The JETi 88 Peripheral Thrombectomy System is intended to: - remove/aspirate fluid and break-up soft emboli and thrombus from the peripheral vasculature, and - subselectively infuse/deliver diagnostics or therapeutics with or without vessel occlusion.
Device Story
JETi 88 Peripheral Thrombectomy System removes/aspirates fluid and breaks up soft emboli/thrombus in peripheral vasculature; also enables subselective infusion of diagnostics/therapeutics. System comprises aspiration catheter, JETi 88 Pump Set, JETi 88 Peripheral Saline Drive Unit (SDU), and accessory cart. Catheter features multi-lumen design; delivers saline stream via displacement pump at distal tip to macerate thrombus while simultaneously aspirating material through primary lumen. Saline/thrombus mixture deposited into vacuum pump receptacle; no saline injected into patient. SDU uses microprocessor-controlled circuit/firmware to monitor motor/vacuum functions; user interface includes LED status indicators and foot pedal control for saline delivery and vacuum pinch valve. Operated by clinicians in clinical setting. Output allows physician to clear peripheral vascular obstructions, potentially improving blood flow and patient outcomes.
Clinical Evidence
No clinical data generated. Substantial equivalence established via bench testing only. Testing included pull, leak, pressure, flow, kink, embolic analysis, saline stream containment, and IEC 60601-1 electrical safety testing. Biocompatibility, sterilization, and shelf-life data were adopted from the predicate device.
Technological Characteristics
Multi-lumen aspiration catheter; hydro-mechanical thrombus disruption via saline stream; vacuum-based aspiration. System includes SDU with microprocessor-controlled circuit/firmware, piston pump, and vacuum pump. Connectivity: RJ45 vacuum sensor connector. Power: 24V external supply. Compliance: IEC 60601-1 (4th ed). Form factor: Cart-mounted SDU with adjustable arm.
Indications for Use
Indicated for removal/aspiration of fluid and break-up of soft emboli and thrombus from peripheral vasculature, and subselective infusion/delivery of diagnostics or therapeutics with or without vessel occlusion.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
Predicate Devices
ClearLumen II Peripheral Thrombectomy System (K172000)
Submission Summary (Full Text)
{0}------------------------------------------------
The full content of this file cannot be displayed with your current PDF viewer. Please update to the latest possible version to view this document.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is an emblem of the Department of Health & Human Services. To the right of the emblem, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION". The logo is simple and professional, and it is easily recognizable.
April 17, 2019
Walk Vascular, LLC % Paul Gasser Medical Device RA/QA Consultant 13612 Rushmore Lane Santa Ana, California 92705
Re: K183403
Trade/Device Name: JETi 88 Peripheral Thrombectomy System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEZ Dated: March 18, 2019 Received: March 19, 2019
Dear Mr. Gasser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Traditional 510(k) Summary
| Submitter: | Walk Vascular, LLC<br>17171 Daimler Street<br>Irvine, CA 92614<br>USA |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Brad Culbert<br>VP, Engineering<br>Walk Vascular LLC<br>Telephone: 949-752-9642<br>Fax: 949-752-9658<br>Email: bsculbert@yahoo.com |
| Date Summary Prepared: | December 5, 2018 |
| Device Trade Name: | JETi 88 Peripheral Thrombectomy System |
| Common Name: | Embolectomy/Thrombectomy Catheter |
| Classification Name: | Embolectomy Catheter (21 CFR 870.5150) |
| Product Code: | QEZ, Aspiration Thrombectomy Catheter |
| Predicate Device: | ClearLumen II Peripheral Thrombectomy System<br>(510(k) K172000) |
# Device Description:
The JETi 88 Peripheral Thrombectomy System consists of one aspiration catheter, which is connected to the JETi 88 Pump Set, one JETi 88 Peripheral Saline Drive Unit (SDU), and an Accessory Cart. A Sterile Accessory Kit and Non-Sterile Accessory Device are supplied. In use, thrombus enters the distal catheter tip via the suction force provided by the vacuum pump. The peripheral SDU and pump set deliver a stream of sterile saline through the secondary lumen to break up and dilute the thrombus within the catheter. The diluted thrombus and saline are drawn back through the primary lumen and deposited into the vacuum pump receptacle. No saline is injected into the patient during normal operation. The peripheral SDU, an adjustable mounting arm, a mounting bracket for vacuum canisters, an IV pole and a vacuum pump are contained on an accessory cart.
The peripheral catheter is a multi-lumen device that allows for simultaneous hydro-mechanical thrombus disruption and thrombus aspiration. It is designed to simultaneously deliver a stream of saline via a displacement pump within the distal tip of the catheter, while aspirating thrombotic material macerated by the saline stream.
{4}------------------------------------------------
The peripheral catheter has the ability to infuse fluid through the aspiration lumen and saline lumen.
For infusion through the aspiration lumen, an appropriate sized syringe filled with fluid is infused to the treatment area by passing through the side arm of the rotating hemostasis valve (RHV) that is attached to the catheter.
For infusion through the saline lumen, a clamp is used on the suction tubing to cut off vacuum.
The proximal end of the pump set consists of a spike and an in-line drip chamber that is used to pierce the saline bag and connect the pump set to the saline source. The cassette contains a piston pump and is mounted onto the peripheral SDU. The cassette is powered by the motor contained in the peripheral SDU. The distal end of the pump set has a connector, which mounts to the proximal end of the multi-port Luer adapter of the peripheral catheter and delivers the saline to the peripheral catheter.
The peripheral SDU is a reusable device. The fork drive of the peripheral SDU is designed to run the piston pump contained in the pump set to deliver the stream of saline to the peripheral catheter, when activated by the foot pedal. Vacuum is achieved with a vacuum pump, turned on/off with the pinch valve on the side of the SDU, which is also controlled by the foot pedal. The vacuum sensor is connected to the vacuum tubing, which is connected to the multi-port connector of the peripheral catheter. The peripheral SDU contains a microprocessor controlled circuit and firmware that monitors various functions of the motor and vacuum to assure that the device is functioning as expected. Various colored LED lights on the front panel indicate to the user what the current status is of the SDU. Energy is provided by a 24 volt external power supply, which is connected to mains power.
#### Indications for Use:
The JETi 88 Peripheral Thrombectomy System is intended to:
- remove/aspirate fluid and break-up soft emboli and thrombus from the peripheral vasculature, and
- subselectively infuse/deliver diagnostics or therapeutics with or without vessel occlusion.
#### Statement of Equivalence:
The subject device and the predicate share intended use and have similar technological characteristics.
Key differences between the subject and predicate devices are reflected in the following table.
{5}------------------------------------------------
| Design Features | Predicate ClearLumen II Peripheral Thrombectomy System | Subject JETi 88 Peripheral Thrombectomy System |
|-------------------------------------|----------------------------------------------------------------------------------|-------------------------------------------------------------|
| Catheter | | |
| Catheter working length (cm) | 115 | 100 |
| Catheter connections and location | Two port "Y" connector with Luer connections on the proximal end of the catheter | Multi-port Luer adapter on the proximal end of the catheter |
| Pump Set | | |
| Saline input tube length (feet) | 4 | 6 |
| Saline Drive Unit | | |
| IEC 60601-1-2 edition complied with | 3rd | 4th |
| On/off control | Stopcock activated (vacuum sensing) | Foot pedal |
| Method to stop flow | Stop cock | Pinch valve |
| Pinch valve control | None | Mode button |
| Aspiration method | Single use, rechargeable 60 mL syringe (e.g., VacLock) | Vacuum pump |
| Location | Pole mounted outside sterile field | Mounted on cart outside sterile field |
| Vacuum sensor connector | Merit Medical custom plug | Standard RJ45 |
| Accessory Cart | Not present | Present |
| Sterile Accessory Kit | Not present | Present |
| Non-Sterile Accessory Device | Not present | Present |
The JETi 88 Peripheral Thrombectomy System is substantially equivalent to the predicate device.
#### Summary of Non-Clinical Performance Data:
Device evaluation consisted of in vitro testing performed pursuant to Walk Vascular's risk analysis. All data met the acceptance criteria and fell within pre-determined product specifications and external standard requirements. The following testing was performed to support the determination of substantial equivalence:
Design Verification Testing:
- Pull test of welded and glued joints Leak testing Pressure testing Flow testing Kink testing Embolic analysis Saline stream containment 60601-1 Product and electrical safety testing
# Biocompatibility Testing:
Biocompatibility testing was not conducted, as the test results from the ClearLumen II Thrombectomy System were adopted for the JETi 88 Peripheral Thrombectomy System.
{6}------------------------------------------------
# Sterilization Testing:
Sterilization testing was not conducted, as the test results from the ClearLumen II Thrombectomy System were adopted for the JETi 88 Peripheral Thrombectomy System.
### Transportation and Shelf Life Testing:
Transportation and shelf life testing was not conducted, as the test results from the ClearLumen II Thrombectomy System were adopted for the JETi 88 Peripheral Thrombectomy System.
The data from the in vitro testing above supports the substantial equivalence of the subject device to the predicate and reference device.
### Summary of Pre-Clinical and Clinical Data:
No pre-clinical or clinical data were generated to establish substantial equivalence. Bench data are considered adequate to support a determination of substantial equivalence.
### Summary:
Based on the intended use, and in vitro performance information provided in this premarket notification, the JETi 88 Peripheral Thrombectomy System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.