3MTM AttestTM Super Rapid Steam Biological Indicator 1592, 3MTM AttestTM Super Rapid Steam Challenge Pack, 3MTM AttestTM Auto- Readers 490 and 490H
K192550 · 3M Company · FRC · Jun 5, 2020 · General Hospital
Device Facts
Record ID
K192550
Device Name
3MTM AttestTM Super Rapid Steam Biological Indicator 1592, 3MTM AttestTM Super Rapid Steam Challenge Pack, 3MTM AttestTM Auto- Readers 490 and 490H
Applicant
3M Company
Product Code
FRC · General Hospital
Decision Date
Jun 5, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
Use the 3M Attest™ Super Rapid Steam Biological Indicator 1592 in conjunction with the 3M Attest™ Auto-reader 490 having software version 4.0.0 or greater or the 3M™ Attest™ Auto-reader 490H having software version 4.0.0 or greater to qualify or monitor the following sterilization cycles: Dynamic-air-removal (pre-vacuum and SFPP) 250°F (121°C) 15 minutes; Dynamic-air-removal (pre-vacuum and SFPP) 250°F (121°C) 20 minutes; Dynamic-air-removal (pre-vacuum and SFPP) 250°F (121°C) 30 minutes; Dynamic-air-removal (pre-vacuum and SFPP) 250°F (121°C) 35 minutes; Gravity Displacement 250°F (121°C) 30 minutes. Use the 3M Attest™ Super Rapid Steam Challenge Pack 51582 in conjunction with the 3M™ Attest™ Auto-reader 490 having software version 4.0.0 or greater or the 3M™ Attest™ Auto-reader 490H having software version 4.0.0 or greater to qualify or monitor dynamic-air-removal (pre-vacuum and SFPP) or gravity displacement steam sterilization cycles of 30 minutes at 250°F (121°C).
Device Story
System comprises self-contained biological indicator (BI) 1592, challenge pack 51582, and auto-readers 490/490H. BI 1592 contains Geobacillus stearothermophilus spores and media ampoule in plastic sleeve. After steam sterilization, BI is incubated in auto-reader at 60°C. Auto-reader detects fluorescence resulting from enzymatic hydrolysis of non-fluorescent substrate by active enzyme in spores; fluorescence indicates sterilization failure. Challenge pack 51582 contains BI 1592 and chemical integrator within medical index card layers to simulate AAMI ST79 16-towel process challenge device. Used by healthcare personnel to verify sterilization efficacy. Output is binary (pass/fail) provided via auto-reader interface. Benefits include rapid 24-minute readout for sterilization monitoring.
Clinical Evidence
Bench testing only. Performance verified against ISO 11138-1:2017, ISO 11138-3:2017, and USP <1035>/<55>. Testing confirmed spore population ≥ 10^6, D-value ≥ 1.5 min at 121°C, and Z-value ≥ 10°C. Reduced incubation time (24 minutes) demonstrated >97% alignment with 7-day conventional incubation. Challenge pack resistance verified as equivalent to AAMI ST79 16-towel PCD.
Technological Characteristics
Plastic-housed self-contained biological indicator; Geobacillus stearothermophilus spores; enzymatic fluorescence detection; 60°C incubation; chemical process indicator (pink to brown); challenge pack uses medical index cards; software-controlled auto-reader (v4.0.0+); compliant with ISO 11138-1/3, ISO 11140-1, and ANSI/AAMI ST79.
Indications for Use
Indicated for healthcare facilities to qualify or monitor steam sterilization cycles (dynamic-air-removal and gravity displacement) at 250°F (121°C) using 3M Attest™ Super Rapid Steam Biological Indicator 1592 or 3M Attest™ Super Rapid Steam Challenge Pack 51582 in conjunction with compatible 3M Attest™ Auto-readers.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
Predicate Devices
3M™ Attest™ Super Rapid Readout Biological Indicator 1492V, 3M™ Attest™ Auto-reader 490, and 3M™ Attest™ Auto-reader 490H (K173437)
3M™ Attest™ Super Rapid 5 Steam-Plus Challenge Pack 41482V, 3M™ Attest™ Auto-reader 490, and 3M™ Attest™ Auto-reader 490H (K173519)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".
June 5, 2020
3M Company Hilary Hovde Regulatory Affairs Specialist Bldg. 275-5W-06 St.Paul, Minnesota 55144
## Re: K192550
Trade/Device Name: 3M™ Attest™ Super Rapid Steam Biological Indicator 1592, 3M™ Attest™ Super Rapid Steam Challenge Pack, 3M™ Attest™ Auto- Readers 490 and 490H Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: May 6, 2020 Received: May 7, 2020
## Dear Hilary Hovde:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Christopher K. Dugard, M.S. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K192550
#### Device Name
3MTM Attest™ Super Rapid Steam Biological Indicator 1592 and 3MTM Attest™ Auto-reader 490 and 490H
#### Indications for Use (Describe)
Use the 3MM Attest™ Super Rapid Steam Biological Indicator 1592 in conjunction with the 3MM AttestTM Auto-reader 490 having software version 4.0.0 or greater or the 3MTM Attest™ Auto-reader 490H having software version 4.0.0 or greater to qualify or monitor the following sterilization cycles:
| Cycle Type | Exposure Temperature | Exposure Time |
|-------------------------------------------|----------------------|---------------|
| Dynamic-air-removal (pre-vacuum and SFPP) | 250°F (121°C) | 15 minutes |
| Dynamic-air-removal (pre-vacuum and SFPP) | 250°F (121°C) | 20 minutes |
| Dynamic-air-removal (pre-vacuum and SFPP) | 250°F (121°C) | 30 minutes |
| Dynamic-air-removal (pre-vacuum and SFPP) | 250°F (121°C) | 35 minutes |
| Gravity Displacement | 250°F (121°C) | 30 minutes |
Type of Use (Select one or both, as applicable)
J Prescription Use (Part 21 CFR 801 Subpart D)
> Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid 0MB number."
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K192550
Device Name
3MTM AttestTM Super Rapid Steam Challenge Pack 51582 and 3M™ Attest™ Auto-reader 490 and 490H
#### Indications for Use (Describe)
Use the 3MM Attest™ Super Rapid Steam Challenge Pack 51582 in conjunction with the 3M™ Attest™ Auto-reader 490 having software version 4.0.0 or greater or the 3M™ Attest™ Auto-reader 490H having software versio n 4.0.0 or greater to qualify or monitor dynamic-air-removal (pre-vacuum and SFPP) or gravity displacement steam sterilization cycles of 30 minutes at 250°F (121°C).
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources , gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.go v
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number. "
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the 3M logo. The logo is red and features the number "3" followed by the letter "M". The number and letter are in a bold, sans-serif font. The logo is simple and recognizable.
# 510(k) Summary for K192550 3MTM Attest™ Super Rapid Steam Biological Indicator 1592, 3M™ Attest™ Super Rapid Steam Challenge Pack 51582, and 3M™ Attest™ Auto-readers 490 and 490H
3M Company 3M Health Care 3M Center, Bldg. 275-5W-06 St. Paul, MN 55144-1000
> Contact: Hilary B. Hovde Regulatory Affairs Specialist Phone Number: (651) 736-0364 FAX Number: (651) 737-5320
Submission Date: August 27, 2019
{5}------------------------------------------------
#### Device Name and Classification- Biological Indicator:
| Trade Name: | 3M™ Attest™ Super Rapid Steam Biological Indicator 1592<br>3M™ Attest™ Auto-reader 490<br>3M™ Attest™ Auto-reader 490H |
|------------------------|------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Biological Indicator (BI) |
| Device Classification: | Class II |
| Classification Name: | Indicator, Biological Sterilization Process<br>[21 CFR § 880.2800(a), FRC] |
### Predicate Device- Biological Indicator:
3MTM Attest™ Super Rapid Readout Biological Indicator 1492V, 3M™ Attest™ Auto-reader 490, and 3MTM Attest™ Auto-reader 490H, K173437
#### Indications for Use- Biological Indicator
Use the 3M™ Attest™ Super Rapid Steam Biological Indicator 1592 in conjunction with the 3MTM Attest™ Auto-reader 490 having software version 4.0.0 or greater or the 3M™ Attest™ Auto-reader 490H having software version 4.0.0 or greater to qualify or monitor the following sterilization cycles:
| Cycle Type | Exposure Temperature | Exposure Time |
|-------------------------------------------|----------------------|---------------|
| Dynamic-air-removal (pre-vacuum and SFPP) | 250°F (121°C) | 15 minutes |
| Dynamic-air-removal (pre-vacuum and SFPP) | 250°F (121°C) | 20 minutes |
| Dynamic-air-removal (pre-vacuum and SFPP) | 250°F (121°C) | 30 minutes |
| Dynamic-air-removal (pre-vacuum and SFPP) | 250°F (121°C) | 35 minutes |
| Gravity Displacement | 250°F (121°C) | 30 minutes |
### Description of Device- Biological Indicator
The 3M™ Attest™ Super Rapid Steam Biological Indicator 1592 is a self-contained biological indicator (BI) specifically designed to be used with the 3M™ Attest™ Auto-reader 490 having software version 4.0.0 or greater or the 3MTM Attest™ Auto-reader 490H having software version 4.0.0 or greater to qualify or monitor dynamic-air-removal and gravity displacement steam sterilization cycles at 250°F (121°C).
The 1592 BI is a single-use device composed of a plastic sleeve containing a spore carrier and media ampoule, enclosed with a cap. On each 1592 cap is a chemical process indicator that changes color from pink to light brown or darker when exposed to steam. The detection of
{6}------------------------------------------------
fluorescence upon incubation of the 1592 in the 490 Auto-reader or the 490H Auto-reader (having software version 4.0.0 or greater) indicates a sterilization failure.
| Feature | Submission Device:<br>(K192550):<br>3MTM AttestTM<br>Super Rapid Steam<br>Biological Indicator 1592 and<br>3MTM AttestTM Auto-reader<br>490 and 490H | Predicate Device<br>(K173437):<br>3MTM AttestTM<br>Super Rapid Readout<br>Biological Indicator 1492V<br>and<br>3MTM AttestTM<br>Auto-reader 490 and 490H | Comparison |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | Use the 3MTM AttestTM Super<br>Rapid Steam Biological<br>Indicator 1592 in conjunction<br>with the 3MTM AttestTM Auto-<br>reader 490 having software<br>version 4.0.0 or greater or the<br>3MTM AttestTM Auto-reader<br>490H having software version<br>4.0.0 or greater to qualify or<br>monitor the following<br>sterilization cycles:<br>Dynamic-air-removal<br>(pre-vacuum and<br>SFPP), 250°F (121°C),<br>15 minutes Dynamic-air-removal<br>(pre-vacuum and<br>SFPP), 250°F (121°C),<br>20 minutes Dynamic-air-removal<br>(pre-vacuum and<br>SFPP), 250°F (121°C),<br>30 minutes Dynamic-air-removal<br>(pre-vacuum and<br>SFPP), 250°F (121°C),<br>35 minutes Gravity Displacement,<br>250°F (121°C),<br>30 minutes | Use the 3MTM AttestTM<br>Super Rapid Readout<br>Biological Indicator 1492V<br>in conjunction with both the<br>3MTM AttestTM Auto-reader<br>490 or the AttestTM Auto-<br>reader 490H having<br>software version 4.0.0 or<br>greater to qualify or monitor<br>dynamic-air-removal steam<br>sterilization cycles of:<br>3 minutes at 270°F<br>(132°C) 4 minutes at 270°F<br>(132°C) 3 minutes at 275°F<br>(135°C) | The 3MTM AttestTM<br>Super Rapid Steam<br>Biological Indicator<br>1592 is intended to<br>qualify or monitor both<br>dynamic-air-removal<br>(pre-vacuum and SFPP)<br>and gravity<br>displacement cycles at<br>various exposure times<br>at 250°F (121°C).<br>Whereas the 3MTM<br>AttestTM Super Rapid<br>Readout Biological<br>Indicator 1492Vis<br>intended to qualify or<br>monitor dynamic-air-<br>removal steam<br>sterilization cycles at<br>various exposure times<br>at 270°F (132°C) and<br>275°F (135°C).<br>Both BIs are intended<br>to be used in<br>conjunction with the<br>3MTM AttestTM Auto-<br>reader 490 and the<br>3MTM AttestTM Auto-<br>reader 490H. |
| Indicator Organism | <i>Geobacillus stearothermophilus</i><br>traceable to ATCCTM 7953 | <i>Geobacillus<br/>stearothermophilus</i><br>traceable to ATCCTM 7953 | Identical |
| Mechanism of<br>Action | When the enzyme that is<br>naturally occurring in the spore<br>is in its active state, it is<br>detected by measuring the | When the enzyme that is<br>naturally occurring in the<br>spore is in its active state, it<br>is detected by measuring the | Identical |
| Feature | Submission Device:<br>(K192550):<br>3MTM Attest™<br>Super Rapid Steam<br>Biological Indicator 1592 and<br>3MTM Attest™ Auto-reader<br>490 and 490H | Predicate Device<br>(K173437):<br>3MTM Attest™<br>Super Rapid Readout<br>Biological Indicator 1492V<br>and<br>3MTM Attest™<br>Auto-reader 490 and 490H | Comparison |
| | fluorescence produced by the<br>enzymatic hydrolysis of a non-<br>fluorescent substrate. The<br>resultant fluorescent by-product<br>is detected by the Auto-reader.<br>The presence of fluorescence<br>upon incubation in the Auto-<br>reader indicates a sterilization<br>process failure. | fluorescence produced by<br>the enzymatic hydrolysis of<br>a non-fluorescent substrate.<br>The resultant fluorescent<br>by-product is detected by<br>the Auto-reader. The<br>presence of fluorescence<br>upon incubation in the<br>Auto-reader indicates a<br>sterilization process failure. | |
| Auto-reader | 3MTM Attest™ Auto-reader 490<br>having software version 4.0.0 or<br>greater or 3MTM Attest™ Auto-<br>reader 490H having software<br>version 4.0.0 or greater | 3MTM Attest™ Auto-reader<br>490 or 3MTM Attest™ Auto-<br>reader 490H having<br>software version 4.0.0 or<br>greater | The 3MTM Attest™<br>Super Rapid Steam<br>Biological Indicator<br>1592 BI has a 24<br>minute readout and is<br>intended to be used<br>with a 490 or 490H<br>having software version<br>4.0.0 or greater.<br>Indications for a 24<br>minute time to result<br>were cleared under<br>K173437. |
| Viable spore<br>population | $\ge$ 1 x 106 | $\ge$ 1 x 106 | Identical |
| Resistance | D121 $\ge$ 1.5 min | D121 is N/A<br>D132 $\ge$ 10 s<br>D135 $\ge$ 8 s | The 3MTM Attest™<br>Super Rapid Steam<br>Biological Indicator<br>1592 BI is indicated for<br>qualifying and<br>monitoring sterilization<br>cycles at 121°C<br>whereas the predicate<br>device is indicated for<br>qualifying and<br>monitoring sterilization<br>cycles at 132°C and<br>135°C. |
| Survival Time | Meets the longer of FDA and<br>ISO 11138-1 and ISO 11138-3<br>requirements | Meets the longer of FDA<br>and ISO 11138-1 and ISO<br>11138-3 requirements | Identical |
| Kill Time | Meets the ISO 11138-1 and ISO<br>11138-3 requirements | Meets the ISO 11138-1 and<br>ISO 11138-3 requirements | Identical |
| Feature | Submission Device:<br>(K192550):<br>3MTM AttestTM<br>Super Rapid Steam<br>Biological Indicator 1592 and<br>3MTM AttestTM Auto-reader<br>490 and 490H | Predicate Device<br>(K173437):<br>3MTM AttestTM<br>Super Rapid Readout<br>Biological Indicator 1492V<br>and<br>3MTM AttestTM<br>Auto-reader 490 and 490H | Comparison |
| Carrier material | Plastic | Plastic | Identical |
| Incubation<br>temperature | 60 ± 2°C | 60 ± 2°C | Identical |
| Readout time | 24 minute final fluorescent<br>result in both the 490 and 490H<br>Auto-readers having software<br>version 4.0.0 or greater.<br>Optional visual pH color<br>change result in 7 days | 24 minute final fluorescent<br>result in both the 490 and<br>490H Auto-readers having<br>software version 4.0.0 or<br>greater.<br>1 hour final fluorescent<br>result in 490 Auto-readers<br>having software versions<br>less than 4.0.0.<br>Optional visual pH color<br>change result in 48 hours. | The 3MTM AttestTM<br>Super Rapid Steam<br>Biological Indicator<br>1592 BI has a 24<br>minute readout and is<br>intended to be used<br>with a 490 or 490H<br>having software version<br>4.0.0 or greater.<br>Indications for a 24<br>minute time to result<br>were cleared under<br>K173437. |
| Chemical indicator | Turns from pink to light brown<br>or darker upon steam exposure | Turns from pink to light<br>brown or darker upon steam<br>exposure | Identical |
| Shelf-life | 6 months | 21 months | Real-time testing is<br>ongoing. |
### Technological Characteristics Comparison Table – Biological Indicator
{7}------------------------------------------------
# TRADITIONAL PREMARKET NOTIFICATION [K192550]
# 3M™ Attest™ Super Rapid Steam Biological Indicator 1592
3MTM Attest™ Super Rapid Steam Challenge Pack 51582
3M™ Attest™ Auto-readers 490 and 490H
{8}------------------------------------------------
# TRADITIONAL PREMARKET NOTIFICATION [K192550]
# 3M™ Attest™ Super Rapid Steam Biological Indicator 1592
3M™ Attest™ Super Rapid Steam Challenge Pack 51582
3MTM Attest™ Auto-readers 490 and 490H
# Summary of Nonclinical Testing- Biological Indicator
Testing was conducted on the biological indicator following the FDA guidance and the standards below:
- Guidance for Industry and FDA Staff, Biological Indicator Premarket Notification [510(k)] Submissions, October 4, 2007
- ISO 11138-1:2017 Sterilization of health care products Biological indicators, Part 1: ● General Requirements
- ISO 11138-3:2017 Sterilization of health care products Biological indicators, Part 3: ● Biological indicators for moist heat sterilization processes
- . United States Pharmacopeia, Chapter <1035> Biological Indicators for Sterilization and Chapter <55> Biological Indicators - Resistance Performance Tests
{9}------------------------------------------------
The performance testing below demonstrate that the subject device met the acceptance criteria of the specified standards:
| Test | Results |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Population (Total Viable Spore Count)<br>Greater than or equal to $10^6$ spores | Passed |
| D-Value<br>Greater than or equal to 1.5 minutes at 121°C | Passed |
| Z-Value<br>Greater than or equal to $10$ °C | Passed |
| Survival Time<br>Meets the longer of FDA and ISO 11138-1 and ISO 11138-3 requirements | Passed |
| Kill Time<br>Meets ISO 11138-1 and ISO 11138-3 requirements | Passed |
| Component Inhibition Studies<br>Components have no impact on the recovery of 10-100 organisms | Passed |
| Hold Time Assessment<br>D-value does not change when activated 7 days post sterilization | Passed |
| Reduced Incubation Time<br>Meets FDA's requirements for Reduced Incubation Time with > 97% alignment with the conventional incubation time of 7 days for the following readout time:<br>• Fluorescent result in 24 minutes | Passed |
| Simulated Use<br>Verification of performance in claimed cycles | Passed |
## 3MT™ Attest™ Auto-reader 490 having software version 4.0.0 or greater and 3MTM Attest™ Auto-reader 490H
The 3M™ Attest™ Auto-reader 490 and the 3M™ Attest™ Auto-reader 490H were cleared for use with the 3M™ Attest™ Super Rapid Readout Biological Indicator 1492V under K173437. This submission extends the use of the 3M™ Attest™ Auto-reader 490 having software version 4.0.0 or greater or the 3M™ Attest™ Auto-reader 490H having software version 4.0.0 or greater to the 3MTM Attest™ Super Rapid Steam Biological Indicator 1592.
The 3M™ Attest Auto-reader 490 having software version 4.0.0 or greater or the 3M™ Attest™ Auto-reader 490H having software version 4.0.0 or greater are designed to incubate at 60°C and automatically read the 1592 BI for a fluorescent result within 24 minutes. The fluorescent readout at 24 minutes meets the FDA's requirement of > 97% alignment with the result after conventional incubation time of 7 days.
{10}------------------------------------------------
### Device Name and Classification- Challenge Pack:
| Trade Name: | 3MTM AttestTM Super Rapid Steam Challenge Pack 51582 |
|------------------------|----------------------------------------------------------------------------|
| Common/Usual Name: | Biological Indicator (BI) Challenge Pack |
| Device Classification: | Class II |
| Classification Name: | Indicator, Biological Sterilization Process<br>[21 CFR § 880.2800(a), FRC] |
### Predicate Device- Challenge Pack:
3MTM Attest™ Super Rapid 5 Steam-Plus Challenge Pack 41482V, 3M™ Attest™ Auto-reader 490, and 3MTM Attest™ Auto-reader 490H, K173519
### Indications for Use- Challenge Pack
Use the 3M™ Attest™ Super Rapid Steam Challenge Pack 51582 in conjunction with the 3M™ Attest™ Auto-reader 490 having software version 4.0.0 or greater or the 3MTM Attest™ Autoreader 490H having software version 4.0.0 or greater to qualify or monitor dynamic-air removal (pre-vacuum and SFPP) or gravity displacement steam sterilization cycles of 30 minutes at 250°F (121°C).
### Description of Device- Challenge Pack
The 3M™ Attest™ Super Rapid Steam Challenge Pack 51582 is specifically designed to qualify or monitor dynamic-air removal (pre-vacuum and SFPP) or gravity displacement steam sterilization cycles of 30 minutes at 250°F (121°C).
The 51582 Challenge Pack consists of multiple layers of medical index cards, some of which are die-cut to contain monitoring products. The pack is overwrapped and secured with a label. The Challenge Packs have the same design as the predicate device but contain different monitoring products. Each 51582 Challenge Pack contains a 3M™ Attest™ Super Rapid Steam Biological Indicator 1592 and a 3MTM Attest™ Steam Chemical Integrator (Type 5 (Category i5) Integrating Indicator as categorized by ISO 11140-1:2014). The 3M™ Attest™ Chemical Integrator offers an immediate ACCEPT or REJECT reading. The 1592 BI is specifically designed for a rapid fluorescent result when used in conjunction with the 3M™ Attest™ Autoreader 490 having software version 4.0.0 or greater or the 3M™ Attest™ Auto-reader 490H having software version 4.0.0 or greater. The detection of fluorescence upon incubation of the 1592 in the 490 Auto-reader or the 490H Auto-reader (having software version 4.0.0 or greater) indicates a sterilization failure. 3M™ Attest™ 1592 BI controls are provided with the Challenge Packs. Each Challenge Pack has a process indicator on the outside of the device that changes from yellow to brown or darker when exposed to steam.
{11}------------------------------------------------
| Feature | Submission Device:<br>(K192550):<br>3MTM AttestTM<br>Super Rapid Steam<br>Challenge Pack 51582 and<br>3MTM AttestTM Auto-reader<br>490 and 490H | Predicate Device<br>(K173519):<br>3MTM AttestTM Super Rapid 5<br>Steam-Plus Challenge Pack<br>41482V<br>and<br>3MTM AttestTM Auto-reader<br>490 and 490H | Comparison |
|--------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>use | Use the 3MTM AttestTM Super<br>Rapid Steam Challenge Pack<br>51582 in conjunction with the<br>3MTM AttestTM Auto-reader<br>490 having software version<br>4.0.0 or greater or the 3MTM<br>AttestTM Auto-reader 490H<br>having software version 4.0.0<br>or greater to qualify or<br>monitor dynamic-air removal<br>(pre-vacuum and SFPP) or<br>gravity displacement steam<br>sterilization cycles of 30<br>minutes at 250°F (121°C). | Use the 3MTM AttestTM Super<br>Rapid 5 Steam-Plus Challenge<br>Pack 41482V in conjunction<br>with the 3MTM AttestTM Auto-<br>reader 490 or 3MTM AttestTM<br>Auto-reader 490H having<br>software version 4.0.0 or<br>greater to qualify or monitor<br>dynamic-air-removal steam<br>sterilization cycles of 4 minutes<br>at 270°F (132°C) and 3 minutes<br>at 275°F (135°C). | The 3MTM AttestTM<br>Super Rapid Steam<br>Challenge Pack 51582 is<br>intended to qualify or<br>monitor both dynamic-<br>air-removal (pre-vacuum<br>and SFPP) and gravity<br>displacement<br>sterilization cycles of 30<br>minutes at 250°F<br>(121°C). Whereas the<br>3MTM AttestTM Super<br>Rapid 5 Steam-Plus<br>Challenge Pack is<br>intended to qualify or<br>monitor dynamic-air-<br>removal steam<br>sterilization cycles at<br>various exposure times<br>at 270°F (132°C) and<br>275°F (135°C). |
| General Design | Layers of medical index<br>cards, some of which are die-<br>cut to contain indicators,<br>overwrapped and secured<br>with a label. | Layers of medical index cards,<br>some of which are die-cut to<br>contain indicators, overwrapped<br>and secured with a label. | Identical |
| Biological<br>Indicator | 3MTM AttestTM Super Rapid<br>Steam Biological Indicator<br>1592 | 3MTM AttestTM Super Rapid<br>Readout Biological Indicator<br>1492V | The 51582 Challenge<br>Pack contains the new<br>1592 BI as proposed in<br>this submission. |
| Biological<br>Indicator<br>Incubation<br>temperature | 60 ± 2°C | 60 ± 2°C | Identical |
| Biological<br>Indicator Readout<br>time | 24 minute final fluorescent<br>result in both the 490 and<br>490H Auto-readers having<br>software versions 4.0.0 or<br>greater. | 24 minute final fluorescent<br>result in both the 490 and 490H<br>Auto-readers having software<br>versions 4.0.0 or greater.<br>1 hour final fluorescent result in<br>490 Auto-readers having | The 3MTM AttestTM<br>Super Rapid Steam<br>Biological Indicator<br>1592 BI has a 24 minute<br>readout and is intended<br>to be used with a 490 or |
| Feature | Submission Device:<br>(K192550):<br>3MTM AttestTM<br>Super Rapid Steam<br>Challenge Pack 51582 and<br>3MTM AttestTM Auto-reader<br>490 and 490H | Predicate Device<br>(K173519):<br>3MTM AttestTM Super Rapid 5<br>Steam-Plus Challenge Pack<br>41482V<br>and<br>3MTM AttestTM Auto-reader<br>490 and 490H | Comparison |
| | | software versions less than<br>4.0.0. | 490H having software<br>version 4.0.0 or greater.<br>Indications for a 24<br>minute time to result<br>were cleared under<br>K173519. |
| Resistance<br>Comparison to<br>the AAMI ST79<br>16 Towel PCD | Equivalent in resistance to the<br>AAMI ST79<br>16 Towel PCD | Equivalent in resistance to the<br>AAMI ST79<br>16 Towel PCD | Identical |
| Chemical<br>Integrator | 3MTM AttestTM Chemical<br>Integrator | 3MTM SteriGageTM Chemical<br>Integrator | Both integrators meet<br>FDA requirements and<br>are Type 5 Integrating<br>Indicators as classified<br>per ISO 11140-1. |
| External<br>Chemical Process<br>Indicator | Turns from yellow to brown<br>or darker upon steam<br>exposure | Turns from yellow to brown or<br>darker upon steam exposure | Identical |
| Shelf-life | 6 months | 21 months | Real-time testing is<br>ongoing. |
# Technological Characteristics Comparison Table – Challenge Pack
{12}------------------------------------------------
#### TRADITIONAL PREMARKET NOTIFICATION [K192550] 3M™ Attest™ Super Rapid Steam Biological Indicator 1592
3M™ Attest™ Super Rapid Steam Challenge Pack 51582
3MTM Attest™ Auto-readers 490 and 490H
### Summary of Nonclinical Testing- Challenge Pack
Testing was conducted on the challenge packs following the FDA guidance and the standards below:
- Guidance for Industry and FDA Staff, Biological Indicator Premarket Notification [510(k)] Submissions, October 4, 2007
- Premarket Notification [510(k)] Submission for Chemical Indicators: Guidance for ● Industry and FDA Staff, December 19, 2003
- ANSI/AAMI ST79:2017, Comprehensive guide to steam sterilization and sterility ● assurance in health care facilities
- ISO 11138-1:2017 Sterilization of health care products – Biological indicators, Part 1: General Requirements
- . ISO 11138-3:2017 Sterilization of health care products – Biological indicators, Part 3: Biological indicators for moist heat sterilization processes
{13}------------------------------------------------
- ISO 11140-1:2014 Sterilization of health care products Chemical indicators, Part 1: ● General Requirements
- United States Pharmacopeia, Chapter <1035> Biological Indicators for Sterilization and ● Chapter <55> Biological Indicators - Resistance Performance Tests
The performance testing below demonstrate that the subject device met the acceptance criteria of the specified standards:
| Test | Acceptance Criteria | Results |
|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Resistance of the<br>Challenge Pack as<br>compared to AAMI 16<br>Towel PCD | Challenge Pack is at least as resistant as the<br>biological indicator AAMI 16 Towel Process<br>Challenge Device (PCD) described in ANSI/AAMI<br>ST79: 2017 | Passed |
| Resistance of the<br>Challenge Pack as<br>compared to the<br>Biological Indicator<br>alone | Challenge Pack provides a greater resistance than<br>the Biological Indicator alone | Passed |
## Conclusion
The conclusions drawn from the non-clinical tests performed, the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate devices, the 3MM Attest™ Super Rapid Readout Biological Indicator 1492V, 3MTM Attest™ Auto-reader 490, and 3MTM Attest™ Auto-reader 490H (cleared under K173437) and the 3M™ Attest™ Super Rapid 5 Steam-Plus Challenge Pack 41482V, 3M™ Attest™ Auto-reader 490, and 3M™ Attest™ Auto-reader 490H (cleared under K173519), Class II (21 CFR 880.2800), product code FRC.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.