The Surgimap software assists healthcare professionals in viewing, storing, and measuring images as well as planning orthopedic surgeries. The device allows service providers to perform generic as well as specialty measurements of the images, and to plan surgical procedures. The device also includes tools for measuring anatomical components for placement of surgical implants, and offer online synchronization of the database with the possibility to share data among Surgimap users. Clinical judgment and experience are required to properly use the software.
Device Story
Surgimap 2.0 is a software application for orthopedic surgical planning and image analysis. It accepts standard image formats (JPEG, TIFF, PNG) and DICOM files. Users perform generic and specialty anatomical measurements, simulate surgical procedures (e.g., osteotomies), and template implants (screws, cages, rods). The software operates on PC/Mac platforms, supporting local and PACS connectivity. It features an online database synchronization tool allowing case sharing among authorized users. Healthcare professionals use the output to guide surgical planning and implant selection; clinical judgment remains mandatory for all interpretations. The device benefits patients by facilitating precise pre-operative planning and implant templating.
Clinical Evidence
Bench testing only. Verification and validation activities were conducted in accordance with IEC 62304:2006 and ANSI/IEEE 829-2008 standards. Software was evaluated at a 'Moderate' level of concern, addressing risks associated with surgical planning errors.
Technological Characteristics
Software-based PACS/orthopedic planning tool. Compatible with Windows and Mac OS. Supports DICOM and standard image formats. Features include generic/specialty measurement modules, implant templating, and online database synchronization. Developed under ISO 14971:2012 risk management and IEC 62304:2006 software lifecycle processes.
Indications for Use
Indicated for healthcare professionals requiring software to view, store, and measure medical images and to plan orthopedic surgical procedures, including anatomical measurements for surgical implant placement.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 19, 2014
Nemaris, Inc. % Mr. Keith Barritt Attorney Fish and Richardson P.C. 1425 K Street NW WASHINGTON DC 20005
Re: K141669
Trade/Device Name: Surgimap 2.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: August 13, 2014 Received: August 14, 2014
Dear Mr. Barritt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D. O'Hara
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K141669
Device Name Surgimap
#### Indications for Use (Describe)
The Surgimap software assists healthcare professionals in viewing, and measuring images as well as planning orthopedic surgeries. The device allows service providers to perform generic as well as specialty measurements of the images, and to plan surgical procedures. The device also includes tools for measuring anatomical components for placement of surgical implants, and offer online synchronization of the database with the possibility to share data among Surgimap users. Clinical judgment and experience are required to properly use the software.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness Nemaris Inc. Surgimap
The following information is provided pursuant to 21 CFR 807.92
### 807.92(a)(1)
(i) Submitter
Nemaris Inc 306 East 15th Street | Suite 1R New York, NY 10003 Phone: (646)-794-8650 Fax: (646)-602-6925
(ii) Submitter Contact
Virginie Lafage, PhD Vice President and Chief Technology Officer Nemaris Inc 306 East 15th Street | Suite 1R New York, NY 10003 Phone: (646)-794-8650 Email: vlafage(@surgimap.com
(iii) Preparation Date May 05, 2014
## 807.92(a)(2)
| Trade Name: | Surgimap |
|---------------|---------------------------------------------------|
| Common Name: | Picture Archiving and Communication System (PACS) |
| Section: | 892.2050 |
| Class | II |
| Product Code: | LLZ |
{4}------------------------------------------------
## 807.92(a)(3)
●
| • | Surgimap Spine | |
|---|----------------|---------------------------------------------------|
| • | Manufacturer: | Nemaris Inc |
| • | Tradename: | Surgimap Spine |
| • | 510K #: | K111019 (September 30, 2011) |
| • | Common Name | Picture Archiving and Communication System (PACS) |
| • | Class: | II |
| • | Product Code: | LLZ |
| • | Traumacad | |
| • | Manufacturer: | Voyant Health formerly known as Orthocrat |
| • | Tradename: | TraumaCad Release 2.0 |
| • | 510K #: | K073714 (March 19, 2008) |
| • | Common Name | Picture Archiving and Communication System (PACS) |
| • | Class: | II |
| • | Product Code: | LLZ |
| | | |
## 807.92(a)(4)
Surgimap is software developed for the medical community. It is intended to be used to view, store and transport images as well as perform generic or specialty measurements and plan or simulate aspects of surgical procedures. The image formats supported encompasses the standard image formats (jpeg, tiff, png, ....) and also DICOM images. Images can be stored in the Surgimap database and measurements (generic or specialty specific) can be overlaid to each image. Surgimap also offers the ability for the end user to plan, or simulate, aspects of certain surgical procedures such as osteotomies and templating implants (including but not limited to screws, interbody cages, rods). Via internet connection during use of the software application, the database can be synchronized online. An optional feature consists in organizing patient information into cases with the possibility to share these cases among Surgimap users.
## 807.92(a)(5)
The Surgimap software assists healthcare professionals in viewing, storing, and measuring images as well as planning orthopedic surgeries. The device allows service providers to perform generic as well as specialty measurements of the images, and to plan surgical procedures. The device also includes tools for measuring anatomical components for placement of surgical implants, and offer online synchronization of the database with the possibility to share data among Surgimap users. Clinical judgment and experience are required to properly use the software.
{5}------------------------------------------------
The Indications for Use statement for Surgimap 2.0 device is not identical to the predicate devices; however, the differences do not affect the safety and effectiveness of the device relative to the predicates. The subject and predicate devices have the same intended use for local images, and in both cases, the mitigation of hazard analysis demonstrates an acceptable level of residual risks.
### 807.92(a)(6)
The tabulated comparison of technological characteristics between Surgimap Spine and its predicated devices is outlined in the table hereafter:
| Feature | TraumaCad 2.0 | Surgimap Spine | Surgimap 2.0 |
|------------------------------------------------------------------------|------------------------------|----------------|------------------------------|
| Computer | PC Compatible | PC Compatible | PC Compatible |
| Operating System | Windows | Windows | Windows + MAC |
| Image Input | Local + PACS<br>connectivity | Local | Local + PACS<br>connectivity |
| Scout feature | Yes | No | Yes |
| Runs on Server | Yes | no | no |
| Trauma Module | Yes | No | No |
| Osteotomy Module | Yes | Yes | Yes |
| Generic Measurements | Yes | Yes | Yes |
| Spine measurements | Yes | Yes | Yes |
| Lower Limbs measurements | Yes | No | Yes |
| Pre-operative planning | Yes | Yes | Yes |
| Templating (custom implant) | Yes | Yes | Yes |
| Templating (vendor specific<br>implant) | Yes | No | Yes |
| Database | Yes | Yes | Yes |
| Online synchronization of<br>database | Yes | No | Yes |
| Case Sharing | Yes | No | Yes |
| Control of life-saving devices | None | None | None |
| Human Intervention for<br>interpretation and<br>manipulation of images | Required | Required | Required |
| Ability to add additional<br>Modules when available | Yes | Yes | Yes |
| Web Content | Yes | Yes | Yes |
{6}------------------------------------------------
## 807.92(b)
Performance data for Surgimap 2.0 consisted in verification activities. The following guidelines and standards were used throughout the software development process:
| EN ISO 14971:2012 | Medical Devices - Application of risk management to<br>medical devices |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 62304:2006 | Medical device software - Software life cycle processes |
| ANSI/IEEE Standard 829-2008 | IEEE Standard for Software and System Test Documentation |
| FDA Guidance | Guidance for Industry and FDA Staff - Guidance for the<br>content of premarket submissions for software contained in<br>medical devices (May 11, 2005) |
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered to be at a "Moderate" level of concern, since a latent design flaw might lead the user to inadequately plan a surgery and thus indirectly injure patients.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.