Use the 3M™ Attest™ Super Rapid Readout Steam Challenge Pack 1496V and the 3M™ Attest™ Super Rapid 5 Steam-Plus Challenge Pack 41482V in conjunction with the 3M™ Attest™ Auto-reader 490 or 3MTM Attest™ Auto-reader 490H having software version 4.0.0 or greater to qualify or monitor dynamic-air-removal steam sterilization cycles of 4 minutes at 270°F (132°C) and 3 minutes at 275°F (135°C).
Device Story
System comprises biological indicator (BI) challenge packs and auto-reader; used to monitor steam sterilization efficacy. Challenge packs contain BI 1492V (and chemical integrator in 41482V); placed in sterilizer to challenge cycle. Post-sterilization, BI is incubated in 3M Attest Auto-reader 490/490H. Auto-reader detects fluorescence indicating presence of active enzymes; fluorescence change signifies sterilization failure. Software version 4.0.0+ enables 24-minute rapid readout at 60°C incubation. Healthcare providers use output to verify sterilization cycle success before releasing instruments for patient use. Benefits include rapid sterility assurance, reducing wait times for surgical instrument availability.
Clinical Evidence
No clinical data included. Bench testing only. Performance validated via resistance testing against AAMI 16 Towel PCDs, temperature stability testing (60±2°C), and electrical safety/EMC compliance (IEC 61010-1, IEC 61010-2-010, IEC 61326-1, FCC Part 15).
Technological Characteristics
Challenge packs consist of medical index cards, die-cut holders, and overwrapped labels. Biological indicator uses fluorescent enzyme detection. Auto-reader provides automated incubation and optical sensing. Connectivity: Standalone. Software: Version 4.0.0+. Sterilization: Steam. Compliance: ISO 11138-1/3, ISO 18472, ANSI/AAMI ST79.
Indications for Use
Indicated for use in healthcare facilities to qualify or monitor dynamic-air-removal steam sterilization cycles (4 min at 270°F/132°C or 3 min at 275°F/135°C) using 3M Attest Super Rapid Readout Biological Indicators.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
Predicate Devices
3M Attest Super Rapid Readout Steam Challenge Pack 1496V, 3M Attest Super Rapid 5 Steam-Plus Challenge Pack 41482V, and 3M Attest Auto-reader 490 (K121593)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 23, 2018
3M Company Hilary Hovde Regulatory Affairs Associate 3M Center, Building 275-5W-06 St Paul, Minnesota 55144
Re: K173519
Trade/Device Name: 3M Attest Super Rapid Readout Steam Challenge Pack 1496V, 3M Attest Super Rapid 5 Steam-Plus Challenge Pack 41482V, 3M Attest Auto-reader 490, 3M Attest Auto-reader 490H Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: March 19, 2018 Received: March 22, 2018
Dear Hilary Hovde:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
# Sincerely, Geeta K. Pamidimukkala -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### Device Name
3MTM Attest™ Super Rapid Readout Steam Challenge Pack 1496V, 3M™ Attest™ Super Rapid 5 Steam-Plus Challenge Pack 41482V, 3M™ Attest™ Auto-reader 490, and 3M™ Attest™ Auto-reader 490H
#### Indications for Use (Describe)
Use the 3M™ Attest™ Super Rapid Readout Steam Challenge Pack 1496V and the 3M™ Attest™ Super Rapid 5 Steam-Plus Challenge Pack 41482V in conjunction with the 3M™ Attest™ Auto-reader 490 or 3MTM Attest™ Auto-reader 490H having software version 4.0.0 or greater to qualify or monitor dynamic-air-removal steam sterilization cycles of 4 minutes at 270°F (132°C) and 3 minutes at 275°F (135°C).
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the 3M logo. The logo consists of the number "3" and the letter "M" in a bold, sans-serif font. The color of the text is red, and the background is white. The logo is simple and recognizable.
# 510(k) Summary for 3M™ Attest™ Super Rapid Readout Steam Challenge Pack 1496V 3M™ Attest™ Super Rapid 5 Steam-Plus Challenge Pack 41482V 3M™ Attest™ Auto-reader 490 3M™ Attest™ Auto-reader 490H (software version 4.0.0 or greater)
3M Health Care 3M Center, Bldg. 275-5W-06 St. Paul, MN 55144-1000
> Contact: Hilary B. Hovde Regulatory Affairs Associate Phone Number: (651) 736-0364 FAX Number: (651) 737-5320
Submission Date: April 19, 2018
K173519
{4}------------------------------------------------
### Device Name and Classification:
| Common or Usual Name: | Biological Indicator |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary Name: | 3MTM AttestTM Super Rapid Readout Steam Challenge Pack 1496V<br>3MTM AttestTM Super Rapid 5 Steam-Plus Challenge Pack 41482V<br>3MTM AttestTM Auto-reader 490<br>3MTM AttestTM Auto-reader 490H |
| Classification Name: | Indicator, Biological Sterilization Process<br>(21 CFR § 880.2800(a)) |
| Device Classification: | Class II |
| Product Code: | FRC |
### Predicate Device:
- 3MTM Attest™ Super Rapid Readout Steam Challenge Pack 1496V, 3MTM . Attest™ Super Rapid 5 Steam-Plus Challenge Pack 41482V, and 3MTM Attest™ Auto-reader 490, K121593
#### Reference Device:
- 3MTM Attest™ Auto-reader 490H, K171003 .
## Description of Device:
The 3M™ Attest™ Super Rapid Readout Steam Challenge Pack 1496V and the 3M™ Attest™ Super Rapid 5 Steam-Plus Challenge Pack 41482V are specifically designed to qualify or monitor dynamic-air-removal steam sterilization cycles of 4 minutes at 270°F (132°C) and 3 minutes at 275°F (135°C).
The 1496V and 41482V Challenge Packs consist of multiple layers of medical index cards, some of which are die-cut to contain monitoring products. The pack is overwrapped and secured with a label. The Challenge Packs have the same design as the predicate device. Each 1496V Challenge Pack contains a 3M™ Attest™ Super Rapid Readout Biological Indicator 1492V while the 41482V Challenge Pack contains a 1492V BI and a 3M™ SteriGage™ Steam Chemical Integrator (Type 5 Integrating Indicators as categorized by ISO 11140-1:2014). The 1492V BI is specifically designed for a rapid fluorescent result when used in conjunction with the 3M™ Attest™ Auto-reader 490 or the 3M™ Attest™ Auto-reader 490H (software version 4.0.0 or greater). A fluorescence change indicates a steam sterilization process failure. 3M™ Attest™ 1492V BI controls are provided with the Challenge Packs. Each Challenge Pack has a process indicator on the outside of the device that changes from vellow to brown or darker when exposed to steam.
{5}------------------------------------------------
This submission is addressing a software change to the 3M™ Attest™ Auto-reader 490 to reduce the final fluorescent readout for the 1492V BI from 1 hour to 24 minutes, and to change the incubation temperature from 56°C to 60°C making the 490 Auto-reader identical to the 490H Auto-reader (software version 4.0.0 or greater). The 3M™ Attest™ Super Rapid Readout Steam Challenge Pack 1496V, the 3M™ Attest™ Super Rapid 5 Steam-Plus Challenge Pack 41482V, and the 3M™ Attest™ Super Rapid Readout Biological Indicator 1492V are the same design as the previously cleared devices of the same model numbers. The ComplyTM SteriGage™ Steam Chemical Integrators contained within the 41482V challenge pack are the same as those in the predicate device. The device has the same materials and fundamental scientific technology.
## Summary of Nonclinical Testing
As the change is to the 3M™ Attest™ Auto-reader software, testing was conducted on the biological indicators and the challenge packs following the FDA guidance and the standards below:
- Guidance for Industry and FDA Staff, Biological Indicator Premarket Notification [510(k)] Submissions, October 4, 2007
- ANSI/AAMI ST79:2017, Comprehensive guide to steam sterilization and sterility assurance in health care facilities
- . ISO 11138-1:2017 Sterilization of health care products – Biological indicators, Part 1: General Requirements
- . ISO 11138-3:2017 Sterilization of health care products – Biological indicators, Part 3: Biological indicators for moist heat sterilization processes
- ANSI/AAMI/ISO 18472:2006 Sterilization of Health Care Products – Biological and Chemical Indicators: Test Equipment
- United States Pharmacopeia, Chapter <1035> Biological Indicators for Sterilization and . Chapter <55> Biological Indicators - Resistance Performance Tests
{6}------------------------------------------------
#### TRADITIONAL PREMARKET NOTIFICATION [510(k)] 3MTM Attest™ Super Rapid Readout Steam Challenge Pack 1496V 3MTM Attest™ Super Rapid 5 Steam-Plus Challenge Pack 41482V 3MTM Attest™ Auto-reader 490 and 490H
The functionality of the 3M™ Attest™ Super Rapid Readout Steam Challenge Pack 1496V and the 3M™ Attest™ Super Rapid 5 Steam-Plus Challenge Pack 41482V in conjunction with the 3M™ Attest™ Auto-reader 490 and the 3M™ Attest™ Auto-reader 490H (software version 4.0.0 or greater) with a final fluorescent readout of 24 minutes and an incubation temperature of 60°C was demonstrated in the following tests:
| Test | Acceptance Criteria | Results |
|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Resistance of the<br>Challenge Pack as<br>compared to AAMI 16<br>Towel PCD | Challenge Pack is at least as resistant as the<br>biological indicator AAMI 16 Towel Process<br>Challenge Device (PCD) described in ANSI/AAMI<br>ST79: 2017 | Passed |
| Resistance of the<br>Challenge Pack as<br>compared to the<br>Biological Indicator<br>alone | Challenge Pack provides a greater resistance than<br>the Biological Indicator alone | Passed |
| Auto-reader Maintenance<br>of Incubation<br>Temperature | Auto-reader maintains incubation temperature of<br>$60 \pm 2$ ℃ over a period of 7 days | Passed |
The results of these evaluations showed that the 3M™ Attest™ Super Rapid Readout Steam Challenge Pack 1496V and the 3MTM Attest™ Super Rapid 5 Steam-Plus Challenge Pack 41482V present a challenge to the sterilization process equivalent to the biological indicator AAMI 16 Towel PCD described in ANSI/AAMI ST79, Comprehensive guide to steam sterilization and sterility assurance in health care facilities.
The Attest™ Auto-readers were tested for safety by Underwriters Laboratory to verify compliance to:
- IEC 61010-1 (2010) 3rd Edition; Safety requirements for electrical equipment for . measurement, control, and laboratory use – Part 1: General requirements, and
- . IEC 61010-2-010 (2014) 3rd Edition; Safety requirements for electrical equipment for measurement, control, and laboratory use – Part 2-010: Particular requirements for laboratory equipment for the heating of materials
In addition the 3M™ Attest™ Auto-readers 490 and 490H have been tested by a certified Testing Laboratory to verify electromagnetic compatibility per:
- USA Title 47, Code of Federal Regulations for:
- Radiated Emissions (FCC Part 15, Subpart B, Class A) O
- o Conducted Emissions (FCC Part 15, Subpart B, Class A), and
- . IEC 61326-1:2012 Electrical Equipment for Measurement, Control, and Laboratory Use – EMC Requirements
{7}------------------------------------------------
No clinical data was included in this premarket application submission.
### Indications for Use
Use the 3M™ Attest™ Super Rapid Readout Steam Challenge Pack 1496V and the 3M™ Attest™ Super Rapid 5 Steam-Plus Challenge Pack 41482V in conjunction with the 3M™ Attest™ Autoreader 490 or 3M™ Attest™ Auto-reader 490H having software version 4.0.0 or greater to qualify or monitor dynamic-air-removal steam sterilization cycles of 4 minutes at 270°F (132°C) and 3 minutes at 275°F (135°C).
{8}------------------------------------------------
## TRADITIONAL PREMARKET NOTIFICATION [510(k)]
#### 3M™ Attest™ Super Rapid Readout Steam Challenge Pack 1496V 3M™ Attest™ Super Rapid 5 Steam-Plus Challenge Pack 41482V 3M™ Attest™ Auto-reader 490 and 490H
# Comparison to Predicate Device
| | Submission Device:<br>3MTM Attest™ | Predicate Device (K121593):<br>3MTM Attest™ |
|--------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | Super Rapid Readout<br>Steam Challenge Pack 1496V and<br>3MTM AttestTM Super Rapid 5 | Super Rapid Readout<br>Steam Challenge Pack 1496V and<br>3MTM Attest™ Super Rapid 5 |
| | Steam-Plus Challenge Pack 41482V | Steam-Plus Challenge Pack 41482V |
| Indications for use | Use the 3MTM Attest™ Super Rapid Readout<br>Steam Challenge Pack 1496V and the 3MTM<br>Attest™ Super Rapid 5 Steam-Plus Challenge<br>Pack 41482V in conjunction with the 3MTM<br>AttestTM Auto-reader 490 or 3MTM AttestTM<br>Auto-reader 490H having software version 4.0.0<br>or greater to qualify or monitor dynamic-air- | Use the 3M™ Attest™ Super Rapid<br>Readout Steam Challenge Pack 1496V<br>and the 3MTM Attest™ Super Rapid 5<br>Steam-Plus Challenge Pack 41482V in<br>conjunction with the 3MTM AttestTM<br>Auto-reader 490 to qualify or monitor<br>dynamic-air-removal (pre-vacuum)<br>steam sterilization cycles 270°F<br>(132°C) at 4 minutes and 275°F<br>(135°C) at 3 minutes. |
| | removal steam sterilization cycles of 4 minutes<br>at 270°F (132°C) and 3 minutes at 275°F<br>(135°C). | The 3MTM Attest™ Super Rapid<br>Readout Biological Indicator 1492V<br>contained in the challenge pack<br>provides a final fluorescent result in 1<br>hour. An optional visual pH color<br>change result is observed in 48 hours. |
| General Design | Layers of medical index cards, some of which<br>are die-cut to contain indicators, overwrapped<br>and secured with a label. | Identical |
| Biological Indicator | 3MTM Attest™ Super Rapid Readout Biological<br>Indicator 1492V | Identical |
| Biological Indicator<br>Incubation<br>temperature | 60 ± 2°C | 56 ± 2°C |
| Biological Indicator<br>Readout time | 24 minute final fluorescent result in both the 490<br>and 490H Auto-readers having software versions<br>4.0.0 or greater.<br>1 hour final fluorescent result in 490 Auto-<br>readers having software versions less than 4.0.0. | 1 hour final fluorescent result in 490<br>Auto-readers. |
| Resistance<br>Comparison to the<br>AAMI ST79<br>16 Towel PCD | Equivalent in resistance to the AAMI ST79<br>16 Towel PCD | Identical |
| Chemical Integrator | The 3M™ Attest™ Super Rapid Readout Steam<br>Challenge Pack 1496V does not contain a<br>chemical integrator.<br>The 3MTM Attest™ Super Rapid 5 Steam-Plus<br>Challenge Pack 41482V contains a 3M™<br>SteriGage™ Chemical Integrator | Identical |
| External Chemical<br>Process Indicator | Turns from yellow to brown or darker upon<br>steam exposure | Identical |
| Shelf-life | 21 months | Identical |
{9}------------------------------------------------
## Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the 3MTM Attest™ Super Rapid Readout Steam Challenge Pack 1496V, the 3M™ Attest™ Super Rapid 5 Steam-Plus Challenge Pack 41482V, the 3MTM Attest™ Auto-reader 490, and the 3MTM Attest™ Autoreader 490H are as safe, as effective, and perform as well as or better than the 3M™ Attest™ Super Rapid Readout Steam Challenge Pack 1496V, the 3M™ Attest™ Super Rapid 5 Steam-Plus Challenge Pack 41482V, and the 3M™ Attest™ Auto-reader 490 (K121593).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.