The ENVOY® Guiding Catheter is intended for use in the peripheral, coronary, and neurovasculature for the intravascular introduction of interventional/diagnostic devices.
Device Story
ENVOY® Guiding Catheter is a single-lumen, braided, pre-shaped catheter; facilitates intravascular passage of interventional/diagnostic devices. Used in peripheral, coronary, and neurovasculature. Operated by clinicians. Device provides a pathway for other devices; does not perform automated analysis or processing. Line extension adds 7 French diameter size to existing 6 French product line.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing and verification/validation of design modifications.
Technological Characteristics
Single-lumen braided catheter; stainless steel reinforcement; PTFE liner. 7Fr diameter (0.092" OD, 0.078" ID). Sterilized via EtO. Complies with ISO 10555-1, ISO 594-1/2, ISO 10993 series, and BS EN ISO 11135-1.
Indications for Use
Indicated for patients requiring intravascular introduction of interventional or diagnostic devices within the peripheral, coronary, or neurovasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
#### K140307
# SECTION 005 510(k) Summary [As Required By 21 CFR 807.92(a)]
# A. Sponsor
ﺮ ﺳﺎ
| Submitter's Name: | Codman & Shurtleff, Inc. |
|-------------------|------------------------------------------|
| Address: | 325 Paramount Drive<br>Raynham, MA 02767 |
| Primary Contact: | Hannah Foley |
|------------------|----------------|
| Telephone: | (305) 265-6810 |
| Fax: | (508) 977-6979 |
| Secondary Contact: | Amarilys Machado |
|--------------------|------------------|
| Telephone: | (305) 265-6869 |
| Fax: | (508) 977-6979 |
- B. Date Prepared: February 6, 2014
#### C. Device Name and Classification:
| Proprietary Name: | ENVOY® Guiding Catheter |
|----------------------|--------------------------------------------------|
| Common/Usual Name: | Catheter, Percutaneous |
| Classification Name: | Percutaneous Catheter (21 CFR 870.1250), Class I |
| Product Code: | DQY |
#### D. Predicate Devices
This 510(k) submission provides pre-market notification for the proposed ENVOY® Guiding Catheter line extension. The line extension will include a new diameter size 7 French (Fr) with same lengths, pre-shaped configurations and XB (extra backup) options as the predicate ENVOY® 6 French (Fr) Guiding Catheter. The proposed line extension has not altered the fundamental technology or the predicate device's intended use.
| Table 1: Prior 510(k) Clearances | | | | | |
|----------------------------------|-----------------|--------------------------|--------------------------|-----------------|------------------------------------------------------------------------------------|
| 510(k)<br>Number | Date<br>Cleared | Name | Manufacturer | Product<br>Code | Predicate For: |
| Predicate<br>K093184 | 11/06/2009 | ENVOY® Guiding Catheters | Codman & Shurtleff. Inc. | DQY | Intended Use<br>Design<br>Materials<br>Manufacturing<br>Sterilization<br>Packaging |
## E. Device Description
The ENVOY® Guiding Catheter is a single lumen, braided catheter with a large non-tapered lumen that facilitates the intravascular passage of interventional/diagnostic devices. The guiding catheter is pre-shaped to facilitate positioning.
{1}------------------------------------------------
#### F. Indications for Use
The ENVOY® Guiding Catheter is intended for use in the peripheral, coronary, and neuro vasculature for the intravascular introduction of interventional/diagnostic devices.
# G. Summary of Technological Characteristics of the Proposed Device to the Predicate Device
The proposed 7Fr ENVOY® Guiding Catheter is substantially equivalent to the predicate ENVOY® Guiding Catheter (K093184). No new technological characteristics are being introduced with the proposed device. The proposed ENVOY® Guiding Catheter has the same intended use, function, mechanism of action, clinical utility, design, manufacturing process and sterilization process as the predicate ENVOY® Guiding Catheter. The proposed ENVOY® Guiding Catheter also has the same materials as the current ENVOY® Guiding Catheter (K093184) except for the orange ink used on the ID Band. The proposed ENVOY® 7Fr Guiding Catheter was shown to be substantially equivalent to the predicate device through comparison of indications for use, function, operating principle, bench testing, biocompatibility, and materials. A summary table including characteristics of the proposed device compared with those of the predicate device is provided in Table 2.
| Table 2: Predicate Comparison Profile | | |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|
| Description | Predicate Device:<br>ENVOY® 6Fr Guiding Catheter<br>(K093184) | This Submission:<br>ENVOY® 7Fr Guiding Catheter |
| Indications for Use | The Envoy Guiding Catheter is intended for use<br>in the peripheral, coronary and neuro<br>vasculature for the intravascular introduction of<br>interventional/ diagnostic devices | Same predicate |
| Device Description | The ENVOY® Guiding Catheter is a single<br>lumen, braided catheter with a large non-tapered<br>lumen that facilitates the intravascular passage of interventional devices. The guiding catheter<br>is preshaped to facilitate positioning. | Same predicate |
| Product Code | DQY | Same as predicate |
| Classification | 21 CFR 870.1250, Class II | Same as predicate |
| Length/ Working<br>Length (cm) | '90cm & 100cm | Same as predicate |
| Catheter Inner<br>Diameter | 0.070" (1.8mm) | 0.078" (2.0mm) |
| Catheter Outer<br>Diameter | 6.0F (0.082"/2.0mm) | 7.0F (0.092"/2.3mm) |
| Shapes | Straight<br>Multi-purpose D (MPD)<br>Multi-purpose C (MPC)<br>Modified Cerebral (Burke)(CBL)<br>Head-hunter 1 (H1)<br>Simmons 2 | Same as predicate |
| Reinforcing Member<br>(Braid) | Stainless Steel | Same as predicate |
| Liner | PTFE Liner | Same as predicate |
| Sterilization | EtO | Same as predicate |
{2}------------------------------------------------
## H. Summary of Nonclinical Testing
. "
The proposed ENVOY® Guiding Catheter has the same intended use, function, mechanism of action, clinical utility, design, manufacturing process and sterilization process as the predicate ENVOY® Guiding Catheter. The proposed ENVOY® Guiding Catheter also has the same materials as the current ENVOY® Guiding Catheter (K093184) except for the orange ink used on the ID Band. The testing conducted to assess the line extension includes performance assessments per the following recognized standards:
| Table 3: Performance Standards | |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Standard/Guidance/Directive | Description |
| BS EN ISO 11607-1: 2009 | Packaging for Terminally Sterilized Medical Devices - Part 1: Requirements for Materials. Sterile<br>Barrier Systems and Packaging Systems |
| BS EN ISO 11135-1: 2007 | Sterilization of Healthcare products Ethylene Oxide: Part I Requirements for development,<br>validation and routine control of a sterilization process for medical devices |
| BS EN ISO 10993-7: 2008 | Biological Evaluation of medical devices: Ethylene oxide sterilization residuals - Part 7 |
| BS EN ISO 10555-1: 2009 | Sterile, single use intravascular catheters Part 1: General requirements; Sterile |
| ISO 594-1: 1986 (E) | Conical fitting with a 6% (Luer) taper for syringes, needles and certain other medical equipment:<br>Part 1 - General Requirements |
| ISO 594-2 : 1998 (E) | Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment:<br>Part 2 - Local fittings |
| AAMI / ANSI HE75 : 2009 | Human Factors Engineering - Design of Medical Devices |
| BS EN ISO 10993-1: 2009 | Biological evaluation of medical devices: Evaluation & Testing - Part 1 |
| BS EN ISO 10993-5: 2009 | Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity |
| BS EN ISO 10993-4: 2009 | Biological evaluation of medical devices Part 4: Selection for tests for interactions with blood |
| USP <661> (2013) | Containers - Plastic, Physicochemical Tests |
| ISO 14971: 2012 | Medical Devices - Application of risk management to medical device |
| BS EN ISO 15223-1: 2012 | Medical Device - Symbols to be used with medical device labels. labeling and information to be<br>supplied |
## Bench Testing
Results of verification and validation testing that was conducted on the proposed ENVOY® Guiding Catheter demonstrates that it performs as designed, is suitable for its intended use, is substantially equivalent to the predicate device and therefore, does not raise any new questions of safety and effectiveness. Appropriate testing was identified based on a review of the products' risk analyses and previous validation and verification testing.
The following Verification and Validation tests were conducted to verify the modified design ENVOY® Guiding Catheter:
- . Visual Inspection
- Catheter Shape ●
- Catheter Dimensional Verification
- Tensile Strength Testing .
- . System Liquid Leakage Testing
- Lateral Stiffness Testing .
- Linear Stiffness Testing ●
{3}------------------------------------------------
- Back-Up Support .
- Track Testing .
- Sheath Introducer Compatibility �
- Multiple In-dwelling device .
The following Biocompatibility Testing was conducted with the proposed ENVOY® Guiding Catheter:
- . Cytotoxicity: MEM Elution
- . Hemocompatibility - In vitro Hemolysis (Direct & Extract)
- . USP Physicochemical Aqueous Extraction Test
#### I. Animal Testing
No animal studies were required as appropriate verification and validation of the ENVOY® Guiding Catheter line extension was achieved based on the similarities of the proposed device to the predicate device, and from results of bench testing.
## J. Summary of Clinical testing:
No clinical studies were required as appropriate verification and validation of the ENVOY® Guiding Catheter line extension was achieved based on the similarities of the proposed device to the predicate device, and from results of bench testing.
#### Conclusion:
Based upon the design, materials, function, intended use, and the non-clinical testing performed by Codman it is concluded that the proposed ENVOY® 7Fr Guiding Catheter is substantially equivalent to the predicate ENVOY® 6Fr Guiding Catheter (K093184), and therefore, does not raise any new questions of safety or effectiveness.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three overlapping wing-like shapes, representing the department's commitment to health, human services, and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular pattern around the eagle symbol. The logo is black and white.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
### April 21, 2014
Codman & Shurtleff, Inc. Hannah Foley 325 Paramount Dr. Raynham, MA 02767-0350 US
Re: K140307
> Trade/Device Name: ENVOY Guiding Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: March 21, 2014 Received: March 23, 2014
Dear Hannah Folcy.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (2) CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm,
Sincerely yours,
# Bram D. Zuckerman -S
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
Indications for Use
K140307 510(k) Number (if Known):
Device Name: ENVOY® Guiding Catheter
Indications for Use:
The ENVOY® Guiding Catheter is intended for use in the peripheral, coronary, and neurovasculature for the intravascular introduction of interventional/diagnostic devices.
Prescription Use: × (Part 21 CFR 801Subpart D) AND/OR
Over-The-Counter Use: (Part 21 CFR 807 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Bram D. Zuc
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.