The Precision Casting Solutions Total Hip System implant components are indicated for use in cementless reconstruction of the articulating surface of femoral and/or acetabular portions of the hip that are severely disabled and/or very painful as a result of: - Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis - Rheumatoid arthritis or traumatic arthritis - Correction of functional deformity - Non-union femoral neck fracture - Trochanteric fractures of the proximal femur with head involvement which is unmanageable using other techniques. The components can be used for primary hip arthroplasty or for revision of a failed total hip arthroplasty.
Device Story
Non-cemented 4-part total hip replacement system; includes femoral stem, femoral head, acetabular poly liner, and acetabular metal shell. Femoral stem (forged Ti 6AL 4V with Ti plasma spray); femoral head (CoCr or BIOLOX delta ceramic); acetabular liner (UHMWPE); acetabular shell (forged Ti 6AL 4V with Ti porous coating). Components restore natural hip rotation, extension, and flexion. Used by orthopedic surgeons in clinical settings for primary or revision hip arthroplasty. System provides mechanical reconstruction of articulating hip surfaces to alleviate pain and restore function in severely disabled patients.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing performed on the predicate devices, including fatigue, disassembly, range of motion, burst strength, rotational stability, and pull-out/push-out tests.
Technological Characteristics
Materials: Forged Ti 6AL 4V (with Ti plasma spray or porous coating), CoCr, BIOLOX delta ceramic, UHMWPE. Components: Femoral stem, femoral head (28, 32, 36 mm), acetabular poly liner (hooded/non-hooded), acetabular shell (no-hole, cluster-hole, revision multi-hole). Energy source: None (mechanical). Connectivity: None. Sterilization: Identical to predicate.
Indications for Use
Indicated for patients with severe hip disability or pain due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, traumatic arthritis, functional deformity, non-union femoral neck fracture, or trochanteric fractures of the proximal femur. Suitable for primary or revision total hip arthroplasty.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
Progressive Orthopaedic Total Hip System (K143314/K151424)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health and Human Services logo is a stylized depiction of a human figure, while the FDA acronym and full name are written in blue, with the acronym in a blue square.
December 17, 2019
Precision Casting Solution, LLC Jim Moore Vice President 200031 Henrici Road OREGON CITY OR 97045
Re: K192450
Trade/Device Name: Precision Casting Solutions Total Hip System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH, LZO Dated: August 30, 2019 Received: September 6, 2019
Dear Jim Moore:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Vesa Vuniqi Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
## K192450
Device Name Precision Casting Solutions Total Hip System
### Indications for Use (Describe)
The Precision Casting Solutions Total Hip System implant components are indicated for use in cementless reconstruction of the articulating surface of femoral and/or acetabular portions of the hip that are severely disabled and/or very painful as a result of:
- · Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis
- · Rheumatoid arthritis or traumatic arthritis
- · Correction of functional deformity
- · Non-union femoral neck fracture
· Trochanteric fractures of the proximal femur with head involvement which is unmanageable using other techniques.
The components can be used for primary hip arthroplasty or for revision of a failed total hip arthroplasty.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the letters "P", "C", and "S" in both blue and pink colors. There is a medical symbol in the background. In the foreground, there are two DNA strands, one in blue and the other in pink.
### 510(k) Summary 1.
The following 510(k) Summary is provided in accordance with 21 CFR 807.92.
#### 510(k) Owner and Registration 1.1
| Owner's Name: | Precision Casting Solutions, INC |
|---------------------------------------|--------------------------------------------|
| Address: | 20031 Henrici Road., Oregon City, OR 97045 |
| Phone Number: | (503) 421-1251 |
| Fax Number: | N/A |
| Date Summary Prepared: | August 30, 2019 |
| Establishment Registration<br>Number: | N/A |
#### 1.2 510(k) Contact
| Contact: | Jim Moore |
|-----------------|---------------------------------------------|
| Address: | 20031 Henrici Road<br>Oregon City, OR 97045 |
| Phone Number: | (503) 421-1251 |
| Fax Number: | N/A |
| Contact Person: | Jim Moore |
#### 1.3 Device Name and Classification
| Device Trade Name: | Precision Casting Solutions Total Hip System |
|---------------------------------------|----------------------------------------------|
| Device Common Name: | Total Hip Replacement |
| Regulation Number and<br>Description: | 21 CFR 888.3558 |
| Device Class: | Class II |
| Product Codes: | LPH, LZO |
| Advisory Panel: | 87 (Orthopedic) |
#### Legally Marketed Predicate 1.4
Precision Casting Solutions is utilizing the Progressive Orthopaedic Total Hip System as the predicate device (K143314/K151424). The Precision Casting Solutions Total Hip System features component designs, materials, indications, manufacturing methods, and contract manufacturer's that are identical to the Progressive Orthopaedic Total Hip System.
#### 1.5 Device Description
The present 510k submission is for a non-cemented primary hip prosthesis that consists of a 4-part total hip replacement system including femoral stem, femoral Head, acetabular poly liner, and acetabular metal (or shell) components. The femoral head component articulates within the poly acetabular component. The poly acetabular component snaps into the metal acetabular component. The design and sizing of the components correspond to natural hip anatomy to restore normal rotation, extension, and flexion.
The femoral stem component is made from forged Ti 6AL 4V, with a Ti plasma spray coating. The uni-polar femoral
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo with the letters "P", "C", and "S" on both the left and right sides. The letters on the left are in blue, while the letters on the right are in pink. In the background, there is a faint image of a caduceus, a symbol associated with medicine. Below the letters and the caduceus, there is a depiction of a DNA double helix, with one strand in blue and the other in pink.
head component is made from CoCr, or BIOLOX® delta ceramic, in 28, 32, and 36 mm sizes.
The acetabular poly liner component is made of standard UHMWPE polyethylene in both hooded and non-hooded options. The acetabular poly liner component is offerent inner and outer diameter combinations to accept various size uni-polar femoral heads and acetabular metal components.
The acetabular metal (or shell) component is made from forged Ti 6AL 4V with a Ti porous coating. It is available in nohole, cluster-hole, and revision multi-hole styles.
#### 1.6 Indications For Use
The Precision Casting Solutions Total Hip System is identical to the predicate Progressive Orthopaedic Total Hip System. The indications for use are also identical.
The Precision Casting Solutions Total Hip System implant components are indicated for use in cementless reconstruction of the articulating surface of femoral and/or acetabular portions of the hip that are severely disabled and/or very painful as a result of:
- · Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis
- Rheumatoid arthritis or traumatic arthritis .
- · Correction of functional deformity
- . Non-union femoral neck fracture
- Trochanteric fractures of the proximal femur with head involvement which is unmanageable using other ● techniques.
The components can be used for primary hip arthroplasty or for revision of a failed total hip arthroplasty.
#### 1.7 Summary of Technological Characteristics
The Precision Casting Solutions Total Hip system is identical to the predicate Progressive Orthopaedic Total Hip System. Both devices are manufactured from identical materials, possess the same the same packaging and sterilization processes. Extensive precimical testing was performed on the predicate devices per K143314 and K151424 and found substantially equivalent. The performance tests are listed below and used herein to establish substantial equivalence (Section Error! Reference source not found. Error! Reference source not found.)
Given that the subject device is identical to the subject system is substantially equivalent to the predicate systems (K143314/K151424).
#### 1.8 Performance Testing
Extensive preclinical performance testing was conducted, and substantial equivalence determined per K143314 and K151424. The components of the subject device are identical to the predicate device, and therefore the predicate device testing demonstrates substantial equivalence for the subject device. The results confirm that all components of the Precision Casting Solutions Total Hip System exhibit the appropriate mechanical characteristics for total hip joint replacement and is substantially equivalent to the predicate devices.
- Fatigue Performance Test for Precision Casting Solutions Stem .
- . Fatigue Performance Test for the Neck Portion of the Precision Casting Solutions Stem
- Disassembly Force Test for Precision Casting Solutions Neck Taper/Femoral Head Interface
- Range of Motion Test for the Precision Casting Solutions Total Hip System
- Burst Strength Test for Precision Casting Solutions Ceramic Femoral Heads (Static Compression)
- . Cyclic Fatigue Test for Precision Casting Solutions Ceramic Femoral Heads (Cyclic Compression)
- . Post-Cvclic Fatigue Burst Test for Precision Casting Solutions Ceramic Femoral Heads (Static Compression)
- . Pull-Off Test for Precision Casting Solutions Ceramic Femoral Heads
- Rotational Stability Test for Precision Casting Solutions Ceramic Femoral Heads
- Torsional Properties Test for Precision Casting Solutions Bone Screws
- Driving Torque Test for Precision Casting Solutions Bone Screws ●
- Axial Pull-Out Strength Test for Precision Casting Solutions Bone Screws .
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with the letters "P", "C", and "S" arranged in a specific pattern. On the left side, the letters "P", "C", and "S" are in blue, while on the right side, they are in pink. Below the letters, there is a depiction of a DNA double helix, with one strand in blue and the other in pink. In the background, there is a medical symbol, suggesting a connection to healthcare or medicine.
- Lever-Out Test for Precision Casting Solutions Acetabular Shell/Liner Assembly .
- Torque-Out Test for Precision Casting Solutions Acetabular Shell/Liner Assembly
- Push-In Test for Precision Casting Solutions Acetabular Shell/Liner Assembly ●
- Push-Out Test for Acetabular Shell/Liner Precision Casting Solutions Assembly ●
#### 1.9 Conclusions
The Precision Casting Solutions Total Hip System is identical to the predicate Progressive Orthopaedic Total Hip System. The subject device has the same design features, materials, and indications for use as the predicate devices. The testing performed for the predicate device indicates that the Precision Casting Solutions Total Hip System is safe for clinical use.
The Precision Casting Solutions Total Hip System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.