K192625 · MicroVention, Inc. A TERUMO Group Company · DQY · Nov 21, 2019 · Cardiovascular
Device Facts
Record ID
K192625
Device Name
PG Pro Microcatheter
Applicant
MicroVention, Inc. A TERUMO Group Company
Product Code
DQY · Cardiovascular
Decision Date
Nov 21, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PG Pro microcatheter is intended for the peripheral vasculature for the infusion of diagnostic and therapentic agents.
Device Story
The PG Pro Microcatheter is a single-lumen, percutaneous catheter designed for navigating small, tortuous peripheral vasculature. It is introduced over a steerable guidewire. The device features a semi-rigid proximal section for steerability and a flexible distal tip for atraumatic navigation. It includes dual radiopaque markers for fluoroscopic visualization and a hydrophilic coating to reduce friction. The catheter is used in clinical settings by physicians to deliver diagnostic or therapeutic agents. The device is supplied sterile for single use. It does not include a shaping mandrel. The device benefits patients by providing a steerable, lubricious pathway for targeted delivery of agents within the peripheral vascular system.
Clinical Evidence
No human clinical data. Evidence includes biocompatibility testing (ISO 10993-1), extensive bench testing (physical attributes, mechanical integrity, flow rates, kink resistance, torque, and radiodetectability), and an acute large animal study (n=3). The animal study demonstrated that the PG Pro Microcatheter's safety profile, compatibility with therapeutic agents, and tissue response were similar to the predicate Headway 27 Microcatheter.
Technological Characteristics
Single-lumen catheter; materials include Pebax, Tungsten, Grilamid, and PTFE liner; Pt/Ir marker bands; Nylon hub. Features hydrophilic coating. Dimensions: 0.027" min ID, 0.038" max OD. Compatible with 0.021" guidewires or smaller. Sterile, single-use. No software or electronic components.
Indications for Use
Indicated for patients requiring infusion of diagnostic and therapeutic agents into the peripheral vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 21, 2019
MicroVention, Inc., Mr. Ganesh Balachandar Regulatory Affairs Specialist 35 Enterprise Aliso Viejo, California 92656
Re: K192625
Trade/Device Name: PG Pro Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY, KRA Dated: September 20, 2019 Received: September 23, 2019
Dear Mr. Balachandar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192625
Device Name PG Pro Microcatheter
Indications for Use (Describe)
The PG Pro microcatheter is intended for the peripheral vasculature for the infusion of diagnostic and therapentic agents.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
## SUBMITTER
MicroVention Inc. 35 Enterprise, Aliso Viejo, CA 92656 Phone: 714-247-8000 Contact Person: Ganesh Balachandar Date Prepared: 20th, September 2019
## DEVICE
Name of the Device: PG Pro Microcatheter Common Name: Percutaneous Catheter Classification Name: Device, Percutaneous Catheter Regulatory Class: Class II, 21 CFR 870.1250 Product Code: DQY Additional Product Code: KRA
## PREDICATE DEVICE
Headway 27 Microcatheter (K142449)
#### REFERENCE DEVICE
Headway 17 Microcatheter (K101542)
#### DEVICE DESCRIPTON
The PG Pro Microcatheter is a single lumen catheter designed to be introduced over a steerable guidewire to access small, tortuous vasculature. The semi-rigid proximal section transitions to a flexible distal tip to facilitate advancement through vessels. Dual radiopaque markers at the distal end facilitate fluoroscopic visualization. The outer surface of the catheter is coated with a hydrophilic polymer to increase lubricity. A luer fitting on the catheter hub is used for the attachment of accessories.
#### INDICATION FOR USE
The PG Pro Microcatheter is intended for the peripheral vasculature for the infusion of diagnostic and therapeutic agents.
{4}------------------------------------------------
# TECHNOLOGICAL CHARACTERISTICS COMPARISON TABLE
| Table 1: Headway Microcatheters and PG Pro Comparison table | | | | |
|-------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Property | Headway 17<br>Microcatheter<br>Reference Device | Headway 27<br>Microcatheter<br>Predicate Device | PG Pro Microcatheter<br>Subject Device | Differences |
| | (K101542) | (K142449) | | |
| Indication For<br>Use | The Headway 17<br>Microcatheter is<br>intended for general<br>intravascular use,<br>including the peripheral,<br>coronary and<br>neurovasculature for the<br>infusion of diagnostic<br>agents, such as contrast<br>media, and therapeutic<br>agents, such as occlusion<br>coils. | The Headway 27<br>Microcatheter is intended<br>for general intravascular<br>use, including the<br>peripheral, coronary and<br>neurovasculature for the<br>infusion of diagnostic<br>agents, such as contrast<br>media, and therapeutic<br>agents, such as occlusion<br>coils. | The PG Pro<br>microcatheter is<br>intended for the<br>peripheral vasculature<br>for the infusion of<br>diagnostic and<br>therapeutic agents. | The PG Pro<br>Microcatheter is<br>intended only<br>for the<br>peripheral<br>vasculature. |
| Principle of<br>Operation | The principle of<br>operation incorporates<br>various design features<br>that allow it to achieve<br>its intended use. A semi-<br>rigid proximal section<br>allows for steerability.<br>Progressively softer<br>durometer distal<br>segments and flexible tip<br>allow for atraumatic<br>navigation through the<br>vasculature. Radiopaque<br>markers allow for<br>fluoroscopic<br>visualization.<br>Hydrophilic polymer<br>coating reduces friction | The principle of operation<br>incorporates various<br>design features that allow<br>it to achieve its intended<br>use. A semi-rigid<br>proximal section allows<br>for steerability.<br>Progressively softer<br>durometer distal segments<br>and flexible tip allow for<br>atraumatic navigation<br>through the vasculature.<br>Radiopaque markers<br>allow for fluoroscopic<br>visualization.<br>Hydrophilic polymer<br>coating reduces friction<br>during navigation through<br>the vasculature. | The principle of<br>operation incorporates<br>various design features<br>that allow it to achieve<br>its intended use. A semi-<br>rigid proximal section<br>allows for steerability.<br>Progressively softer<br>durometer distal<br>segments and flexible<br>tip allow for atraumatic<br>navigation through the<br>vasculature. Radiopaque<br>markers allow for<br>fluoroscopic<br>visualization.<br>Hydrophilic polymer<br>coating reduces friction | Identical |
| Table 1: Headway Microcatheters and PG Pro Comparison table | | | | |
| Property | Headway 17<br>Microcatheter<br>Reference Device<br>(K101542) | Headway 27<br>Microcatheter<br>Predicate Device<br>(K142449) | PG Pro Microcatheter<br>Subject Device | Differences |
| | during navigation<br>through the vasculature. | | during navigation<br>through the vasculature. | |
| Material(s) | Catheter body: Pebax,<br>stainless steel, Grilamid,<br>inner liner of PTFE and<br>barium sulfate<br>incorporated in distal<br>Pebax segment. | Catheter body: Pebax,<br>stainless steel, Grilamid,<br>inner liner of PTFE | Catheter body: Pebax,<br>Tungsten, Grilamid,<br>inner liner of PTFE | PG Pro has a<br>Tungsten coil<br>versus a<br>Stainless-steel<br>coil for the<br>Headway 27 and<br>Headway 17<br>Microcatheters. |
| | Marker band: Pt/Ir | Marker band: Pt/Ir | Marker band: Pt/Ir | Identical |
| | Hub: Nylon | Hub: Nylon | Hub: Nylon | Identical |
| | Strain relief: Pebax | Strain relief: Pebax | Strain relief: Pebax | Identical |
| | Introducer: Pebax | Introducer: Pebax | Introducer: Pebax | Identical |
| | Shaping mandrel:<br>Stainless steel | Shaping mandrel:<br>Stainless steel | Shaping mandrel: N/A | PG Pro does not<br>come with a<br>Shaping<br>mandrel. |
| Proximal<br>ID/OD | ID = .0170" min<br>OD = .031" | ID = .027" min<br>OD = .040" | ID = .027" min<br>OD = .038"MAX | PG Pro has a<br>Max Proximal<br>O.D of .038"<br>which is in-<br>between<br>Headway 17 and |
| Table 1: Headway Microcatheters and PG Pro Comparison table | | | | |
| Property | Headway 17<br>Microcatheter<br>Reference Device<br>(K101542) | Headway 27<br>Microcatheter<br>Predicate Device<br>(K142449) | PG Pro Microcatheter<br>Subject Device | Differences |
| Distal ID/OD | ID = .0170" min<br>OD = .022" | ID = .027" min<br>OD = .034" | ID = .027" min<br>OD = .038" MAX | 27<br>Microcatheter<br>proximal OD.<br>PG Pro has a<br>Max Distal O.D<br>of .038" |
| No. of Marker<br>bands | 2 | 2 | 2 | Identical |
| Coating | Hydrophilic Coating | Hydrophilic Coating | Hydrophilic Coating | Identical |
| Effective<br>Length | 150 cm | 150 cm and 156 cm | 140 cm and 165 cm | PG Pro comes in<br>two lengths –<br>140 and 165 cm. |
| Tip<br>Configuration | Preshaped tips – with the<br>option of secondary<br>shaping for proper<br>adjustment to the<br>anatomy prior to use. | Straight - Steam<br>Shapeable by physician<br>prior to use | Straight configuration | PG Pro is<br>available only in<br>the straight<br>configuration. |
| Guidewire<br>Compatibility | 0.014 "wires or smaller | 0.014 "wires or smaller | 0.021 "wires or smaller | PG Pro is<br>compatible with<br>0.021" wires or<br>smaller |
| Method of<br>supply | Sterile and single use | Sterile and single use | Sterile and single use | Identical |
| Accessories | Shaping mandrel and<br>Introducer sheath | Shaping mandrel and<br>Introducer sheath | Introducer sheath | PG Pro does not<br>include shaping<br>mandrel |
{5}------------------------------------------------
K192625
Page 3 of 6
{6}------------------------------------------------
K192625
Page 4 of 6
{7}------------------------------------------------
# PERFORMANCE DATA
# Biocompatibility testing
Biocompatibility evaluation for the PG Pro Microcatheter was conducted in accordance with ISO10993-1 Biological Evaluation of Medical Devices-Part 1: Guidance on Selection of Tests. The microcatheter is considered a limited (<24hour) circulating blood contacting device.
The battery of testing for the device included the following tests:
- Cytotoxicity ●
- Sensitization
- Irritation or Intracutaneous Reactivity ●
- Acute Systemic Toxicity
- Hemocompatibility
# Bench testing
Bench testing conducted for the PG Pro Microcatheter included the following:
- Surface Contamination ●
- Physical Attributes ●
- Force at break (Hub and distal portion of the catheter) ●
- Liquid leakage
- Air leakage
- Static burst
- Durability and lubricity of hydrophilic coating
- Simulated use in flow model ●
- Dynamic burst
- Catheter flow rate
- Kink resistance
- Stiffness testing
- Catheter flexural fatigue
- Particulate testing
- Catheter dead space ●
- Torque strength
- Hub testing
- Corrosion testing ●
- Radio detectability ●
# Animal Study
The PG Pro Microcatheter was evaluated in an acute large animal study (n=3) The safety profile as assessed through device safety, compatibility with therapeutic agents and tissue response showed that the PG Pro Microcatheter performed similar to the predicate device. Headway 27 Microcatheter.
{8}------------------------------------------------
# CONCLUSION
The PG Pro Microcatheter is substantially equivalent to the identified predicate regarding performance, intended use, design, materials, principle of operation and overall technological characteristics. The nonclinical data supports the substantial equivalence of the subject device and the verification and validation testing demonstrate that the subject device should perform as intended when used as instructed in the instructions for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.