The intended use of the Tyber Medical Anatomical Plating System is to draw two or more aligned bone fragments together to facilitate healing in an adult patient
Device Story
System comprises metallic bone plates and screws for orthopedic fixation; used in foot and ankle reconstruction, fracture stabilization, and osteotomies. Plates available in various anatomical shapes (e.g., T-shaped, Y-shaped, cloverleaf, straight); screws include cortical and locking variants. Operated by orthopedic surgeons in clinical/surgical settings. Device provides mechanical stabilization of bone fragments to facilitate healing. Ancillary instrumentation included. Offered sterile or non-sterile for single-use.
Clinical Evidence
No clinical data provided; nonclinical testing not provided as the device is identical to the predicate.
Technological Characteristics
Materials: Medical Grade CP Titanium (ASTM F67) and 6-4 Alloyed Titanium (ASTM F136). Surface treatment: Anodized Type II. Components: Locking and standard cortical screws (2.0mm–4.0mm), various plate geometries. Single-use, sterile or non-sterile. Mechanical fixation principle.
Indications for Use
Indicated for adult patients for fixation of small, medium, and large bone fragments in the foot (phalanges, metatarsals, cuneiform, cuboid, navicular, talus, calcaneus) and ankle (distal tibia/fibula) for stabilization of fractures, joint fusions, osteotomies, nonunions, malunions, reconstruction, revision surgeries, and replantations. Not for spinal use.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
January 20, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Tyber Medical LLC Jessica Stigliano Regulatory Affairs Specialist 83 South Commerce Way, Suite 310 Bethlehem, Pennsylvania 18017
Re: K193222
Trade/Device Name: Tyber Medical Anatomical Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: November 22, 2019 Received: November 22, 2019
Dear Jessica Stigliano:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K193222
Device Name
Tyber Medical Anatomical Plating System
#### Indications for Use (Describe)
The intended use of the Tyber Medical Anatomical Plating System is to draw two or more aligned bone fragments together to facilitate healing in an adult patient and is composed of the following bone plate categories:
#### I. Forefoot System:
The Tyber Medical Forefoot Plating System is Indicated for Use in fixation of small bone fragments in the foot (Phalanges and Metatarsals) for stabilization of fractures, joint fusions, osteotomies, malunions, reconstruction of small bones, revision surgeries and replantations in an adult patient. The Tyber Medical Forefoot System is not for Spinal Use.
II. Mid & Hindfoot System:
The Tyber Medical Mid & Hindfoot Plating System is Indicated for Use in fixation of medium/large bones and medium/ large bone multi-fragments in the foot (Cuneiform, Cuboid, Navicular, Talus and Calcaneus) for stabilization of fractures, joint fusions, osteotomies, nonunions, reconstruction of medium/large bones, revision surgeries and replantations in an adult patient.
The Tyber Medical Mid & Hindfoot System is not for Spinal Use.
III. Ankle Fracture System:
The Tyber Medical Ankle Fracture System is Indicated for Use in:
I). Fixation of fractures of the distal tibia included, but not limited to, ankle fractures, corrective osteotomies, non-unions, intra- and extra- articular and distal tibia fractures with a shaft extension, and malleolar fractures; 2). In intra- and extra articular fractures, osteotomies, medial malleolar fractures and non-unions of the metaphyseal and diaphyseal region of the distal fibula;
3). In distal tibia/fibula long bones which include the metaphyseal regions of the tibia and fibula in the ankle.
The Tyber Medical Ankle Fracture System is not for Spinal Use.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## Traditional 510(k) Summary
as required by section 807.92(c).
# Tyber Medical Anatomical Plating System
## K193222
| Submitted | 11/20/2019 |
|-----------------------------------|--------------------------------------------------------------------------------------------------|
| Submitter: | Tyber Medical LLC |
| | 83 South Commerce Way, Suite 310 |
| | Bethlehem, PA 18017 |
| Contact Person | Primary Contact: |
| | Jessica Stigliano |
| | Regulatory Affairs Specialist |
| | Phone: 866-761-0933 (ext. 424) Fax: (866) 889-9914 |
| | Email: jstigliano@tybermed.com |
| | Secondary Contact: |
| | Mark Schenk |
| | V.P. Regulatory |
| | Phone: 610-849-0645 |
| | Email: mschenk@tybermed.com |
| Trade Name | Tyber Medical Anatomical Plating System |
| Common Name | Plate, Fixation, Bone; Screw, Fixation, Bone |
| Device Class | Class II |
| Classification Name<br>and Number | Single/multiple component metallic bone fixation appliances and<br>accessories (21 CFR 888.3030) |
| | Smooth or threaded metallic bone fixation fastener (21 CFR 888.3040) |
| Classification Panel: | Orthopedic |
| Product Code | HRS/HWC |
| Predicate Devices | BioPro Foot Plating Systems – K162674 |
| Device Description | The Tyber Medical Anatomical Plating System consists of the following<br>categories: |
| | 1. Forefoot System |
| | 2. Mid & Hindfoot System |
| | 3. Ankle Fracture System |
{5}------------------------------------------------
A brief and concise description of each system is as follows:
### Forefoot System
The Tyber Medical Forefoot System is designed to address a variety of indications in forefoot reconstruction fixation/osteotomy surgery. The system is composed of many locking plate types which include: 1). Straight Fracture Plates, 2). T-Shaped Fracture Plates, 3). Y-Shape Fracture Plates, 4). L-Shaped Fracture Plates, 5). Cloverleaf Fracture Plates, 6). TMT 1 Medial Fusion Plates, 7). Open Wedge Fusion Plates, 8). MTP Fusion Plates, 9). MTP Fusion Revision Plates, and 10). Dorsal TMT 1 Step Fusion Plates, - in various plate lengths, Right & Left versions. The System will incorporate both Cortical Locking Screws and standard Cortical Screws in 2.0mm, 2.5mm, 3.0mm and 3.5mm sizes in various lengths. All plates are composed of Medical Grade CP Titanium material (to ASTM F67) with an Anodized Type II surface treatment. All screws are composed of Medical Grade 6-4 Alloyed Titanium material (to ASTM F136) color anodized for sizing.
A full set of Ancillary Instrumentation is available with the system. The Tyber Medical Forefoot System is offered both sterile and non-sterile for single-use.
### Mid and Hindfoot System
The Tyber Medical Mid & Hindfoot System is designed to address a variety of indications in midfoot & hindfoot reconstruction fixation/osteotomy surgery. The system is composed of many locking plate types which include:
1). Straight Fracture/Fusion Plates, 2). T-Shaped Fracture/Fusion Plates, 3). L-Shaped Fracture/Fusion Plates, 4). Cloverleaf Fracture/Fusion Plates, 5). X-Shaped Fracture/Fusion Plates, 6). Retangular Fracture/Fusion Plates, 7). Cotton Osteotomy Plates, 8). Dwyer Step Plates, 9). Evans Osteotomy Plates, 10.) Plantar TMT Plates, 11). Distal Medial & 3 - 4 Column Plates and 11). ORIF & Standard Calcaneal Plates - in various plate lengths, Right & Left versions.
The System will incorporate both Cortical Locking Screws and standard Cortical Screws in 3.0mm, 3.5mm and 4.0mm sizes in various lengths. All plates are composed of Medical Grade CP Titanium material (to ASTM F67) with an Anodized Type II surface treatment. All screws are composed of Medical Grade 6-4 Alloyed Titanium material (to ASTM F136) color anodized for sizing.
A full set of Ancillary Instrumentation is available with the system. The
{6}------------------------------------------------
| Tyber Medical Mid and Hindfoot System is offered both sterile and<br>non- sterile for single-use. |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Ankle Fracture System</b> |
| The Tyber Medical Ankle Fracture System is designed to address a<br>variety of indications in ankle reconstruction mid-shaft and distal<br>tibia/fibula fixation surgery. The system is composed of many locking<br>plate types which include:<br>1). Distal Fibula Plates, 2). Medial Distal Tibia Plates, 3). AnteroLateral<br>Distal Tibia Plates, 4). Posterior Distal Tibia T-Plates, 5). Straight Low<br>Contact Plates, and 6). Straight 1/3 Tubular Plates – in various plate<br>shapes, lengths, Right & Left versions. The System will incorporate both<br>Cortical Locking Screws and standard Cortical Screws in 2.8mm, 3.0mm,<br>3.5mm and 4.0mm sizes in various lengths. All plates are composed of<br>either Medical Grade CP Titanium (to ASTM F67) or 6-4 Alloyed Titanium<br>(to ASTM F136) materials with an Anodized Type II surface treatment.<br>All screws are composed of Medical Grade 6-4 Alloyed Titanium<br>material (to ASTM F136) color anodized for sizing. |
| A full set of Ancillary Instrumentation is available with the system. The<br>Tyber Medical Ankle Fracture System is offered both sterile and non-<br>sterile for single-use. |
| Intended Use | The intended use of the Tyber Medical Anatomical Plating System<br>is to draw two or more aligned bone fragments together to<br>facilitate healing in an adult patient |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | I. Forefoot System:<br>The Tyber Medical Forefoot Plating System is Indicated for Use in<br>fixation of small bones and small bone fragments in the foot<br>(Phalanges and Metatarsals) for stabilization of fractures, joint<br>fusions, osteotomies, nonunions, malunions, reconstruction of<br>small bones, revision surgeries and replantations in an adult<br>patient.<br>The Tyber Medical Forefoot Plating System is not for Spinal Use.<br><br>II. Mid & Hindfoot System:<br>The Tyber Medical Mid & Hindfoot Plating System is Indicated for<br>Use in fixation of medium/large bones and medium/large bone<br>multi-fragments in the foot (Cuneiform, Cuboid, Navicular, Talus<br>and Calcaneus) for stabilization of fractures, joint fusions,<br>osteotomies, nonunions, malunions, reconstruction of |
{7}------------------------------------------------
| Ankle Fracture System:<br>III.<br>The Tyber Medical Ankle Fracture System is Indicated for Use in:<br>1). Fixation of fractures of the distal tibia included, but not<br>limited to, ankle fractures, perarticular fractures, corrective | medium/large bones, revision surgeries and replantations in an<br>adult patient.<br>The Tyber Medical Mid & Hindfoot Plating System is not for Spinal<br>Use. |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| osteotomies, non-unions, intra- and extra- articular and<br>distal tibia fractures with a shaft extension, and malleolar<br>fractures;<br>2). In intra- and extra articular fractures, osteotomies, medial<br>malleolar fractures and non-unions of the metaphyseal and<br>diaphyseal region of the distal fibula;<br>3). In distal tibia/fibula long bones which include the<br>metaphyseal and diaphyseal regions of the tibia and fibula<br>in the ankle.<br>The Tyber Medical Ankle Fracture System is not for Spinal Use. | |
| Statement of | The Tyber Medical Anatomical Plating System is the same in |
|------------------------|------------------------------------------------------------|
| Technological | design, material, technological characteristics and |
| Comparison and | indications when compared to the predicate, BioPro Foot |
| Fundamental Scientific | Plating Systems (K162674). |
| Technology | |
| Nonclinical Testing<br>Summary | There have been no changes to the design, material,<br>technological characteristics or indications of the plating<br>systems since receiving clearance in K162674. Therefore,<br>nonclinical testing is not provided in this submission. |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Clinical Test Summary | n/a |
| Conclusion | The Tyber Medical Anatomical Plating System is the same in design, material, technological characteristics, and indications when compared to the BioPro Foot Plating Systems (K162674). |
|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.