K193028 · Bayer Medical Care, Inc. · DXT · Dec 5, 2019 · Cardiovascular
Device Facts
Record ID
K193028
Device Name
MEDRAD Mark 7 Arterion Injection System
Applicant
Bayer Medical Care, Inc.
Product Code
DXT · Cardiovascular
Decision Date
Dec 5, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1650
Device Class
Class 2
Indications for Use
The MEDRAD® Mark 7 Arterion Injection System is indicated to be used specifically for the purposes of injecting contrast medium and common flushing solutions into humans for angiographic studies. The ISI option is indicated for the specific purpose of allowing a MEDRAD® Mark 7 Arterion Injection System to interface with an angiography imaging system.
Device Story
Software-controlled automated injection system; uses 150 ml disposable syringe to deliver contrast medium and flushing solutions. User programs volume, flow rate, and pressure limits via LCD touch screen. System features include syringe heat maintainer, multi-phase protocol memory, and manual/foot switch controls. Used in clinical settings for angiographic procedures. ISI module enables interface with angiography imaging systems to synchronize injection and imaging. Output is controlled delivery of contrast; assists clinicians in performing diagnostic angiographic studies.
Clinical Evidence
No clinical data. Bench testing from previous 510(k) submissions remains applicable; no new verification testing required for administrative update.
Technological Characteristics
Software-controlled angiographic injector; 150 ml disposable syringe; color LCD touch screen interface; syringe heat maintainer (35°C ± 5°C); pressure limit 100-1200 psi; flow rate 0.1-45.0 ml/sec. Connectivity includes Imaging System Interface (ISI) for angiography system integration. Class II device.
Indications for Use
Indicated for humans undergoing angiographic studies requiring injection of contrast medium and flushing solutions. Includes use of Imaging System Interface (ISI) for interfacing with angiography imaging systems.
Regulatory Classification
Special Controls
*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 5, 2019 Bayer Medical Care Inc. Kelly Frank Project Manager 1 Bayer Drive Indianola, Pennsylvania 15051
Re: K193028
Trade/Device Name: MEDRAD Mark 7 Arterion Injection System Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector And Syringe Regulatory Class: Class II Product Code: DXT Dated: November 4, 2019 Received: November 5, 2019
Dear Kelly Frank:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193028
Device Name MEDRAD Mark 7 Arterion Injection System
Indications for Use (Describe)
MEDRAD® Mark 7 Arterion Injection System
The MEDRAD® Mark 7 Arterion Injection System is indicated to be used specifically for the purposes of injecting contrast medium and common flushing solutions into humans for angiographic studies.
Imaging System Interface (ISI)
The ISI option is indicated for the specific purpose of allowing a MEDRAD® Mark 7 Arterion Injection System to interface with an angiography imaging system.
Type of Use (Select one or both, as applicable)
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the Bayer company logo. The logo consists of the word "BAYER" stacked vertically, with each letter in a bold, sans-serif font. The word is surrounded by a circular design, with the top half of the circle in green and the bottom half in blue. The logo is simple and recognizable, and it is often used to represent the Bayer company.
# 510(k) Summary
| Submitter: | Bayer Medical Care, Inc.<br>1 Bayer Drive<br>Indianola, PA 15051 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Kelly Frank<br>Project Manager<br>Phone: (412) 576-7583<br>Email: kelly.frank@bayer.com |
| Date Prepared: | October 24, 2019 |
| Device Trade Name: | MEDRAD® Mark 7 Arterion Injection<br>System |
| Common Name: | Angiographic Injector and Syringe |
| Classification Name: | Injector and Syringe, Angiographic [21 CFR<br>870.1650] |
| Product Code: | DXT |
| Classification: | Class II |
| Predicate Device: | The subject device is substantially<br>equivalent to the following device:<br>MEDRAD® Mark 7 Arterion Injection<br>System, K132928. |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the Bayer company logo. The logo consists of the word "BAYER" stacked vertically in a bold, sans-serif font. The letters are in a dark blue color. The wordmark is encircled by two curved lines, one in green and the other in blue, forming a ring around the name.
### Device Description:
The MEDRAD Mark 7 Arterion Iniector is a software-controlled medical device used to inject contrast agents from a 150 ml disposable syringe. Commonly referred to as an automated injection system, it is designed to allow a user to fill a disposable syringe and perform an injection with a user programmed volume and flow rate.
#### Indications for Use:
#### MEDRAD Mark 7 Arterion Injection System
The MEDRAD Mark 7 Arterion Injection System is indicated to be used specifically for the purposes of injecting contrast medium and common flushing solutions into humans for angiographic studies.
#### Imaging System Interface (ISI)
The ISI option is indicated for the specific purpose of allowing a MEDRAD Mark 7 Arterion Injection System to interface with an angiography imaging system.
#### Comparison to the Predicate Device:
The modification is an administrative update to include mention of the ISI (Imaging System Interface) module in the Indications for Use for the MEDRAD Mark 7 Arterion Injection System. This update does not involve any changes to the MEDRAD Mark 7 Arterion Injection System Syringe Kit. Table 1 provides a comparison of the proposed MEDRAD Mark 7 Arterion Injection System with the predicate device, the MEDRAD Mark 7 Arterion Injection System cleared per K132928.
The fundamental scientific technology, principle of operation, and intended use are unchanged from the predicate device, the MEDRAD Mark 7 Arterion Injection System cleared per K132928. The differences between the proposed device and the predicate device do not introduce new issues of safety and effectiveness.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the Bayer company logo. The logo consists of the word "BAYER" stacked vertically in a bold, sans-serif font. The letters are a dark blue color. The wordmark is encircled by a ring that is green on the top half and blue on the bottom half.
Table 1. Comparison of MEDRAD Mark 7 Arterion Injection System (proposed device) with predicate device.
| PARAMETER | PREDICATE<br>DEVICE | PROPOSED<br>DEVICE | RATIONALE FOR<br>CHANGE |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| | Mark 7 Arterion<br>Injection System<br>(K132928) | Mark 7 Arterion<br>Injection System | |
| Indications for<br>Use | The MEDRAD Mark 7<br>Arterion Injection<br>System is intended to<br>be used specifically<br>for the purposes of<br>injecting contrast<br>medium and common<br>flushing solutions into<br>humans for<br>angiographic studies. | The MEDRAD Mark<br>7 Arterion Injection<br>System is indicated<br>to be used<br>specifically for the<br>purposes of injecting<br>contrast medium and<br>common flushing<br>solutions into<br>humans for<br>angiographic studies.<br>The ISI option is<br>indicated for the<br>specific purpose of<br>allowing a MEDRAD<br>Mark 7 Arterion<br>Injection System to<br>interface with an<br>angiography imaging<br>system. | Administrative<br>update to add ISI<br>module specific<br>indications for use<br>statement |
| Fill Volume | 1 – 150 ml in 1 ml<br>increments | Same | No change |
| Fill Speed (user) | 1 — 10 ml/sec | Same | No change |
| PARAMETER | PREDICATE<br>DEVICE<br>Mark 7 Arterion<br>Injection System<br>(K132928) | PROPOSED<br>DEVICE<br>Mark 7 Arterion<br>Injection System | RATIONALE FOR<br>CHANGE |
| configurable for<br>AutoFill) | | | |
| Fill Speed<br>(manual control<br>by user) | 1 - 20 ml/sec | Same | No change |
| Fixed Flow Rate | 0.1 to 45.0 ml/sec in<br>0.1 ml increments<br>(Single and Phased<br>protocols)<br>0.1 to 59.9 ml/min in<br>0.1 ml/min<br>increments (ml/m<br>protocol) | Same | No change |
| Variable Flow<br>Rate | 1.0 – 10.0 ml/sec in<br>0.1 ml/sec<br>increments | Same | No change |
| Flow Rate Rise<br>Time | 0.0 to 9.9 seconds in<br>0.1sec increments | Same | No change |
| Delay | 0.0 to 9.9 seconds in<br>0.1sec increments | Same | No change |
| Pressure Limit<br>(150 ml syringe) | 100-1200 psi or 689-<br>8273 kPa in<br>increments of 1 psi<br>(kPa) | Same | No change |
| Syringe Heat<br>Maintainer | 35°C + 5°C | Same | No change |
| Protocol Memory | 40 protocols | Same | No change |
| Injection History | 50 injections | Same | No change |
| PARAMETER | PREDICATE<br>DEVICE<br>Mark 7 Arterion<br>Injection System<br>(K132928) | PROPOSED<br>DEVICE<br>Mark 7 Arterion<br>Injection System | RATIONALE FOR<br>CHANGE |
| Memory | | | |
| Information<br>Display | Color LCD | Same | No change |
| Programming<br>Keys | Software-generated<br>via an LCD touch<br>screen | Same | No change |
| Touch Screen | Yes | Same | No change |
| Multi-Phase | 4 phases per protocol | Same | No change |
| Arming Modes | Single and Multi-<br>arming modes | Same | No change |
| Syringe System | Single (150 ml)<br>syringe | Same | No change |
| Manual Retract<br>Control | Yes | Same | No change |
| Check for Air<br>Control | Yes | Same | No change |
| Variable Flow<br>Option | Yes | Same | No change |
| Hand Switch<br>(start switch) | Yes | Same | No change |
| Foot Switch<br>(start switch) | Yes | Same | No change |
| Variable Flow<br>Hand Controller | Yes | Same | No change |
| Splash Guard | Yes | Same | No change |
| Wiper Seal | Optional | Same | No change |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the Bayer company logo. The logo consists of the word "BAYER" stacked vertically in a bold, sans-serif font. The letters are arranged in the center of a circle that is divided into two halves. The top half of the circle is green, and the bottom half is blue.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the Bayer company logo. The logo consists of the word "BAYER" stacked vertically in a bold, sans-serif font. The letters are arranged in the center of a circle that is divided into two halves. The top half of the circle is green, and the bottom half is blue.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the Bayer company logo. The logo consists of the word "BAYER" stacked vertically, with each letter in a bold, sans-serif font. The word is enclosed in a circular shape, with the top half of the circle being green and the bottom half being blue. The logo is simple and modern, and the colors are bright and eye-catching.
# Performance Data:
Verification testing was not necessary to demonstrate that the modified indication for use remains safe and effective. Performance testing performed with MEDRAD Mark 7 Arterion Injection System which was provided in previous 510(k)s remains applicable.
No performance standards have been established under Section 514 of the Food, Drug and Cosmetic Act for angiographic injectors and syringes.
# Clinical Testing:
No clinical testing was required or performed to support this Special 510(k) Premarket Notification.
## Conclusion:
Bayer Medical Care Inc. considers the MEDRAD Mark 7 Arterion Injection System to be substantially equivalent to the predicate device listed above. The fundamental scientific technology, principle of operation, and intended use are unchanged from the predicate device. The differences between the proposed device and the predicate device do not introduce new issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.