MEDRAD MARK 7 ARTERION INJECTION SYSTEM / SYRINGE, TWIST & GO SYRINGE
K132928 · Medrad, Inc. / Bayer Medical Care, Inc. · DXT · Jan 23, 2014 · Cardiovascular
Device Facts
Record ID
K132928
Device Name
MEDRAD MARK 7 ARTERION INJECTION SYSTEM / SYRINGE, TWIST & GO SYRINGE
Applicant
Medrad, Inc. / Bayer Medical Care, Inc.
Product Code
DXT · Cardiovascular
Decision Date
Jan 23, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1650
Device Class
Class 2
Indications for Use
The MEDRAD Mark 7 Arterion Injection System is intended to be used specifically for the purposes of injecting contrast medium and common flushing solutions into humans for angiographic studies. The MEDRAD Mark 7 Arterion Syringe, Quick Fill Tube and other MEDRAD supplied disposables are specifically intended for single use only with the MEDRAD Mark 7 Arterion Injection System for angiographic studies. The MEDRAD Twist & Go Syringe, Quick Fill Tube and other MEDRAD supplied disposables are specifically intended for single use only with the MEDRAD Mark 7 Arterion Injection System for angiographic studies.
Device Story
Automated injection system for contrast media and flushing solutions in angiographic studies; comprises software-controlled injector head, 150 ml disposable syringe, and optional VFlow hand controller. Operator programs volume and flow rate via LCD touch screen. VFlow hand controller allows precise, instantaneous variable flow rate adjustment during injection or puffing; enables clinician to step away from radiation source. Syringes are front-loading, polycarbonate for improved air visualization; Twist & Go variant allows single-handed tubing attachment. System provides automated delivery, enhancing procedural control and clinician safety.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench and laboratory testing, including performance, safety, EMC, package integrity, environmental, and pressure/burst testing.
Technological Characteristics
Software-controlled angiographic injector; 150 ml polycarbonate syringes with silicone coating; polyisoprene plunger cover. Variable flow rate (1.0-10.0 ml/sec) and fixed flow rate (0.1-45.0 ml/sec) modes. Connectivity via LCD touch screen. Sterilization via Gamma radiation (syringes). Pressure limit up to 1200 psi.
Indications for Use
Indicated for use in humans undergoing angiographic studies requiring the injection of contrast media and flushing solutions.
Regulatory Classification
Special Controls
*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the Bayer company logo. The logo consists of the word "BAYER" arranged in a cross shape, with the letters "B", "A", "Y", "E", and "R" forming the arms of the cross. The entire logo is enclosed within a double-lined circle.
MEDRAD, INC.
(412) 767-2400
www.ri.bayer.com
U.S.A.
One Medrad Drive
Indianola, PA 15051
K132928
JAN 23 2014
# 510(k) Summary
#### Submitter:
MEDRAD, Inc. / Bayer Medical Care, Inc. One Medrad Drive Indianola, PA 15051
Contact Person:
Lisa A. Ewing Principal Regulatory Affairs Specialist Phone: (412) 406-3780 Fax: (412) 406-4052 Email: lisa.ewing@bayer.com
Date Prepared:
Device Trade Name:
Common Name:
Classification Name:
Product Code:
Classification:
Predicate Device(s):
September 16, 2013
MEDRAD Mark 7 Arterion Injection System MEDRAD Mark 7 Arterion Syringe MEDRAD Twist & Go Syringe
Angiographic Injector and Syringe
Injector and Syringe, Angiographic [21 CFR 870.1650]
DXT
Class II
The subject devices are equivalent to the following devices: MEDRAD Mark 7 Arterion Injection System K113133, December 15, 2011 O
K112086, October 14, 2011 o
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the Bayer company logo. The logo consists of the word "BAYER" stacked vertically inside of a circle. The letters are bolded and in a sans-serif font. The circle has a double-line border.
#### Device Description:
### MEDRAD Mark 7 Arterion Injector
The MEDRAD Mark 7 Arterion Injector is a software-controlled medical device used to inject contrast agents from a 150 ml disposable syringe. Commonly referred to as an automated injection system, it is designed to allow a user to fill a disposable syringe and perform an injection with a user-programmed volume and flow rate.
#### MEDRAD VFlow Hand Controller
For the Mark 7 Arterion Injector System, the MEDRAD VFlow Hand Controller is provided sterile for single patient use only. The hand controller provides the operator the ability to precisely control and instantaneously adjust variable flow rates of contrast during contrast injection and puffing operations.
When the injector is in the variable rate injection mode, the flow rate increases incrementally as the hand controller plunger is depressed, and decreases as the hand controller is released. In the fixed rate injection mode, the hand controller acts as a start switch, and release of the device ceases all flow. The VFlow Hand Controller requires the same amount of force to depress the plunger regardless of the set flow rate, volume, and pressure limit or the viscosity of the delivered contrast media. Additionally, the VFlow Hand Controller allows the clinician to step away from the radiation source during an injection.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the Bayer company logo. The logo consists of the word "BAYER" written twice, once horizontally and once vertically, intersecting at the "Y" to form a cross shape. The logo is enclosed within a circle.
### Mark 7 Arterion Syringe and Twist & Go Syringe
The Twist & Go Syringe function is the same as the Mark 7 Arterion Syringe. The syringes are both supplied sterile and are designed to be loaded into the automated injector head from the front of the injector (front loading). The clear polycarbonate material used in both syringes improves visualization of residual air compared to opaque polypropylene. The Twist & Go Syringe enables single-handed attachment of the Twist & Go High Pressure Connector Tubing to the syringe.
The MEDRAD Mark 7 Arterion Injection System is intended to be used specifically for the purposes of injecting contrast medium and common flushing solutions into humans for angiographic studies.
The MEDRAD Mark 7 Arterion Syringe, Quick Fill Tube and other MEDRAD supplied disposables are specifically intended for single use only with the MEDRAD Mark 7 Arterion Injection System for angiographic studies.
The MEDRAD Twist & Go Syringe, Quick Fill Tube and other MEDRAD supplied disposables are specifically intended for single use only with the MEDRAD Mark 7 Arterion Injection System for angiographic studies.
#### Intended Use:
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the Bayer company logo. The logo consists of the word "BAYER" arranged vertically in a circle. The letters are stacked on top of each other, with the "A" in the center. The logo is black and white.
### Comparison to Predicate:
Design changes were made to the predicate device which included: updating injector software to include a new variable flow rate option, adding a variable flow rate ("VFIow") hand controller and adding a new configuration of the Arterion Syringe, which is called the Twist and Go Syringe.
#### Performance Data:
Bench and laboratory testing were performed to support a determination of substantial equivalence to the predicate devices. Results from the testing provide assurance that the proposed devices conform to the requirements for their intended use. This included the following testing:
- . Pérformance
- Safety .
- EMC .
- . Package Integrity
- Environmental .
- . Pressure, burst, wet friction testing (syringe)
- System level testing with a MEDRAD . automated injector (syringe)
{4}------------------------------------------------
## Comparison Tables
.
Table 1. Comparison of Features in MEDRAD Mark 7 Arterion Injection System (predicate) and the modified MEDRAD Mark 7 Arterion Injection System (proposed).
| Specification /<br>Feature | Predicate Device:<br>MEDRAD Mark 7<br>Arterion Injection<br>System (K113133) | Proposed Device:<br>MEDRAD Mark 7<br>Arterion Injection<br>System | Rationale for<br>Change |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use /<br>Intended Use | The MEDRAD Mark<br>7 Arterion Injection<br>System is intended<br>to be used<br>specifically for the<br>purposes of injecting<br>contrast medium and<br>common flushing<br>solutions into<br>humans for<br>angiographic studies. | The MEDRAD Mark<br>7 Arterion Injection<br>System is intended<br>to be used<br>specifically for the<br>purposes of injecting<br>contrast medium and<br>common flushing<br>solutions into<br>humans for<br>angiographic studies. | No change. |
| Fill Volume | 1 - 150 ml in 1 ml<br>increments | 1 - 150 ml in 1 ml<br>increments | No change. |
| Fill Speed (user<br>configurable for<br>AutoFill) | 1 - 10 ml/sec | 1 - 10 ml/sec | No change. |
| Fill Speed (manual<br>control by user) | 1 - 20 ml/sec | 1 - 20 ml/sec | No change. |
| Fixed Flow Rate | 0.1 to 45.0 ml/sec in<br>0.1 ml increments<br>(single and phased<br>protocols)<br>0.1 to 59.9 ml/min in<br>0.1 ml/min<br>increments (single<br>ml/min protocol) | 0.1 to 45.0 ml/sec in<br>0.1 ml increments<br>(Single and Phased<br>protocols)<br>0.1 to 59.9 ml/min in<br>0.1 ml/min<br>increments (ml/m<br>protocol) | No change. |
| Variable Flow Rate | N/A - not available | Variable: 1.0 - 10.0<br>ml/sec in 0.1 ml/sec<br>increments | Variable Flow Rate<br>is being added as an<br>additional, optional<br>feature to inject a<br>set volume of<br>contrast at a flow<br>rate determined by<br>the hand controller. |
| Flow Rate Rise Time | 0.0 to 9.9 seconds in<br>0.1sec increments | 0.0 to 9.9 seconds in<br>0.1 sec increments | No change. |
| Delay Time | 0.0-99.9 seconds in<br>0.1 sec increments | 0.0-99.9 seconds in<br>0.1 sec increments | No change. |
| Pressure Limit (150<br>ml syringe) | 100-1200 psi or 689-<br>8273 kPa in<br>increments of 1 psi<br>(kPa) | 100-1200 psi or 689-<br>8273 kPa in<br>increments of 1 psi<br>(kPa) | No change. |
| | | | |
| Syringe Heat<br>Maintainer | $35°C ± 5°C$ | $35°C ± 5°C$ | No change. |
| Protocol Memory | 40 Protocols | 40 Protocols | No change. |
| Injection History<br>Memory | 50 Injections | 50 Injections | No change. |
| Information Display | Color LCD | Color LCD | No change. |
| Programming Keys | Software generated<br>via an LCD touch<br>screen | Software generated<br>via an LCD touch<br>screen | No change. |
| Touch Screen | Yes | Yes | No change. |
| Multi-Phase | 4 phases per<br>protocol | 4 phases per<br>protocol | No change. |
| Arming Modes | Single and Multi<br>arming modes | Single and Multi<br>arming modes | No change. |
| Syringe System | Single (150 ml)<br>syringe | Single (150 ml)<br>syringe | No change. |
| Manual Retract<br>Control | Yes | Yes | No change. |
| Check for Air Control | Yes | Yes | No change. |
| Variable Flow Option | No | Yes | Variable Flow Rate<br>is being offered as<br>an additional,<br>optional feature to<br>inject a set volume<br>of contrast at a flow<br>rate determined by<br>the hand controller. |
| Hand Switch<br>(start switch) | Yes | Yes | No change. |
| Foot Switch<br>(start switch) | Yes | Yes | No change. |
| Variable Flow Hand<br>Controller | No | Yes | Hand controller is<br>being offered as an<br>additional, optional<br>accessory for use<br>with the new<br>variable flow rate<br>software feature. |
| Splash Guard | Yes | Yes | No change. |
| Wiper Seal | Yes | Optional | Wiper seal is a<br>redundant feature<br>and is not a required |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the Bayer company logo. The logo consists of the word "BAYER" arranged vertically in a circle. The letters are in a bold, sans-serif font and are black. The circle is also black and has a thin outline.
.
:
·
{6}------------------------------------------------
.
100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000
Image /page/6/Picture/1 description: The image shows the Bayer company logo. The logo consists of the word "BAYER" arranged vertically inside of a circle. The letters are in a bold, sans-serif font and are black in color. The circle is also black and has a slightly textured appearance.
Table 2. Comparison of Features in MEDRAD Mark 7 Arterion Syringe (predicate), MEDRAD Mark 7 Arterion Syringe (proposed) and the MEDRAD Twist & Go Syringe (proposed).
. . . .
| | Component / Feature | Predicate Device:<br>MEDRAD Mark 7<br>Arterion Syringe<br>(K112086, K113133) | Proposed Device:<br>MEDRAD Mark 7<br>Arterion Syringe | Proposed Device:<br>MEDRAD Twist &<br>Go Syringe | Rationale for Change |
|----------|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Labeling | Intended Use | The MEDRAD Mark 7<br>Arterion Syringe, Quick<br>Fill Tube and other<br>MEDRAD supplied<br>disposables are<br>specifically intended for<br>single use only with the<br>MEDRAD Mark 7<br>Arterion Injection<br>System in the x-ray<br>angiography<br>environment. | The MEDRAD Mark 7<br>Arterion Syringe,<br>Quick Fill Tube and<br>other MEDRAD<br>supplied disposables<br>are specifically<br>intended for single use<br>only with the<br>MEDRAD Mark 7<br>Arterion Injection<br>System for<br>angiographic studies. | The MEDRAD Twist<br>& Go Syringe, Quick<br>Fill Tube and other<br>MEDRAD supplied<br>disposables are<br>specifically intended<br>for single use only<br>with the MEDRAD<br>Mark 7 Arterion<br>Injection System for<br>angiographic studies. | No change in intended<br>use, only updated syringe<br>product name. |
| | Syringe Volume | 150 mL | 150 mL | 150 mL | No change. |
| | Syringe Length | 8.61 inches | 8.61 inches | 8.61 inches | No change. |
| | Barrel OD | 1.81 inches | 1.81 inches | 1.81 inches | No change. |
| | Barrel ID | 1.60 inches | 1.60 inches | 1.60 inches | No change. |
| | FasTurn Nut at<br>Syringe Tip | Yes | Yes | No | FasTurn Nut is not<br>required for Twist & Go<br>connection |
| | Construction | Dust Cover | Yes (sized to fit syringe<br>tip with FasTurn nut) | Yes (sized to fit<br>syringe tip with<br>FasTurn nut) | Yes (sized to fit<br>syringe tip without<br>FasTurn nut) |
| | | Tubing Set<br>Compatibility | Connects to female luer<br>on high pressure<br>connector tubing | Connects to female<br>luer on high pressure<br>connector tubing | Connects to Twist &<br>Go luer connector on<br>the MEDRAD Twist &<br>Go High Pressure<br>Connector Tubing |
| | | Detents | Yes, molded in to<br>syringe tip | Yes, molded in to<br>syringe tip | No |
:
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the Bayer company logo. The logo consists of the word "BAYER" arranged vertically inside a circle. The letters are stacked on top of each other, with the "A" in the middle being slightly larger than the other letters.
| Materials | Syringe Barrel | Polycarbonate | Polycarbonate | Polycarbonate | No change. |
|------------|------------------------------------|----------------------------------------|----------------------------------------|----------------------------------------|-------------------------------------------------------------|
| | Syringe Barrel ID.<br>Coating | Silicone | Silicone | Silicone | No change. |
| | Plunger | Polycarbonate | Polycarbonate | Polycarbonate | No change. |
| | Plunger Rubber Cover | Polyisoprene | Polyisoprene | Polyisoprene | No change. |
| | Plunger Coating | Silicone | Silicone | Silicone | No change. |
| | FasTurn Nut | Polycarbonate | Polycarbonate | N/A | FasTurn Nut is not<br>required for Twist & Go<br>connection |
| | | | | | |
| Packaging | Dust Cover | Polypropylene | Polypropylene | Polypropylene | No change. |
| | Type | Tyvek lid covering<br>polystyrene tray | Tyvek lid covering<br>polystyrene tray | Tyvek lid covering<br>polystyrene tray | No change. |
| | | | | | |
| Biological | Sterilization | E-beam radiation | Gamma radiation | Gamma radiation | Updated mode of<br>sterilization; SAL is<br>unchanged. |
| | Sterility Assurance<br>Level (SAL) | 10-6 | 10-6 | 10-6 | No change. |
| | Shelf Life | 2 years | 2 years | 2 years | No change. |
| | Pyrogenicity | Non-Pyrogenic Fluid<br>Path | Non-Pyrogenic Fluid<br>Path | Non-Pyrogenic Fluid<br>Path | No change. |
| | Latex content | Not made with natural<br>rubber latex | Not made with natural<br>rubber latex | Not made with<br>natural rubber latex | No change. |
| | Performance | Pressure rating | 1200 psi | 1200 psi | 1200 psi |
| | | | | | |
Conclusion:
MEDRAD considers the MEDRAD Mark 7 Arterion Injection System, Mark 7 Arterion Syringe and Twist & Go Syringe to be substantially equivalent to the predicate devices listed above. This conclusion is based upon the devices' similarities in functional design, materials, indications for use, fundamental scientific technology, and principles of operation.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all caps. The text is in a bold, sans-serif font. The words are arranged on a single line and are centered horizontally.
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 23, 2014
MEDRAD, Inc. C/O Ms. Lisa A. Ewing Principal Regulatory Affairs Specialist One Medrad Drive Indianola, PA 1505 I
Re: K132928
Trade/Device Name: MEDRAD Mark 7 Arterion Injection System Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector and Syringe Regulatory Class: II Product Code: DXT Dated: December 23, 2013 Received: December 24, 2013
Dear Ms. Ewing:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{9}------------------------------------------------
Page 2 - Ms. Lisa A. Ewing
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note
the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
## Indications for Use Statement
510(k) Number:
K132928
Device Name:
MEDRAD Mark 7 Arterion Injection System MEDRAD Mark 7 Arterion Syringe MEDRAD Twist & Go Syringe
#### Indications for Use:
The MEDRAD Mark 7 Arterion Injection System is intended to be used specifically for the purposes of injecting contrast medium and common flushing solutions into humans for angiographic studies.
The MEDRAD Mark 7 Arterion Syringe, Quick Fill Tube and other MEDRAD supplied disposables are specifically intended for single use only with the MEDRAD Mark 7 Arterion Injection System for angiographic studies.
The MEDRAD Twist & Go Syringe, Quick Fill Tube and other MEDRAD supplied disposables are specifically intended for single use only with the MEDRAD Mark 7 Arterion Injection System for angiographic studies.
Prescription Use 2 (21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M.A. Hillerman
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.