K142449 · MicroVention, Inc. · DQY · Mar 26, 2015 · Cardiovascular
Device Facts
Record ID
K142449
Device Name
Headway 27 Microcatheter
Applicant
MicroVention, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Mar 26, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Headway Microcatheter is intended for general intravascular use, including the peripheral, coronary and neuro vasculature for the infusion of diagnostic agents, such as contrast media, and therapeutic agents, such as occlusion coils.
Device Story
Single lumen microcatheter; designed for navigation through small, tortuous vasculature over steerable guidewire. Features semi-rigid proximal section transitioning to flexible distal tip; dual radiopaque markers for fluoroscopic visualization. Supplied with introducer sheath and shaping mandrel. Used by physicians in clinical settings (e.g., interventional radiology/cardiology). Physician manually shapes distal tip via steam; advances catheter under fluoroscopic guidance. Infuses contrast media or therapeutic agents (occlusion coils) to treat vascular conditions. Benefits include access to distal, complex anatomy for targeted delivery of diagnostic or therapeutic materials.
Clinical Evidence
Bench testing only. Testing included surface contamination, dimensional attributes, force at break, leakage (liquid/air), burst pressure, coating durability/lubricity, tip shape retention, simulated use, flow rate, kink resistance, and fatigue testing. Biocompatibility testing performed per ISO 10993 standards (cytotoxicity, sensitization, irritation, hemocompatibility, systemic toxicity, pyrogenicity). All tests passed.
Indicated for patients requiring intravascular infusion of diagnostic agents (contrast media) or therapeutic agents (occlusion coils) within peripheral, coronary, or neuro vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is placed to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 26, 2015
MicroVention, Inc. Ms. Cynthia Valenzuela Sr. International Regulatory Affairs 1311 Valencia Avenue Tustin, California 92780
Re: K142449 Trade/Device Name: Headway 27 Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: January 19, 2015 Received: February 27, 2015
Dear Ms. Valenzuela,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142449
Device Name Headway 27 Microcatheter
Indications for Use (Describe)
The Headway Microcatheter is intended for general intravascular use, including the peripheral, coronary and neuro vasculature for the infusion of diagnostic agents, such as contrast media, and therapeutic agents, such as occlusion coils.
Type of Use (Select one or both, as applicable)
| <div> <input checked="true" type="checkbox"/> Research Use (Part 21 CFR 312) or Screening Purposes </div> |
|------------------------------------------------------------------------------------------------------------------|
| <div> <input type="checkbox"/> Compassionate Use (21 CFR 312.310) </div> |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 1.3. 510(k) Summary
| Trade Name: | Headway 27 Microcatheter |
|-------------------|-------------------------------------------------------------------------|
| Generic Name: | Percutaneous Catheter |
| Classification: | Class II, 21 CFR 870.1250 (DQY), 21CFR 870.1200 (DQO) |
| Submitted By: | MicroVention, Inc.<br>1311 Valencia Avenue<br>Tustin, California U.S.A. |
| Contact: | Cynthia Valenzuela (714) 247 8053 or (949) 413-0071 |
| Date: | 2014 AUG 06 |
| Predicate Device: | Headway 27 Microcatheter (K110813) |
## 510(k) Summary
### Device Description:
The Headway 27 Microcatheter is a single lumen catheter designed to be introduced over a steerable guidewire to access small, tortuous vasculature. The semi-rigid proximal section transitions to a flexible distal tip to facilitate advancement through vessels. Dual radiopaque markers at the distal end facilitate fluoroscopic visualization. An introducer sheath and shaping mandrel are also provided.
### Indications For Use:
The Headway Microcatheter is intended for general intravascular use, including the peripheral, coronary and neuro vasculature for the infusion of diagnostic agents, such as contrast media, and therapeutic agents, such as occlusion coils.
| | Headway 27 (predicate) | Headway 27 (proposed) |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Intended Use | The Headway microcatheter is intended<br>for general intravascular use, including the<br>peripheral, coronary and neurovasculature-<br>for the infusion of diagnostic and<br>therapeutic agents. | Same |
| Material | | Same |
| | Catheter Body<br>Outer layer of polyurethane elastomer<br>(Polyblend and Pellethane), polyether block<br>amide (Pebax) and polyamide (Grilamid); inner<br>layer of stainless steel braid/coil, PTFE and<br>polyolefin elastomer | |
| | Marker<br>Platinum/Iridium | |
| | Hub<br>Nylon | |
| | Strain Relief<br>Pebax | |
| | Introducer<br>Pebax | |
| Shaping Mandrel | Stainless steel | |
#### Technological Comparison:
{4}------------------------------------------------
| Proximal ID/OD | ID = .027" min<br>OD = .040" | ID = .0278" min<br>OD = .040" |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|
| Distal ID/OD | ID = .027" min<br>OD = .034" | ID = .027" min<br>OD = .035" |
| Effective Length | 150 cm | 156 cm |
| Coating | Hydrophilic coating<br>(Hydak® - same) | Same |
| Tip Configuration | Straight - Steam Shapeable by physician<br>prior to use | Same |
| Guidewire Compatibility | 0.018" wires or smaller | 0.014"/0.016"/0.018"/0.021" |
| Accessories | Introducer sheath and shaping mandrel | Same |
| Method of Supply | Sterile and single use | Same |
| Sterilization Method | Ethylene Oxide | Same |
| Packaging Configuration | Catheter placed into a HDPE dispenser tube.<br>Dispenser tube, introducer and shaping mandrel<br>placed on a polyethylene packaging card that is<br>inserted into a Tyvek® pouch. Pouch and IFU<br>placed in bleached sulfate carton box. | Same |
#### Technological Comparison (conti.):
## Verification and Test Summary:
| Bench Testing | Acceptance Criteria | Result | |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Surface Contamination | • Liquid on surface<br>• Particulate on external<br>surface<br>• Surface defects/sharp<br>edges | • Free from uncured<br>hydrophilic coating.<br>• No surface<br>particulate > .02 mm²<br>per tappi chart<br>• Free from surface<br>defect/no sharp edges<br>• Embedded particulate<br>acceptable if OD is in<br>specification<br>• Free from damage | Passed |
| Dimensional Attributes | • Catheter effective<br>length<br>• Catheter lumen<br>• Catheter outer<br>diameter<br>• Length of distal OD<br>(2.2Fr section) | • 150 ± 2 cm<br>• .027" (0.69 mm)<br>• nominal .040"/.034-<br>.028" (1.0/.86--.71<br>mm)<br>• ≥ 6cm | Passed |
| Force at break | Device shall not break<br>during use | ≥ 1.12 lbs (5.0 N) for<br>outer diameters from<br>.030" to .045" (.76 to<br>1.1 mm) | Passed |
{5}------------------------------------------------
| Bench Testing | Acceptance Criteria | Result | |
|-----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|--------|
| Freedom from Leakage<br>(Liquid) | (low pressure - long<br>duration)<br>Device shall not leak<br>fluids | No liquid leaking from<br>hub and catheter shaft at<br>46 psi (317.2 kPa) for<br>30 second duration | Passed |
| Freedom from Leakage<br>(air) | (high pressure - short<br>time)<br>Device shall not leak<br>fluids | No liquid leaking from<br>hub and catheter shaft at<br>300 psi/2068 kPa (rated<br>burst pressure) for 10<br>second duration | Passed |
| Burst Pressure of Catheter | Air shall not leak into<br>device | No air leaking into<br>syringe for 15 seconds | Passed |
| Dynamic Burst Pressure | Microcatheter will not<br>burst statically below<br>rated burst pressure. | Microcatheter: will not<br>burst below 300 psi<br>(2068 kPa) | Passed |
| Durability and Lubricity of Hydrophilic Coating-<br>Verification that hydrophilic coating does not delaminate<br>during use | | Rated 3 or higher<br>(simulated use) | Passed |
| Tip Shape and Tip Retention | | Tip retain better than<br>55% of its original<br>shape | Passed |
| Simulated Use | | Rated 3 or higher in<br>tested categories | Passed |
| Compatibility with agents | | Rated 3 or higher in<br>tested applicable<br>categories | Passed |
| Flow Rate | Reference data | N/A | |
| Kink Resistance | | Equivalent to or better<br>kink resistance than<br>competitive | Passed |
| Catheter Stiffness | | Document stiffness<br>using Tinius Olsen -<br>reference data only | N/A |
| Catheter Flexural Fatigue | The catheter must have<br>acceptable results per<br>the following<br>conditions:<br>- Flexural fatigue:<br>simulated use, tip<br>shaping testing<br>- Hoop stress fatigue:<br>flow rate, dynamic<br>burst, liquid leakage, | Passed | |
{6}------------------------------------------------
| Bench Testing | Acceptance Criteria | Result |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------|--------|
| Catheter Particle Testing | Per USP <788> - less<br>than 25 particles greater<br>than 10 microns and less<br>than 3 particles greater<br>than 25 micron | Passed |
| Dead Space | Reference data | N/A |
| Torque Test | 50 rotations without<br>catheter breakage or<br>equivalent to<br>competitive product<br>catheters. | Passed |
| DMSO Test | Functional performance<br>and chemical stability | Passed |
| Biocompatibility | Results | Conclusion |
|-----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|-----------------------------------------------------|
| Cvtotoxicity ISO 10993-5<br>MEM Elution Test | Cell culture tested with test article<br>exhibited slight reactivity (Grade 1) | Non-toxic |
| Cytotoxicity ISO 10993-5<br>Cell Culture Agar Overlay | Grade 2: zone limited to under<br>specimen | Non-toxic |
| Sensitization ISO 10993-10<br>Guinea Pig Maximization Test | Grade 0: No visible change | Non-irritant |
| Irritation ISO 10993-10<br>Intracutaneous Reactivity Evaluation<br>Test | Comparative between control and test<br>article <1.0 | Non-irritant |
| Hemocompatibility - Rabbit Blood<br>Direct Contact (ISO10993-4) | 0.1% hemolysis | Non-hemolytic |
| Hemocompatibility - Unactivated<br>Partial Thromboplastin Time Assay<br>Direct Contact (ISO10993-4) | Test article = 12.6 seconds<br>(Normal PT clotting range = 10-14<br>seconds) | No effect on coagulation |
| Hemocompatibility - Complement<br>Activation Assay (ISO10993-4) | C3a = 17.7%<br>SC5b-9= 1.6 % | No effect on<br>complement activation |
| Hemocompatibility -<br>Thrombogenicity Study in Dogs<br>(ISO10993-4) | Minimal to moderate thrombus<br>formation, clotting ability was not<br>compromised | Similar<br>thromboresistance<br>compared to control |
| Systemic toxicity - Systemic Injection<br>Test<br>(ISO10993-11) | No decrease in body weight of >10%.<br>No reaction found. | Non-toxic effects |
| Systemic toxicity - Rabbit Pyrogen<br>Test (ISO10993-11) | Temperature increases was 0.0℃ from<br>baseline. | Non-pyrogenic |
{7}------------------------------------------------
### Summary of Substantial Equivalence:
The data presented in this submission demonstrates the technological similarity and equivalency of the Headway 27 Microcatheter when compared with the predicate device, Headway 27 Microcatheter (K110813).
The devices,
- " Have the same intended use,
- Use the same operating principle,
- Incorporate the same basic design,
- . Are packaged and sterilized using same methods.
In summary, the Headway 27 Microcatheter described in this submission was found to have a safety and effectiveness profile that is similar to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.