K143218 · Penumbra, Inc. · HCG · Mar 18, 2015 · Neurology
Device Facts
Record ID
K143218
Device Name
Penumbra Smart Coil
Applicant
Penumbra, Inc.
Product Code
HCG · Neurology
Decision Date
Mar 18, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5950
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Smart Coil is indicated for the embolization of: · Intracranial aneurysms · Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae · Arterial and venous embolizations in the peripheral vasculature
Device Story
The Penumbra Smart Coil is a neurovascular embolization device used to occlude targeted vascular segments. It consists of a coil implant, a detachment pusher, and a detachment handle. The device is operated by a physician in a clinical setting. The coil is delivered via a microcatheter to the target site; once positioned, the physician uses the detachment handle to release the coil from the pusher. The coil packs the aneurysm or vessel to achieve occlusion. The device provides a mechanical means to treat neurovascular and peripheral vascular abnormalities, potentially reducing the risk of aneurysm rupture or abnormal blood flow. It is a single-use, sterile device.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility (ISO 10993), mechanical design verification (fatigue, torsion, tensile, friction), sterilization (ISO 11135-1), shelf life (ASTM D4169, F2096), and MR compatibility (ASTM F2182, F2052, F2213, F2119). All tests met predetermined acceptance criteria.
Technological Characteristics
Materials: Platinum/Tungsten, Nitinol, Polymer, Adhesives. Components: Coil implant, detachment pusher, detachment handle. Dimensions: Secondary diameter 1-18 mm, length 1-60 cm. Sterilization: EtO (EN ISO 11135-1:2007). MR Conditional. Mechanical performance verified via fatigue, torsion, and tensile testing.
Indications for Use
Indicated for patients requiring embolization of intracranial aneurysms, neurovascular abnormalities (AVMs, AV fistulae), and peripheral arterial/venous embolizations.
Regulatory Classification
Identification
A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Re:
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 18, 2015
Penumbra, Inc. % Charles DeNault Regulatory Affairs Specialist 1351 Harbor Bay Parkway Alameda, California 94502
> K143218 Trade/Device Name: Penumbra Smart Coil Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: Class II Product Code: HCG, KRD Dated: January 23, 2015 Received: January 26, 2015
Dear Mr. Charles DeNault,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -S 同公
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143218
Device Name Penumbra Smart Coil
Indications for Use (Describe)
The Smart Coil is indicated for the embolization of:
· Intracranial aneurysms
- · Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae
- · Arterial and venous embolizations in the peripheral vasculature
Type of Use (Select one or both, as applicable):
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 11 510(k) Summary
(as required by 21 CFR 807.92)
Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra Inc. is providing the summary of Substantial Equivalence for the Penumbra Smart Coil™.
# 11.1 Sponsor/Applicant Name and Address
Penumbra, Inc. 1351 Harbor Bay Parkway Alameda, CA 94502 USA
#### 11.2 Sponsor Contact Information
Charles DeNault Regulatory Affairs Specialist Phone: (510) 748-3302 Fax: (510) 217-6414 Email: cdenault@penumbrainc.com
#### 11.3 Date of Preparation of 510(k) Summary
March 18, 2015
# 11.4 Device Trade or Proprietary Name
Penumbra Smart Coil™
### 11.5 Primary Device Classification
| Regulatory Class: | II |
|-----------------------|-----------------------------------|
| Classification Panel: | Neurology |
| Classification Name: | Neurovascular embolization device |
| Regulation Number: | 21 CFR 882.5950 |
| Product Code: | HCG |
#### 11.6 Secondary Device Classification
| Regulatory Class: | II |
|-----------------------|------------------------------|
| Classification Panel: | Cardiovascular |
| Classification Name: | Vascular embolization device |
| Regulation Number: | 21 CFR 870.3300 |
| Product Code: | KRD |
{4}------------------------------------------------
### 11.7 Predicate Devices
| 510(k) Number | Clearance Date | Name of Predicate Device | Name of<br>Manufacturer |
|---------------|------------------|--------------------------|-------------------------|
| K103305 | January 26, 2011 | Penumbra Coil System | Penumbra, Inc. |
| K120330 | April 02, 2012 | Penumbra Coil System | Penumbra, Inc. |
# 11.8 Predicate Comparison
| | Penumbra Coil System | Penumbra Smart Coil |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| Classification | Class II, HCG, KRD | Same |
| Indications for Use | Indicated for the embolization of:<br>• Intracranial aneurysms<br>• Other neurovascular abnormalities<br>such as arteriovenous malformations<br>and arteriovenous fistulae<br>• Arterial and venous embolizations in<br>the peripheral vasculature | Same |
| Materials | | |
| Coil Implant | Platinum/Tungsten, Nitinol. Adhesive | Platinum/Tungsten, Nitinol, Polymer, Adhesives |
| Inner Coil | Stainless Steel, Polymer, Platinum/Tungsten, Nitinol | Same |
| Detachment Handle | Plastic | Same |
| Dimensions/Shape | | |
| Coil Secondary Diameter | 2-32 mm | 1-18 mm |
| Coil Length | 1-60 cm | Same |
| Coil Secondary Shape | Complex, Helical (Curve), J | Complex, Helical (Curve) |
| Pusher Length | 175 cm | 185 cm |
| Sterilization | | |
| Sterilization Method | EtO | Same |
# 11.9 Device Description
The Smart Coil functions to selectively embolize targeted segments of the vasculature by packing a sufficient quantity of bare platinum coils to achieve occlusion in an identical fashion to the Penumbra Coil System. The Smart Coil System consists of three components: a Coil Implant attached to a Detachment Pusher and a Detachment Handle.
# 11.10 Indications for Use
The Penumbra Smart Coil System is indicated for the embolization of:
- . Intracranial aneurysms
{5}------------------------------------------------
- . Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae
- . Arterial and venous embolizations in the peripheral vasculature
# 11.11 Summary of Non-Clinical Data
Included in this section is a description of the testing, which substantiates the safe and effective performance of the Smart Coil and Smart Coil Detachment Handle as well as its substantial equivalence to the predicate devices:
- Biocompatibility
- Design Verification (Bench-Top Testing) ●
- . MR Environment
- Sterilization ●
- Shelf Life
The subject Smart Coil met all predetermined requirements.
# 11.11.1 Biocompatibility Testing
Non-clinical testing found the Smart Coil and Smart Coil Detachment Handle to be biocompatible according to the requirements of EN ISO 10993mbra Smart Coil should perform as intended in the specified use conditions. requirements. Studies were selected in accordance with EN ISO 10993-1:2009 guidelines (Biological Evaluation of Medical Devices). All studies were conducted pursuant to 21 CFR, Part 58, Good Laboratory Practices. The following tests were performed:
| Test | Acceptance Criteria | Results | Conclusions |
|----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| <i>In vitro</i> Cytotoxicity<br>(MEM Elution) | Sample extracts must yield cell<br>lysis grade $\u2264$ 2 | Grade: 0 | Non-toxic |
| Sensitization<br>(Magnusson-Kligman<br>Method) | Test Group shall yield Grade < 1<br>score on Magnusson and Kligman<br>scale (provided Control Grade < 1) | Grade: 0 (No Sensitization<br>response) | Non-<br>sensitizing |
| Irritation<br>(Intracutaneous<br>Reactivity (ISO)) | The difference in the mean test<br>article and mean control score must<br>be grade 1.0 or lower | Grade 0.0 difference<br>(saline extract) and Grade<br>0.2 difference (sesame oil<br>extract) | Non-irritant |
| Implant Study | | | |
| Test | Acceptance Criteria | Results | Conclusions |
| Implant Study | Overall interpretation of<br>biocompatibility exhibited by the<br>test article based on the clinical<br>observations, and gross and<br>microscopic analysis comparing<br>the test article to the control, shall<br>show the test article to be a non-<br>irritant. | Test article is considered a<br>non-irritant | Non-irritant |
| Systemic Toxicity<br>(Acute) | | | |
| Systemic Injection<br>(ISO) | Sample extracts must not cause<br>significant biological reaction<br>greater than control. That is:<br>Death in 2 or more animals<br>●<br>Toxic signs (i.e. convulsions,<br>●<br>prostration)<br>Weight loss > 10% in 3 or<br>●<br>more animals | No evidence of systemic<br>toxicity from sample<br>extracts:<br>No deaths<br>●<br>No toxic signs<br>●<br>No weight loss > 10%<br>●<br>in any animals | Non-toxic |
| Material Mediated<br>Pyrogen | Sample extracts must not cause a<br>total rise in body temperature of ≥<br>0.5° C | Non-pyrogenic: no animals<br>had a temperature rise ≥<br>0.5°C | Non-<br>pyrogenic |
| Sub-Chronic Toxicity<br>(Sub-Acute Toxicity) | | | |
| Sub-Chronic/Sub-<br>Acute Toxicity: 14<br>day / 14 dose IV study<br>(mice) | The following parameters will be<br>used as signs of toxicity; however<br>final evaluation of the test article<br>will involve correlation of all data<br>for patterns of toxicity:<br>Death of more than one animal<br>●<br>in group<br>Mean body weight change in<br>group relative to control<br>●<br>Clinical signs of toxicity in<br>more than one animal in group<br>●<br>Hematology and clinical<br>chemistry values for animal<br>groups will be reviewed for<br>patterns of toxicity<br>Histopathology of tissues will<br>●<br>be evaluated for patterns of<br>toxicity | Negative for systemic<br>toxicity:<br>No deaths<br>●<br>No statistically<br>significant differences<br>between the test and<br>control mice body<br>weights<br>No abnormal clinical<br>●<br>signs of toxicity<br>Hematology showed<br>●<br>no findings indicative<br>of a pattern of toxicity<br>Histopathology<br>●<br>showed no adverse<br>findings attributed to<br>the test article | Non-toxic |
| Hemocompatibility | | | |
| Thrombosis (Dog<br>Thrombogenicity) | Device must be non-thrombogenic<br>after 4 hours in vivo when<br>compared to a predicate device | Test device was not<br>thrombogenic when<br>compared to the predicate.<br>The device exhibited<br>acceptable interaction with<br>blood | Non-<br>hemolytic |
| Coagulation (PT) | Coagulation times must be similar<br>to the predicate or negative control<br>values using analysis of variance. | Test article coagulation<br>times are statistically<br>similar to the predicate (p-<br>value=1.000) | |
| Test | Acceptance Criteria | Results | Conclusions |
| Coagulation (PTT) | Coagulation times must be similar<br>to the predicate or negative control<br>values using analysis of variance. | Test article coagulation<br>times are statistically<br>similar to the predicate (p-<br>value=0.105) | |
| Hematology<br>(Hemolysis) - Direct<br>Contact | Sample extracts must be<br>nonhemolytic (≤2% hemolytic<br>index) | Nonhemolytic:<br>Hemolytic index = 1.25%<br>Corrected hemolytic index<br>= 0.40% | |
| Hematology<br>(Hemolysis) - Indirect<br>Contact | Sample extracts must be<br>nonhemolytic (≤2% hemolytic<br>index) | Nonhemolytic:<br>Hemolytic index = 0.37%<br>Corrected hemolytic index<br>= 0.00% | |
| Complement<br>Activation | The concentrations of C3a and<br>SC5b-9 in the test samples are<br>statistically similar to or lower than<br>the predicate | Concentrations of test<br>article compared to<br>predicate:<br>C3a = similar at 30, 60, and<br>90 minutes<br>SC5b-9 = lower at 30 and<br>60 minutes / similar at 90<br>minutes | |
| Genotoxicity | | | |
| Ames Mutagenicity | Sample extracts must be non-<br>mutagenic | Non-mutagenic | |
| Mouse Lymphoma | Sample extracts must be non-<br>mutagenic | Non-mutagenic | Non-<br>mutagenic |
| <i>In vivo</i> Mouse<br>micronucleus | Sample extracts must be non-<br>mutagenic | Non-mutagenic | |
Biocompatibility Testing Summary – Coil Implant/Detachment Pusher
{6}------------------------------------------------
{7}------------------------------------------------
Biocompatiblity Testing Summary – Detachment Handle
| Test | Acceptance Criteria | Results | Conclusion |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|-----------------|
| In vitro Cytotoxicity<br>(MEM Elution) | Sample extracts must yield<br>cell lysis grade ≤ 2 | Grade: 0 | Non-toxic |
| Sensitization<br>(Magnusson-Kligman<br>Method) | Test Group shall yield Grade<br>< 1 score on Magnusson and<br>Kligman scale (provided<br>Control Grade < 1) | Grade: 0 (No<br>Sensitization<br>response) | Non-sensitizing |
| Irritation<br>(Intracutaneous<br>Reactivity (ISO)) | The difference in the mean<br>test article and mean control<br>score must be grade 1.0 or<br>lower | Grade 0.0 difference<br>(saline extract) and<br>Grade 0.2 difference<br>(sesame oil extract) | Non-irritant |
# 11.11.2 Bench-top Testing
The physical, mechanical and performance testing of the Smart Coil components demonstrate that the devices are substantially equivalent to the currently marketed predicate devices.
{8}------------------------------------------------
Design Verification testing was conducted to evaluate the physical and mechanical properties of Smart Coil components. The tests performed on the Smart Coil and Smart Coil Detachment Handle components included:
| Attribute | Specification | Results |
|-----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Dimensional / Visual Inspection | These evaluations confirm that the units used in this Design Verification testing meet all inspection criteria for release of finished goods (clinically acceptable) product. | Pass |
| Simulated Use/Detachment Reliability | These evaluations confirm that the units used in this Design Verification testing meet all inspection criteria for release of finished goods (clinically acceptable) product. | Pass |
| Fatigue Testing | The Coil Implant shall retain its secondary shape after being cycled into/out of introducer sheath 5 times | Pass |
| Torsion Resistance | The Coil Implant can be subjected to 8 turns minimum | Pass |
| Corrosion Resistance | No visible corrosion after testing | Pass |
| Friction Testing | Push/pull friction acceptable through a 0.0165 in. ID microcatheter | Pass |
| Distal System Tensile Test | Coil Implant and Detachment Pusher joints has sufficient tensile strength | Pass |
| Coil Dimensional Inspection (after simulated use testing) | These evaluations confirm that the units used in this Design Verification testing withstood the simulated use testing | Pass |
| Distal System Tensile Test | Coil Implant and Detachment Pusher joints has sufficient tensile strength | Pass |
All testing met specification. The results of the tests appropriately address the physical and mechanical performance expectations of the device. Based on these overall results, the physical and mechanical properties of the Smart Coil and Smart Coil Detachment Handle are acceptable for the intended use and substantially equivalent to the predicate devices.
# 11.11.3 Sterilization
The Smart Coil and Smart Coil Detachment Handle were tested to be sterile using identical acceptance criteria and testing methods as the predicate device. Testing was accordance with EN ISO 11135-1:2007.
#### 11.11.4 Shelf Life
The Smart Coil and Smart Coil Detachment Handle were tested to have a minimum one year shelf life using identical acceptance criteria and testing methods as the predicate device. Testing was done in accordance to ASTM D4169:2009 and ASTM F2096:2004.
{9}------------------------------------------------
### 11.11.5 MR Environment
The Smart Coil was tested in the MR environment to advise the MR Conditional statement in the IFU. Testing was done in accordance to standards ASTM F2182-11, ASTM F2052-06, ASTM F2213-06, and ASTM F2119-07 (R 13).
### 11.12 Summary of Substantial Equivalence
The Penumbra Smart Coil was found to have a safety and effectiveness profile that is similar to the predicate device.
# 11.13 Conclusion
The non-clinical data supports the safety of the device and the verification and validation demonstrate that the Penumbra Smart Coil should perform as intended in the specified use conditions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.