MRIaudio PREM System with MRIview

K193102 · Mriaudio, Inc. · LNH · Jan 6, 2020 · Radiology

Device Facts

Record IDK193102
Device NameMRIaudio PREM System with MRIview
ApplicantMriaudio, Inc.
Product CodeLNH · Radiology
Decision DateJan 6, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1000
Device ClassClass 2

Indications for Use

The MRIaudio PREM System with MRIview is intended to provide audio and visual entertainment to patients off 'trekkeyg r clegveqo o wplecyq in MRI environments up to, and including, 3.0 Tesla. The product is not intended for medical diagnosis or treatment. Technologist control units are intended to be used outside of the MRI scan room.

Device Story

MRIaudio PREM System with MRIview provides audio/visual entertainment, patient communication, and hearing protection during MRI scans. System operates in MRI environments up to 3.0 Tesla. Technologist control units located outside scan room. Device transforms audio/video inputs for patient consumption; allows technologists to view patient. Benefits include patient comfort and reduced anxiety during imaging. Healthcare providers use system to facilitate communication and provide entertainment; does not affect clinical diagnostic decision-making.

Clinical Evidence

Bench testing only. Evaluated magnetic field interactions at 3-Tesla (no interaction). Compliance with IEC 60601-1 and IEC 60601-1-2 standards. Sample clinical images in axial, sagittal, and coronal planes demonstrated no adverse effect on MR image production in worst-case environment.

Technological Characteristics

MRI-conditional audio/visual system for 3.0 Tesla environments. Components include technologist control units (external) and patient-facing hardware. Evaluated per IEC 60601-1 (General Safety) and IEC 60601-1-2 (EMC).

Indications for Use

Indicated for patients undergoing MRI procedures in environments up to 3.0 Tesla to provide audio/visual entertainment and facilitate communication. Not for medical diagnosis or treatment.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ January 6, 2020 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. MRIaudio, Inc. % Mr. Joseph Caruso Operations Manager 5909 Sea Lion Place, Suite F CARLSBAD CA 92010 Re: K193102 Trade/Device Name: MRIaudio PREM System with MRIview Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH Dated: November 6, 2019 Received: November 8, 2019 Dear Mr. Caruso: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) {1}------------------------------------------------ regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) Device Name MRIaudio PREM System with MRIview Indications for Use (Describe) The MRIaudio PREM System with MRIview is intended to provide audio and visual entertainment to patients off 'trekkeyg r clegveqo o wplecyq in MRI environments up to, and including, 3.0 Tesla. The product is not intended for medical"diagnosis or treatment. Technologist control units are intended to be used outside of the MRI scan room. | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------| | <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for MRIaudio. The text "MRIaudio" is in bold, black letters. To the right of the text is a circular graphic in teal, divided into four sections by white lines. K193102 Traditional 510(k) Premarket Notification MRIaudio PREM System with MRIview ## 510(k) Summary In accordance with Title 21 of the Code of Federal Regulations, Part 807, and 21 CFR §807.92, the following summary of information is provided: #### A. Submitted by: Joseph Caruso Operations Manager MRlaudio, Inc. 5909 Sea Lion Place, Suite F Carlsbad, CA 92010 Telephone: (858) 427-0679 Date Prepared: December 16, 2019 #### B. Device Name Trade or Proprietary Name: Classification Name: Classification Regulation: Classification Panel: Device Class: Product Code: MRIaudio PREM System with MRIview Magnetic Resonance Diagnostic Device 21 CFR § 892.1000 Radiology Class II LNH #### C. Predicate Device Trade or Proprietary Name: Maufacturer: 510(k) Clearance: Classification Regulation: Classification Name: Classification Panel: Device Class: Product Code: MRIaudio PREM System MRlaudio K180100 21 CFR § 892.1000 Magnetic Resonance Diagnostic Device Radiology Class II LNH #### D. Reference Device Trade or Proprietary Name: Maufacturer: 510(k) Clearance: Classification Regulation: Classification Name: Classification Panel: Device Class: Product Code: M.R. VISION 2000 ULTRA AUDIO VISUAL SYSTEM / COMMANDER X6 SYSTEM Resonance Technology, Inc. K994351 21 CFR § 892.1000 Magnetic Resonance Diagnostic Device Radiology Class II LNH {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for MRLaudio. The logo consists of the text "MRLaudio" in a bold, sans-serif font, with the letters "MRL" being larger than the letters "audio". To the right of the text is a circular graphic that is divided into four quadrants by two intersecting lines. The graphic is teal in color and has a white circle in the center. #### E. Device Description The MRIaudio PREM System with MRIview is an MRI conditional audio and video solution that provides MRI patients with music, direct communication, and hearing protection. With the addition of the MRIview technology, the system provides MRI patients with video and allows technologists to view the patient. #### F. Indications for Use The MRIaudio PREM System with MRIview is intended to provide audio and visual entertainment to patients and facilitate patient communication in MRI environments up to, and including, 3.0 Tesla. The product is not intended for medical diagnosis or treatment. Technologist control units are intended to be used outside of the MRI scan room. #### G. Technological Characteristics As was established in this submission, the subject MRIaudio PREM System with MRlview is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have the same technological characteristics to its predicate device through comparison in areas including design, intended use, material composition, and function. #### H. Performance Data Testing was performed to demonstrate that the subject MRIaudio PREM System with MRIview is substantially equivalent to other predicate devices. The following bench testing was performed: | Test | Result | |----------------------------------------|----------------| | Magnetic field interactions at 3-Tesla | No interaction | The subject MRIaudio PREM System with MRIview has also been evaluated to the following standards: | Electrical Safety/Electromagnetic Compatibility | | |-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | IEC 60601-1 | Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance | | IEC 60601-1-2 | Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests | The results of performance testing demonstrate that the subject MRIaudio PREM System with MRIview presents no adverse effect within the intended environment, and the subject device was therefore found to be substantially equivalent to the predicate. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for MRLaudio. The logo consists of the text "MRLaudio" in a bold, sans-serif font, followed by a circular graphic. The graphic is divided into four quadrants, each filled with a teal color, creating a stylized, modern design. Futher, sample clinical images submitted exhibit a mix of pulse sequences and imaging options in the axial, sagittal and coronal planes as recommended in FDA guidance, Submission Of Premarket Notifications for Magnetic Resonance Diagnostic Devices, issued November 18, 2016. No adverse events were reported; therefore, the subject MRIaudio PREM System with MRIview does not adversely affect MR image production in the worstcase environment. #### l. Conclusions Based on the indications for use, technological characteristics, and comparison to the predicate device, the subject MRIaudio PREM System with MRIview has been shown to be substantially equivalent to legally marketed predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...