K192584 · Mermaid Care A/S · BZK · Dec 18, 2019 · Anesthesiology
Device Facts
Record ID
K192584
Device Name
BEACON Caresystem Model 00002144
Applicant
Mermaid Care A/S
Product Code
BZK · Anesthesiology
Decision Date
Dec 18, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1850
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The intended use of the BEACON Caresystem is to provide respiratory mechanics monitoring in adult patients in the Intensive Care Unit (ICU) on-demand ventilator-dependent open-loop advice in mechanically ventilated adult patients as prescribed by the clinician Patients should be hemodynamically stable. The device is intended for use by properly trained clinicians. BEACON Caresystem is for prescription use, only.
Device Story
BEACON Caresystem is an ICU ventilation assist system providing on-demand, open-loop advice for ventilator settings. Inputs include real-time airway flow, CO2/O2 gas fractions (via BEACON Gas Module), SpO2 (via NONIN XPOD), data from connected third-party mechanical ventilators, and manual clinician entries. The system processes these inputs to generate ventilator setting recommendations displayed on a tablet. Used in the ICU by trained clinicians, the device acts as an adjunct to mechanical ventilators (e.g., Maquet Servo I, Dräger Evita XL, Hamilton G5). Clinicians review the advice to inform ventilator adjustments, potentially optimizing patient respiratory support and breathing effort. The system integrates spirometry, gas monitoring, and pulse oximetry into a single interface.
Clinical Evidence
Five clinical studies submitted. Three studies demonstrated high clinical agreement between BEACON advice and clinical practice. One study using esophageal catheterization as a reference technique confirmed the device appropriately responded to over/under support in spontaneous breathing patients. A fifth study involved a blinded comparison of BEACON advice versus active clinicians. Results support the device's ability to provide appropriate ventilator setting recommendations.
Technological Characteristics
System comprises a tablet display, gas module, power adaptor, and SpO2 analyzer. Complies with IEC 60601-1, IEC 60601-1-2, IEC 80601-2-61, and AIM 7351731. Software lifecycle per IEC 62304; usability per IEC 62366-1. Connectivity via USB and serial ports to ventilators. Measurement principles: pulse oximetry, airway flow/volume, and gas analysis (CO2/O2).
Indications for Use
Indicated for respiratory mechanics monitoring and on-demand ventilator-dependent open-loop advice for hemodynamically stable adult patients in the ICU undergoing mechanical ventilation. For use by trained clinicians; prescription only.
Regulatory Classification
Identification
A monitoring spirometer is a device used to measure continuously a patient's tidal volume (volume of gas inhaled by the patient during each respiration cycle) or minute volume (the tidal volume multiplied by the rate of respiration for 1 minute) for the evaluation of the patient's ventilatory status.
Predicate Devices
Philips NM3 Respiratory Monitor with VentAssist (K103578)
Reference Devices
GE Healthcare Datex-Ohmeda S/5TM Compact Airway Module (model family E-CAIOVX) E-CAIOVX, E-CAIOV, E-CAIO, E-COVX, E-COV, E-CC (K051092)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 18, 2019
Mermaid Care A/S Claus Lindholt Chief Technology Officer Hedelund 1 Norresundby, DK DK-0400
Re: K192584
Trade/Device Name: BEACON Caresystem, Model 00002144 Regulation Number: 21 CFR 868.1850 Regulation Name: Monitoring Spirometer Regulatory Class: Class II Product Code: BZK Dated: September 16, 2019 Received: September 19, 2019
Dear Claus Lindholt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James J. Lee Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
K192584
Device Name
BEACON Caresystem, Model 00002144
Indications for Use (Describe)
The intended use of the BEACON Caresystem is to provide
- · respiratory mechanics monitoring in adult patients in the Intensive Care Unit (ICU)
- · on-demand ventilator-dependent open-loop advice in mechanically ventilated adult patients as prescribed by the clinician
Patients should be hemodynamically stable.
The device is intended for use by properly trained clinicians. BEACON Caresystem is for prescription use, only.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K192584 510(K) SUMMARY
# MERMAID CARE
#### 510(k) Summary in accordance with 21 CFR 807.92
| Submitted by: | Mermaid Care A/S<br>Hedelund 1<br>DK-9400 Noerresundby<br>Denmark | | |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|---------------------|
| | Tel.: +45 70 23 70 15 | | |
| Contact Person: | Claus Lindholt, CTO | | |
| Date of Summary: | September 16, 2019 | | |
| Device Trade Name: | BEACON Caresystem, Model 00002144 | | |
| Common or Usual<br>Name: | Monitoring Spirometer (Product code: BZK) | | |
| Classification Name: | Spirometer, Monitoring (W/Wo Alarm) [21 C.F.R. §868.1850] | | |
| Class: | II | | |
| Predicate Device: | Philips NM3 Respiratory Monitor with VentAssist (K103578) | | |
| Reference Device and<br>accessories: | GE Healthcare Datex-Ohmeda S/5TM Compact Airway Module (model<br>family E-CAIOVX) E-CAIOVX, E-CAIOV, E-CAIO, E-COVX, E-COV, E-CC<br>(K051092) | | |
| Reference Product<br>Codes: | Code: | Regulatory Description: | Regulation Number: |
| | BZL | Oxygen uptake computer | 21 C.F.R. §868.1730 |
| | DQA | Oximeter | 21 C.F.R. §870.2700 |
| | CCK | Carbon dioxide gas analyzer | 21 C.F.R. §868.1400 |
| | CCI | Oxygen gas analyzer | 21 C.F.R. §868.1720 |
#### Device Description:
BEACON Caresystem is an Intensive Care Unit (ICU) ventilation assist system. BEACON Caresystem will provide ICU clinicians with on-demand ventilator-dependent open-loop advice for ventilating adult patients.
BEACON Caresystem uses inputs measured by BEACON Caresystem and data inputs from the ventilator and manual data entry by the clinician into a computer program that has been designed to provide advice for changes to the ventilator settings for the individual patient undergoing mechanical ventilation.
BEACON Caresystem is an adjunct to third party legally marketed ICU mechanical ventilators. Currently supported mechanical ventilators are:
- Maquet Servo I [K073179] ●
- Dräger Evita XL [K083050]
- Dräger Infinity V500 [K093633] .
- Puritan Bennett PB840 [K001646]
- Puritan Bennett PB940 [K131252] ●
- Hamilton G5 [K070513]. ●
{4}------------------------------------------------
BEACON Caresystem consists of the following components:
- BEACON Display Unit
- BEACON Gas Module
- BEACON Power Adaptor ●
- NONIN XPOD SpO2 Analyzer .
BEACON Caresystem utilizes the following cleared accessories:
- BEACON Flow Sensor (Adult) private label version of Metaphor E-Z Flow Sensor, Size Adult ● [K0930801
- BEACON Water Trap (Adult) [K182075 & K171292] .
- NONIN Finger Clip SpO2 sensor NONIN 8000AA1 [K071285] .
- NONIN Ear Clip SpO2 sensor- NONIN 8000Q2 [K080255]
- NONIN Forehead Reflectance SpO2 sensor NONIN 8000R [K071285] ●
The BEACON Display Unit is a standard off-the-shelf tablet touch-screen computer. The BEACON display unit connects through a USB port to the NONIN XPOD SpO2 analyzer for pulse oximetry measurements and serial data ports for connection to the BEACON Gas Module and the connected ICU ventilator. The BEACON Display unit includes the BEACON Caresystem Software. The BEACON Gas Module performs measurement of patient airway flow and inspired and expired fractions of CO2 and O2 in the airway.
As described above, BEACON Caresystem combines the functionalities of a spirometric monitor, respiratory gas monitor, and pulse oximeter. Each of these functions, respectively, is intended for:
- . Respiratory mechanics monitoring (i.e. airway flow/volume)
- . Measurement of expired and inspired breathing gases (CO2, O2) and respiration rate
- . Non-invasive measurement of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate.
The BEACON Caresystem Software utilizes the monitored parameters above with data from the connected ICU ventilator and clinician entered values to generate on demand, open-loop advice for changes to the ventilator settings for an adult patient undergoing mechanical ventilation.
#### Indications for Use:
The intended use of the BEACON Caresystem is to provide
- respiratory mechanics monitoring in adult patients in the Intensive Care Unit (ICU)
- on-demand ventilator-dependent open-loop advice in mechanically ventilated adult patients as prescribed by the clinician
Patients should be hemodynamically stable.
The device is intended for use by properly trained clinicians. BEACON Caresystem is for prescription use, only.
The Indications for Use statement for the BEACON Caresystem is not identical to the predicate device; however, the differences do not alter the intended clinical use of they affect the substantial equivalence to the predicate.
#### Comparison of the technological characteristics with the Predicate Device and Reference Device:
BEACON Caresystem and the predicate device both continuously measure pulse oximetry, inspiratory and expiratory air flow and gas and provide on-demand advice on ventilator settings based on the patient specific measurements on a display unit.
{5}------------------------------------------------
At a high level, the subject and predicate devices are based on the following same technological elements:
- . Pulse oximetry
- . Respiratory flow and carbon dioxide gas measurements and signal processing
- . Advice generation of mechanical ventilator settings
- Graphical display of measurements and advice for the user ●
The following technological differences exist between the subject and predicate devices:
- Subject device measures respiratory Oxygen gas fraction
- Subject device requires connection to a supported mechanical ventilator for advice generation, . whereas ventilator connectivity is optional in predicate device
The Reference Device has been included to account for the addition of Oxygen measurements in the subject device. The subject and reference devices uses similar measurement technologies and testing has shown comparable performance.
#### Non-Clinical Performance Tests Submitted:
BEACON Caresystem has been designed and tested for compliance to the following recognized standards:
- . IEC 60601-1:2012 Ed. 3.1, Medical electrical equipment-Part1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014, ed.4. Medical electrical equipment-Part 1-2: General requirements for basic safety and essential performance-Collateral standard: Electromagnetic compatibility Requirements and tests.
- . IEC 80601-2-61:2017 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment.
- . AIM Standard 7351731 Rev. 2.00 2017-02-23. Medical Electrical Equipment and System Electromagnetic Immunity Test for Exposure to Radio Frequency Identification Readers
- IEC 62304 Edition 1.1 2015-06, Medical Device Software Software Life Cycle Processes. .
- IEC 62366-1 Edition 1.0 2015-02, Medical Devices Part 1: Application Of Usability Engineering To ● Medical Devices [Including Corrigendum 1 (2016)].
- . ISTA 3A 2008, Packaged-Products For Parcel Delivery System Shipment 70 Kg (150 Lb) Or Less.
A head-to-head test comparing the advice presented by BEACON and the predicate device was performed on a patient simulator and the results showed that the BEACON advice generation is comparable to the predicate device.
Gas and flow measurements in BEACON has been evaluated in bench test comparisons with predicate and reference devices. End-tidal Carbon Dioxide (EtCO2), Alveolar Ventilation (VA), Respiratory Rate (RR) measurements were tested against the predicate device. Oxygen Uptake (VO2), Carbon Dioxide Production (VCO2) and Energy Expenditure (EE) measurements were tested against the reference device. All tests showed that the measurements of BEACON is comparable to the predicate or reference devices in the full ranges specified.
BEACON Caresystem has two patient contacting accessories, which are both FDA-cleared and legally marketed, so no new biocompatibility testing was performed.
#### Clinical Tests Submitted:
Five clinical studies where the advice generation of BEACON Caresystem were tested have been submitted.
{6}------------------------------------------------
Three studies evaluated the clinical agreement of BEACON Caresystem advice and the effects of the applied advice to the patients' respiration and oxygenation. The studies concluded that BEACON Caresystem generated advice with high clinical practice agreement.
One study evaluated the ability of BEACON Caresystem advice to find appropriate breathing effort levels for spontaneous breathing patients. The reference technique esophageal catheter was used to estimate breathing effort. The study concluded that BEACON Caresystem responded appropriately to over and under support.
The fifth study evaluated blindly the BEACON Caresystem advice and active clinicians in patient cases.
#### Conclusion:
The BEACON Caresystem utilizes equivalent technology and has similar technical specifications as the predicate device and the reference device, the Philips NM3 Respiratory Monitor with VentAssist and Reference Device, the GE Healthcare Datex-Ohmeda S/5™ Compact Airway Module (model family E-CAiOVX and accessories), respectively.
The detailed comparison and the submitted non-clinical performance testing and clinical data demonstrate that the BEACON Caresystem performs comparably to the predicate and reference device and is therefore substantially equivalent to the legally marketed predicate device, Philips NM3 Respiratory Monitor with VentAssist [K103578].
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.