K193054 · Sectra AB · QKQ · Mar 31, 2020 · Pathology
Device Facts
Record ID
K193054
Device Name
Sectra Digital Pathology Module
Applicant
Sectra AB
Product Code
QKQ · Pathology
Decision Date
Mar 31, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.3700
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Sectra Digital Pathology Module device is a software intended for viewing and management of digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. It is an aid to the pathologist to review and interpret these digital images for the purposes of primary diagnosis. Sectra Digital Pathology Module is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. It is the responsibility of the pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images using Sectra Digital Pathology Module. Sectra Digital Pathology Module is intended for use with Leica’s Aperio AT2 DX scanner and Dell MR2416 monitor.
Device Story
Sectra Digital Pathology Module is a software-only device for viewing/managing digital pathology images. Input: digital images of H&E stained FFPE tissue slides from Aperio AT2 DX scanner. Operation: images imported via Sectra Pathology Import Server (SPIS), stored in Sectra Core, and accessed via Sectra Workstation IDS7. Pathologist uses Pathology Image Window (PIW) to view, zoom, pan, adjust focus, annotate, and measure distances/areas. Used in clinical settings by pathologists for primary diagnosis. Output: visual display of pathology slides on Dell MR2416 monitor. Device facilitates clinical decision-making by enabling digital review of slides equivalent to conventional light microscopy. Benefits: enables remote/digital workflow for primary diagnosis.
Clinical Evidence
No clinical data. Substantial equivalence is based on non-clinical bench testing, including color reproducibility, measurement accuracy, and turnaround time (loading/panning latency), conducted per FDA guidance for digital pathology WSI devices. Human factors and usability testing confirmed the interface is safe and effective for intended users.
Technological Characteristics
Software-only device; web application (PIW) integrated with Sectra PACS. Operates on Windows 10/7 (64-bit) workstations with NVIDIA Quadro graphics. Connectivity via 1 Gbit/s LAN. Uses standard DICOM-like image management workflows. No specific ML/AI architecture; relies on standard image rendering and manipulation algorithms.
Indications for Use
Indicated for pathologists to review and interpret digital images of FFPE surgical pathology slides for primary diagnosis. Not for use with frozen section, cytology, or non-FFPE hematopathology specimens. Intended for use with Leica Aperio AT2 DX scanner and Dell MR2416 monitor.
Regulatory Classification
Identification
The whole slide imaging system is an automated digital slide creation, viewing, and management system intended as an aid to the pathologist to review and interpret digital images of surgical pathology slides. The system generates digital images that would otherwise be appropriate for manual visualization by conventional light microscopy.
Special Controls
A whole slide imaging system must comply with the following special controls: (1) Premarket notification submissions must include the following information: (i) The indications for use must specify the tissue specimen that is intended to be used with the whole slide imaging system and the components of the system. (ii) A detailed description of the device and bench testing results at the component level, including for the following, as appropriate: (A) Slide feeder; (B) Light source; (C) Imaging optics: (D)Mechanical scanner movement; (E) Digital imaging sensor; (F) Image processing software; (G)Image composition techniques; (H)Image file formats; (I) Image review manipulation software; (J) Computer environment; (K)Display system. (iii)Detailed bench testing and results at the system level, including for the following, as appropriate: (A)Color reproducibility; (B) Spatial resolution; (C) Focusing test; (D) Whole slide tissue coverage; (E) Stitching error: (F) Turnaround time. (iv) Detailed information demonstrating the performance characteristics of the device, including, as appropriate: (A)Precision to evaluate intra-system and inter-system precision using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included. (B) Reproducibility data to evaluate inter-site variability using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included. (C) Data from a clinical study to demonstrate that viewing, reviewing, and diagnosing digital images of surgical pathology slides prepared from tissue slides using the whole slide imaging system is non-inferior to using an optical microscope. The study should evaluate the difference in major discordance rates between manual digital (MD) and manual optical (MO) modalities when compared to the reference (e.g., main sign-out diagnosis). (D) A detailed human factors engineering process must be used to evaluate the whole slide imaging system user interface(s). (2) Labeling compliant with 21 CFR 809.10(b) must include the following: The intended use statement must include the information described in paragraph (i) (1)(i) of this section, as applicable, and a statement that reads, "It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using this device." (ii) A description of the technical studies and the summary of results, including those that relate to paragraph (1)(ii) and (1)(iii) of this section, as appropriate. (iii) A description of the performance studies and the summary of results, including those that relate to paragraph (1)(iv) of this section, as appropriate. (iv) A limiting statement that specifies that pathologists should exercise professional judgment in each clinical situation and examine the glass slides by conventional microscopy if there is doubt about the ability to accurately render an interpretation using this device alone.
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the following information:
(i) The indications for use must specify the tissue specimen that is intended to be used with the whole slide imaging system and the components of the system.
(ii) A detailed description of the device and bench testing results at the component level, including for the following, as appropriate:
(A) Slide feeder;
(B) Light source;
(C) Imaging optics;
(D) Mechanical scanner movement;
(E) Digital imaging sensor;
(F) Image processing software;
(G) Image composition techniques;
(H) Image file formats;
(I) Image review manipulation software;
(J) Computer environment; and
(K) Display system.
(iii) Detailed bench testing and results at the system level, including for the following, as appropriate:
(A) Color reproducibility;
(B) Spatial resolution;
(C) Focusing test;
(D) Whole slide tissue coverage;
(E) Stitching error; and
(F) Turnaround time.
(iv) Detailed information demonstrating the performance characteristics of the device, including, as appropriate:
(A) Precision to evaluate intra-system and inter-system precision using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(B) Reproducibility data to evaluate inter-site variability using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(C) Data from a clinical study to demonstrate that viewing, reviewing, and diagnosing digital images of surgical pathology slides prepared from tissue slides using the whole slide imaging system is non-inferior to using an optical microscope. The study should evaluate the difference in major discordance rates between manual digital (MD) and manual optical (MO) modalities when compared to the reference (
*e.g.,* main sign-out diagnosis).(D) A detailed human factor engineering process must be used to evaluate the whole slide imaging system user interface(s).
(2) Labeling compliant with 21 CFR 809.10(b) must include the following:
(i) The intended use statement must include the information described in paragraph (b)(1)(i) of this section, as applicable, and a statement that reads, “It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using this device.”
(ii) A description of the technical studies and the summary of results, including those that relate to paragraphs (b)(1)(ii) and (iii) of this section, as appropriate.
(iii) A description of the performance studies and the summary of results, including those that relate to paragraph (b)(1)(iv) of this section, as appropriate.
(iv) A limiting statement that specifies that pathologists should exercise professional judgment in each clinical situation and examine the glass slides by conventional microscopy if there is doubt about the ability to accurately render an interpretation using this device alone.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
INSTRUMENT ONLY
## I. Background Information:
A. 510(k) Number
K193054
B. Applicant
Sectra AB
C. Proprietary and Established Names
Sectra Digital Pathology Module
D. Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QKQ | Class II | 21 CFR 864.3700 | Pathology |
## II. Submission/Device Overview:
A. Purpose for Submission:
New device
B. Type of Test:
Not applicable – software only device
## III. Intended Use/Indications for Use:
A. Intended Use(s):
See Indications for Use below.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
B. Indication(s) for Use:
For In Vitro Diagnostic Use
Sectra Digital Pathology Module device is a software intended for viewing and management of digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. It is an aid to the pathologist to review and interpret these digital images for the purposes of primary diagnosis. Sectra Digital Pathology Module is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. It is the responsibility of the pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images using Sectra Digital Pathology Module. Sectra Digital Pathology Module is intended for use with Leica’s Aperio AT2 DX scanner and Dell MR2416 monitor.
C. Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
IV. Device/System Characteristics:
A. Device Description:
Sectra Digital Pathology Module (version 2.2) is a software-only device for viewing and manipulating digital pathology images of glass slides obtained from the Aperio AT2 DX scanner on the Dell MR2416 monitor. Sectra Digital Pathology Module may only be used in combination with Sectra Picture Archiving and Communication System (PACS), which consists of Sectra Workstation IDS7 and Sectra Core. Sectra Digital Pathology Module consists of four elements: Pathology Image Window (PIW), Sectra Pathology Server (SPS), Database Engine (SPDB), and Sectra Pathology Import Server (SPIS). The PIW is a web application embedded in Sectra Workstation IDS7 for end users to view and manipulate digital pathology images. The SPS is the main server component for the PIW client. The SPDB stores the metadata required for the SPS. The SPIS is used for importing digital pathology images from the Aperio AT2 DX scanner.
The subject device is operated as follows:
1. The image acquisition is performed using the predicate device, Aperio AT2 DX scanner. The operator performs quality control of the digital slides using ImageScope DX and Dell MR2416 per the instructions of Aperio AT2 DX to determine if re-scans are necessary.
2. The digital slides are imported by SPIS and automatically assembled as groups of cases. The unaltered images are then sent to the external storage Sectra Core. Sectra Core offers an application programming interface allowing communication between the Sectra Healthcare Server (SHS), which is part of Sectra Core, and the SPS, which is part of the subject device. The image metadata (e.g. pixel size) is stored locally in the SPDB to improve operational performance.
3. The laboratory technician performs a quality control by using PIW to verify that all images have been imported.
K193054 - Page 2 of 10
{2}
4. The reading pathologist selects a case (patient) from a worklist external to the subject device such as the laboratory information system (LIS) whereby the SPS fetches the associated images from the Sectra Core.
5. The reading pathologist uses the subject device to view the images and is able to perform the following actions, as needed:
a. Zoom and pan the image
b. Adjust focus (when different focus depths are available)
c. Measure distances and areas in the image
d. Annotate images
e. Choose to view multiple images side by side in a synchronized fashion
After viewing all images belonging to a particular case (patient), the pathologist will make a diagnosis. The diagnosis is documented in another system such as the LIS.
## Minimum System Requirements - Computer Environment
The system requirements are given in Tables 1 through 4 below.
Table 1: Requirements for PIW
| Workstation (Host for subject device) | IDS7 |
| --- | --- |
| Operating System | MS Windows 10 or MS Windows 7 (64 bit) |
| CPU | 4-core CPU, 3.6 GHz |
| RAM | 8 GB |
| Display (Image Window) | Dell MR2416 monitor |
| Graphics board | NVIDIA 2 GB Quadro P600 or 4 GB Quadro P1000 |
| Network | 1 Gbit/s LAN connection |
| 3D mouse | 3Dconnexion SpaceMouse Pro |
Table 2: Requirements for SPDB and SPS
| Operating System & Database Engine | • Windows Server 2016 Std Ed, with SQL Server 2016 SP2 Std Ed
• Windows Server 2012 R2 Std Ed, with SQL Server 2012 SP4 Std Ed Windows |
| --- | --- |
| .NET Framework | .NET Framework 4.7.1, 4.7.2 or 4.8 |
K193054 - Page 3 of 10
{3}
Table 3: Requirements for SPIS
| Operating System | · Windows Server 2016 · Windows Server 2012 R2 (upgrades only) · Windows Server 2008 R2 (upgrades only) · Windows 10 · Windows 8.1 (upgrades only) · Windows 7 (upgrades only) |
| --- | --- |
| .NET Framework | .NET Framework 4.7.1, 4.7.2 or 4.8 |
Table 4: Recommended Configurations
| SPDB and SPS | |
| --- | --- |
| CPU | 8-core CPU, e.g., Intel Xeon 4110 CPU |
| RAM | 24 GB RAM |
| Disk | 2x 146 GB system disk (mirrored) |
| Network | Gigabit LAN connection |
| SPIS | |
| CPU | 8-core CPU, e.g. Intel Xeon 4110 CPU |
| RAM | 8 GB RAM |
| Disk | 146 GB system disk |
| Network | Gigabit LAN connection |
B. Instrument Description Information:
| Modes of Operation | Yes | No |
| --- | --- | --- |
| Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device? | ☑ | ☐ |
| Does the applicant's device transmit data to a computer, webserver, or mobile device using wireless transmission? | ☐ | ☑ |
| Software | | |
| FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types. | ☑ | ☐ |
1. Instrument Name:
Sectra Digital Pathology Module
2. Specimen Identification:
The Sectra Digital Pathology Module uses digital pathology images obtained from the Aperio AT2 DX scanner of Hematoxylin and Eosin (H&E) stained glass slides. The reading pathologist selects a case (patient) from a worklist external to the subject device whereby the subject device fetches the associated images from the external image storage. The scanned images are identified based on the previously assigned specimen identifier.
K193054 - Page 4 of 10
{4}
K193054 - Page 5 of 10
3. Specimen Sampling and Handling:
Specimen sampling and handling are performed upstream and independent of the use of the subject device. Specimen sampling includes biopsy or resection specimens which are processed using histology techniques. The FFPE tissue section is H&E stained. Digital images are then obtained from these glass slides using the Aperio AT2 DX scanner.
4. Calibration:
Not applicable
5. Quality Control:
The subject device receives quality-controlled images from the scanner. The subject device specific quality control measures are as follows:
- Connect scanner - This test should be performed before connecting the scanner to the subject device in order to verify the on-site integration between the Aperio AT2 Dx scanner and the Sectra Workstation.
- View pathology images - Every pathologist should perform this test on review workstation before reading pathology images using the subject device to ensure that all scanned slide images have been imported and for every case, view the thumbnails in the pathology image window to verify that each slide that should be in the case is present (manually verifying tissue block and staining information from LIS).
Additional details of the quality control procedures are provided in the device User's Guide and the Installation Guide.
V. Substantial Equivalence Information:
A. Predicate Device Name(s):
Aperio AT2 DX System
B. Predicate 510(k) Number(s):
K190332
C. Comparison with Predicate(s):
| Device & Predicate Device(s): | K193054 | K190332 |
| --- | --- | --- |
| Device Trade Name | Sectra Digital Pathology Module | Aperio AT2 DX System |
| General Device Characteristic Similarities | | |
{5}
| Intended Use | Sectra Digital Pathology Module device is a software intended for viewing and management of digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. It is an aid to the pathologist to review and interpret these digital images for the purposes of primary diagnosis. Sectra Digital Pathology Module is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. It is the responsibility of the pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images using Sectra Digital Pathology Module. Sectra Digital Pathology Module is intended for use with Leica's Aperio AT2 DX scanner, Sectra Workstation IDS7 and Dell MR2416 monitor. | The Aperio AT2 DX System is an automated digital slide creation and viewing system. The Aperio AT2 DX System is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. The Aperio AT2 DX System is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. The Aperio AT2 DX System is composed of the Aperio AT2 DX scanner, the ImageScope DX review application and Display. The Aperio AT2 DX System is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using the Aperio AT2 DX System. |
| --- | --- | --- |
| Specimen Type | Surgical pathology slides prepared from FFPE tissue (scanned digital images is the starting point) | Surgical pathology slides prepared from FFPE tissue |
| Image Storage | Images are stored in an end | Same |
K193054 - Page 6 of 10
{6}
| | user provided image storage attached to the local network | |
| --- | --- | --- |
| Image Manipulation Functions | Panning, zooming, gamma function, annotations, and measurements (distance & area) | Same |
| General Device Characteristic Differences | | |
| Device Components | Sectra Digital Pathology Module which consists only of the image viewing software, Sectra Workstation IDS7, Sectra Core | WSI scanner (Aperio AT2 DX scanner), Image Management System (ImageScope DX application), and color monitor display |
| Principle of Operation | After WSI images are successfully acquired by using Aperio At2 DX scanner, WSI images are imported by SPIS as groups of cases. The laboratory technician performs a quality control by using PIW to verify that all images have been imported. During review, the pathologist selects a patient case and opens WSI images from the Sectra Core to read the WSI images and make a diagnosis. | After conducting Quality Control (QC) on the glass slides per laboratory standards (e.g., staining, coverslipping, barcode placement, etc.), the technician loads the slides into the Aperio AT2 DX scanner. The scanner scans the slides and generates WSI image for each slide. The technician performs QC on scanned WSI images by checking image data and image quality. When QC is failed, the slide will be re-scanned. The acquired WSI images are stored in an end user provided image storage attached to the local network. During review, the pathologist opens WSI images acquired with the Aperio AT2 DX scanner from the image storage, performs further QC to ensure image quality and reads WSI images of the slides to make a diagnosis. |
| End User's Interface | Pathology Image Window | Aperio ImageScope DX |
K193054 - Page 7 of 10
{7}
K193054 - Page 8 of 10
| | (the client component of Sectra Digital Pathology Module) | |
| --- | --- | --- |
| Type of Software Application | Web application Native MS Windows | Native MS Windows |
VI. Standards/Guidance Documents Referenced:
1. Guidance for Industry “Technical Performance Assessment of Digital Pathology Whole Slide Imaging Devices”, dated April 20, 2016.
2. Special controls under 21 CFR Sec. 864.3700 Whole slide imaging system.
3. EN ISO 14971:2012 – Medical devices - Application of risk management to medical devices (same as ISO 14971:2007, Corrected version 2007-10-01), Recognition Number: 5-40.
4. EN 62304:2006 – Medical device software - software life-cycle processes, Recognition Number: 13-79.
5. EN ISO 62366-1:2015 – Application of usability engineering to medical devices, Recognition Number: 5-114.
VII. Performance Characteristics (if/when applicable):
A. Analytical Performance:
1. Precision/Reproducibility:
Not applicable
2. Linearity:
Not applicable
3. Analytical Specificity/Interference:
Not applicable
4. Accuracy (Instrument):
Not applicable
5. Carry-Over:
Not applicable
B. Other Supportive Instrument Performance Characteristics Data:
Technical performance testing for Sectra Digital Pathology Module device was performed. The new device was compared to the image management software component of the predicate device. The following testing was performed:
{8}
a. Pixel-wise comparison with the predicate device including zooming and panning operations across multiple tiles and vertical/horizontal stitching seams
The equivalence between the subject and predicate image review manipulation software (IRMS, as defined in the FDA guidance titled "Guidance for Industry "Technical Performance Assessment of Digital Pathology Whole Slide Imaging Devices", dated April 20, 2016 [TPA guidance, IV(A)(9)] was supported by bench testing data based on pixel-level comparison. The subject IRMS was tested as operating with the intended components, including the scanner, image management system and display. The scanner color profile embedded in the whole slide image (WSI) file as well as the display color profile associated with the intended display were included in the tests. WSI files from 30 FFPE tissue glass slides from different anatomic locations that were H&E stained were used as the test input. For each region of interest (ROI), the differences between the views generated by the subject and predicate IRMS were evaluated with the 1976 International Commission on Illumination (CIE) color difference metric $\Delta E$ for each corresponding pixel pair. The two views generated by the subject and predicate IRMS were adjusted and registered by using only the graphical user interface without image processing. The test cases of ROIs included relevant biological features at different magnification levels such as $20\mathrm{x}$ and $40\mathrm{x}$. Horizontal/vertical stitching seams between the tiles were included in the ROIs when possible. For each ROI, zooming to an odd magnification level and panning across multiple tiles that require interpolated rendering were also tested. The average color difference of all pixels within each ROI was reported. The image data of all ROIs were also provided for verification. The test results demonstrated that the $99^{\text{th}}$ percentile of the difference is below $2.06\Delta E$. The subject device has been found to adequately reproduce digital pathology images at the pixel level with respect to its intended use.
b. Turnaround Time for the image to be fully loaded while opening a slide from a medical device data system (MDDS) and for panning/zooming
Comparison with predicate device: The predicate device ImageScopeDX has the following reported turnaround times: "should not take longer than 7 secs till the image is fully loaded while opening a slide from a MDDS." and "Panning on the image should not take longer than 0.5 sec in order to load and show all image data." The turnaround times for opening an image and panning have been measured in the subject device. The results are similar to that of the predicate device ImageScopeDX.
Turnaround time zooming and panning: Long-time running performance test simulating normal usage: A long-time running client test simulating normal end user behavior, measuring performance and executing for at least 6 hours. This test showed no signs of any memory leak or performance issues. The subject device has been found to have an acceptable turnaround time with respect to its intended use.
c. Measurements – area and distance
Measurement accuracy has been verified using a test image containing objects of known sizes. The following set of tests were used to validate the measurement accuracy of the subject device.
Comparison with predicate device: The same measurements are performed using the subject device and using the Predicate device ImageScopeDX. The results are verified to correspond.
K193054 - Page 9 of 10
{9}
Measurements verified using calibration slide with known distances: An image of a calibration slide with known distances was used to verify the measurement accuracy of the subject device. Test results showed that the subject device performed accurate measurements with respect to its intended use.
d. Human Factors Testing
Human factors study designed around critical user tasks and use scenarios performed by representative users were conducted. A systematic evaluation of task-based usability including critical tasks required for operation of the device were evaluated at multiple sites using multiple users. All tasks associated with reviewing and reporting results for cases including confirmation that all slides belonging to specific cases are reviewed before reporting results, were included in the study. After having completed all the tasks, all the participants were asked a set of overall satisfaction questions followed by the System Usability Scale (SUS) questionnaire. Overall, the results of the human factors testing were acceptable.
VIII. Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K193054 - Page 10 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.