The Datex-Ohmeda S/5™ Compact Airway Module, E-CAiOVX family is intended to be used with Datex-Ohmeda S/5 patient monitors for monitoring respiratory, ventilatory and gas exchange parameters of hospitalized patients.
Device Story
The E-CAiOVX is a plug-in module for Datex-Ohmeda S/5 patient monitors. It utilizes an AOVA infrared sensor for CO2, N2O, and anesthetic agents, a paramagnetic sensor for O2, and side-stream spirometry for airway pressure, volume, and flow. It calculates gas exchange parameters (VO2, VCO2, RQ, EE). The device connects to the patient via sampling lines, water traps, and D-lite airway adapters. Operated by qualified medical personnel in clinical settings, it provides real-time numeric, waveform, and loop data on the host monitor. Auditory and visual alarms alert clinicians to parameter limit breaches. The system assists clinicians in assessing patient respiratory and ventilatory status to guide clinical decision-making and patient management.
Clinical Evidence
Bench testing only. The device was verified and validated against international standards including IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 9918 (capnometers), ISO 7767 (oxygen monitors), and ISO 11196 (anaesthetic gas monitors).
Technological Characteristics
Double-width plug-in module. Sensing: AOVA infrared (CO2, N2O, anesthetic agents), paramagnetic (O2), side-stream spirometry. Connectivity: Proprietary interface for Datex-Ohmeda S/5 monitors. Software: Embedded firmware. EMC: Shielded mechanics with foam. Sterilization: Not applicable (external accessories). Standards: IEC 60601-1, ISO 9918, ISO 7767, ISO 11196, ASTM F-1456.
Indications for Use
Indicated for monitoring hospitalized patients' respiration (CO2, O2, N2O, anesthetic agents, agent identification, respiration rate), ventilation (airway pressure, volume, flow), and gas exchange status (VO2, VCO2, RQ, EE). Contraindicated for gas exchange monitoring in presence of N2O+O2 mixtures. For use by qualified medical personnel only.
Regulatory Classification
Identification
An oxygen gas analyzer is a device intended to measure the concentration of oxygen in respiratory gases by techniques such as mass spectrometry, polarography, thermal conductivity, or gas chromatography. This generic type of device also includes paramagnetic analyzers.
{0}------------------------------------------------
MAY 2 5 2005
KOS1092
# Premarket Notification 510(k) Summary As required by section 807.92 Datex-Ohmeda S/5TM Compact Airway Module (model family E-CAIOVX) E-CAiOVX, E-CAiOV, E-CAiOV, E-CAiO, E-COVX, E-COV, E-CO and accessories.
# GENERAL COMPANY INFORMATION as required by 807.92(a)(1)
#### COMPANY NAME/ADDRESS/PHONE/FAX:
GE Healthcare 86 Pilgrim Road Needham, MA 02492 USA Tel: 781-449-8685 Fax: 781-433-1344
#### NAME OF CONTACT.
Mr. Joel Kent
DATE:
April 19, 2005
## DEVICE NAME as required by 807.92(a)(2)
#### TRADE NAME:
Datex-Ohmeda S/5™ Compact Airway Module (model family E-CAiOVX) E-CAiOVX, E-CAiOV, E-CAiO, E-COVX, E-COV, E-CO and accessories.
#### COMMON NAME:
Airway gas, pressure and volume, anesthetic agent and agent identification and gas exchange monitor. Airway gas and Patient Spirometry accessories.
{1}------------------------------------------------
#### CLASSIFICATION NAME:
## The following Class II classifications appear applicable:
| Product Code | Classification Name | CFR Section |
|--------------|-------------------------------------------------------------|-------------|
| CCK | Analyzer, Gas, Carbon-Dioxide, Gaseous-phase | 868.1400 |
| CCL | Analyzer, Gas, Oxygen, Gaseous-phase | 868.1720 |
| BZK | Spirometer, Monitoring (W/WO alarm) | 868.1850 |
| CAP | Monitor, Airway Pressure (Includes gauge and/or alarm) | 868.2600 |
| CBR | Analyzer, Gas, Nitrous-Oxide, Gaseous-phase (Anesthetic co) | 868.1700 |
| BZL | Computer, Oxygen-uptake | 868.1730 |
| CBQ | Analyzer, Gas, Enflurane, Gaseous-phase (Anesthetic conc.) | 868.1500 |
| CBS | Analyzer, Gas, Halothane Gaseous-phase (Anesthetic conc.) | 868.1500 |
| NHO | Analyzer, Gas, Desflurane, Gaseous-phase (Anesthetic conc.) | 868.1500 |
| NHQ | Analyzer, Gas, Isoflurane Gaseous-phase (Anesthetic conc.) | 868.1500 |
| NHP | Analyzer, Gas, Sevoflurane, Gaseous-phase (Anesthetic conc) | 868.1500 |
## NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)
The Datex-Ohmeda S/5™ Compact Airway Module family, E-CAiOVX is substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda M-CAiOVX Module (K001814).
#### DEVICE DESCRIPTION as required by 807.92(a)(4)
The Datex-Ohmeda S/5™ Compact Airway Module, E-CAiOVX is a double-width plug-in parameter module for monitoring respiratory (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiration rate), ventilatory (airway pressure, volume and flow) and gas exchange parameters (Oxygen Consumption VO2, Carbon Dioxide production VCO2, Respiratory Quotient RQ, and Energy Expenditure, EE). The E-CAiOVX is a Compact Airway measuring module for a modular monitoring system.
The intended use for the modified device is the same as for the predicate, Datex-Ohmeda M-CAiOVX module (K001814) The indications for use are also the same.
There has been no change to the basic technology from the predicate. The E-CAiOVX module is a facelifted version of the predicate M-CAiOVX module. The module cover and mechanics have changed, but the fundamental scientific technology is the same as in the predicate device M-CAiOVX (K001814). The electronic measurement boards are the same in E-CAiOVX and M-CAiOVX, except for one component type change (EEPROM), and the software is basically the same, although both the software code and revision has changed during the five years since the predicate submission. Software changes include mostly improvements simplifying manufacturing and tests during manufacturing. There are also some enhancements in the mechanics of the E-CAiOVX modulc compared with the predicate M-CAiOVX, e.g. EMC cover is better implemented into the mechanics. The EMC specifications for the E-CAiOVX and M-CAiOVX modules are the same. The Datex-Ohmeda Compact Airway Module module, E-CAiOVX can be used with the following Datex-Olımeda modular monitors with any monitor software (*)
S/5TM Anesthesia Monitor (AM) with monitor software S-STD93 or newer (*)
. S/5™ Compact Anesthesia Monitor (CAM) with monitor software S-STD93 or newer (*)
. S/5TM Critical Care Monitor (CCM) with monitor software S-ICU97 or newer (*)
S/5TM Compact Critical Care Monitor (CCCM).with monitor software S ICU97 or newer (*)
(*) The Gas exchange measurement works only with software versions 99 or newer. All monitors can be upgraded to software version 99 by using the U-LIFE U-xxx99(A).
{2}------------------------------------------------
The E-CAiOVX module consists of:
- AOVA infrared measuring sensor for measuring CO2, N2O and anaesthetic agents
- Paramagnetic O2 sensor .
- Side-Stream Spirometry measurement .
- Gas Exchange measurement
. Halothane, Sevoflurane, Isoflurane and Desflurane).
Harothalle, Sevonurane, Isofturane and Decritical).
The E-CAiOVX module uses the same measurement technology and accessories as the predicate I he E-CAIOVX (K001814). The main accessories include airway gas sampling lines, D-fend device, M- CAIO + A (1001011). In tubing and D-lite sensors. The sampling line and the water traps, Spironicily measurement table connectors. The monitor is switched on and the gas spirometry tube and the spirometry tube is attached to the D-lite™ airway adapter. The D-lite™ is sampling the and the spironionly tace is that and moisture exchanger (HIME) of the patient's intubation tube. The monitor displays measurements from the E-CAiOVX's and subtype modules intubation table. The moment and loops. All the calculated paraneters can be selected on in the form of numeric values, our res, and view on the E-CAiOVX moduels are taken the usplay, and action. And for wear interface for alarms in Datex-Ohmeda S/5 patient care of by the nost monitor and forlow the asser merines. There are auditory and visual alarms and user adjustable limits for the gas measurement variables.
## INTENDED USE as required by 807.92(a)(5)
#### Intended use:
The Datex-Ohmeda S/5™ Compact Airway Module, E-CAiOVX family is intended to be used TIE Datex-Ohmeda 07-5 - Oct plant in monitors for monitoring respiratory, ventilatory and gas exchange parameters of hospitalized patients.
#### Indications for use:
The Datex-Ohmeda S/5M Compact Airway Module, E-CAiOVX family is indicated for The Date.x-Onlineda 075 - Comparation (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiration rate) ventilation (airway pressure, volume and flow) and gas luchange status (Oxygen Consumption VO2, Carbon Dioxide production VCO2, Respiratory Cxchange status (Oxygon Octountiples, EE). Gas exchange status monitoring is not indicated in the Quottent RQ, and Energy Expendhadevice is indicated for use by qualified medical personnel only.
{3}------------------------------------------------
## SUMMARY OF TECHNOLOGICAL CHARACTERITICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6)
The Datex-Ohmeda S/5™ Compact Airway Module family, E-CAiOVX is substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda M-CAiOVX Module (K001814).
The E-CAiOVX module has the following similarities compared to the predicate M-CAiOVX (K001814):
- identical intended use and indications for use .
- identical fundamental scientific technology .
- same electronic measurement board (EEPROM type changed) .
- same module software (version changed from 891977-3.2 to 8001805-4.5) .
- same algorithms for respiratory, ventilatory and gas exchange parameter calculation .
- use the same operating principle ●
- use largely the same accessories ●
- have the same user interface at the monitor and alarms (can be used with the same . monitor software)
- the Customer and parameter specifications are the same, except for one minor . modification
- have the same safety and effectiveness .
- are manufactured using the same processes t
The main differences between the new E-CAiOVX and the predicate M-CAiOVX (K001814) is primarily due to fact that the new E-CAiOVX module has the following changes:
- new color and shape and thus differing mechanics .
- The front panel and labeling have changed .
- The connector for gas return (front panel and accessories) have changed to be ◆ incompatible with female Luer Lock connectors.
- The EEPROM type on the electronic measurement boards have changed .
- The module software has been enhanced and revised several times. Most new revisions . include enhancements for manufacturing and testing during manufacturing.
- The EMC is better implemented into the mechanics, e.g. some module materials changed . and shielding EMC foam was used.
Based on the analysis and other documentation included in this 510(k) notification and Dused on the and you and of the main features and indications for use of the Datex-Ohmeda S/STM Compact Airway Module, E-CAiOVX are substantially equivalent to the predicate Datex-Ohmeda M-CAiOVX Module (K001814).
{4}------------------------------------------------
## SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)
The Datex-Ohmeda S/5TM Compact Airway Module (model family E-CAiOVX) E-CAiOVX, E-The Datex-Onlineda S75 - Compact In-CO and accessories have been assessed against the CAIOV, L-CAIO, L-CO P.R. B OC T, B OO T, E oo oughly tested through validation and verification of specifications.
- FDA/DCRND Reviewer Guidance for Premarket Notification Submissions, November 1993 .
- I DriDCRISD 110101 1:1991 + Amdt. 2:1995 (Part 1: General requirements for safety) .
- IEC 00001-1:1990+ A1:1993 + A13:1996 + A2:1995 (identical to IEC60601-1:1988 + Amdt. . 1:1991 + Amdt. 2:1995)
- 1:1994 Aind. 2:1993601.1-M90 + S1:1994 (Canadian deviations to IEC 60601-1:1988 + ◆ Amdt. 1:1991) + S2:1998 (=IEC Amdt 2:1995)
- Amdt. 1:1971) SE: 27997 (U.S. deviations to IEC 60601-1:1988 + Amdt. 1:1991+ Amdt. . 2:1995)
- IEC 60601-1-2:2001 (Electromagnetic compatibility Requirements and tests) .
- IEC 60601-1-4:2000 (Programmable medical systems) ◆
- ISO 9918 (1993) / EN 864 (1996): Capnometers for use with Humans Requirements .
- ISO 7767:1997 Oxygen monitors for monitoring breathing mixtures. .
- EN 12598:1999 Oxygen monitors for monitoring breathing mixtures. .
- ISO 11196:1995 + Corr. 1:1997 /EN ISO 19996 (1997) Anaesthetic gas monitors ◆
- ASTM F-1456 (2001) Standard specification for Capnometers .
- ADAM I 150 (2007) Datent of Premarket Submission for Software Contained in Medical . Devices. Version 1.0. (May, 29, 1998)
#### CONCLUSION:
The summary above shows that there are no new questions of safety and effectiveness for the The summal y a 010 client module (model family E-CAiOVX) E-CAiOVX) E-CAiOVX, E-Datex-Olmedal 8-5-COVX, E-COV, E-CO and accessories as compared to the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird symbol, composed of three curved lines that suggest the shape of a bird in flight. The logo is encircled by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 5 2005
Mr. Joel Kent Manager, Quality and Regulatory Affairs GE Healthcare 86 Pilgrim Road Needham, Massachusetts 02492
Re: K051092
Trade/Device Name: Datex-Ohmeda S/5™ Compact Airway Module (model family E-CAiOVX) E-CAiOVX, E-CAiOV, E-CAiOV, E-CAiO, E-COVX, E-COV, E-CO and Accessories Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: II Product Code: CCL, BZK, CAP, CBR, BZL, CBQ, CBS, NHO, NHQ, NHP Dated: April 27, 2005 Received: April 28, 2005
Dear Mr. Kent:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvin uppression (the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 Nr.), it hay of base of Federal Regulations, Title 21, Parts 800 to 898. In your device ear or roublish further announcements concerning your device in the Eederal Register.
{6}------------------------------------------------
### Page 2 -- Mr. Kent
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. or the Fet of all the Act's requirements, including, but not limited to: registration 1 od intilet 021 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and instills (21 er retries in the quality systems (QS) regulation (21 CFR Part 820); and if requirements as bet ronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) This tetter "Hiffication. The FDA finding of substantial equivalence of your device to a premaired predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), If you desire specifical of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may obtain of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
🚗
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Datex-Ohmeda S/5™ Compact Airway Module (model family E-CAiOVX) E-CAiOVX, E-CAiOV, E-CAiO, E-CAiO, E-COVX, E-COV, E-CO and accessories.
Indications for Use:
The Datex-Ohmeda S/5™ Compact Airway Module, E-CAiOVX family is indicated for monitoring hospital patient´s respiration (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiration rate) ventilation (airway pressure, volume and flow) and gas exchange status (Oxygen Consumption VO2, Carbon Dioxide production VCO2, Respiratory Quotient RQ, and Energy Expenditure, EE). Gas exchange status monitoring is not indicated in the presence of N2O+O2 mixtures. The device is indicated for use by qualified medical personnel only.
Over-The-Counter Use Prescription Use ___ × AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Cluis Ls
Page _ of
(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Device
510(k) Number: K051042
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.