K193022 · Jiangsu Caina Medical Co.,Ltd · MEG · Jul 20, 2020 · General Hospital
Device Facts
Record ID
K193022
Device Name
Retractable Safety Insulin Syringe
Applicant
Jiangsu Caina Medical Co.,Ltd
Product Code
MEG · General Hospital
Decision Date
Jul 20, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Retractable Safety Insulin Syringe is a sterile, single use, disposable and non-reusable, automatically retractable safety insulin syringe intended for injection of U-100 insulin into the body, while reducing the risk of sharps injuries and the potential for insulin syringe reuse.
Device Story
Retractable Safety Insulin Syringe is a sterile, single-use, disposable syringe for U-100 insulin delivery. Device comprises plunger, piston, barrel, needle cap, sealing plug, slide bushing, needle tube, spring, plunger lid, barrel barb, and hub. Operation mimics conventional syringes; however, after injection, user pushes plunger to bottom, engaging barrel barb and convex point to prevent reuse. This action simultaneously triggers spring-loaded retraction, pulling contaminated needle into plunger for sharps injury protection. Used in home or hospital settings by patients or clinicians. Benefits include accurate insulin dosing and reduced risk of accidental needle sticks and syringe reuse.
Clinical Evidence
No clinical study included. Bench testing only. Device performance verified against ISO 7864, ISO 9626, ISO 8537, and ISO 23908. Biocompatibility testing performed per ISO 10993 series (cytotoxicity, irritation, sensitization, systemic toxicity) and USP <151>/<85> (pyrogen/endotoxin). Mechanical testing confirmed activation force (10.48-20.91N) and plunger separation force (>70N).
Indicated for injection of U-100 insulin into the body for patients requiring insulin therapy. Intended for use in home or hospital settings.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" stacked on top of each other.
July 20, 2020
Jiangsu Caina Medical Co.,Ltd. Xinyan Ruan Quality Engineer No.23, Huanxi Road, Zhutang Town Jiangyin, 214415 Cn
Re: K193022
Trade/Device Name: Retractable Safety Insulin Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: MEG Dated: June 18, 2020 Received: June 22, 2020
## Dear Xinyan Ruan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name
Indications for Use (Describe)
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Tab 2 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K193022
- 1. Date of Preparation: 04/29/2020
- 2. Sponsor Identification
Jiangsu Caina Medical Co., Ltd. No.23, Huanxi Road, Zhutang Town, Jiangyin, Jiangsu, 214415, China
Establishment Registration Number: 3005670221
Contact Person: Jun Lu Position: General Manager Tel: +86-510-86205183 Fax: +86-510-86215183 Email: jun.lu@cainamed.com
- 3. Designated Submission Correspondent
Mr. Jianwei pan (Primary Contact Person) Email: jianwei.pan@cainamed.com Ms. Xinyan Ruan (Alternative Contact Person) Email: sherry.ruan@cainamed.com Tel: +86-510-86866666-8027 Fax: +86-510-86866666-8009
- 4. Identification of Proposed Device
Trade Name: Retractable Safety Insulin Syringe
Regulatory Information Classification Name: Syringe, Anitstick Classification: II Product Code: MEG Regulation Number: 21 CFR 880.5860 Review Panel: General Hospital
{4}------------------------------------------------
Indications for Use Statement:
The Retractable Safety Insulin Syringe is a sterile, single use, disposable and non-reusable, automatically retractable safety insulin syringe intended for injection of U-100 insulin into the body, while reducing the risk of sharps injuries and the potential for insulin syringe reuse.
#### ર. Device Description
The Retractable Safety Insulin Syringe has the same technological characteristics as the Retractable Safety Syringe(manufacture by Jiangsu Caina Medical Co., Ltd.) as cleared in K191490 at date 2019/10/03. The difference between the Retractable Safety Insulin Syringe and the 1ml Retractable Safety Syringe is only the scale.
The Retractable Safety Insulin Syringe works like a conventional insulin syringe except for its ability to retract the contaminated needle inside of the syringe immediately after patient injection. Needle retraction is activated by the syringe user. Because the contaminated needle is automatically withdrawn into the syringe plunge, the syringe user is protected from accidental needle sticks. These accidental needle sticks would occur between removing the needle from the patient and disposing of the syringe in a sharps disposable container.
The Retractable Safety Insulin Syringe is available in various specifications. The proposed device consists of eleven components: (1) plunger. (2) piston. (3) barrel. (4) needle cap. (5) sealing plug. (6) slide bushing, (7) needle tube, (8) spring, (9) plunger lid, (10) barrel barb, (11) hub.
The proposed device is provided sterile. The product is sterilized by Ethylene Oxide Gas to achieve a SAL of 10-6 and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of five years.
- న్. Identification of Predicate Device
Predicate Device 510(k) Number: K152808 Product Name: U&U Insulin Syringe with/without Safety Retractable Device
#### 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- A ISO 10993-7:2008 Biological evaluation of medical devices- Part 7: Ethylene Oxide Sterilization
{5}------------------------------------------------
Residuals
- > ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Test for in vitro cytotoxicity
- ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and A delayed-type hypersensitivity
- > ISO 10993-11:2017 Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity
- > USP <151> Pyrogen Test
- > ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials
- > ISO 10993-4:2017 Biological evaluation of medical devices -- Part 4: Selection of tests for interactions with blood
- > ASTM F1929-15 Standard test method for detecting seal leaks in porous medical packaging by dye penetration.
- > ASTM F88/F88M-15 standard method for seal strength of flexible barrier materials
- A ASTM F1886 / F1886M-16, Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
- > USP <85> Bacterial Endotoxins Test
- > ISO 7864:2016 Sterile hypodermic needles for single use-requirements and test method.
- A ISO 9626:2016 Stainless steel needle tubing for the manufacture of the medical devices-requirements and test method
- > ISO 8537:2016 Sterile Single-use syringes, with or without needle ,for insulin
- > ISO 23908:2011 Sharps injury protection -- Requirements and test methods -- Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling
The structure principle, safety function and operation method of the proposed devices are identical with Retractable Safety Syringes(manufacture by Jiangsu Caina Medical Co., Ltd.) as cleared in K191490 at date 2019/10/03.Simulated clinical study was not conducted on proposed device.
- 7. Clinical Test Conclusion
No clinical study is included in this submission.
- 8. Substantially Equivalent (SE) Comparison
| ITEM | Proposed Device | Predicate Device<br>K152808 | Comment |
|-------------------|-----------------|-----------------------------|---------|
| Product code | MEG | MEG and FMF | Same |
| Regulation<br>No. | 21 CFR 880.5860 | 21 CFR 880.5860 | Same |
Table 1 Comparison of Technology Characteristics with K152808
{6}------------------------------------------------
| Class | II | II | Same |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for use | The Retractable Safety<br>Insulin Syringe is a sterile,<br>single use, disposable and<br>non-reusable, automatically<br>retractable safety insulin<br>syringe intended for<br>injection of U-100 insulin<br>into the body, while reducing<br>the risk of sharps injuries<br>and the potential for insulin<br>syringe reuse. | The U&U Insulin Syringe<br>with Safety Retractable<br>Device a sterile, single use,<br>disposable and non-reusable,<br>manual retractable safety<br>insulin syringe intended for<br>injection of U-100 insulin<br>into the body, while reducing<br>the risk of sharps injuries<br>and the potential for insulin syringe<br>reuse.<br>The U&U insulin syringe is<br>a sterile, single use,<br>disposable and non-reusable,<br>insulin syringe intended for<br>injection of U-100 insulin<br>into the body. | Same |
| Configuration | Piston<br>Plunger<br>Barrel<br>Needle Hub<br>Needle<br>Safety feature | Piston<br>Plunger<br>Barrel<br>Needle Hub<br>Needle<br>Safety feature | Same |
| Sterility<br>condition | EO Sterilized | EO Sterilized | Same |
| Environment<br>of use | Home or Hospital | Home or Hospital | Same |
| Safety feature<br>principle | Spring retraction | manual retraction | See Comment 1 |
| Volume | 1ml | 1ml | Same |
| Needle gauge | 27-31G (0.25mm~0.40mm) | 27-31G | Same |
| Needle length | 8mm,10mm,13mm | Unknown | See Comment 2 |
| Single use | Yes | Yes | Same |
| Operation<br>mode | For Manual Use Only, For<br>Single Use only | For Manual Use Only, For<br>Single Use only | Same |
| Label/labeling | Complied with 21 CFR part<br>801 | Complied with 21 CFR part<br>801 | Same |
| Materials | PP, Polysoprene, TPE,<br>Stainless Steel, MABS and | PP,TPE,Stainless<br>Steel Silicone Oil | See Comment 3 |
| | | | |
| | Polydimethysiloxane | | |
| Performance<br>Testing | Conform to: | Conform to: | Same |
| | ISO 7864 | ISO 7864 | |
| | ISO 9626 | ISO 9626 | |
| | ISO 8537 | ISO 8537 | |
| | ISO 23908 | ISO 23908 | |
| | Force to activation: not be<br>more than 25N;<br>test value 10.48-20.91 | Test value 10.18-20.45N; | Same |
| | Force to move plunger: not<br>be less than 70N;<br>test value 83.67-116.97 | NA; | Comment 4 |
| Safety Feature<br>Performance | Reuse prevention:<br>It is achieved through the<br>barrel barb and convex point<br>are stuck when the plunger is<br>pushed to the bottom. | Reuse prevention:<br>It is achieved through<br>pulling back the plunger and<br>snapped off. | |
| | Force to separate plunger<br>lid: not be less than 20N;<br>test value 25.14-32.67 | Test value 25.18-32.92N; | Same |
| | Biocompatibil<br>ity | Cytotoxicity | Conforms to ISO10993<br>(Part1:Evaluation and<br>testing, Part 4: Selection of<br>tests for interactions with<br>blood, Part 5: Tests for in<br>vitro cytotoxicity, Part 7:<br>Ethylene oxide sterilization<br>residuals,Part 10: Tests for<br>irritation and delayed-type<br>hypersensitivity, Part 11:<br>Tests for systemic toxicity) |
| | | Sensitization | |
| | | Irritation | |
| Acute systemic toxicity | | | |
| Material mediated<br>pyrogenicity | | | |
| Hemocompatibility | | | |
| Subchronic toxicity | | | |
| Particulate matters | | | |
{7}------------------------------------------------
### Comment 1
Differences in safety feature principle between the predicate and subject device were addressed through ISO 23908 Safety Feature Performance Test.The safety feature principle of subject devicer is identical with Retractable Safety Syringes(manufacture by Jiangsu Caina Medical Co., Ltd.) as cleared in K191490 at date 2019/10/03.
### Comment 2
Differences in needle length between the predicate and subject device were addressed through ISO 8537 performance testing.
{8}------------------------------------------------
### Comment 3
Differences in materials between the predicate and subject device were addressed through Biocompatibility Statement and copy of test report.
### Comment 4
Differences in reuse prevention, for the predicate device this is achieved through pulling back the plunger and snapped off, for the proposed device a convex point is designed on the plunger and a barrel barb is designed in the barrel. When the plunger is pushed to the bottom, the barb and convex point are stuck, the plunger can not be re-moved to prevent reuse. At the same time, the front end of the slide bushing opens the barrel to release the needle hub and activate the safety function. Under the spring, the needle hub falls back into the plunger, and the needle tube is shielded inside the plunger. Force to re-move plunger have be tested in Safety Feature Performance Test Report, which is more than 70N. Proposed Device and Predicate Device are different designs, however performance demonstrated the differences do not raise any new safety concerns.
#### 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.