K192439 · Walk Vascular, LLC · QEZ · Dec 9, 2019 · Cardiovascular
Device Facts
Record ID
K192439
Device Name
JETi Peripheral Thrombectomy System
Applicant
Walk Vascular, LLC
Product Code
QEZ · Cardiovascular
Decision Date
Dec 9, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The JETi Peripheral Thrombectomy System is intended to: - remove/aspirate fluid and break-up soft emboli and thrombus from the peripheral vasculature, and - subselectively infuse/deliver diagnostics or therapeutics with or without vessel occlusion.
Device Story
System removes/aspirates fluid and breaks up soft emboli/thrombus from peripheral vasculature; also enables subselective infusion of diagnostics/therapeutics. Components include multi-lumen aspiration catheter, Saline Drive Unit (SDU), peristaltic pump, and accessory stand. Operation: thrombus enters distal catheter tip via suction; SDU/pump set delivers sterile saline stream through secondary lumen to macerate/dilute thrombus; diluted material aspirated through primary lumen into pump receptacle. No saline injected into patient. Used in clinical settings by physicians. SDU contains microprocessor-controlled circuit monitoring motor/pressure functions; status indicated via front panel LEDs. Foot pedal controls saline delivery and aspiration. Output allows physician to clear peripheral vascular obstructions, potentially improving blood flow and patient outcomes.
Clinical Evidence
No clinical data generated. Substantial equivalence supported by in vitro bench testing, including design verification, software/functional performance validation, peristaltic pump life cycle testing, aspiration source comparison, and clot removal efficacy.
Technological Characteristics
Multi-lumen aspiration catheter; peristaltic pump for aspiration; piston pump for saline delivery; microprocessor-controlled SDU; 24V external power supply. Connectivity: RJ45 for vacuum sensor. Complies with IEC 60601-1 and IEC 60601-1-2 (4th ed) electrical safety/EMC standards. Software: firmware-based monitoring of motor and pressure.
Indications for Use
Indicated for removal/aspiration of fluid and soft emboli/thrombus from peripheral vasculature and subselective infusion/delivery of diagnostics or therapeutics in patients requiring peripheral vascular intervention.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 9, 2019
Walk Vascular, LLC % Mr. Paul Gasser Medical Device RA/QA Consultant 13612 Rushmore Lane Santa Ana, California 92705
## Re: K192439
Trade/Device Name: JETi Peripheral Thrombectomy System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: QEZ Dated: November 19, 2019 Received: November 21, 2019
Dear Mr. Gasser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192439
Device Name JETi Peripheral Thrombectomy System
#### Indications for Use (Describe) The JETi Peripheral Thrombectomy System is intended to:
- remove/aspirate fluid and break-up soft emboli and thrombus from the peripheral vasculature, and
- subselectively infuse/deliver diagnostics or therapeutics with or without vessel occlusion.
Type of Use (Select one or both, as applicable):
| <span></span> | <div></div> <span></span> <label>☑ Prescription Use (Part 21 CFR 801 Subpart D)</label> | <span></span> | <div></div> <span></span> <label>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
|---------------|-----------------------------------------------------------------------------------------|---------------|----------------------------------------------------------------------------------------|
|---------------|-----------------------------------------------------------------------------------------|---------------|----------------------------------------------------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Traditional 510(k) Summarv
| Submitter: | Walk Vascular, LLC<br>17171 Daimler Street<br>Irvine, CA 92614<br>USA |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Brad Culbert<br>VP, Engineering<br>Walk Vascular LLC<br>Telephone: 949-752-9642<br>Fax: 949-752-9658<br>Email: bsculbert@yahoo.com |
| Date Summary Prepared: | September 6, 2019 |
| Device Trade Name: | JETi Peripheral Thrombectomy System |
| Common Name: | Embolectomy/Thrombectomy Catheter |
| Classification Name: | Embolectomy Catheter (21 CFR 870.5150) |
| Product Code: | QEZ |
| Predicate Device: | JETi 88 Peripheral Thrombectomy System<br>(510(k) K183403) |
## Device Description:
The JETi Peripheral Thrombectomy System consists of one aspiration catheter, which is connected to the JETi Pump Set, one JETi Peripheral Saline Drive Unit (SDU), one Peristaltic Pump, and an Accessory Stand. A Sterile Accessory Kit and Non-Sterile Accessory Device are supplied. In use, thrombus enters the distal catheter tip via the suction force provided by the peristaltic pump. The peripheral SDU and pump set deliver a stream of sterile saline through the secondary lumen to break up and dilute the thrombus within the catheter. The diluted thrombus and saline are drawn back through the primary lumen and deposited into the peristaltic pump receptacle. No saline is injected into the patient during normal operation. The peripheral SDU, peristaltic pump, an adjustable mounting arm, an accessory basket, and an IV pole are contained on an accessory stand.
The peripheral catheter is a multi-lumen device that allows for simultaneous hydro-mechanical thrombus disruption and thrombus aspiration. It is designed to simultaneously deliver a stream of saline via a displacement pump within the distal tip of the catheter, while aspirating thrombotic material macerated by the saline stream.
{4}------------------------------------------------
The peripheral catheter has the ability to infuse fluid through the aspiration lumen and saline lumen.
For infusion through the aspiration lumen, an appropriate sized syringe filled with fluid is infused to the treatment area by passing through the side arm of the rotating hemostasis valve (RHV) that is attached to the catheter.
For infusion through the saline lumen, a clamp is used on the suction tubing to cut off aspirate flow.
The proximal end of the pump set consists of a spike and an in-line drip chamber that is used to pierce the saline bag and connect the pump set to the saline source. The cassette contains a piston pump and is mounted onto the peripheral SDU. The cassette is powered by the motor contained in the peripheral SDU. The distal end of the pump set has a connector, which mounts to the proximal end of the multi-port Luer adapter of the peripheral catheter and delivers the saline to the peripheral catheter.
The peripheral SDU is a reusable device. The fork drive of the peripheral SDU is designed to run the piston pump contained in the pump set to deliver the stream of saline to the peripheral catheter, when activated by the foot pedal. Aspiration is achieved with a peristaltic pump that is connected to the SDU and is also turned on/off with the foot pedal. The pressure sensor is connected to the suction tubing, which is connected to the multi-port connector of the peripheral catheter. The peripheral SDU contains a microprocessor controlled circuit and firmware that monitors various functions of the motor and pressure to assure that the device is functioning as expected. Various colored LED lights on the front panel indicate to the user what the current status is of the SDU. Energy is provided by a 24 volt external power supply, which is connected to mains power.
## Indications for Use:
The JETi Peripheral Thrombectomy System is intended to:
- remove/aspirate fluid and break-up soft emboli and thrombus from the peripheral vasculature, and
- subselectively infuse/deliver diagnostics or therapeutics with or without vessel occlusion.
#### Statement of Equivalence:
The subject device and the predicate share the same intended use and have similar technological characteristics.
Key differences between the subject and predicate devices are reflected in the following table.
{5}------------------------------------------------
| Design Features | Predicate JETi 88 Peripheral<br>Thrombectomy System | Subject JETi Peripheral<br>Thrombectomy System |
|----------------------------------------|----------------------------------------------------------------|----------------------------------------------------|
| Catheter | | |
| Catheter working length<br>(cm) | 100 | Same |
| Catheter connections and<br>location | Multi-port Luer adapter on the<br>proximal end of the catheter | Same |
| Pump Set | | |
| Saline input tube length<br>(feet) | 6 | Same |
| Saline Drive Unit | | |
| IEC 60601-1-2 edition<br>complied with | 4th | Same |
| On/off control | Foot pedal | Same |
| Method to stop flow | Pinch valve | Peristaltic pump |
| Pinch valve control | Mode button | N/A |
| Location | Mounted on cart outside sterile field | Same |
| Vacuum sensor connector | Standard RJ45 | Same |
| Aspiration | | |
| Aspiration Method | Vacuum pump | Peristaltic pump |
| Aspiration Flow Rate | 650 mL/min | 450 mL/min |
| Pressure – Free Flow | - 23.6 inHg | - 12.9 inHg |
| Pressure - Engaged in<br>Clot | < - 27 inHg | Same |
| Accessory Stand | Present | Re-designed to accommodate the<br>peristaltic pump |
| Sterile Accessory Kit | Four components | Three components |
| Non-Sterile Accessory<br>Device | Present | Tubing section added |
The JETi Peripheral Thrombectomy System is substantially equivalent to the predicate device.
## Summary of Non-Clinical Performance Data:
Device evaluation consisted of in vitro testing performed pursuant to Walk Vascular's risk analysis. All data met the acceptance criteria and fell within pre-determined product specifications and external standard requirements. The following testing was performed to support the determination of substantial equivalence:
Design Verification Testing:
SDU/peristaltic pump software and functional performance SDU/peristaltic pump/accessory stand validation Peristaltic pump life cycle Aspiration source comparison Clot removal Software validation 60601-1 Product and electrical safety
{6}------------------------------------------------
## Biocompatibility Testing:
Biocompatibility testing was not conducted, as the catheter and pump set are identical to the predicate device.
## Sterilization Testing:
Sterilization testing was not conducted, as the catheter and pump set are identical to the predicate device.
## Transportation and Shelf Life Testing:
Transportation and shelf life testing was not conducted on the catheter and pump set, as the catheter and pump set are identical to the predicate device. Transportation /shelf life testing was not repeated for the SDU, sterile accessory kit, non-sterile disposable accessory device, and accessory stand, as the design differences between the predicate and subject devices are not considered sufficiently significant to justify repeating those studies. Shelf life testing was conducted for the peristaltic pump.
The data from the in vitro testing above supports the substantial equivalence of the subject device to the predicate device.
## Summary of Pre-Clinical and Clinical Data:
No pre-clinical or clinical data were generated to establish substantial equivalence. Bench data are considered adequate to support a determination of substantial equivalence.
## Summary:
Based on the intended use, and in vitro performance information provided in this premarket notification, the JETi Peripheral Thrombectomy System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.