The U&U Insulin Syringe with Safety Retractable Device a sterile, single use, disposable and non-reusable, manual retractable safety insulin syringe intended for injection of U-100 insulin into the body, while reducing the risk of sharps injuries and the potential for insulin syringe reuse. The U&U insulin syringe is a sterile, single use, disposable and non-reusable, insulin syringe intended for injection of U-100 insulin into the body.
Device Story
Manual piston syringe for U-100 insulin injection; available with or without integrated manual retractable safety mechanism. Safety version allows needle retraction into barrel after injection to prevent sharps injuries; plunger is snapped off to render device unusable. Used in clinical or home settings by patients or healthcare providers. Provides clear visualization of injection site. Benefits include accurate insulin delivery and reduced risk of accidental needle sticks and syringe reuse.
Clinical Evidence
Bench testing only. No clinical data provided. Performance evaluated against ISO 8537, ISO 7886-1, ISO 7864, ISO 9626, and ISO 23908. Biocompatibility testing performed per ISO 10993 (parts 1, 4, 5, 7, 10, 11).
Technological Characteristics
Materials: Polypropylene (PP) barrel/plunger/hub/sheath, TPE piston, stainless steel needle. Lubricant: Silicone oil. Sizes: 0.3cc/ml to 1cc/ml; 27G to 31G needles. Principle: Manual piston syringe with optional manual retractable safety mechanism. Sterilization: Ethylene oxide. Standards: ISO 8537, ISO 7886-1, ISO 7864, ISO 9626, ISO 23908, ISO 10993.
Indications for Use
Indicated for patients requiring subcutaneous injection of U-100 insulin. Intended for single-use, manual administration.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, representing the department's focus on people and health.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 25, 2016
U&U Medical Technology Co., Ltd. Black Wang General Manager Dongzhou Village, Hengshanqiao Changzhou, Jiangsu 213119 CHINA
Re: K152808
Trade/Device Name: U&U Insulin Syringe with/without Safety Retractable Device Regulation Number: 21 CFR 880.5860 Regulation Name: Piston syringe Regulatory Class: II Product Code: MEG and FMF Dated: February 18, 2016 Received: February 19, 2016
Dear Black Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image shows a signature of Tina Kiang. The signature is written in black ink on a white background. There is also a faint FDA logo in the background. The signature is followed by the letter S.
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152808
Device Name
U&U Insulin Syringe with/without Safety Retractable Device
Indications for Use (Describe)
The U&U Insulin Syringe with Safety Retractable Device is a sterile, single use, disposable and non-reusable, manual retractable safety insulin syringe intended for injection of U-100 insulin into the body, while reducing the risk of sharps injuries and the potential for insulin syringe reuse.
The U&U Insulin Syringe is a sterile, single use, disposable, insulin syringe intended for injection of U-100 insulin into the body.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
| <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K152808 510(K) Summary
Date Prepared: March 9, 2016
The assigned 510(k) Number: K152808
## 1. Submitter Name and Address:
| Owner Name: | U&U Medical Technology Co., Ltd |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Address: | Dongzhou Village,Hengshanqiao,Changzhou,Jiangsu,China<br>RM C1-D 6/F WING HING IND BLDG 14 HING YIP ST KWUN TONG KLN<br>HONG KONG |
| Contactor Name: | Xuebo Wang |
| TEL: | +86-13564751751 |
| E-mail: | Blackwang@tkmedical.com |
## Contract Manufacturer Name: ANHUI TIANKANG MEDICAL PRODUCTS CO., LTD.
| Address: | No 20 south renhe road tianchang, CHINA 239300 |
|----------|-------------------------------------------------|
| Web: | www.tkmedical.com |
### US Agent:
| Name: | CARELIFE (USA) INC. |
|------------------|----------------------------------------------|
| Address: | 1580 Boggs Rd, Suite 500/600 Duluth GA 30096 |
| TEL: | 404-661-2228 |
| Contact person : | Ms. LI QIAN liqian@shanghaicarelife.com |
# 2. Submission Devices Information:
Trade/Proprietary Name: U&U Insulin Syringe with/without Safety Retractable Device Common Name: Syringe, Antistick Piston Syringe Classification name: Piston Syringe. Class: 2 Panel: 80 Product codes: MEG and FMF Submission Type: 510(k) Traditional Regulation Number: 880.5860
## 3. Predicate Devices Information:
| Trade Name: | InsoSAFE BakSNAP Retractable Insulin Safety Syringe |
|----------------|-----------------------------------------------------|
| 510(K) Number: | K050131 |
| Company: | M.K. Meditech Co., Ltd |
| Trade Name: | Disposable Insulin Syringe |
| 510(K) Number: | K151949 |
| Company: | Jiangyin Caina Technology Co., Ltd. |
# 4. Devices Description:
The U&U Insulin Syringe with/without Safety Retractable Devices are similarl devices except one has a safety feature and the other is a standard insulin syringe.
The U&U Insulin Syringe with/without Safety Retractable Device is an integrated needle and piston syringe with an anti-needle-stick mechanism. The mechanism allows clear visualization of
{4}------------------------------------------------
the injection site at all times. The mechanism shows the needle is contained within the syringe barrel. After standard techniques for injection, the plunger is withdrawn and snapped off which renders the needle unusable and prevents accidental needle sticks. The used syringe is then discarded into a sharps container.
The subject devices are sterile, single use, disposable and non-reusable, insulin syringe intended for injection of U-100 insulin into the body.
The U&U Insulin Syringe with/without Safety Retractable Device will be available in numerous sizes and combinations between the smallest (0.3cc/ml + 31G) and the largest (1cc/ml + 27G )
| Ref<br>Number | Model<br>Number | Description | Size |
|---------------|-----------------|---------------------------------------------------|----------------------------|
| UUIS001 | UUIS | Insulin Syringe | 0.3cc/ml Needle 27G to 31G |
| UUIS002 | UUIS | Insulin Syringe | 0.5cc/ml Needle 27G to 31G |
| UUIS003 | UUIS | Insulin Syringe | 1cc/ml Needle 27G to 31G |
| UUSIS001 | UUSIS | Insulin Syringe with Safety<br>Retractable Device | 1cc/ml Needle 27G to 31G |
# 5. Intended Use:
The U&U Insulin Syringe with Safety Retractable Device a sterile, single use, disposable and non-reusable, manual retractable safety insulin syringe intended for injection of U-100 insulin into the body, while reducing the risk of sharps injuries and the potential for insulin syringe reuse.
The U&U insulin syringe is a sterile, single use, disposable and non-reusable, insulin syringe intended for injection of U-100 insulin into the body.
### 6. Technological Characteristics:
Through comparisons between the submitted devices with the predicate devices as follows tables. We believe the applicant devices are substantially equivalent with the predicate devices.
### Comparison Table
| Element of Comparison | Submission Device | Predicate Device<br>K050131 | Predicate Device<br>K151949 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The U&U Insulin Syringe<br>with Safety Retractable<br>Device a sterile, single use,<br>disposable and non-<br>reusable, manual retractable<br>safety insulin syringe<br>intended for injection of U-<br>100 insulin into the body,<br>while reducing the risk of<br>sharps injuries and the<br>potential for insulin syringe<br>reuse.<br><br>The U&U insulin syringe is a<br>sterile, single use,<br>disposable and non-<br>reusable, insulin syringe<br>intended for injection of U-<br>100 insulin into the body. | The InsoSAFE BakSNAP<br>Retractable Insulin Safety<br>Syringe a sterile, single-<br>use, disposable and non-<br>reusable, manual<br>retractable safety insulin<br>syringe intended for<br>injection of insulin into<br>the body, while reducing<br>the risk of sharps injuries<br>and the potential for<br>insulin syringe reuse. | The disposable insulin syringe<br>is intended for medical<br>purposes for the manual<br>aspiration of insulin, and<br>for the injection of insulin into<br>parts of the body below the<br>surface skin. |
| Principle of Operation | Normal | Normal | Normal |
{5}------------------------------------------------
U&U Medical Technology Co., Ltd Dongzhou Village,Hengshanqiao,Changzhou,Jiangsu,China U&U (HONGKONG) Medical Technology Co., Limited RM C1-D 6/F WING HING IND BLDG 14 HING YIP ST KWUN TONG KLN HONG KONG [Insulin Syringe]
| Syringe Capacity | 0.3cc/ml Needle 27G to 31G<br>0.5cc/ml Needle 27G to 31G<br>1cc/ml Needle 27G to 31G<br><br>U&U Insulin Syringe with Safety Retractable Device:<br>1cc/ml only needle 27G to 31G | Various Sizes | 0.3ml<br>0.5ml<br>1ml |
|------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|--------------------------------------------------------|
| Nozzle Type | N.A | N.A | N.A |
| Lubricant for Barrel | Silicone Oil | Silicone Oil | Polydimethylsiloxane |
| Barrel Transparency | Transparent and Clear | Transparent and Clear | Transparent and Clear |
| Gradations Legibility | Legible | Legible | Legible |
| Materials | | | |
| Barrel<br>Plunger<br>Piston<br>Needle Hub<br>Needle<br>Needle Sheath | PP<br>PP<br>TPE<br>PP<br>Stainless Steel<br>PP | PP<br>PP<br>Silicone<br>PP<br>Stainless Steel<br>PP | PP<br>PP<br>Polysoprene<br>PP<br>Stainless Steel<br>PE |
| Needle Gauge and Length | Various Sizes | Various Sizes | Various Sizes |
| Lubricant for Needle | Silicone Oil | Silicone Oil | Polydimethylsiloxane |
| U&U Insulin Syringe with<br>Safety Retractable Device:<br>Sharps Injury Prevention<br>Features | Manual Retractable<br>Conforms to ISO 23908 | Manual Retractable<br>Conforms to ISO 23908 | N.A |
| Performances | Conforms to ISO7864<br>ISO8537 | Conforms to ISO7864<br>ISO8537 | Conforms to ISO7864 ISO8537 |
| Biocompatibility | Conforms to ISO10993 (Part<br>1: Evaluation and testing,<br>Part 4: Selection of tests for<br>interactions with blood, Part<br>5: Tests for in vitro<br>cytotoxicity, Part 7: Ethylene<br>oxide sterilization residuals,<br>Part 10: Tests for irritation<br>anddelayed-type<br>hypersensitivity, Part 11:<br>Tests for systemic toxicity) | Conforms to ISO10993 | Conforms to ISO10993 |
| Labeling | Meet the requirements of 21<br>CFR Part 801 | Meet the requirements of<br>21 CFR Part 801 | Meet the requirements of 21<br>CFR Part 801 |
# 7. Non-Clinical Tests performed on the subject device.
The proposed devices were tested per the following standards, to evaluate its performance.
• ISO 8537 Second edition 2007-10-01 Sterile single-use syringes, with or without needle, for insulin.
• ISO7886-1:1993 Sterile hypodermic syringes for single use - Part 1: Syringes for manual use
· ISO7864:1993 Sterile hypodermic needles for single use
For the U&U Insulin Syringe with Safety Retractable Device Only: · ISO 9626:1991 Stainless steel needle tubing for the manufacture of medical devices. ISO 23908 Sharps Injury Prevention
Guidance document.
Guidance for Industry and FDA Staff: Medical Devices with Sharps Injury Prevention Features:
{6}------------------------------------------------
U&U Medical Technology Co., Ltd Dongzhou Village,Hengshanqiao,Changzhou,Jiangsu,China U&U (HONGKONG) Medical Technology Co., Limited RM C1-D 6/F WING HING IND BLDG 14 HING YIP ST KWUN TONG KLN HONG KONG [Insulin Syringe]
# http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/uc m071663.htm
## The test items include following.
- Insulin Syringe Physical Performance Needle length
- Insulin Syringe Physical Performance Diameter
- · Insulin Syringe Physical Performance Gauge
- · Insulin Syringe Physical Performance Hub/needle bond strength
- · Insulin Syringe Physical Performance Cover strength
- Insulin Syringe Physical Performance Patency of Needle lumen
- · Insulin Syringe Physical Performance Resistance to breakage
- · Insulin Syringe Physical Performance Stiffness
- Insulin Svringe Physical Performance Resistance to corrosion
- · Insulin Syringe Physical Performance Maximum Dead Space
- · Insulin Syringe Physical Performance Freedom from leakage at needle
- Insulin Syringe Physical Performance Freedom Liquid and air leakage past piston
- Insulin Syringe Physical Performance Force required to Operate plunger
- · Sharps Injury Prevention Features Force to activate safety feature
- Insulin Syringe Physical Performance Tolerance on graduated capacity (Dose accuracy)
#### Biocompatibility
Conforms to ISO10993
- Part 1: Evaluation and testing.
- Part 4: Selection of tests for interactions with blood,
- Part 5: Tests for in vitro cytotoxicity,
- Part 7: Ethylene oxide sterilization residuals,
- Part 10: Tests for irritation and delayed- type hypersensitivity,
- Part 11: Tests for systemic toxicity
### 8. Conclusion:
The materials, performance, and operational features of both the subject device and the predicate devices are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.