The Progressive Orthopaedic Total Hip System implant components are indicated for use in cementless reconstruction of the articulating surface of femoral and acetabular portions of the hip that are severely disabled and/or very painful as a result of: - Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis - Rheumatoid arthritis or traumatic arthritis - Correction of functional deformity - Non-union femoral neck fracture - Trochanteric fractures of the proximal femur with head involvement which is unmanageable using other techniques. The components can be used for primary hip arthroplasty or for revision of a failed total hip arthroplasty.
Device Story
The Progressive Orthopaedic Total Hip System consists of orthopedic implant components, specifically additional screw sizes and an apical hole plug, designed for use with the previously cleared Progressive Orthopaedic Total Hip System (K143314). These components are used in cementless hip arthroplasty to reconstruct femoral and acetabular articulating surfaces. The device is intended for patients suffering from severe hip pain or disability due to degenerative joint disease, arthritis, or fractures. Surgeons perform the implantation procedure in an operating room setting. The components provide structural support and fixation for the hip joint, aiming to restore function and alleviate pain. The system is an extension of an existing cleared platform, utilizing established materials and design principles to ensure compatibility with the primary system.
Clinical Evidence
No clinical testing was required. Substantial equivalence is supported by bench testing, including mechanical characterization, torsion, driving torque, and pull-out testing.
Technological Characteristics
Components are manufactured from Ti6Al4V alloy per ASTM 1472. The system includes screws (6.5mm OD x 15-45mm length) and an apical hole plug (8.52mm OD, 5/16 x 24 UNF 2A thread). The device is a cementless, semi-constrained porous-coated hip prosthesis.
Indications for Use
Indicated for patients requiring cementless hip arthroplasty due to severe disability or pain from non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, traumatic arthritis, functional deformity, non-union femoral neck fracture, or proximal femur trochanteric fractures with head involvement. Suitable for primary or revision total hip arthroplasty.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
Progressive Orthopaedic Total Hip System (K143314)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other to create a sense of depth and unity.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 27, 2016
The Progressive Orthopaedic Company, LLC Mr. Thomas Smith Quality/Regulatory Consultant 801 US Highway 1, Suite B North Palm Beach, Florida 33408
Re: K151424 Trade/Device Name: The Progressive Orthopaedic Total Hip System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, LZO Dated: June 24, 2016 Received: June 27, 2016
Dear Mr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K151424
Device Name
The Progressive Orthopaedic Total Hip System
#### Indications for Use (Describe)
The Progressive Orthopaedic Total Hip System implant components are in cementless reconstruction of the articulating surface of femoral and acetabular portions of the hip that are severely disabled and/or very painful as a result of:
- · Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis
- · Rheumatoid arthritis or traumatic arthritis
- · Correction of functional deformity
- · Non-union femoral neck fracture
- · Trochanteric fractures of the proximal femur with head involvement which is unmanageable using other techniques.
The components can be used for primary hip arthroplasty or for revision of a failed total hip arthroplasty.
| Type of Use (Select one or both, as applicable) | <table><tr><td><div style="display:inline-block;">☒</div> Production (Part 21 CFR 601 Subpart D)</td></tr><tr><td><div style="display:inline-block;">☐</div> For Test Article (21 CFR 601 Subpart E)</td></tr></table> | <div style="display:inline-block;">☒</div> Production (Part 21 CFR 601 Subpart D) | <div style="display:inline-block;">☐</div> For Test Article (21 CFR 601 Subpart E) |
|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <div style="display:inline-block;">☒</div> Production (Part 21 CFR 601 Subpart D) | | | |
| <div style="display:inline-block;">☐</div> For Test Article (21 CFR 601 Subpart E) | | | |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for The Progressive Orthopaedic Company. The logo consists of a blue hexagon with a white caduceus symbol inside. To the right of the hexagon is the company name, with "The" above "PROGRESSIVE" and "Orthopaedic Company" below.
# 510(k) Summary
| Applicant/Sponsor: | The Progressive Orthopaedic Company, LLC.<br>801 US Highway 1, Suite B<br>North Palm Beach, FL, 33408<br>(561) 440-4460 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Thomas Smith<br>Quality and Regulatory Consulting<br>801 US Highway 1, Suite B<br>North Palm Beach, FL, 33408<br>(203) 641-3936 |
| Proposed Trade Name: | The Progressive Orthopaedic Total Hip System |
| Common Name: | Hip Joint Prosthesis |
| Classification Name: | Hip joint metal/polymer/metal semi-constrained porous-<br>coated uncemented prosthesis per 21 CFR 888.3358. This<br>falls under the Orthopedics panel/87 as a Class II device. |
| Device Product Code: | LPH, LZO |
| Predicate Devices: | Progressive Orthopaedic Total Hip System (K143314)<br>NovoSource NovoHip Total Hip System (K140701) |
# Device Description:
The present 510k The Progressive Orthopaedic additional screw sizes and apical hole plug are components to be used with the Progressive Orthopaedic Total Hip System, previously cleared via K143314.
| Screw Sizes | 6.5mm OD x 15-45mm Length |
|-------------------------|---------------------------|
| Material | Ti6Al4V per ASTM 1472 |
| Apical Hole Plug OD | $\u00d8$ 8.52mm OD |
| Apical Hole Plug Thread | 5/16 X 24 UNF 2A |
| Material | Ti6Al4V |
The Progressive Orthopaedic Company 801 US Highway 1, Suite B, North Palm Beach, FL, 33408 Office: (561)440-4460 Fax: (304)777-4790
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for The Progressive Orthopaedic Company. On the left side of the logo is a blue hexagon with a white caduceus symbol inside. To the right of the hexagon is the company name, with "The" above "PROGRESSIVE" and "Orthopaedic Company" below.
# Intended Use:
Hip joint arthroplasty
### Indications for Use:
The Progressive Orthopaedic Total Hip System implant components are indicated for use in cementless reconstruction of the articulating surface of femoral and acetabular portions of the hip that are severely disabled and/or very painful as a result of:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis
- Rheumatoid arthritis or traumatic arthritis
- Correction of functional deformity
- Non-union femoral neck fracture
- Trochanteric fractures of the proximal femur with head involvement which is unmanageable using other techniques.
The components can be used for primary hip arthroplasty or for revision of a failed total hip arthroplasty.
## Substantial Equivalence
The Progressive Orthopaedic Total Hip System and NovoSource Total Hip System have similar intended use, indications for use and the same manufacturing materials as the predicate device. The range of sizes of the Progressive Ortho Total Hip System is similar to the predicate devices.
## Non-Clinical Testing
Mechanical testing has demonstrated the device's ability to perform under expected conditions. Testing included mechanical characterization testing, torsion, driving torque and pull out. In all testing, the subject device shows equivalent properties when compared to predicate products.
Testing has determined that the device is substantially equivalent to the predicate device in material, construction, and performance characteristics.
## Clinical Performance Data Summary
No clinical testing was required.
The Progressive Orthopaedic Company 801 US Highway 1, Suite B, North Palm Beach, FL, 33408 Office: (561)440-4460 Fax: (304)777-4790
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.