cherry angioma · spider nevi Treatment of Pigmented Lesions: (1064nm,532nm) · café-au-lait birthmarks (532nm) · solar lentiginos (532nm) · senile lentiginos (532nm) · becker's nevi (532nm) · freckles (532nm) • nevus spilus (532nm ) • nevus of ota (1064nm) Device Description: The Helios III laser system is based on the Nd:YAG (1064 nm) and frequency doubled KTP Nd:YAG (532 nm) laser technology. Three basic elements of operations are as follows: 1) A Nd:YAG crystal is used as a gain medium which produces a laser beam. 2) A resonator then amplifies the beam. 3) A lamp that contains Xe gas is used, as a pumping
{5}------------------------------------------------
| | light source. The lamp requires a high-pressure power source device for operation. When the electric energy generated from the high-pressure power source is induced into the electrode of the lamp, it converts into light energy. This converted light energy pumps the Nd–YAG crystal – a gain medium – and the light exhausted from the crystal is amplified into a specific wavelength light. As it passes between the resonant gases, laser beam radiates to an output unit. |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The regulation of laser output and repetition rate can be set by the user via GUI (Graphic User Interface) and controlled by microprocessor, which interfaces with the power supply. |
| Performance Data: | Non-clinical testing of Helios III included visual and mechanical inspection, electrical and mechanical safety testing, functional performance testing, etc., in bench. Some of these tests include visual inspection, earthbond testing, software testing, transit testing, leakage current testing and measurements. Test reports have been submitted. |
| Substantial Equivalence | Helios III's intended use is identical to the intended use of the predicate devices: Cynosure, Inc.'s Revlite Q-Switched Nd:YAG Laser System (K133254) and Laseroptek Co. Ltd.'s Helios II Q-Switched Nd:YAG Laser System (K083203) |
| | Comparison of indications for use and performance specifications are provided in the following tables. |
{6}------------------------------------------------
## 510(k) Summary Helios III Q-Switched Nd:YAG Laser
| | Predicate Device | Predicate Device | Proposed Device |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) Number | K133254 | K083203 | K152856 |
| Manufacturer | Cynosure (HOYA ConBio) | Laseroptek Co. Ltd. | Laseroptek Co. Ltd. |
| Device Name | RevLite Q-Switched Nd:YAG Laser System | Helios II Q-Switched Nd:YAG Laser System | Helios III Q-Switched Nd:YAG Laser System |
| Intended Use /<br>Indications for Us: | Tattoo Removal (dark ink: blue and black) | Tattoo Removal (dark ink: blue and black) | Tattoo Removal (dark ink: blue and black) |
| | Dermal pigmented lesions, including, but<br>not limited to: Nevus of Ota, Lentigines,<br>Nevi, Melasma and Café-au-lait | Dermal pigmented legions: Nevus of Ota,<br>Café-au-lait birthmarks, solar lentiginos,<br>senile lentiginos, becker's nevi, freckles,<br>and nevus spilus | Dermal pigmented legions: Nevus of Ota,<br>Café-au-lait birthmarks, solar lentiginos,<br>senile lentiginos, becker's nevi, freckles,<br>and nevus spilus |
| | Removal or lightening of hair with or<br>without adjuvant preparation | Removal or lightening of hair with or<br>without adjuvant preparation | Removal or lightening of hair with or<br>without adjuvant preparation |
| | Skin resurfacing for acne scars and wrinkles | Incision, excision, ablation, vaporization of<br>soft tissue for general dermatology | Incision, excision, ablation, vaporization of<br>soft tissue for general dermatology |
| | Benign cutaneous lesions: including, but<br>not limited to; striae and scars (excludes<br>the 650 nm wavelength) | Tattoo Removal (light ink: red, sky blue,<br>green) | Tattoo Removal (light ink: red, sky blue,<br>green) |
| | Reduction of red pigmentation in<br>hypertrophic and deloid scars where<br>vascularity is an integral part of the scar<br>(excludes the 650 nm wavelength) | Treatment of vascular lesions: port wine,<br>birthmarks, telangiectasias, spider<br>angioma, cherry angioma, and spider nevi | Treatment of vascular lesions: port wine,<br>birthmarks, telangiectasias, spider<br>angioma, cherry angioma, and spider nevi |
| | Tattoo Removal (light ink: red, sky blue,<br>green) | | |
| | Vascular lesions including but not limited<br>to: port wine, birthmarks, telangiectasias,<br>spider angioma, cherry angioma, spider<br>nevi | | |
| | Epidermal pigmented lesions; including,<br>but not limited to: café-au-lait birthmarks,<br>solar lentiginos, senile lentiginos, Becker's<br>nevi, Freckles, Nevus spilus, seborrheic<br>keratosis | | |
| | Skin resurfacing for acne scars and wrinkles | | |
| | Benign cutaneous lesions; including, but<br>not limited to: striae and scars (excludes<br>the 650 nm wavelength) | | |
| | Reduction of red pigmentation in<br>hypertrophic and deloid scars where<br>vascularity is an integral part of the scar<br>(excludes the 650 nm wavelength) | | |
Table 1: Approved Indication of Uses for Helios III Predicate Devices
{7}------------------------------------------------
| | Predicate Device | Predicate Device | Proposed Device |
|-----------------------------|-----------------------------------------|------------------------------------------|-----------------------------------------------------------------|
| 510(K) Number | K133254 | K083203 | K152856 |
| Manufacturer | Cynosure (HOYA ConBio) | Laseroptek Co. Ltd. | Laseroptek Co. Ltd. |
| Device Name | RevLite Q-Switched Nd:YAG Laser System | Helios II Q-Switched Nd:YAG Laser System | Helios III Q-Switched Nd:YAG Laser System |
| Clearance Date: | 3/5/2014 | 4/28/2009 | Not Available |
| Classification / Regulation | 21 CFR 878.4810 (GEX) | 21 CFR 878.4810 (GEX) | 21 CFR 878.4810 (GEX) |
| Laser Medium | Nd:YAG | Nd:YAG | Nd:YAG |
| Operating Parameters | Q-Switched | Q-Switched | Q-Switched |
| Wavelength | 1064 nm / 532 nm | 1064 nm / 532 nm | 1064 nm / 532 nm |
| Pulse Characteristics: | | | |
| Maximum Pulse Duration | 7 – 20 ns | 8 ns | 10 ns |
| Energy Delivered | 1.6 J | 1 J (0.5 J @ 532 nm) / pulse | 1.3 J (0.5J @ 532 nm) / pulse |
| Fluence | 1 – 8 J/cm2 @ 3 – 8 mm spot size | 1 – 8 J/cm2 @ 1 to 7 mm spot size | 1 - 8 J/cm2 @ 1 to 8 mm spot size |
| Spot Sizes | 2 – 8.5 mm range with 0.1 mm increments | 1 – 7 mm | (1064) 5 mm<br>(532) 4 mm<br>(Collimator) 8 mm<br>(Zoom) 1~7 mm |
| Repetition Rate | Single shot, 1 - 10 Hz | Single Shot, 1 - 10 Hz | Single Shot, 1 - 10 Hz |
| Average Power (Max): | 10 W (5 W @ 532 nm) | 10 W (5 W @ 532 nm) | 13 W (5W @ 532 nm) |
| Physical Characteristics: | | | |
| System Dimensions | 31.8"(H) X 12" (W) X 28.5" (D) | 37.2"(H) X 13"(W) X 37.2 (D) | 36.8"(H) X 11.7"(W) X 32.2" (D) |
| System Weight | 131 lbs. | 154 lbs. | 176 lbs. |
| Electrical Requirements | AC 230 V, 50/60 Hz | AC 230 V, 50/60 Hz | AC 230 V, 50/60 Hz |
| Maximum Power | 20W | 20 W | 20W |
Table 2: Performance Specification Comparison with Predicate Devices
Conclusion
The Helios III Q-Switched Nd:YAG Laser System is substantially equivalent to other existing systems in commercial distribution for treatment of the intended use indications share very similar / exactly same performance specification parameters as noted in above table.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.