CEREBASE DA Guide Sheath, 95cm, CEREBASE DA Guide Sheath, 90cm, CEREBASE DA Guide Sheath, 80cm, CEREBASE DA Guide Sheath, 70cm
K192804 · Medos International SARL · QJP · Feb 27, 2020 · Cardiovascular
Device Facts
Record ID
K192804
Device Name
CEREBASE DA Guide Sheath, 95cm, CEREBASE DA Guide Sheath, 90cm, CEREBASE DA Guide Sheath, 80cm, CEREBASE DA Guide Sheath, 70cm
Applicant
Medos International SARL
Product Code
QJP · Cardiovascular
Decision Date
Feb 27, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The CEREBASE DA Guide Sheath is indicated for the introduction of interventional devices into the neuro vasculature.
Device Story
The CEREBASE DA Guide Sheath is a single-lumen, stainless steel braided catheter designed for neurovascular access. It features a lubricious PTFE-lined inner lumen, a hydrophilic outer coating on the distal end, and a radiopaque marker band. The device is provided with a dilator and a hemostasis valve. It is used by clinicians to facilitate the intravascular passage of interventional devices. The sheath is flushed with heparinized saline and inserted into the vasculature using standard techniques. By providing a stable conduit, it enables the delivery of therapeutic or diagnostic devices to the neurovasculature, potentially benefiting patients undergoing neurointerventional procedures.
Clinical Evidence
No clinical studies were performed. Evidence consists of bench testing (ISO 10555-1, ISO 11070), biocompatibility testing (ISO 10993-1), and in vivo testing in a porcine model. Bench tests included visual inspection, dimensional verification, leak testing, tensile strength, lubricity, kink resistance, and backup support. Biocompatibility testing confirmed the device is non-cytotoxic, non-sensitizing, non-irritating, non-toxic, non-pyrogenic, and hemocompatible.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 27, 2020
Medos International SARL Michael Liao Senior Regulatory Affairs Program Lead Chemin-Blanc 38 Le Locle, 2400, Switzerland
Re: K192804
Trade/Device Name: CEREBASE DA Guide Sheath Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP Dated: January 29, 2020 Received: January 30, 2020
Dear Michael Liao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Xiaolin Zheng, Ph.D., M.S. Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K192804
Device Name CEREBASE DA Guide Sheath
Indications for Use (Describe)
The CEREBASE DA Guide Sheath is indicated for the introduction of interventional devices into the neuro vasculature.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Medos International, SARL I. Submitter Chemin-Blanc 38 2400 Le Locle, Switzerland
Tel: (510) 770-5376
Contact Person: Michael Liao Email of Contact Person: mliao7(@jts.jnj.com Date Prepared: January 28, 2020
#### II. Device
| Table 1. Device | |
|---------------------------|------------------------------------------------------|
| Device Proprietary Name | CEREBASE DA Guide Sheath |
| Common or Usual Name | Catheter, Percutaneous, Neurovasculature |
| Classification Name | Catheter, Percutaneous, Class II, 21 C.F.R. 870.1250 |
| Regulatory Classification | II |
| Product Code | QJP |
#### III. Predicate Device
IV. Device
Description
The predicate device is listed below in Table 2.
| =10/1-) | |
|---------|--|
| Table 2. Prior 510(k) Clearance | | | |
|---------------------------------|--------------|---------------------------|-----------------------------|
| 510(k)<br>Number | Date Cleared | Name | Manufacturer |
| K111380 | 07/19/2011 | Neuron MAX System | Penumbra, Inc. |
| K140080 | 04/24/2014 | ENVOY DA Guiding Catheter | Medos International<br>SARL |
| K140307 | 04/21/2014 | ENVOY Guiding Catheter | Codman & Shurtleff,<br>Inc. |
The CEREBASE DA Guide Sheath is a single lumen, stainless steel braided catheter of variable stiffness with a large non-tapered lumen that facilitates the intravascular passage of interventional devices. The guide sheath is a straight shaped catheter that comes in four lengths; 70 cm, 80 cm, 90 cm, and 95 cm. The lubricious PTFE-lined inner lumen is designed to facilitate delivery of the interventional devices. It has a radiopaque marker band on the distal end and a luer hub at the proximal end. The guide sheath has an outer hydrophilic coating at the distal end that reduces friction during use. A hemostasis valve and a dilator are provided with the CEREBASE DA Guide Sheath within its sterile packaging. The dilator is a single lumen radiopaque catheter with a tapered distal end and a luer hub at the proximal end. The dilator is compatible with the CEREBASE DA Guide Sheath and up to 0.038 inch diagnostic guidewires. The hemostasis valve with side port is an off the shelf component used for flushing and insertion of catheters.
{4}------------------------------------------------
| IV. Device<br>Description,<br>Continued | The CEREBASE DA Guide Sheath's lumen must be flushed and outer body wet with<br>heparinized saline solution prior to use. If using as a primary access sheath with the<br>dilator, the dilator must be soaked and flushed with heparinized saline solution as well<br>prior to use. After being flushed with heparinized saline solution the dilator can then<br>be inserted and advanced until the tapered distal tip is beyond the distal end of the<br>CEREBASE DA Guide Sheath. Then the hemostasis valve is closed around the dilator<br>and primary access can be gained to the vasculature using a standard technique of<br>choice. |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| V. Indications<br>for Use | The CEREBASE DA Guide Sheath is indicated for the introduction of<br>interventional devices into the neuro vasculature. |
#### VI.
Comparison of Technological Characteristics with Predicate Device
Table 3 below provides comparison of technological characteristics with the predicate device. Based on design verification and validation testing, the minor differences in characteristics do not raise different questions of safety and effectiveness.
| Table 3. Technological Characteristic Comparison | | |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Description | CEREBASE DA Guide Sheath - Subject<br>Device | Neuron MAX<br>(K111380) |
| Indications | The CEREBASE DA Guide Sheath is indicated<br>for the introduction of interventional devices<br>into the neuro vasculature. | The Neuron MAXX System is<br>indicated for the introduction of<br>interventional devices into the<br>peripheral, coronary, and neuro<br>vasculature |
| Lengths | 70 cm<br>80 cm<br>90 cm<br>95cm | 80 cm<br>90 cm<br>-- |
| Shapes | Straight | Straight, MP |
| Outside Diameter | 8F | 8F |
| Inside Diameter | 0.090" | 0.088" |
| Tip Length | 1 mm | 2mm |
| Hub Material | Polycarbonate (yellow) | Nylon, Polyurethane |
| Shaft Material | Nylon Segment(s)<br>Pebax (nylon blend) Segment(s)<br>Polyurethane Segment(s) | Nylon, Polyurethane |
| Inner Lining | PTFE | PTFE |
| Outer Coating | Hydrophilic Coating | Hydrophilic Coating |
| Shaft Braid | 304 Stainless Steel | Stainless Steel |
| Outer Jacket<br>Material Segment | 11 outer jacket material segments of different<br>durometer | 5 outer jacket material segments of<br>different durometer |
| Distal Tip<br>Radiopaque Marker | Metal Marker Band | Radiopaque Filler and Metal Marker<br>band |
| Hemostasis Valve | Yes | Yes |
| Dilator | Yes | Yes |
| Dilator Material | Polyethylene | Polyethylene |
| Packaging | PET/ LDPE Tyvek Pouch<br>Polypropylene Tubing<br>SBS Mounting Card<br>SBS Carton | PET/PE/ Tyvek Pouch<br>HPDE Tubing<br>Mounting Card<br>Carton |
| Sterilization Method | EtO | EtO |
| Shelf Life | 1 year | 3 years |
{5}------------------------------------------------
VII. Non-Clinical Data Performance
The design verification and validation testing activities were conducted on the CEREBASE DA Guide Sheath to ensure that design outputs met design inputs. Table 4 below provides a description and result of each of these performance tests.
| Table 4. Performance Testing | | |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
| Test | Test Method Summary | Results |
| Visual Inspection | The purpose of this test was to verify that the test samples meet the visual requirements of ISO 10555-1. | Pass. Samples passed established acceptance criteria. |
| Catheter Internal Diameter (ID) | The purpose of this test was to verify that the catheter internal diameter meets requirements. The inner diameter of each sample was measured. | Pass. Samples passed established acceptance criteria. |
| Catheter Outer Diameter (OD) | The purpose of this test was to verify that the catheter outer diameter meets requirements. The outer diameter of each sample was measured. | Pass. Samples passed established acceptance criteria. |
| Catheter Working Length | The purpose of this test was to ensure the working length meets requirements. The length of the catheter was measured. | Pass. Samples passed established acceptance criteria. |
| Distal Tip Length | The purpose of this test was to ensure the distance of the distal edge of the marker band to the tip of the catheter meets requirements. | Pass. Samples passed established acceptance criteria. |
| Hub Luer Taper | The purpose of this test was to verify that the catheter hub luer taper fits standard luer fittings using a taper device. | Pass. Samples passed established acceptance criteria. |
| Air Leak Test | The purpose of this test was to verify that there are no air leaks into the hub subassembly. The samples were inspected to ensure no air bubbles that may indicate a hub air leak. | Pass. Samples passed established acceptance criteria. |
| Liquid Leak Test | The purpose of this test was to ensure that the catheter joint strength meets the freedom from leakage requirements (liquid during pressurization) requirements of ISO 10555-1. | Pass. Samples passed established acceptance criteria. |
| Static Burst | The purpose of this test was to verify the maximum hydrostatic pressure meets requirements of ISO 10555-1. | Pass. Samples passed established acceptance criteria. |
| Hub Pull Testing | The purpose of this test was to verify that the strength of the catheter hub meets the requirements of ISO 10555-1. A tensile tester measured the tensile strength of the joint. | Pass. Samples passed established acceptance criteria. |
| Shaft Tensile Strength | The purpose of this test was to verify that the strength of the catheter shaft meets the requirements of ISO 10555-1. A tensile tester measured the tensile strength of the distal section of the catheter shaft. | Pass. Samples passed established acceptance criteria. |
| Particulate Count | The purpose of this test was to quantify the particulate count generated by simulated use of the test samples. | Pass. Samples passed established acceptance criteria. |
| Coating Lubricity | The purpose of this test was to measure the lubricity of the coating. The test samples were put through a lubricity tester to measure frictional force. | Pass. Samples passed established acceptance criteria. |
| Coating Durability | The purpose of this test was to measure the durability of the lubricious coating layer. | Pass. Samples passed established acceptance criteria. |
| Coating Length | The purpose of this test was to verify that the catheter hydrophilic coating length meets the design requirement. | Pass. Samples passed established acceptance criteria. |
| Kink Resistance | The purpose of this test was to confirm the catheter met requirement for stability and did not kink during use. | Pass. Samples passed established acceptance criteria. |
| Delamination of PTFE Liner | The purpose of this test was to verify that the PTFE is appropriately adhered to the inner lumen of the catheter with braid reinforcement. | Pass. Samples passed established acceptance criteria. |
{6}------------------------------------------------
| VII. Non- |
|---------------|
| Clinical Data |
| Performance, |
| Continued |
| Table 4. Performance Testing, Continued | | |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
| Test | Test Method Summary | Results |
| Backup Support | The purpose of this test was to measure the stability of the test sample while advancing other interventional devices through its lumen.<br>A wire was pushed through the lumen of each test sample and the force at which catheter support failed was measured. | Pass. Samples passed established acceptance criteria. |
| Tip Linear Stiffness | The purpose of this test was to evaluate the flexibility of the catheter relative to other devices of similar design. | Pass. Samples passed established acceptance criteria. |
| Dilator Visual Inspection | The purpose of this test was to verify that the dilator was free of extraneous matter per ISO 11070. | Pass. Samples passed established acceptance criteria. |
| Dilator Inner Diameter (ID) | The purpose of this test was to verify that the dilator internal diameter meets requirements. | Pass. Samples passed established acceptance criteria. |
| Dilator Hub Luer Taper | The purpose of this test was to verify that the dilator fits standard luer fittings using a tapered device. | Pass. Samples passed established acceptance criteria. |
| Insertion Forces | The purpose of this test was to measure the forces required to insert the dilator and catheter into a simulated model. | Pass. Samples passed established acceptance criteria. |
| Dilator Outer Diameter (OD) | The purpose of this test was to verify that the dilator outer diameter meets requirements. | Pass. Samples passed established acceptance criteria. |
| Dilator Working Length (WL) | The purpose of this test is to verify the working length of the dilator to ensure compatibility with the catheter. | Pass. Samples passed established acceptance criteria. |
| Dilator Tensile Strength | The purpose of this test was to verify that the dilator joint strength meets the requirements of ISO 11070. | Pass. Samples passed established acceptance criteria. |
| Torque (Turns to Failure) | The purpose of this test is to count the number of turns the catheter can withstand without separating. | Pass. Samples passed established acceptance criteria. |
| Trackability | The purpose of this test is the measure of the force required to advance the catheter through a simulated vascular model. | Pass. Samples passed established acceptance criteria. |
| In Vitro Modeling and In Vivo Testing | The purpose of In Vitro modeling and In Vivo testing in a porcine model were to ensure that design outputs meet the customer requirements. | Pass. Samples passed established acceptance criteria. |
#### VII. Non-Clinical Data Performance, Continued
One year accelerated aging was successfully performed on the CEREBASE DA Guide Sheath. Additionally, through review of package integrity testing and previous testing with a sterile pouch made of the same material combination, the sterile pouch is confirmed to have a shelf life of at least one year. The shelf life of the CEREBASE DA Guide Sheath is established as one year.
### Packaging Qualification
Shelf Life
The CEREBASE DA Guide Sheath was sterilized, packaged and subjected to conditioning and simulation testing for worst case conditions per ISTA 3A: 2018, Packaged-Products for parcel Delivery System Shipment 70Kg (150 lb) or less. The samples were used to validate the package integrity of the CEREBASE DA Guide Sheath per EN ISO 11607-1:2009, Packaging for terminally sterilized medical devices – Part 1: Requirements for materials, sterile barrier systems and packaging systems.
{7}------------------------------------------------
VII. Non-Clinical Data Performance, Continued
The following packaging qualification testing was completed as part of this evaluation:
| Table 5. Packaging Qualification | |
|----------------------------------|--------|
| Test | Result |
| Visual Inspection | PASS |
| Dye Leak Penetration | PASS |
| Seal Strength | PASS |
| Shelf Life | PASS |
| Biocompatibility | PASS |
#### Performance Testing - Clinical
No clinical studies were required as appropriate verification and validation of the catheter and packaging modifications were achieved based on the similarities of the proposed device to the predicate device, and from results of bench testing.
#### Biocompatibility Testing
The CEREBASE DA Guide Sheath was assessed for biocompatibility in accordance with International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation of Testing within a Risk Management Process." and FDA Guidance for Industry and FDA Staff: Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing' (Issued June 16, 2016). The subject device is considered an externally communicating medical device with circulating blood contact for less than 24 hours. The following Biocompatibility Testing was completed as part of this evaluation:
{8}------------------------------------------------
VII. Non-Clinical Data Performance, Continued
| Table 6. Biocompatibility Test Results | | | |
|----------------------------------------|--------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| | Test | Test Summary | Conclusion |
| | <b>CEREBASE DA Guide Sheath</b> | | |
| | Cytotoxicity - ISO Elution<br>Method | Test articles exhibited no discrete intracytoplasmic<br>granules, no cell lysis, no reduction of cell growth,<br>resulting in a reactivity grade = 0 per ISO 10993-5. | PASS: Non-<br>cytotoxic |
| | Sensitization - ISO Guinea Pig<br>Maximization Sensitization | No abnormal dermal reactions and no sensitization<br>was observed. | PASS: Non-<br>sensitizing |
| | ISO Intracutaneous Reactivity<br>Study | Injection sites appeared normal and no difference in<br>erythema and edema scores from control. | PASS: Non-<br>irritating |
| | ISO Acute Systemic Toxicity in<br>Mice | Animals appeared clinically normal throughout the<br>study upon observation. No deaths and no significant<br>changes in body weight. | PASS: Non-toxic |
| | USP Rabbit Material-Mediated<br>Pyrogenicity | Out of 3 animals, no single animal showed a<br>total temperature rise of ≥ 0.5 °C, within USP<br>requirements. | PASS: Non-<br>pyrogenic |
| Hemocompatibility | ASTM Hemolysis | Hemolysis % was not higher than control in samples<br>either in direct contact with the catheter or in<br>samples extracted after contact with the catheter. | PASS: Non-<br>hemolytic |
| | SC5b-9 Complement<br>Activation Assay | SC5b-9 concentration of test article samples were<br>not statistically higher than both negative control and<br>activated normal human serum (NHS) controls. | PASS: Not a<br>potential activator<br>of the complement<br>system |
| | In Vivo Thromboresistance<br>Study | Thrombus formation on the catheter was comparable<br>to that on control article in similar animal<br>vasculature. | PASS: Acceptable<br>Thromboresistance |
| | <b>Hemostatic Valve</b> | | |
| | Cytotoxicity - ISO Elution<br>Method | Test articles exhibited no discrete intracytoplasmic<br>granules, no cell lysis, no reduction of cell growth,<br>resulting in a reactivity grade = 0 per ISO 10993-5. | PASS: Non-<br>cytotoxic |
| | Sensitization - ISO Guinea Pig<br>Maximization Sensitization | No abnormal dermal reactions and no sensitization<br>was observed. | PASS: Non-<br>sensitizing |
| | Irritation or Intracutaneous<br>Reactivity | Injection sites appeared normal. Mean erythema and<br>edema scores of test articles had differences of <1<br>from control. | PASS: Non-<br>irritating |
| | ISO Acute Systemic Toxicity in<br>Mice | Animals appeared clinically normal throughout the<br>study upon observation. No deaths and no significant<br>changes in body weight. | PASS: Non-toxic |
| | USP Rabbit Material-Mediated<br>Pyrogenicity | Out of 3 animals, no single animal showed a<br>total temperature rise of ≥ 0.5 °C, within USP<br>requirements. | PASS: Non-<br>pyrogenic |
| | Hemocompatibility:<br>ASTM Hemolysis | Hemolysis % was not higher than control in samples<br>either in direct contact with the catheter or in<br>samples extracted after contact with the valve. | PASS: Non-<br>hemolytic |
| | (Table continued on next page) | | |
{9}------------------------------------------------
#### VII. Non-Clinical Data Performance, Continued
| Table 6. Biocompatibility Test Results, Continued | | | |
|--------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Test | Test Summary | Conclusion | |
| Dilator | | | |
| Cytotoxicity - ISO Elution<br>Method | Test articles exhibited no discrete intracytoplasmic<br>granules, no cell lysis, no reduction of cell growth,<br>resulting in a reactivity grade = 0 per ISO 10993-5. | PASS: Non-<br>cytotoxic | |
| Sensitization - ISO Guinea Pig<br>Maximization Sensitization | No abnormal dermal reactions and no sensitization<br>was observed. | PASS: Non-<br>sensitizing | |
| ISO Intracutaneous Reactivity<br>Study | Injection sites appeared normal. Mean erythema and<br>edema scores of test articles had differences of <1<br>from control. | PASS: Non-<br>irritating | |
| ISO Acute Systemic Toxicity in<br>Mice | Animals appeared clinically normal throughout the<br>study upon observation. No deaths and no significant<br>changes in body weight. | PASS: Non-toxic | |
| USP Rabbit Material-Mediated<br>Pyrogenicity | Out of 3 animals, no single animal showed a<br>total temperature rise of ≥ 0.5 °C, within USP<br>requirements. | PASS: Non-<br>pyrogenic | |
| Hemocompatibility | ASTM Hemolysis | Hemolysis % was not higher than control in samples<br>either in direct contact with the dilator or in samples<br>extracted after contact with the dilator. | PASS: Non-<br>hemolytic |
| | SC5b-9 Complement<br>Activation Assay | SC5b-9 concentration of test article samples were<br>not statistically higher than both negative control and<br>activated normal human serum (NHS) controls. | PASS: Not a<br>potential activator<br>of the complement<br>system |
| | In Vivo Thromboresistance<br>Study | Thrombus formation on the dilator was comparable<br>to that on control article in similar animal<br>vasculature. | PASS: Acceptable<br>Thromboresistance |
| Packaging | | | |
| Cytotoxicity - ISO Elution<br>Method | Test articles exhibited no discrete intracytoplasmic<br>granules, no cell lysis, no reduction of cell growth,<br>resulting in a reactivity grade = 0 per ISO 10993-5 | PASS: Non-<br>cytotoxic | |
{10}------------------------------------------------
single use only.
| VII. Non-Clinical Data<br>Performance, Continued | <b>Sterilization</b><br>The CEREBASE DA Guide Sheath, as packaged with included accessories, is sterilized using a validated 100% Ethylene Oxide sterilization process to ensure sterility assurance level (SAL) of 10-6 in accordance with ISO 11135-1: 2014 Sterilization of health-care products. Ethylene oxide. Requirements for the development, validation and routine control of a sterilization process for medical devices. The CEREBASE DA Guide Sheath and all accessories meet EO residuals per EN ISO 10993-7: 2009 Biological evaluation of medical devices. Ethylene oxide sterilization residuals for a limited contact delivery system – externally communicating. The CEREBASE DA Guide Sheath and all accessories are for |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
- VIII. Conclusion Based upon the intended use, design, materials, function, side-by-side in-vitro testing and animal testing, and passing biocompatibility test results, it is concluded that the subject device, CEREBASE DA Guide Sheath is substantially equivalent to the predicate device, Neuron MAX System (K111380, cleared 19 July 2011). The fundamental scientific technology is the same as the predicate device. The differences in verbiage in the Indications for Use statement only limits the locations that the subject device can be used and do not raise any questions regarding the safety and effectiveness of the device. Risk assessment of the CEREBASE DA and verification/validation testing confirmed the device is as safe and effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.