The MRlaudio PREM system is intended to provide audio entertainment and facilitate patient communication in MRI environments, up to, and including, 3.0 Tesla. The product is not intended for medical diagnosis or treatment. Technologist control units are intended to be used outside of the MRI scan room.
Device Story
MRIaudio PREM system provides audio entertainment, patient-technologist communication, and hearing protection for patients in MRI environments up to 3.0 Tesla. System consists of patient-side audio components and technologist control units located outside the MRI scan room. Device facilitates communication and patient comfort during scanning; does not perform medical diagnosis or treatment. Technologist operates control unit to manage audio input and communication. System provides 29 dBA NRR noise reduction. Device is MR conditional.
Clinical Evidence
Bench testing included noise reduction (29 dBA NRR) and MR conditional safety evaluation (no field interaction up to 2 feet from bore/1 foot from side). Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2. Biocompatibility evaluated per ISO 10993-1. Clinical data submitted demonstrated no adverse effects on MR image production across various pulse sequences and imaging planes (axial, sagittal, coronal).
Technological Characteristics
MR conditional audio system for 3.0 Tesla environments. Provides 29 dBA NRR noise reduction. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and ISO 10993-1 (biocompatibility).
Indications for Use
Indicated for patients undergoing MRI procedures (up to 3.0 Tesla) requiring audio entertainment, communication, or hearing protection. Not for medical diagnosis or treatment.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Patient Communication and Entertainment System (K133670)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the logo for the Department of Health & Human Services, which features a stylized human figure. To the right is the logo for the U.S. Food & Drug Administration (FDA). The FDA logo includes the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with "ADMINISTRATION" written below in a smaller font.
June 14, 2018
MRIaudio Joe Caruso Chief Operations Officer 2720 Loker Ave., Suite N CARLSBAD, CA 92010
Re: K180100
Trade/Device Name: MRIaudio PREM System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH Dated: May 14, 2018 Received: May 21, 2018
Dear Joe Caruso:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
# Page 2 - Joe Caruso
K180100
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jeff Radigan
for Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
### 510(k) Number K180100
Device Name
MRlaudio PREM system
Indications for Use
The MRlaudio PREM system is intended to provide audio entertainment and facilitate patient communication in MRI environments, up to, and including, 3.0 Tesla. The product is not intended for medical diagnosis or treatment. Technologist control units are intended to be used outside of the MRI scan room.
| Type of Use |
|-------------|
|-------------|
● Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for MRlaudio. The logo consists of the text "MRlaudio" in a bold, sans-serif font, followed by a stylized circular graphic. The graphic is divided into four sections, each with a slightly different shade of teal, creating a sense of depth and movement.
## 510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and 21 CFR §807.92, the following summary of information is provided:
#### A. Submitted by:
Joe Caruso Chief Operations Officer MRlaudio, Inc. 2720 Loker Avenue West, Suite N Carlsbad, CA 92010 Telephone: (858) 914-4217 Date Prepared: May 14, 2018
#### B. Device Name
| Trade or Proprietary Name: | |
|----------------------------|--|
| Classification Name: | |
| Classification Regulation: | |
| Classification Panel: | |
| Device Class: | |
| Product Code: | |
MRIaudio PREM System Magnetic Resonance Diagnostic Device 21 CFR § 892.1000 Radiology Class II LNH
#### C. Predicate Devices
Trade or Proprietary Name: Maufacturer: 510(k) Clearance: Classification Regulation: Classification Name: Classification Panel: Device Class: Product Code:
Patient Communication and Entertainment System NeoCoil, LLC K133670 21 CFR § 892.1000 Magnetic Resonance Diagnostic Device Radiology Class II LNH
#### D. Device Description
The MRIaudio PREM system is an MRI conditional audio solution that provides MRI patients with music, direct communication, and hearing protection.
#### E. Indications for Use
The MRlaudio PREM system is intended to provide audio entertainment and facilitate patient communication in MRI environments, up to, and including, 3.0 Tesla. The product is not intended for medical diagnosis or treatment. Technologist control units are intended to be used outside of the MRI scan room.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for MRlaudio. The logo consists of the text "MRlaudio" in a bold, sans-serif font, with the "MRl" portion being larger than the "audio" portion. To the right of the text is a circular graphic composed of four teal-colored segments, resembling a stylized speaker or sound wave.
#### F. Technological Characteristics
As was established in this submission, the subject MRIaudio PREM system is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have the same technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, and function.
#### G. Performance Data
Testing was performed to demonstrate that the subject MRIaudioPREM system is substantially equivalent to other predicate devices. The following bench and clincal testing was performed:
| Test | Result |
|--------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| Noise Reduction to ANSI S3.19-1974 | 29 dBA NRR |
| Magnetic field attraction/MR Conditional rating (3-Tesla field strength) | No field interaction up to two (2) feet from bore and one (1) foot from the side of magnet |
The subject MRIaudioPREM system has also been evaluated to the following standards:
| Electrical Safety/Electromagnetic Compatibility | |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1 | Medical electrical equipment - Part 1. General requirements for basic safety and essential performance |
| IEC 60601-1-2 | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests |
| Biocompatibility | |
| ISO 10993-1 | Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process |
The results of performance testing demonstrate that the subject MRIaudio PREM system presents no adverse effect within the intended environment, and the subject device was therefore found to be substantially equivalent to the predicate.
Futher, clinical data submitted exhibits a mix of pulse sequences and imaging options in the axial, sagittal and coronal planes as recommended in the FDA guidance, Guidance for the Submission Of Promarket Notifications for Magnetic Resonance Diagnostic Devices, issued November 14, 1998. No adverse events were reported; therefore, the subject MRIaudioPREM system does not adversely affect MR image production in the worst-case environment.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for MRlaudio. The logo consists of the text "MRlaudio" in a bold, sans-serif font, followed by a stylized circular graphic. The graphic is divided into four sections, each with a slightly different shade of teal, creating a sense of depth and movement.
#### H. Conclusions
Based on the indications for use, technological characteristics, and comparison to predicate devices, the subject MRIaudioPREM system has been shown to be substantially equivalent to legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.