K162674 · Biopro, Inc. · HRS · Jun 16, 2017 · Orthopedic
Device Facts
Record ID
K162674
Device Name
BioPro Foot Plating Systems
Applicant
Biopro, Inc.
Product Code
HRS · Orthopedic
Decision Date
Jun 16, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the BioPro Foot Plating Systems is to draw two or more aligned bone fragments together to facilitate healing in an adult patient and is composed of the following bone plate categories: I. Forefoot System: The BioPro Forefoot Plating System is Indicated for Use in fixation of small bones and small bone fragments in the foot (Phalanges and Metatarsals) for stabilization of fractures, joint fusions, osteotomies , nonunions, malunions, reconstruction of small bones, revision surgeries and replantations in an adult patient. The Forefoot System is not for Spinal Use. II. Mid & Hindfoot System: The BioPro Mid & Hindfoot Plating System is Indicated for Use in fixation of medium/large bones and medium/large bone multi-fragments in the foot (Cuneiform, Cuboid, Navicular, Talus and Calcaneus) for stabilization of fractures, joint fusions, osteotomies, malunions, reconstruction of medium/large bones, revision surgeries and replantations in an adult patient. The Mid & Hindfoot System is not for Spinal Use. III. Ankle Fracture System: The BioPro Ankle Fracture System is Indicated for Use in: 1). Fixation of fractures of the distal tibia included, but not limited to, ankle fractures, perarticular fractures, corrective osteotomies, non-unions, intra- and extra- articular and distal tibia fractures with a shaft extension, and malleolar fractures; 2). In intra- and extra articular fractures, osteotomies, medial malleolar fractures and non-unions of the metaphyseal and diaphyseal region of the distal fibula; 3). In distal tibia/fibula long bones which include the metaphyseal and diaphyseal regions of the tibia and fibula in the ankle. The Ankle Fracture System is not for Spinal Use.
Device Story
BioPro Foot Plating Systems consist of metallic bone plates and locking/standard cortical screws used for internal fixation of foot and ankle fractures, osteotomies, and fusions. Operated by orthopedic surgeons in clinical settings; devices provide mechanical stabilization of bone fragments to facilitate healing. System includes ancillary instrumentation (drill guides, screwdrivers, bending irons, etc.) to assist in surgical implantation. Plates and screws are provided either EO sterile or non-sterile for single-use. Clinical benefit derived from rigid fixation of bone segments, promoting union and structural reconstruction.
Clinical Evidence
No clinical data submitted. Substantial equivalence supported by non-clinical engineering analysis, including plate bending strength, bending stiffness, screw insertion/removal torque, yield torque, axial pullout strength, and biocompatibility risk assessment. Pyrogen levels evaluated via LAL method.
Technological Characteristics
Materials: Medical Grade CP Titanium (ASTM F67) and 6-4 Alloyed Titanium (ASTM F136). Surface: Anodized Type II. Components: Locking and standard cortical screws (2.0mm–4.0mm). Form factor: Various plate geometries (straight, T, Y, L, cloverleaf, etc.). Sterilization: EO sterile or non-sterile. Energy source: None (mechanical). Connectivity: None.
Indications for Use
Indicated for adult patients requiring fixation of small, medium, or large bones in the foot and ankle (phalanges, metatarsals, cuneiform, cuboid, navicular, talus, calcaneus, distal tibia, distal fibula) for stabilization of fractures, joint fusions, osteotomies, nonunions, malunions, reconstruction, revision surgeries, and replantations. Contraindicated for spinal use.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Mondeal-Extremity Bone Fixation - M2 System (K072740)
OrthoSolutions – Extremity Fixation Implants for Osteosynthesis - FPS System (K111678)
Synthes – Medial Distal Tibia Plate System (K001945)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of three human profiles facing right, stacked on top of each other. The profiles are rendered in black and have a flowing, wave-like appearance. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the profile graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 16, 2017
Biopro, Inc. % Al Lippincott Biomedical Engineer & Consultant To Biopro, Inc. Engineering Consulting Services, Inc. 3150 E. 200th St. Prior Lake, Minnesota 55372
Re: K162674 Trade/Device Name: BioPro Foot Plating Systems Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: May 3, 2017 Received: May 12, 2017
Dear Al Lippincott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Vincent J. Devlin -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
# 510(k) NUMBER: K162674
## DEVICE NAME: BioPro Foot Plating Systems
The intended use of the BioPro Foot Plating Systems is to draw two or more aligned bone fragments together to facilitate healing in an adult patient and is composed of the following bone plate categories:
# I. Forefoot System:
The BioPro Forefoot Plating System is Indicated for Use in fixation of small bones and small bone fragments in the foot (Phalanges and Metatarsals) for stabilization of fractures, joint fusions, osteotomies , nonunions, malunions, reconstruction of small bones, revision surgeries and replantations in an adult patient. The Forefoot System is not for Spinal Use.
# II. Mid & Hindfoot System:
The BioPro Mid & Hindfoot Plating System is Indicated for Use in fixation of medium/large bones and medium/large bone multi-fragments in the foot (Cuneiform, Cuboid, Navicular, Talus and Calcaneus) for stabilization of fractures, joint fusions, osteotomies, malunions, reconstruction of medium/large bones, revision surgeries and replantations in an adult patient.
The Mid & Hindfoot System is not for Spinal Use.
# III. Ankle Fracture System:
The BioPro Ankle Fracture System is Indicated for Use in:
1). Fixation of fractures of the distal tibia included, but not limited to, ankle fractures, perarticular fractures, corrective osteotomies, non-unions, intra- and extra- articular and distal tibia fractures with a shaft extension, and malleolar fractures;
2). In intra- and extra articular fractures, osteotomies, medial malleolar fractures and non-unions of the metaphyseal and diaphyseal region of the distal fibula;
3). In distal tibia/fibula long bones which include the metaphyseal and diaphyseal regions of the tibia and fibula in the ankle.
The Ankle Fracture System is not for Spinal Use.
Prescription Use XXXXXXXXXX Over-The-Counter-Use (Per 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
K162674 Page 1/4
# 510(k) Summary
## In Accordance with 21 CFR 807.92 of the Federal Code of Regulations
| NAME OF FIRM: | BioPro, Inc. |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 2929 Lapeer Road |
| | Port Huron, MI 48060 |
| | www.bioproimplants.com |
| 510(k) FIRM CONTACT: | Al Lippincott |
| | Engineering Consulting Services, Inc. |
| | 3150 E. 200th St. |
| | Prior Lake, MN 55372 |
| | Tel. No. 952-492-5858 |
| | e-mail: allippincott@msn.com |
| DATE: | May 3, 2017 |
| TRADE NAME: | BioPro Foot Plating Systems |
| COMMON NAME: | Plate, Fixation, Bone; Screw, Fixation, Bone. |
| CLASSIFICATION: | Single/multiple component metallic bone fixation applieances and<br>accessories, Class II (21CFR, Sec. 888.3030)<br>Smooth or threaded metallic bone fixation faxtener, Class II<br>(21CFR, Sec. 888.3040) |
| DEVICE PRODUCT CODE: | HRS |
### HWC SUBSEQUENT PRODUCT CODE:
### SUBSTANTIALLY EQUIVALENT DEVICES:
| | Forefoot System | Mid & Hindfoot System | Ankle Fracture System |
|--------------------|------------------------------------------------------------|--------------------------------------------------------------------------------------------|-------------------------------------------------------|
| Primary Predicates | K072740 Mondeal-<br>Extremity Bone<br>Fixation - M2 System | K111678 OrthoSolutions –<br>Extremity Fixation Implants for<br>Osteosynthesis - FPS System | K001945 Synthes – Medial<br>Distal Tibia Plate System |
| Additional Predicates | K091479 & K142581 | K091479 & K142581 | K073460 & K083213 |
|-----------------------|------------------------------------------------------------|----------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| | Medartis – APTUS<br>Foot System | Medartis - APTUS Foot<br>System | Synthes – LCP Distal Fibula<br>Plate System |
| | K140376 & K141992 | K140376 & K141992 Stryker | K063672 I.T.S. – Fibula |
| | Stryker – VariAx 2<br>Foot Locking Plate<br>System | -VariAx 2 Foot Locking Plate<br>System | Plate Prolock with Angular<br>Stability System |
| | K090692 Wright –<br>ORTHOLOC 2.0/2.4<br>Plate System | K100776 Synthes – LCP<br>Variable Angle Midfoot System | K092812 & K013248<br>Synthes - LCP Anterolateral<br>Distal Tibia Plate System |
| | K100776 Synthes –<br>LCP Variable Angle<br>Forefoot System | K061808 Dacrco International,<br>Inc. - Darco Locking Bone Plate<br>System | |
{4}------------------------------------------------
| Additional Predicates | Forefoot System | Mid & Hindfoot System | Ankle Fracture System |
|-----------------------|--------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Continued: | | | |
| | K141784<br>OrthoSolutions -<br>ULTOS Plating<br>System | K121651, K120359, K090692<br>Wright - ORTHOLOC 3Di<br>Midfoot - Cotten, Evans, MTP,<br>Lapidus, Ankle Fusion Plating<br>Systems | K013248, K082807<br>Synthes – LCP Distal Tibia<br>Plate & LCP Compression<br>Plate Systems & Ankle<br>Arthrodesis Plates |
| | K001945 Synthes. –<br>Medial Distal Tibia<br>Plates | K123000 Integra - Total Foot<br>System - MOWO & Rearfoot | |
| | K090675 Smith &<br>Nephew - VLP Foot<br>Plating/Screw System | K110908 Medartis Ag - Aptus<br>Foot 3.5 System | |
| | K121651 Wright -<br>ORTHOLOC 3Di<br>Flatfoot System | | |
| Reference Predicates: | K922500 BioPro -<br>PSL Total Hip System<br>K132510 BioPro -<br>Infinity Plate Anchor<br>System |
|-----------------------|-------------------------------------------------------------------------------------------------|
|-----------------------|-------------------------------------------------------------------------------------------------|
DEVICE DESCRIPTION: The BioPro Foot Plating Systems consists of Predicate Foot and Ankle Bone Plate and Locking Screw implant components found with many companies with orthopedic markets in the United States. These Foot Bone Plate Systems fracture fixation & osteotomy implant devices consist of the following categories:
- 1. Forefoot System
- 2. Mid & Hindfoot System
- 3. Ankle Fracture System
A brief and concise description of each system is as follows:
1. The BioPro Forefoot System is designed to address a variety of indications in forefoot reconstruction fixation/osteotomy surgery. The system is composed of many locking plate types which include: 1). Straight Fracture Plates, 2). T-Shaped Fracture Plates, 3). Y-Shape Fracture Plates, 4). L-Shaped Fracture Plates, 5). Cloverleaf Fracture Plates, 6). TMT 1 Medial Fusion Plates, 7). Open Wedge Fusion Plates, 8). MTP Fusion Plates, 9). MTP Fusion Revision Plates, and 10). Dorsal TMT 1 Step Fusion Plates, - in various plate lengths, Right & Left versions. The System will incorporate both Cortical Locking Screws and standard Cortical Screws in 2.0mm, 2.5mm, 3.0mm and 3.5mm sizes in various lengths. All plates are composed of Medical Grade CP Titanium material (to ASTM F67) with an Anodized Type II surface treatment. All screws are composed of Medical Grade 6-4 Alloyed Titanium material (to ASTM F136) color anodized for sizing.
{5}------------------------------------------------
A full set of Ancillary Instrumentation (Depth Gauge, Drill Guides, Drill DEVICE DESCRIPTION: Bits, Screwdrivers, Retractors, Bending Irons, K-wire with stop, Distraction/ Compression Device and MTP Fusion Reamers) is available with the system. The BioPro Forefoot System is offered both EO STERILE and Non-
sterile for single-use.
2. The BioPro Mid & Hindfoot System is designed to address a variety of indications in midfoot & hindfoot reconstruction fixation/osteotomy surgery. The system is composed of many locking plate types which include:
1). Straight Fracture/Fusion Plates, 2). T-Shaped Fracture/Fusion Plates, 3). L-Shaped Fracture/Fusion Plates, 4). Cloverleaf Fracture/Fusion Plates, 5). X-Shaped Fracture/Fusion Plates, 6). Retangular Fracture/Fusion Plates, 7). Cotton Osteotomy Plates, 8). Dwyer Step Plates, 9). Evans Osteotomy Plates, 10.) Plantar TMT Plates, 11). Distal Medial & 3 - 4 Column Plates and 11). ORIF & Standard Calcaneal Plates - in various plate lengths, Right & Left versions.
The System will incorporate both Cortical Locking Screws and standard Cortical Screws in 3.0mm, 3.5mm and 4.0mm sizes in various lengths. All plates are composed of Medical Grade CP Titanium material (to A.STM F67) with an Anodized Type II surface treatment. All screws are composed of Medical Grade 6-4 Alloyed Titanium material (to ASTM F136) color anodized for sizing.
A full set of Ancillary Instrumentation (Depth Gauge, Drill Guides, Drill Bits, Screwdrivers, Retractors, Reduction Forcepts, Bending Pliers, Osteotomes, K-Wires with stop and Distraction/Compression Device) is available with the system.
The BioPro Mid & Hindfoot System is offered both EO STERILE and Non-sterile for single-use.
3. The BioPro Ankle Fracture System is designed to address a variety of indications in ankle reconstruction mid-shaft and distal tibia/fibula fixation surgery. The system is composed of many locking plate types which include:
1). Distal Fibula Plates, 2). Medial Distal Tibia Plates, 3). AnteroLateral Distal Tibia Plates, 4). Posterior Distal Tibia T-Plates, 5). Straight Low Contact Plates, and 6). Straight 1/3 Tubular Plates - in various plate shapes, lengths, Right & Left versions. The System will incorporate both Cortical Locking Screws and standard Cortical Screws in 2.8mm, 3.0mm, 3.5mm and 4.0mm sizes in various lengths. All plates are composed of either Medical Grade CP Titanium (to ASTM F67) or 6-4 Alloyed Titanium (to ASTM F136) materials with an Anodized Type II surface treatment. All screws are composed of Medical Grade 6-4 Alloyed Titanium material (to ASTM F136) color anodized for sizing. A full set of Ancillary Instrumentation (Depth Gauge, Drill Guides, Drill Bits, Screwdrivers, Retractors, Reduction Forcepts, Bending Pliers, Osteotomes, K-Wires with stop, and Distraction/Compression Device) is available with the system. The BioPro Ankle Fracture System is offered both EO STERILE and
Non-sterile for single-use.
{6}------------------------------------------------
The intended use of the BioPro Foot Plating Systems is to draw two or INTENDED USE: more aligned bone fragments together to facilitate healing in an adult patient and is composed of the following bone plate categories:
### I. Forefoot System:
The BioPro Forefoot Plating System is Indicated for Use in fixation of small bones and small bone fragments in the foot (Phalanges and Metatarsals) for stabilization of fractures, joint fusions, osteotomies, nonunions, malunions, reconstruction of small bones, revision surgeries and replantations in an adult patient.
The Forefoot Plating System is not for Spinal Use.
### II. Mid & Hindfoot System:
The BioPro Mid & Hindfoot Plating System is Indicated for Use in fixation of medium/large bones and medium/large bone multi-fragments in the foot (Cuneiform, Cuboid, Navicular, Talus and Calcaneus) for stabilization of fractures, joint fusions, osteotomies, nonunions, malunions, reconstruction of medium/large bones, revision surgeries and replantations in an adult patient.
The Mid & Hindfoot Plating System is not for Spinal Use.
### III. Ankle Fracture System:
The BioPro Ankle Fracture System is Indicated for Use in: 1). Fixation of fractures of the distal tibia included, but not limited to, ankle fractures, perarticular fractures, corrective osteotomies, non-unions, intra- and extra- articular and distal tibia fractures with a shaft extension, and malleolar fractures; 2). In intra- and extra articular fractures, osteotomies, medial malleolar fractures and non-unions of the metaphyseal and diaphyseal region of the distal fibula; 3). In distal tibia/fibula long bones which include the metaphyseal and diaphyseal regions of the tibia and fibula in the ankle. The Ankle Fracture System is not for Spinal Use. NON-CLINICAL TESTING: An Engineering Analysis was performed to evaluate plate bending strength and bending stiffness. Also, an Engineering Analysis was performed to evaluate screw insertion/removal torque, yield torque, and axial pullout strength. A Biocompatability Risk Assessment was performed. Pyrogen bacterial endotoxin levels were evaluated using the LAL method. CLINICAL TESTING: Clinical data was not submitted. SUMMARY OF TECH-The BioPro Foot Plating Systems is similar in Material, Design NOLOGICAL and Indications to Predicate Systems legally marketed in the US. CHARACTERISTICS
CONCLUSIONS: Based on the similarity in Material, Geometry, Design, and Indications for Use, as well as all Engineering Analysis, the BioPro Foot Plating Systems have been demonstrated to be Substantially Equivalent (SE) to the Predicate Devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.