NuVasive NuvaLine is a medical device software application intended to assist healthcare professionals in capturing, viewing, measuring, and storage and distribution of spinal alignment assessment images at various time points in patient care. Online synchronization of the database allows healthcare professionals and service providers to conveniently perform and review spinal alignment assessments of images by featuring measurement tools on various platforms. Clinical judgment and experience are required to properly use the software.
Device Story
NuVasive NuvaLine is a software application for spinal alignment assessment; inputs include spinal X-ray images transferred from PACS; software calculates spinal, pelvic, lumbar, thoracic, and cervical parameters; outputs include quantifiable measurements for pre-operative and post-operative assessment; used by healthcare professionals in clinical settings; operates on mobile devices (iOS) and web interfaces; cloud-based database synchronization allows multi-platform access and data storage; assists clinicians in evaluating spinal deformity and surgical correction; supports clinical decision-making by correlating measurements with health-related quality of life (HRQOL) scores.
Clinical Evidence
Bench testing only. No clinical data provided. Verification and validation testing performed according to software requirements, including cloud connectivity validation and measurement library verification. Measurement accuracy verified to be within ±3° for angles and ±1 cm for offsets.
Technological Characteristics
Software-based PACS application; web-based and mobile (iOS) platforms; cloud-synchronized database; supports DICOM image transfer (converted to JPEG for app use); utilizes measurement tools for spinal parameters; connectivity via HTTPS/VPN; system requirements include Windows 10, 3GHz processor, 8GB RAM, and HTML5.2/JavaScript ES7 compatible browsers.
Indications for Use
Indicated for healthcare professionals to capture, view, measure, store, and distribute spinal alignment assessment images for patient care. No specific age or gender restrictions; intended for use in spinal deformity assessment and correction monitoring.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
NucleusHealth Nucleus Image Management System (K171130)
Submission Summary (Full Text)
{0}------------------------------------------------
September 26, 2019
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.
NuVasive, Incorporated % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive Suite 510k SAINT PAUL MN 55114
Re: K192435
Trade/Device Name: NuVasive® NuvaLine® Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: September 3, 2019 Received: September 5, 2019
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K192435
Device Name NuVasive® NuvaLine®
### Indications for Use (Describe)
Nu Vasive NuvaLine is a medical device software application intended to assist healthcare professionals in capturing, viewing, measuring, and storage and distribution of spinal alignment assessment images at various time points in patient care. Online synchronization of the database allows healthcare professionals and service providers to conveniently perform and review spinal alignment assessments of images by featuring measurement tools on various platforms. Clinical judgment and experience are required to properly use the software.
Type of Use (Select one or both, as applicable):
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Nuvasive. The logo consists of a stylized leaf shape on the left, with the word "NUVASIVE" in gray letters to the right of the leaf. The leaf shape is purple on the top and gray on the bottom.
K192435
## 510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations. Part 807, and in particular 21 CFR \$807.92, the following summary of information is provided:
#### Submitted by: A.
Manthan Damani Lead Regulatory Affairs Specialist NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 909-1800 Date Prepared: September 25, 2019
#### B. Device Name
| Trade or Proprietary Name: | <i>NuVasive</i> ® <i>NuvaLine</i> ® |
|----------------------------|---------------------------------------------|
| Common or Usual Name: | Picture archiving and communications system |
| Classification Name: | Picture archiving and communications system |
| Device Class: | Class II |
| Classification: | §892.2050 |
| Product Code: | LLZ |
#### C. Predicate Devices
The subject NuVasive NuvaLine is substantially equivalent to the primary predicate NuVasive NuvaLine Mobile App (K162647) and additional reference devices NuVasive NVM5 System (K152942), Nemaris Surgimap 2.0 (K141669), and NucleusHealth Nucleus Image Mangement System (K171130).
#### D. Device Description
NuVasive NuvaLine is a medical device software application used to calculate the spinal pelvic, lumbar, thoracic, and cervical parameters for pre-operative and post-operative assessment of spinal x-ray images. These measured parameters provide a quantifiable way to assess a patient's spinal deformity and correction correlated to health related quality of life (HRQOL) scores.
The purpose of this premarket notification is to gain clearance of the previously cleared NuvaLine app to communicate with cloud server for online synchronization of database to transfer and store assessment data to allow for use of the NuvaLine app on different platforms (e.g.: mobile, web interface, desktop) by healthcare professionals and service providers.
NuVasive NuvaLine is designed per recommendations provided in the following FDA guidance documents:
- Guidance for the Content of Premarket Submissions for Software Contained in Medical ● Devices
- Off-The-Shelf Software Use in Medical Devices
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Nuvasive logo. The logo consists of a purple and gray leaf-like shape on the left, followed by the word "NUVASIVE" in gray letters. The logo is simple and modern, and the colors are muted and professional.
- Content of Premarket Submissions for Management of Cybersecurity in Medical Devices
#### E. Indications for Use
NuVasive NuvaLine is a medical device software application intended to assist healthcare professionals in capturing, viewing, measuring, and storage and distribution of spinal alignment assessment images at various time points in patient care. Online synchronization of the database allows healthcare professionals and service providers to conveniently perform and review spinal alignment assessments of images by featuring measurement tools on various platforms. Clinical judgment and experience are required to properly use the software.
#### F. Technological Characteristics
As was established in this submission, the subject NuvaLine is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have equivalent technological characteristics to its predicate device through comparison in areas including design, intended use, and functions.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Nuvasive. The logo consists of a stylized leaf shape in purple and gray, followed by the company name "NUVASIVE" in gray, sans-serif font. The leaf shape is split into two colors, with the left side in purple and the right side in gray.
1
| Table 1 – Comparison of Technical Characteristics | | | | | |
|---------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Specification<br>/<br>Property | Predicate Device<br>NuVasive NuvaLine<br>Mobile App (K162647) | Reference Device<br>NucleusHealth Nucleus Image<br>Management System (K171130) | Nemaris Surgimap 2.0<br>(K141669) | Subject Device<br>NuVasive NuvaLine | Discussion |
| Intended Use<br>/ Indications<br>for Use | The NuVasive NuvaLine<br>Mobile App is a medical<br>device software mobile<br>application intended to<br>assist healthcare<br>professionals in<br>capturing, measuring,<br>and storing spinal<br>alignment assessment<br>images at various time<br>points in patient care.<br>The device allows the<br>healthcare professional<br>to conveniently perform<br>and review spinal<br>alignment assessments<br>of images by featuring<br>measurement tools on<br>their mobile device. | The Nucleus Image Management System<br>(Nucleus IMS) is a software based PACS, to<br>be used by radiologists and other medical<br>personnel. The Nucleus IMS is comprised of<br>software modules that provide image receipt,<br>diagnostic viewing, storage, distribution,<br>enhancement, sharing, manipulation, and<br>networking of medical 2D/3D images at<br>distributed locations. All modules of the<br>Nucleus IMS are web-based and can operate<br>on off-the-shelf hardware, as needed. The<br>Nucleus IMS consists of the following<br>primary components: Nucleus Viewer with<br>image streaming technology for use by<br>medical professionals for diagnostic and<br>clinical image review—Nucleus Image<br>Exchange (iX) for image acceptance, transfer,<br>and sharing with hospitals/clinics as well as<br>between facilities—and Nucleus Image Store<br>for secure cloud based image storage and<br>management through HIPAA compliant<br>encryption. Nucleus iX integrates with the<br>Nucleus Viewer and Nucleus Image Store.<br>Nucleus.io, a web based class I PaaS device,<br>provides the basis for the system and for<br>vendor neutral applications.<br>The Nucleus IMS interfaces with health<br>information systems (HIS) using industry-<br>standard image transfer and data exchange<br>protocols—such as DICOM, HL7, and<br>HTML—through web-based networked<br>gateways and local and wide area networks.<br>The Nucleus IMS is compatible with<br>modalities including: Computed Tomography<br>(CT), Magnetic Resonance Imaging (MR),<br>Ultrasound (US), Computed Radiography<br>(CR), Digital Radiography (DX), Nuclear | The Surgimap software assists<br>healthcare professionals in<br>viewing, storing, and measuring<br>images as well as planning<br>orthopedic surgeries. The device<br>allows service providers to<br>perform generic as well as<br>specialty measurements of the<br>images, and to plan surgical<br>procedures. The device also<br>includes tools for measuring<br>anatomical components for<br>placement of surgical implants,<br>and offer online synchronization of<br>the database with the possibility to<br>share data among Surgimap users.<br>Clinical judgment and experience<br>are required to properly use the<br>software. | NuVasive NuvaLine is a<br>medical device software<br>application intended to assist<br>healthcare professionals in<br>capturing, viewing, measuring,<br>and storage and distribution of<br>spinal alignment assessment<br>images at various time points in<br>patient care. Online<br>synchronization of the database<br>allows healthcare professionals<br>and service providers to<br>conveniently perform and<br>review spinal alignment<br>assessments of images by<br>featuring measurement tools on<br>various platforms. Clinical<br>judgment and experience are<br>required to properly use the<br>software. | Same |
| Specification<br>/<br>Property | Predicate Device<br>NuVasive NuvaLine<br>Mobile App (K162647) | Reference Device<br>NucleusHealth Nucleus Image<br>Management System (K171130) | Nemaris Surgimap 2.0<br>(K141669) | Subject Device<br>NuVasive NuvaLine | Discussion |
| | | Medicine (NM), Positron Emission<br>Tomography (PT), and X-Ray Angiography<br>(XA). When appropriate, the Nucleus IMS<br>provides and installs software and, optionally,<br>server hardware at client facilities to facilitate<br>secure, web-based connections for image<br>transmission to and from hospital central<br>servers via the Internet. Additionally, industry<br>standard HTTPS, VPNs, and other encryption<br>methodologies are utilized to allow for<br>optimal, secure, rapid streaming of images.<br>Lossless image compression and encryption<br>adhere to standard industry protocols. The<br>Nucleus IMS can be used as a full featured<br>PACS or as an independent viewer in clinical<br>settings.<br><br>Nucleus Image Management System is not<br>intended for display of mammography imaging<br>for diagnosis. | | | |
| Device Class | II | II | II | II | Same |
| Product Code | LLZ | LLZ | LLZ | LLZ | Same |
| Regulation<br>Number<br>(21CFR) | 21 CFR 892.2050 | 21 CFR 892.2050 | 21 CFR 892.2050 | 21 CFR 892.2050 | Same |
| Device<br>Classification<br>Name | Picture archiving and<br>communications system | Picture archiving and communications<br>system | Picture archiving and<br>communications system | Picture archiving and<br>communications system | Same |
| Software<br>Functionaliti<br>es/<br>Modalities | Spinal alignment<br>assessments of images | PACS image transfer, sharing, storage,<br>management | Spinal alignment assessments of<br>images | Spinal alignment assessments<br>of images | Same |
| Specification<br>/<br>Property | Predicate Device<br>NuVasive NuvaLine<br>Mobile App (K162647) | Reference Device<br>NucleusHealth Nucleus Image<br>Management System (K171130) | Nemaris Surgimap 2.0<br>(K141669) | Subject Device<br>NuVasive NuvaLine | Discussion |
| Algorithms | Various spinal<br>assessment algorithms | N/A | Various spinal assessment<br>algorithms | Various spinal assessment<br>algorithms | Same |
| User<br>Interface | Mobile device | PC or mobile device or web interface | PC or mobile device | PC or mobile device or web<br>interface | Same |
| Obtaining an<br>image | Mobile Device Camera | N/A | Transferred from other devices | Transferred from PACS | Same |
| Online<br>synchronizati<br>on of<br>database | N/A | N/A | Yes | Yes | Same |
| PACS<br>connectivity | N/A | Yes | Yes | Yes | Same |
| DICOM | N/A | Yes | Yes | Yes (DICOM images from<br>PACS converted to jpeg for use<br>in NuvaLine) | Same |
| Supported<br>Platforms | Mobile application<br>supported on devices<br>running iOS version<br>10.0 or later. | N/A | N/A | Mobile application supported<br>on devices running iOS version<br>10.0 or later.<br><br>Web client is supported for the<br>following minimum system<br>specifications:<br>• Windows 10<br>• 3GHz processor<br>• 18GB RAM<br>• Modern browser<br>supporting HTML5.2 and<br>JavaScript ES7 or better.<br>• 1920x1200 display<br>resolution | Same |
| Specification<br>/<br>Property | Predicate Device | Reference Device | | Subject Device | Discussion |
| | NuVasive NuvaLine<br>Mobile App (K162647) | NucleusHealth Nucleus Image<br>Management System (K171130)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.