The study used a retrospective cohort of 2,045 clinical cases to compare the diagnostic performance of the Aperio AT2 DX System (Whole Slide Imaging Review) against traditional light microscopy (Microscope Slide Review) using the original clinical sign-out diagnosis as the reference standard.
Multi-center clinical study for Aperio AT2 DX System; Retrospective multi-center non-inferiority study; Follow-up/Duration: Not applicable; Study Period: Cases were at least one year old at the time of curation
Surgical pathology cases (FFPE tissue) with existing sign-out diagnoses; Sample Size: 2045 cases (5849 slides); Number of Sites: 5
Microscope Slide Review (MSR) using conventional light microscopy
Major discordance rate between WSI review and reference diagnosis
Indications for Use
The Aperio AT2 DX System is an automated digital slide creation and viewing system. The Aperio AT2 DX System is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. The Aperio AT2 DX System is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. The Aperio AT2 DX System is composed of the Aperio AT2 DX scanner, the ImageScope DX review application and Display. The Aperio AT2 DX System is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using the Aperio AT2 DX System.
Device Story
Aperio AT2 DX System is an automated digital slide creation and viewing system for primary diagnosis. System components include an autoloader-equipped scanner (400-slide capacity), a viewing workstation, and ImageScope DX software. The scanner uses a high numeric aperture 20x objective and linear-array scanning to digitize glass slides into seamless digital images. Pathologists use the ImageScope DX software on a calibrated monitor to view, annotate, and interpret digital slides, mimicking conventional light microscopy. The system supports remote intranet access. By digitizing slides, the device enables pathologists to perform clinical assessments and make diagnostic decisions remotely or within a clinical setting, potentially improving workflow efficiency while maintaining diagnostic accuracy comparable to manual microscopy.
Clinical Evidence
Multi-center clinical study (5 sites, 2045 cases, 5849 slides) compared Whole Slide Imaging Review (WSIR) to Light Microscope Slide Review (MSR). Primary endpoint: major discordance rate difference between modalities. Observed major discordance: 3.73% (WSIR) vs 3.28% (MSR). Modeled difference: 0.44% (95% CI: -0.15%, 1.03%). Upper bound of CI was below the 4% non-inferiority margin. Study concluded WSIR is non-inferior to MSR.
Technological Characteristics
System includes an automated scanner with 400-slide autoloader, viewing workstation, and calibrated monitor (Dell MR2416). Employs linear-array scanning technique with high numeric aperture 20x objective. Software includes Console DX, Controller DX, and ImageScope DX. Connectivity via local network/intranet. System is designed for FFPE tissue slides.
Indications for Use
Indicated for use by pathologists to review and interpret digital images of surgical pathology slides prepared from FFPE tissue. Not indicated for frozen section, cytology, or non-FFPE hematopathology specimens.
Regulatory Classification
Identification
The whole slide imaging system is an automated digital slide creation, viewing, and management system intended as an aid to the pathologist to review and interpret digital images of surgical pathology slides. The system generates digital images that would otherwise be appropriate for manual visualization by conventional light microscopy.
Special Controls
A whole slide imaging system must comply with the following special controls: (1) Premarket notification submissions must include the following information: (i) The indications for use must specify the tissue specimen that is intended to be used with the whole slide imaging system and the components of the system. (ii) A detailed description of the device and bench testing results at the component level, including for the following, as appropriate: (A) Slide feeder; (B) Light source; (C) Imaging optics: (D)Mechanical scanner movement; (E) Digital imaging sensor; (F) Image processing software; (G)Image composition techniques; (H)Image file formats; (I) Image review manipulation software; (J) Computer environment; (K)Display system. (iii)Detailed bench testing and results at the system level, including for the following, as appropriate: (A)Color reproducibility; (B) Spatial resolution; (C) Focusing test; (D) Whole slide tissue coverage; (E) Stitching error: (F) Turnaround time. (iv) Detailed information demonstrating the performance characteristics of the device, including, as appropriate: (A)Precision to evaluate intra-system and inter-system precision using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included. (B) Reproducibility data to evaluate inter-site variability using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included. (C) Data from a clinical study to demonstrate that viewing, reviewing, and diagnosing digital images of surgical pathology slides prepared from tissue slides using the whole slide imaging system is non-inferior to using an optical microscope. The study should evaluate the difference in major discordance rates between manual digital (MD) and manual optical (MO) modalities when compared to the reference (e.g., main sign-out diagnosis). (D) A detailed human factors engineering process must be used to evaluate the whole slide imaging system user interface(s). (2) Labeling compliant with 21 CFR 809.10(b) must include the following: The intended use statement must include the information described in paragraph (i) (1)(i) of this section, as applicable, and a statement that reads, "It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using this device." (ii) A description of the technical studies and the summary of results, including those that relate to paragraph (1)(ii) and (1)(iii) of this section, as appropriate. (iii) A description of the performance studies and the summary of results, including those that relate to paragraph (1)(iv) of this section, as appropriate. (iv) A limiting statement that specifies that pathologists should exercise professional judgment in each clinical situation and examine the glass slides by conventional microscopy if there is doubt about the ability to accurately render an interpretation using this device alone.
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the following information:
(i) The indications for use must specify the tissue specimen that is intended to be used with the whole slide imaging system and the components of the system.
(ii) A detailed description of the device and bench testing results at the component level, including for the following, as appropriate:
(A) Slide feeder;
(B) Light source;
(C) Imaging optics;
(D) Mechanical scanner movement;
(E) Digital imaging sensor;
(F) Image processing software;
(G) Image composition techniques;
(H) Image file formats;
(I) Image review manipulation software;
(J) Computer environment; and
(K) Display system.
(iii) Detailed bench testing and results at the system level, including for the following, as appropriate:
(A) Color reproducibility;
(B) Spatial resolution;
(C) Focusing test;
(D) Whole slide tissue coverage;
(E) Stitching error; and
(F) Turnaround time.
(iv) Detailed information demonstrating the performance characteristics of the device, including, as appropriate:
(A) Precision to evaluate intra-system and inter-system precision using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(B) Reproducibility data to evaluate inter-site variability using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(C) Data from a clinical study to demonstrate that viewing, reviewing, and diagnosing digital images of surgical pathology slides prepared from tissue slides using the whole slide imaging system is non-inferior to using an optical microscope. The study should evaluate the difference in major discordance rates between manual digital (MD) and manual optical (MO) modalities when compared to the reference (
*e.g.,* main sign-out diagnosis).(D) A detailed human factor engineering process must be used to evaluate the whole slide imaging system user interface(s).
(2) Labeling compliant with 21 CFR 809.10(b) must include the following:
(i) The intended use statement must include the information described in paragraph (b)(1)(i) of this section, as applicable, and a statement that reads, “It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using this device.”
(ii) A description of the technical studies and the summary of results, including those that relate to paragraphs (b)(1)(ii) and (iii) of this section, as appropriate.
(iii) A description of the performance studies and the summary of results, including those that relate to paragraph (b)(1)(iv) of this section, as appropriate.
(iv) A limiting statement that specifies that pathologists should exercise professional judgment in each clinical situation and examine the glass slides by conventional microscopy if there is doubt about the ability to accurately render an interpretation using this device alone.
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
A. 510(k) Number:
K190332
B. Purpose for Submission:
New Whole Slide Imaging (WSI) system
C. Measurand:
Not applicable
D. Type of Test:
Digital pathology whole slide imaging
E. Applicant:
Leica Biosystems Imaging, Inc.
F. Proprietary and Established Names:
Aperio AT2 DX System
G. Regulatory Information:
1. Regulation section:
21 CFR 864.3700
2. Classification:
Class II (special controls)
3. Product code:
PSY
4. Panel:
88 - Pathology
{1}
H. Intended Use:
1. Intended use(s):
The Aperio AT2 DX System is an automated digital slide creation and viewing system. The Aperio AT2 DX System is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. The Aperio AT2 DX System is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.
The Aperio AT2 DX System is composed of the Aperio AT2 DX scanner, the ImageScope DX review application and Display. The Aperio AT2 DX System is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using the Aperio AT2 DX System.
2. Indication(s) for use:
Same as Intended Use.
3. Special conditions for use statement(s):
For in vitro diagnostic (IVD) use only
For prescription use only
4. Special instrument requirements:
Aperio AT2 DX digital slide scanner (for software Aperio Console DX 102.0.4.46 and Aperio Controller DX 102.0.4.85_x64)
Viewing Workstation executing ImageScope DX viewer software (version 1.0.0.5018)
Display (MR2416)
I. Device Description:
The Aperio AT2 DX System is an automated digital slide creation and viewing system. The system is composed of the following components:
- Aperio AT2 DX digital slide scanner
- Viewing Workstation executing ImageScope DX viewer software
{2}
- Display (MR2416)
The Aperio AT2 DX digital slide scanner digitizes microscope slides at diagnostic resolutions: 0.5 µm/pixel (20x objective equivalent) and 0.25 µm/pixel (40x objective equivalent), to create digital WSI images. The Aperio AT2 DX scanner has a capacity of 400 1x3 inch glass slides. After the slides are loaded the scanning process begins when the operator starts the Aperio AT2 DX scanner and finishes when the scanner has completed scanning of all loaded slides. The image acquisition components include a white LED light source, Olympus Plan Apo 20x lens with a Numeric Aperture (NA) of 0.75, a 10/90 beam splitter, a 2D area-scan camera and a 1D line-scan camera which is the primary image capture device. The 10/90 beam splitter sends 10% of the photons toward the 2D area-scan camera and remaining 90% of the photons toward the 1D line-scan camera. Two motorized stages are used with one stage for moving the slide to be scanned and the other stage for moving the objective lens for focusing purposes. During scanning with a 20x objective lens, the motion of the slide stage combined with the speed of the 1D line-scan camera, the field of view of the objective lens and the number of pixels of the camera result in a resolution of 0.5 µm/pixel for 20x magnification. 40x magnification is achieved with a magnification doubler, with 0.25 µm/pixel resolution. Focus during the scan is maintained using a triangulated focus map built from individual focus points determined in a separate step before scanning is started. Proprietary software is used for image processing during acquisition. Leica's proprietary format, SVS, is used to store and transmit the images between the Aperio AT2 DX digital slide scanner and the ImageScope DX viewer on viewing station. Software Aperio Controller DX runs on the AT2 DX workstation as a system service and controls the operations of the scanner. Software Aperio Console DX is a desktop program that runs on the AT2 DX workstation with a display monitor (Dell MR2416) and provides user interface for an operator to control operations of AT2 DX scanner.
The ImageScope DX application is a software only subsystem to be used with the display monitor (Dell MR2416) and runs on a separate viewing workstation PC. The ImageScope DX software opens the WSI images acquired with Aperio AT2 DX scanner from the image storage attached to local network and uses the color profile to render the image data to the calibrated display monitor to deliver the image view at the appropriate magnification and color. Functionalities of ImageScope DX include navigating the slide similar to using an optical microscope, panning across the slide to examine all tissue regions, zooming in and out to allow review at different magnification levels, and allowing pathologist annotations during review.
The different subsystems of the Aperio AT2 DX System are connected over an IT network at the user site. The IT hardware/software that supports the ImageScope DX Application Server & Storage software is not provided as part of the Aperio AT2 DX System, but may be located in a central server room separate from the workstation with the ImageScope DX viewing software and Display. The communication of data between Aperio AT2 DX scanner and ImageScope DX application is via a customer provided wired network or a direct connected cable between these subsystems. Aperio AT2 DX System includes a display that has been validated as part of the pivotal clinical study.
3
{3}
The Aperio AT2 DX System allows pathologists to view and evaluate digital images of formalin-fixed, paraffin-embedded (FFPE) tissue slides that would otherwise be appropriate for manual visualization by conventional brightfield (light) microscopy. The Aperio AT2 DX System does not include any automated image analysis applications that would constitute computer aided detection or diagnosis.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
Philips IntelliSite Pathology Solution (PIPS)
2. Predicate 510(k) number(s):
DEN160056
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | The Aperio AT2 DX System is an automated digital slide creation and viewing system. The Aperio AT2 DX System is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. The Aperio AT2 DX System is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.The Aperio AT2 DX System is composed of the Aperio AT2 DX scanner, the ImageScope DX review application and Display. The Aperio AT2 DX System is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy. It is the responsibility of a qualified | The Philips IntelliSite Pathology Solution (PIPS) is an automated digital slide creation, viewing, and management system. The PIPS is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. The PIPS is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.The PIPS comprises the Image Management System (IMS), the Ultra Fast Scanner (UFS) and Display. The PIPS is for creation and viewing of digital images of scanned glass slides that would otherwise be |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using the Aperio AT2 DX System. | appropriate for manual visualization by conventional light microscopy. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using PIPS. |
| Specimen type | Surgical pathology slides prepared from FFPE tissue | Same |
| Principle of operation | After conducting Quality Control (QC) on the glass slides per laboratory standards (e.g., staining, cover-slipping, barcode placement, etc.), the technician loads the slides into the Aperio AT2 DX scanner. The scanner scans the slides and generates WSI image for each slide. The technician performs QC on scanned WSI images by checking image data and image quality. When QC is failed, the slide will be re-scanned. The acquired WSI images are stored in an end user provided image storage attached to the local network. During review, the pathologist opens WSI images acquired with the Aperio AT2 DX scanner from the image storage, performs further QC to ensure image quality and reads WSI images of the slides to make a diagnosis. | Same |
| Device Components | WSI scanner (Aperio AT2 DX scanner), Image Management System (ImageScope DX application), and color monitor display | Same (PIPS Ultra Fast Scanner, Image Management System and color monitor display) |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Whole Slide Imaging Scanner | Aperio AT2 DX scanner with loading capacity of 400 slides | Ultra Fast Scanner with loading capacity of 300 slides |
| Graphical User Interface | ImageScope DX | Image Management System |
| Monitor Display | Dell MR2416 | PS27QHDCR |
{5}
K. Standard/Guidance Document Referenced (if applicable):
Guidance for Industry “Technical Performance Assessment of Digital Pathology Whole Slide Imaging Devices”, dated April 20, 2016.
Special controls under 21 CFR Sec. 864.3700 Whole slide imaging system.
L. Test Principle:
The Aperio AT2 DX System is an automated system designed for scanning and digitizing surgical pathology slides prepared from FFPE tissue. These digitized images can then be reviewed and interpreted by pathologists for clinical (patient care) purposes.
Prior to scanning the slide on the Aperio AT2 DX scanner, the technician conducts quality control of the slides per the laboratory’s standards (e.g., staining, cover-slipping, barcode placement, etc.). The technician then places the slides into slide feeder (Autoloader) which is a component of the Aperio AT2 DX scanner as a batch. With AutoLoader, the scanner automatically mounts each slide on the scanning stage, generates a macro image that includes the slide label and a low magnification image of the entire slide. The system then locates specimen tissues within the slide scan area, which are subsequently scanned at high resolution (0.50 µm/pixel for 20x and 0.25 µm/pixel for 40x). After the slide is scanned, it is returned to the same slot of the same rack from which it was originally obtained.
The images scanned in the Aperio AT2 DX scanner are compressed and saved using proprietary SVS file format. The SVS image files are then transmitted to customer provided Medical Device Data System (MDDS). The SVS image files acquired on Aperio AT2 DX scanner can be reviewed through the supplied ImageScope DX review application only. ImageScope DX application allows users to open a case, review the WSI images of the slides (eSlides) by navigating around and viewing the WSI image at the desired magnifications. The pathologist is responsible for ensuring the validity of the interpretation of the digital images obtained from the Aperio AT2 DX System.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The objective of this study was to evaluate intra-system, inter-system/site, within-pathologist and between-pathologist precision for the Aperio AT2 DX System.
The precision of the device was based on five reading pathologists’ assessments and
{6}
identification of specific histopathologic "features" that are observed in FFPE Hematoxylin and Eosin (H&E) stained slides. Twenty-three (23) primary features were selected for the analytical studies. The selected primary features were evaluated at their relevant magnifications with twelve (12) primary features evaluated at 20x magnification level and eleven (11) primary features evaluated at 40x magnification level.
The Precision Panel Slide Curator (curator) identified study slides by conducting a Lab Information System (LIS) search and reviewing the microscope slides from consecutive cases that potentially contain the study features (Table 1). A representative H&E stained slide or a re-cut slide containing the feature(s) of interest was obtained from each case. After the curator selected and marked the study features on the slides, the slides were scanned at magnification levels designated to the study features, either $20\mathrm{x}$ or $40\mathrm{x}$ magnification. The curator reviewed the glass slides through the microscope while reviewing the WSI images on the monitor to extract Field of Views (FOVs) containing the study features from the WSI images using a lossless compression type. For each subsequent WSI images of the same slide, the annotations were manually replicated by a technician using the image viewing software to extract and randomly rotate the FOVs by 90, 180 or 270 degrees to minimize recall bias. A trained study personnel verified the FOV filenames and also that the FOV extraction areas were correct and enrolled the FOVs in the study. During precision study, the extracted FOV images were restricted for both the area of tissue visible to the reading pathologists and for the viewing magnification. The reading pathologists were not allowed to navigate the WSI image or change magnification or get access to any clinical information associated with the study slides. Additional slides that may or may not include any of the study features were included as "wild card" slides to reduce recall bias. The "wild card" FOVs were extracted by following the same procedures as the study FOVs.
As shown in Table 1, each primary study feature was represented in three (3) FFPE sections of different organ types. From each slide, two (2) or three (3) FOVs containing either one primary study feature, or a primary study feature and secondary study feature(s) in each FOV were selected. A total of 202 FOVs were selected from 69 glass slides. Out of 202 FOVs, 46 FOVs (from 24 slides) contained multiple histologic features and 156 FOVs (from 62 slides) contained one primary histologic feature. In addition, 36 wild card FOVs were selected from 12 glass slides.
Table 1: Primary Histologic Study Features in Precision Study
| Magnification Level | Primary Feature | Organ Type |
| --- | --- | --- |
| 20x | Chondrocytes | Toe |
| | | Femoral Head |
| | | Osteosarcoma of humerus |
| | Fat Cells (adipocytes) | Axillary lymph node |
| | | Femoral head |
| | | Prostate |
{7}
| | Foreign Body Giant Cells | Left knee synovium |
| --- | --- | --- |
| | | Shoulder |
| | | Sigmoid colon |
| | Goblet Cells in Intestinal Mucosa or Intestinal Metaplasia | Gastroesophageal junction |
| | | Sigmoid colon |
| | | Tubular adenoma (intestine) |
| | Granulomas | Colon |
| | | Iliac crest (bone) |
| | | Cervical lymph node |
| | Infiltrating or metastatic lobular carcinoma | Iliac crest (bone) |
| | | Jejunum |
| | | Left breast |
| | Intraglandular Necrosis | Left lung |
| | | Liver |
| | | Right colon |
| | Osteoclasts | Sacrum |
| | | Toe |
| | | Paget's disease of spine |
| | Osteocytes | Foot |
| | | Maxilla |
| | | Osteosarcoma of femur |
| | Pleomorphic nucleus of malignant cell | Left 11th rib |
| | | Sacrum |
| | | Vertebra |
| | Serrated intestinal epithelium (for example sessile serrated polyp) | Appendix |
| | | Ascending colon polyp |
| | | Sigmoid colon |
| | Skeletal Muscle fibers | Right lower leg |
| | | Shoulder |
| | | Spine |
| 40x | Asteroid Bodies | Axillary lymph node |
| | | Liver |
| | | Synovium |
| | Clear Cells (renal cell carcinoma) | Humerus |
| | | Retroperitoneal lymph node |
| | | Right kidney |
| | Foreign bodies (for example plant material or foreign debris) | Distal femur |
| | | Foot |
| | | Wrist |
| | Hemosiderin (pigment) | Knee synovium |
| | | Liver |
| | | Osteosarcoma of femur |
| | Megakaryocytes | Cervical Spine |
| | | Femur (margin of sarcoma) |
| | | Tibia |
{8}
| | Necrosis | Femoral head |
| --- | --- | --- |
| | | Left para-aortic lymph node |
| | | Right leg |
| | Nerve Cell Bodies (for example ganglion cells) | Ganglioneuroma |
| | | Small bowel |
| | | Stomach |
| | Nuclear Grooves | Cervical lymph node (papillary thyroid carcinoma) |
| | | Iliac crest (bone) (Langerhan's cell granuloma) |
| | | Ovary (Brenner tumor) |
| | Osteoid Matrix | Femur |
| | | Humerus |
| | | Lung |
| | Psammoma bodies | Cervical lymph node (metastatic papillary carcinoma of thyroid) |
| | | Fallopian tube (papillary ovarian carcinoma) |
| | | Left ventral cranial region (meningioma) |
| | Reed-Sternberg Cell | Axillary lymph node |
| | | Neck mass |
| | | Spleen |
A total of five reading pathologists reviewed the FOVs in precision study. Three pathologists reviewed for within- and between-pathologist precision, one pathologist reviewed for intra-system precision and one pathologist reviewed for intersystem/inter-site precision. For each magnification (20x or 40x), a separate checklist containing eleven (11) features (for 40x) or twelve (12) features (for 20x) was developed. For each FOV, the reading pathologist recorded the presence or absence of each of 11 features (for 40x) or 12 features (for 20x) on a checklist. Only the recorded results from study FOVs were used for statistical analysis. The results of wild card FOVs were not included in the statistical analysis.
# Intra-System Precision
The panel of 69 study slides was split equally among the three (3) sites (i.e., each site scanned a separate subset of slides). Each site has a single scanning system. For each system, 23 slides were scanned once on each of three (3) days, producing three (3) sets of scans for each slide. FOVs were extracted from scanned WSI images. The FOVs from three scanning sessions were evaluated by one reading pathologist over three different reading sessions with at least a two-week washout period in between the reading sessions. In each reading session, the reading pathologist evaluated all 202 study FOVs from each of three scanning sessions (across three scanning systems), plus unique "wild card" FOVs to assist in preventing recall bias between reading sessions.
{9}
For each system, agreements between scan 1 versus scan 2, scan 1 versus scan 3 and scan 2 versus scan 3 were analyzed. The overall intra-system precision was based on the pooled data of all three systems. A bootstrap approach was used to calculate 95% Confidence Intervals (CIs) with the following exception. When the agreement estimate was 100%, the Arcsine (variance stabilizing transformation) approach that corrected for the continuity was used to calculate CIs.
Table 2: Intra-System Precision Study Results
| System | Number of Pairwise Agreements | Number of Comparison Pairs | Agreement Rate and 95% CI | |
| --- | --- | --- | --- | --- |
| | | | % Agreement | 95% CI |
| System 1 | 193 | 201 | 96.0 | (91.0, 100) |
| System 2 | 201 | 201 | 100 | (98.2, 100) |
| System 3 | 199 | 204 | 97.5 | (93.6, 100) |
| Overall | 593 | 606 | 97.9 | (95.9, 99.5) |
# Inter-System/Inter-Site Precision
At each of three study sites, a study technician scanned the entire set of 69 study slides once. FOVs from three scanning sessions (resulting from one session from each of the three sites) were extracted. From all the scanning systems, 606 FOVs were transferred to a single reading pathologist for evaluation in three separate reading sessions. In each reading session, the reading pathologist evaluated all 202 study FOVs from a single site, plus unique "wild card" FOVs to assist in preventing recall bias between reading sessions.
Agreements between systems at Site 1 versus Site 2, systems at Site 1 versus Site 3 and systems at Site 2 versus Site 3 were analyzed. The overall inter-system/inter-site precision is based on the pooled data from each system to system comparison. A bootstrap approach was used to calculate $95\%$ CIs.
Table 3: Inter-System/Inter-Site Precision Study Results
| System | Number of Pairwise Agreements | Number of Comparison Pairs | Agreement Rate and 95% CI | |
| --- | --- | --- | --- | --- |
| | | | % Agreement | 95% CI |
| Site 1 vs. Site 2 | 195 | 202 | 96.5 | (94.1, 99.0) |
| Site 1 vs. Site 3 | 194 | 202 | 96.0 | (93.1, 98.5) |
| Site 2 vs. Site 3 | 193 | 202 | 95.5 | (92.6, 98.0) |
| Overall | 582 | 606 | 96.0 | (93.6, 98.2) |
{10}
Within-Pathologist and Between-Pathologist Precision
The entire set of 69 slides were scanned once at one site. FOVs were extracted and saved in three different orientations. From all the orientations, 606 FOVs were presented to each of three reading pathologists for evaluation in three reading sessions. In each reading session, each reading pathologist evaluated all 202 study FOVs plus unique "wild card" FOVs to assist in preventing recall bias between reading sessions.
For within-pathologist precision, agreements between FOVs in orientation 1 versus orientation 2, orientation 1 versus orientation 3, and orientation 2 versus orientation 3 were analyzed for each reading pathologist. The overall within-pathologist precision was based on the pooled data from each reading pathologist. A bootstrap approach was used to calculate $95\%$ CIs.
Table 4: Within-Pathologist Precision Study Results
| Pathologist | Number of Pairwise Agreements | Number of Comparison Pairs | Agreement Rate and 95% CI | |
| --- | --- | --- | --- | --- |
| | | | % Agreement | 95% CI |
| Pathologist 1 | 561 | 606 | 92.6 | (89.6, 95.7) |
| Pathologist 2 | 595 | 606 | 98.2 | (96.3, 99.7) |
| Pathologist 3 | 571 | 606 | 94.2 | (91.4, 96.9) |
| Overall | 1727 | 1818 | 95.0 | (92.9, 96.8) |
For between-pathologist precision, agreements between pathologist 1 versus pathologist 2, pathologist 1 versus pathologist 3, and pathologist 2 versus pathologist 3 were analyzed. The overall between-pathologist precision was based on the pooled data from each reading pathologist. A bootstrap approach was used to calculate $95\%$ CIs.
Table 5: Between-Pathologist Precision Study Results
| Pathologist Comparison | Number of Pairwise Agreements | Number of Comparison Pairs | Agreement Rate and 95% CI | |
| --- | --- | --- | --- | --- |
| | | | % Agreement | 95% CI |
| Pathologist 1 vs. Pathologist 2 | 572 | 606 | 94.4 | (91.6, 96.9) |
| Pathologist 1 vs. Pathologist 3 | 562 | 606 | 92.7 | (89.9, 95.4) |
| Pathologist 2 vs. Pathologist 3 | 579 | 606 | 95.5 | (93.1, 97.7) |
| Overall | 1713 | 1818 | 94.2 | (91.7, 96.4) |
b. Linearity/Assay reportable range:
{11}
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Not applicable
d. Detection limit:
Not applicable
e. Analytical specificity:
Not applicable
f. Assay cut-off:
Not applicable
2. Technical studies:
Multiple studies were conducted to evaluate the performance assessment data associated with the technical evaluation of the Aperio AT2 DX System.
a. Slide feeder
Information was provided on the configuration of the slide feed mechanism, including a physical description of the slide, the number of slides in queue (carrier), and the class of automation. Information was provided on the user interaction with the slide feeder, including hardware, software, feedback mechanisms, and Failure Mode and Effects Analysis (FMEA).
b. Light source
Descriptive information associated with the lamp and the condenser was provided. Testing information was provided to verify the spectral distribution of the light source as part of the color reproduction capability of the Aperio AT2 DX scanner.
c. Imaging optics
An optical schematic with all optical elements identified from slide (object plane) to digital image sensor (image plane) was provided. Descriptive information regarding the microscope objective, the auxiliary lenses, and the magnification of imaging optics was provided. Testing information regarding the relative irradiance, optical distortions, and lateral chromatics aberrations was provided.
12
{12}
d. Mechanical scanner movement
Information and specifications on the configuration of the stage, method of movement, control of movement of the stage, and FMEA was provided. Test data to verify the repeatability of the stage movement and to verify the mechanism that the stage movement stays within limits during operations was provided.
e. Digital imaging sensor
Information and specifications on the sensor type, pixel information, responsivity specifications, noise specifications, readout rate, and digital output format. Test data to determine the correct functioning of the digital image sensor that converts optical signals of the slide to digital signals which consist of a set of numerical values corresponding to the brightness and color at each point in the optical image was provided.
f. Image processing software
Information and specifications on the exposure control, white balance, color correction, sub-sampling, pixel-offset correction, pixel-gain or flat-field correction, and pixel-defect correction were provided.
g. Image composition
Information and specifications on the scanning method, the scanning speed, and the number of planes at the Z-axis to be digitized was provided. Test data to analyze the image composition performance was provided.
h. Image files format
Information and specifications on the compression method, compression ratio, file format, and file organization were provided.
i. Image review manipulation software
Information and specifications on continuous panning and pre-fetching, continuous zooming, discrete Z-axis displacement, ability to compare multiple slide simultaneously on multiple windows, image enhancement and sharpening functions, color manipulation, annotation tools, and digital bookmarks were provided.
j. Computer environment
Information and specifications on the computer hardware, operating system, graphics card, graphics card driver, color management settings, color profile, and display interface were provided.
{13}
k. Display
Information and specifications on the technological characteristics of the display device, physical size of the viewable area and aspect ratio, backlight type and properties, frame rate and refresh rate, pixel array, pitch, pixel aperture ratio and subpixel matrix scheme, subpixel driving to improve grayscale resolution, supported color spaces, display interface, user controls of brightness, contrast, gamma, color space, power-saving options, etc., via the on-screen display menu, ambient light adaptation, touchscreen technology, color calibration tools, and frequency and nature of quality-control tests was provided. Test data to verify the performance of the display was provided.
l. Color reproducibility
Test data to evaluate the color reproducibility of the system was provided.
m. Spatial resolution
Test data to evaluate the composite optical performance of all components in the image acquisition phase was provided.
n. Focusing test
Test data to evaluate the technical focus quality of the system was provided.
o. Whole slide tissue coverage
Test data to demonstrate that the entire tissue specimen on the glass slide is detected by the tissue detection algorithms, and that all the tissue specimens are included in the digital image file was provided.
p. Stitching error
Test data to evaluate the stitching errors and artifacts in the reconstructed image was provided.
q. Turnaround time
Test data to evaluate the turnaround time of the system was provided.
3. Human factors studies:
Human factors studies designed around critical user tasks and use scenarios performed by representative users for both histopathology technicians and pathologists were conducted. Information included a list of all critical user tasks and a description of the process that was followed to identify them. A systematic evaluation involving simulated use by
14
{14}
representative users performing all tasks (including critical tasks) required for operation of the device, and subjective assessment of failure was provided. All participants were able to perform all tasks (including the critical tasks) and no critical task failures were observed. There were some occasional difficulties that are generally expected with any piece of new software and instrument but learnability and ease of use appeared to be considerably high. All difficulties observed were of little influence on the perception of the usability, and no difficulties or failures were observed on tasks that could lead to patient harm. In all instances both pathologists and histopathology technicians were able to easily identify cases and ensure that everything was complete.
4. Clinical studies:
A multi-center study was conducted to demonstrate that viewing, reviewing, and diagnosing digital images of surgical pathology FFPE tissue slides using the Aperio AT2 DX System is non-inferior to using optical (light) microscopy. The primary endpoint was the difference in major discordance rates between WSI review modality (WSIR) and microscope slide review modality (MSR) when compared to the reference (main) diagnosis, which is based on the original sign-out pathologic diagnosis rendered at the institutions, using an optical (light) microscope.
Five sites were used in the study. At each site, a case curation pathologist or lab technician searched the site's LIS to identify candidate cases with a sign-out pathologist diagnosis. Candidate cases were identified consecutively, starting with cases at least one (1) year old (sign-out diagnosis was made at least one year before) and continuing with sequentially older cases. The case curation pathologist then checked whether the candidate case met the study eligibility criteria. The case curation pathologist reviewed the sign-out diagnosis and supporting documentation available at the time of the diagnosis (i.e., pathology report). By reviewing the microscopic slides used to make the sign-out diagnosis, the slide(s) that were representative of the sign-out diagnosis for the case were identified. The selected slides included H&E, immunohistochemistry (IHC), special stains and any slides that were not critical to diagnosis but supported the final surgical pathology report (e.g., margins, lymph node status, vascular and neural invasion), and any slides representing required elements of the College of American Pathologists (CAP) applicable cancer protocols. In the case of IHCs and special stains, the inclusion of control slides was required to fulfill the quality checks according to general clinical practice. The site's curation verification pathologist verified that the selected slide(s) reflected the main diagnosis for the case as well as required ancillary information for cancer cases, met all inclusion criteria and none of the exclusion criteria, and then the case was enrolled.
Across five sites, a total of 2045 cases (5849 slides) consisting of multiple organ and tissue types were enrolled. At each site (with exception of Site 1 that had three pathologists), four pathologists read all the cases assigned to the site using both MSR and WSIR modalities in an alternating fashion and randomized order and with a washout period of 31 days or longer between the MSR and WSIR reads, resulting in a total of 7781 planned WSIR reads and 7781 planned MSR reads. The reading pathologists were
15
{15}
provided with all representative slide(s) of a case at once. An electronic case report form (eCRF) was completed to document the reading pathologist's diagnosis. Three adjudicators reviewed the reader diagnoses against the corresponding sign-out diagnoses and determined whether the reader diagnoses were concordant, minor discordant or major discordant. A major discordance was defined as a difference in diagnosis that would be associated with a clinically important difference in patient management. A minor discordance was defined as a difference in diagnosis that would not be associated with a clinically important difference in patient management. For each case, two adjudication pathologists independently reviewed the eCRFs to determine whether or not the diagnosis was consistent with the sign-out diagnosis. In the event that there was a disagreement between the two adjudicators, a third adjudication pathologist reviewed the case to achieve majority vote. In cases where all three adjudicators had a different opinion, consensus was arrived at in an adjudication panel meeting consisting of the same three adjudication pathologists.
Inclusion criteria:
- All glass slides with human tissue obtained via surgical pathology of the original case were available.
- The original sign-out diagnosis and ancillary supporting information were available.
- The selected slide(s) for the main diagnosis and the control slide(s) (e.g., controls for IHC stained slides) matched the study requested organ type and fulfilled the quality checks according to general clinical practice.
- For cases with multiple diagnoses in a single sign-out pathology report, only the main diagnosis and ancillary information were included.
Exclusion criteria:
- The H&E stained slides that were used for the original sign-out diagnosis were not available at the site and re-cuts were not available.
- If applicable, the IHC or special stain slides that were used for the original sign-out diagnosis were not available at the site or had unacceptable artifacts. Control slides for IHC or special stain were not available.
- Cases for which slides needed to support the original sign-out diagnosis required either a special light source (e.g., a mercury lamp for fluorescence microscopy) or special filters (e.g., for polarized light).
- Cases with slides containing unremovable markings or were damaged.
- Cases that contained frozen sections or gross specimens only.
- Cases that were missing significant clinical and ancillary information that was available at the time the case was originally interpreted (e.g., X-rays).
- Cases that were signed out less than one year prior to the date of curation.
- Only one case could be enrolled per patient.
For the primary objective of demonstrating the WSIR major discordance rate to be non-inferior to the MSR major discordance rate, a Generalized Linear Mixed Model
16
{16}
(GLIMMIX) logistic regression was conducted. For each reading result the dependent variable was the major discordance status and independent variables included modality as a fixed effect (WSIR vs. MSR) and site, reader, and case as random effects. A two-sided $95\%$ CI for the modality effect, i.e., the overall major discordance rate difference (WSIR minus MSR), was constructed from this analysis. If the upper bound of the $95\%$ CI was less than the non-inferiority margin of $4\%$ , WSIR would be considered non-inferior to MSR.
A total of 2045 cases were included in the Full Analysis Set with 5849 slides and 7781 planned diagnoses for each modality. During the study, reading pathologists deferred cases by either modality per the site's standard procedure, mimicking the intended use. 172 cases were deferred in both modalities, 87 cases (1.1% of total planned reads) were deferred only in MSR modality and 100 cases (1.3% of total planned reads) were deferred only in WSIR modality. 7509 WSIR diagnoses and 7522 MSR diagnoses that were adjudicated by the adjudication panel and had consensus scores, were included in the statistical analyses. Deferred diagnoses were excluded from the data analysis as missing data.
The observed overall major discordance rate, i.e., over all sites, reading pathologists and organs, was $3.73\%$ for WSIR modality and $3.28\%$ for MSR modality. The WSIR-MSR difference in major discordance rate was $0.45\%$ . The major discordance rates as estimated by the GLIMMIX logistic model ("modeled") resulted in similar, slightly lower, proportions, i.e., $3.64\%$ for WSIR modality and $3.20\%$ for MSR modality. The WSIR-MSR difference in major discordance rate was $0.44\%$ , with derived two-sided $95\%$ CI of $[-0.15\%, 1.03\%]$ . The upper limit of this confidence interval was less than the pre-specified non-inferiority margin of $4\%$ , and therefore, the WSIR modality using the Aperio AT2 DX System was demonstrated to be non-inferior to the MSR modality using light microscopy with respect to major discordance rate with comparing to the main diagnosis. Thus, the study met the primary objective as shown in Table 6.
Table 6: Clinical Study Results Based on Major Discordance Rates
| | Whole Slide Imaging Review (WSIR) | | | Light Microscope Slide Review (MSR) | | | Difference (WSIR - MSR) | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Total Reads | % discordant | 95% CI | Total Reads | % discordant | 95% CI | % discordant | 95% CI |
| Observed | 7509 | 3.73 | - | 7522 | 3.28 | - | 0.45 | - |
| Model | | 3.64 | (3.21, 4.12) | | 3.20 | (2.80, 3.65) | 0.44 | (-0.15, 1.03) |
The observed differences in major discordance rates by organ types between WSIR and MSR are shown in Table 7.
Table 7: Major Discordance Rates by Organ
| Organ | Major Discordance Rate (%) | | % Difference in Major Discordance Rates (WSIR - MSR) |
| --- | --- | --- | --- |
| | Whole Slide Imaging Review (WSIR) | Light Microscope Slide Review (MSR) | |
| Observed | 7509 | 3.73 | 0.45 |
| Model | 3.64 | (3.21, 4.12) | 0.44 |
{17}
| Anus/perianal | 3.95 | 2.79 | 1.16 |
| --- | --- | --- | --- |
| Appendix | 0.00 | 0.00 | 0.00 |
| Bladder | 10.40 | 9.47 | 0.93 |
| Brain/neuro | 3.09 | 2.54 | 0.55 |
| Breast | 4.29 | 3.53 | 0.76 |
| Colorectal | 2.46 | 2.46 | 0.00 |
| Endocrine | 4.04 | 4.57 | -0.53 |
| Gastroesophageal junction | 3.69 | 3.16 | 0.54 |
| Gallbladder | 0.00 | 0.00 | 0.00 |
| Gynecological | 4.28 | 3.18 | 1.10 |
| Hernia/peritoneal | 0.00 | 0.00 | 0.00 |
| Kidney | 1.14 | 1.69 | -0.56 |
| Liver/bile duct | 1.59 | 0.53 | 1.06 |
| Lung | 5.24 | 3.68 | 1.55 |
| Lymph node | 1.09 | 1.87 | -0.78 |
| Prostate | 3.00 | 3.44 | -0.44 |
| Salivary gland | 0.55 | 1.69 | -1.14 |
| Skin | 4.74 | 2.72 | 2.03 |
| Soft tissue | 4.23 | 4.83 | -0.60 |
| Stomach | 3.15 | 2.09 | 1.06 |
Three (3) organ types, appendix, gallbladder and hernial/peritoneal, had major discordance rates of $0.00\%$ for both modalities. For all organ types, the absolute values of the observed differences in the major discordance rates between WSIR diagnosis and MSR diagnosis (relative to the reference diagnosis) were $\leq 2.03\%$ . When examining the major discordance rates, 'Bladder' had the highest major discordance rate of $10.40\%$ for WSIR and $9.47\%$ for MSR. However, statistical model showed that the modality-by-organ interaction was not statistically significant.
Secondary analysis of agreement between the WSIR and MSR diagnosis showed that there was also strong agreement between the two modalities. There were 7423 pairs of readings with both adjudication outcomes for WSIR and MSR. Results showed $96.1\%$ of the paired readings resulted in no major discordance for both modalities $(94.7\%)$ or major discordance for both modalities $(1.47\%)$ .
The modality-by-organ interaction was not statistically significant. The clinical study was not powered to analyze the results by individual organ site or diagnosis. The number of unbalanced diagnoses in which the WSIR modality had major discordance and the MSR modality had concordance with the reference diagnosis in each tissue/organ type category ranged from 0 cases for kidney to 15 cases in breast, corresponding to $0.0\%$ to $4.9\%$ of the total cases in each organ type group. Similarly, the number of unbalanced diagnoses in which the MSR modality had major discordance and the WSIR modality had concordance with the reference diagnosis in each tissue/organ type category ranged from 0 cases in skin to 13 cases in breast, corresponding to $0.0\%$ to $4.3\%$ of the total cases in each organ type group. Thus, unbalanced diagnoses represented a small percentage of the
{18}
overall cases and occurred with similar frequency for both modalities. When the specific diagnoses were examined by tissue/organ type, there were no apparent trends in any tissue/organ type that suggested WSIR is more inherently prone to major discrepancies compared to MSR. When examining this data by reading pathologist, none of the reading pathologists showed a tendency to have more types of specific major discrepancies with WSIR as compared to MSR. In every case the observed rates were within the known and established rate of inter-pathologist variation in diagnosis as reported in literature.
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Not applicable
N. Instrument Name:
Aperio AT2 DX System
O. System Descriptions:
1. Modes of Operation:
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ☐ X or No ☐
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ or No ☐ X
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
3. Specimen Identification:
Glass slides are identified by barcodes on slide label.
19
{19}
4. Specimen Sampling and Handling:
Glass slides are loaded manually into the slide feeder (Autoloader) which is a component of the Aperio AT2 DX scanner as a batch. With AutoLoader, the scanner automatically loads each slide in the batch to the stage and scans the glass slide to generate WSI image. After the scanning of all slides in the batch is completed, the slides are unloaded from the scanner and stored according to the lab procedure.
5. Calibration:
The Aperio AT2 DX scanner is calibrated and verified at the factory before shipment. No further manual calibration or verification is required by the customer. Annual preventive maintenance performed by Leica Biosystems Technical Services is recommended to ensure the scanner is operating correctly. The scanner provides automatic calibration and verification every time a slide is scanned by performing calibration during the scan.
The monitors provided with the Aperio AT2 DX System have been calibrated at the factory and may require periodic re-calibration in the field by authorized Leica Biosystems technicians. The user should not attempt to change the monitor settings as doing so may interfere with the monitor calibration.
The ImageScope DX application does not require calibration.
6. Quality Control:
It is the responsibility of the laboratory staff to conduct and maintain quality control of the slides per their laboratory standards (e.g., staining, cover-slipping, barcode placement) prior to loading the slides into the Aperio AT2 DX scanner. After completing a scan, the lab technician checks image data and image quality using the Aperio ConsoleDX application installed on the scanner workstation per the instructions for use. Before review, the pathologist performs quality control on the WSI images of the slide per instructions for use.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Not applicable
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 801 and 809, as applicable, and the special controls for this device type under 21 CFR 864.3700.
R. Conclusion:
{20}
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
21
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.