K192264 · Hivox Biotek, Inc. · NGX · Nov 8, 2019 · Physical Medicine
Device Facts
Record ID
K192264
Device Name
HIVOX Spopad EMS SP-911, SP-921
Applicant
Hivox Biotek, Inc.
Product Code
NGX · Physical Medicine
Decision Date
Nov 8, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 890.5850
Device Class
Class 2
Indications for Use
These Electrical Muscle Stimulation units are indicated for the improvement of muscle tone and firmness, for strengthening muscles in arms, abdomen, thighs and buttocks areas. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.
Device Story
HIVOX Spopad EMS (SP-911, SP-921) are battery-operated, 1-channel electrical muscle stimulators. Device generates symmetrical biphasic electrical pulses delivered to skin via integrated electrodes (2 for SP-911; 4 for SP-921). Used for muscle toning, firming, and strengthening. Operated by user in home/non-clinical setting. Device features automatic shut-off and user-controlled intensity. Output signals are delivered through skin to induce muscle contraction. Benefits include non-medication muscle training and self-exercise. Healthcare providers do not interact with output; device is for consumer self-use.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including shelf-life testing, performance verification, and usability validation.
Technological Characteristics
Powered by 3V CR2032 battery. Silicone housing. Symmetrical biphasic rectangular waveform. Pulse duration 400 μs. Frequency 3/4/5 Hz (SP-911) or 2/4/25 Hz (SP-921). Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 60601-2-10. Microprocessor-controlled.
Indications for Use
Indicated for improvement of muscle tone and firmness, and strengthening muscles in arms, abdomen, thighs, and buttocks. Not for therapy or treatment of medical conditions or diseases.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 8, 2019
Hivox Biotek Inc. Marx Lee Regulatory Affairs Representative 5F., No. 123, Xinde Rd. Sanchong Dist. New Taipei City, TW 24158
Re: K192264
Trade/Device Name: HIVOX Spopad EMS SP-911, SP-921 Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: October 7, 2019 Received: October 9, 2019
Dear Marx Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for: Vivek Pinto, Ph.D. Acting Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192264
Device Name HIVOX Spopad EMS SP-911, SP-921
#### Indications for Use (Describe)
These Electrical Muscle Stimulation units are indicated for the improvement of muscle tone and firmness, for strengthening muscles in arms, abdomen, thighs and buttocks areas. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Section 6
# 510(k) Summary
{4}------------------------------------------------
# 510(k) SUMMARY
| 5.1 Type of Submission | Special |
|------------------------|---------------------------------------------------------------------------------|
| 5.2 Date of Summary | November 8, 2019 |
| 5.3 Submitter | HIVOX BIOTEK INC. |
| Address: | 5F., No. 123, Xingde Rd., Sanchong Dist., New Taipei City 24158, Taiwan, R.O.C. |
| Phone: | +886-2-8511-2668 |
| Fax: | +886-2-8511-2669 |
| Contact: | Marx Lee |
| | (Marx.Lee@hivox-biotek.com) |
## 5.4 Identification of the Subject Device
| Proprietary/Trade name: | HIVOX Spopad EMS |
|------------------------------|-------------------------|
| Models: | SP-911, SP-921 |
| Classification product code: | NGX |
| Regulation number: | 890.5850 |
| Regulation description: | Power muscle stimulator |
| Review panel: | Physical Medicine |
| Device class: | II |
## 5.5 Identification of the Predicate Device
| Proprietary/Trade name: | HIVOX Spopad EMS |
|------------------------------|-------------------------|
| Models: | SP-910, SP-920 |
| 510(k) number: | K141921 |
| Classification product code: | NGX |
| Regulation number: | 890.5850 |
| Regulation description: | Power muscle stimulator |
| Review panel: | Physical Medicine |
| Device class: | II |
{5}------------------------------------------------
## 5.6 Intended Use / Indication for Use of the Device
These Electrical Muscle Stimulation units are indicated for the improvement of muscle tone and firmness, for strengthening muscles in arms, abdomen, thighs and buttocks areas. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.
#### 5.7 Description of the Device
EMS, Electrical Muscle Stimulator, which improves, tones, firms & strengthens muscle and relaxes stiff muscle through the skin. It is recognized as a clinically proven, effective, non-medication method of training muscle from certain causes. It manages muscle strengthen, toning and firming. It is also free from side effects when used properly, and can also be used as a simple means of self-training.
HIVOX Spopad EMS SP series, SP-911 and SP-921 are the proposed devices for the 510(k) submission. These Electrical Muscle Stimulation units are indicated for the improvement of muscle tone and firmness, for strengthening muscles in arms, abdomen, thighs, and buttocks areas. Not intended for use in therapy or for the treatment of any medical conditions or diseases.
SP-911 and SP-921 are 1-channel battery-operated-user-friendly muscle stimulation system specifically designed to exercise the muscles. Each device comprises namely an electronic stimulator module which generates the required stimulation signals. SP-911 comprises 2 electrodes, which connects the signals from the stimulator to the skin. SP-921 comprises 4 electrodes, which connect the signals from the stimulator to the skin. Power is supplied from one battery, CR2032, located in a compartment protected by a removable battery cover. The user cannot access the wiring or connectors.
#### 5.8 Non-clinical Testing
A series of safety and performance tests were conducted on the subject device, HIVOX Spopad EMS SP-911 and SP-921.
- Shelf-Life testing
- Performance verification
- Usability validation
All the test result demonstrate HIVOX Spopad EMS SP-911 and SP-921 meet the requirement of its pre-defined acceptance criteria and intended use, and is substantially equivalent to the predicate device.
{6}------------------------------------------------
# 5.9 Clinical Testing
No clinical test data was used to support the decision of substantial equivalence.
# 5.10 Comparison of Differences and Substantial Equivalence Determination 5.10.1 Comparison between SP-910 and SP-911
| Item | Subject device | Predicate device | Substantial<br>equivalence<br>determination | |
|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| Device name | HIVOX Spopad<br>EMS | HIVOX Spopad<br>EMS | N/A | |
| Model name | SP-911 | SP-910 | N/A | |
| 510(k) number | K192264 | K141921 | N/A | |
| Product code | NGX | NGX | Identical | |
| Classification name | Powered Muscle<br>Stimulator | Powered Muscle<br>Stimulator | Identical | |
| Regulation number | 21 CFR 890.5850 | 21 CFR 890.5850 | Identical | |
| Indication for use | Indicated for the<br>improvement of<br>muscle tone and<br>firmness, for<br>strengthening<br>muscles in arms,<br>abdomen, thighs,<br>and buttocks<br>areas.<br>Not intended for<br>use in any<br>therapy or for the<br>treatment of any<br>medical<br>conditions or<br>diseases. | Indicated for the<br>improvement of<br>muscle tone and<br>firmness, for<br>strengthening<br>muscles in arms,<br>abdomen, thighs,<br>and buttocks<br>areas.<br>Not intended for<br>use in any<br>therapy or for the<br>treatment of any<br>medical<br>conditions or<br>diseases. | Identical | |
| Technology | Electrical Muscle<br>Stimulation | Electrical Muscle<br>Stimulation | Identical | |
| Size (L x W x H, inch) | 8.15 × 4.06 × | 9.41 × 2.76 × | Different but | |
| | 0.51 | 0.45 | does not<br>adversely<br>impact safety<br>and<br>effectiveness of<br>subject device | |
| Weight (g) | 24.2 | 35.8 | Different but<br>does not<br>adversely<br>impact safety<br>and<br>effectiveness of<br>subject device | |
| Power source | 3 V battery × 1 | 3 V battery × 1 | Identical | |
| Method of Line Current Isolation | Battery supply | Battery supply | Identical | |
| Patient Leakage Current<br>Normal Condition (μΑ) | 2.0 | 2.0 | Identical | |
| Patient Leakage Current<br>Single Fault Condition (μΑ) | 2.1 | 2.1 | Identical | |
| Method of channel isolation | Single channel | Single channel | Identical | |
| Average DC current through<br>electrodes when device is on but<br>no pulses are being applied (μΑ) | 0 | 0 | Identical | |
| Number of output modes | 1 | 1 | Identical | |
| Regulated current or regulated<br>voltage? | Voltage | Voltage | Identical | |
| Software / Firmware /<br>Microprocessor control? | Yes | Yes | Identical | |
| Automatic overload trip? | No | No | Identical | |
| Automatic no-load trip? | No | No | Identical | |
| Automatic shut-off? | Yes | Yes | Identical | |
| User overrides control? | Yes | Yes | Identical | |
| Indicator display –<br>On / Off Status | No | No | Identical | |
| Indicator display – | No | No | Identical | |
| Low battery? | | | | |
| Indicator display –<br>Voltage / Current level | No | No | Identical | |
| Timer range (minutes) | 20 | 20 | Identical | |
| Compliance with voluntary<br>standards? | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-10 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-10 | Identical | |
| Compliance with 21 CFR 898? | Yes | Yes | Identical | |
| Housing material and construction | Silicone | Silicone | Identical | |
| Output waveform | Symmetrical<br>biphasic | Symmetrical<br>biphasic | Identical | |
| Shape | Rectangular | Rectangular | Identical | |
| Duration of primary<br>(depolarizing) phase | 0 | 0 | Identical | |
| Pulse duration (μs) | 400 | 400 | Identical | |
| Maximum output voltage<br>(V, ±10%) at 500 Ω | 52 | 52 | Identical | |
| Maximum output voltage<br>(V, ±10%) at 2k Ω | 102 | 102 | Identical | |
| Maximum output voltage<br>(V, ±10%) at 10k Ω | 150 | 150 | Identical | |
| Maximum output current<br>(mA, ±10%) at 500 Ω | 104 | 104 | Identical | |
| Maximum output current<br>(mA, ±10%) at 2k Ω | 51 | 51 | Identical | |
| Maximum output current<br>(mA, ±10%) at 10k Ω | 15 | 15 | Identical | |
| Frequency (Hz) | 3/4/5 | 3/4/5 | Identical | |
| Net charge per pulse at 500 Ω<br>(μC) | 0.416 | 0.416 | Identical | |
| Maximum charge at 500 Ω (μC) | 41.6 | 41.6 | Identical | |
| Conductive surface area (cm²) | 62.20<br>(Two electrodes) | 87.58<br>(Two electrodes) | Identical | |
| Maximum current density at 500 | 1.672 | 1.187 | Different but | |
| Ω (mA/cm²) | | | | does not<br>adversely<br>impact safety<br>and<br>effectiveness of<br>subject device |
| Maximum average power density<br>at 500 Ω (W/cm²) | | 0.0869 | 0.0617 | Different but<br>does not<br>adversely<br>impact safety<br>and<br>effectiveness of<br>subject device |
| | | | | |
| Burst<br>mode | A. Pulse per burst | N/A | N/A | Identical |
| | B. Burst per second | N/A | N/A | Identical |
| | C. Burst duration (sec) | N/A | N/A | Identical |
| | D. Duty cycle | N/A | N/A | Identical |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
## 5.10.2 Comparison between SP-920 and SP-921
| Item | Subject device | Predicate device | Substantial<br>equivalence<br>determination | |
|----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|-----------|
| Device name | HIVOX Spopad<br>EMS | HIVOX Spopad<br>EMS | N/A | |
| Model name | SP-921 | SP-920 | | |
| 510(k) number | K192264 | K141921 | | |
| Product code | NGX | NGX | Identical | |
| Classification name | Powered Muscle<br>Stimulator | Powered Muscle<br>Stimulator | Identical | |
| Regulation number | 21 CFR 890.5850 | 21 CFR 890.5850 | Identical | |
| Indication for use | Indicated for the<br>improvement of<br>muscle tone and<br>firmness, for<br>strengthening | Indicated for the<br>improvement of<br>muscle tone and<br>firmness, for<br>strengthening | Identical | |
| | | | | |
| | muscles in arms,<br>abdomen, thighs,<br>and buttocks<br>areas.<br>Not intended for<br>use in any<br>therapy or for the<br>treatment of any<br>medical<br>conditions or<br>diseases. | muscles in arms,<br>abdomen, thighs,<br>and buttocks<br>areas.<br>Not intended for<br>use in any<br>therapy or for the<br>treatment of any<br>medical<br>conditions or<br>diseases. | | |
| Technology | Electrical Muscle<br>Stimulation | Electrical Muscle<br>Stimulation | Identical | |
| | | | Different but<br>does not<br>adversely<br>impact safety<br>and<br>effectiveness of<br>subject device | |
| Size (L × W × H, inch) | $7.17 × 6.81 ×0.51$ | $6.69 × 6.69 ×0.51$ | | |
| | | | Different but<br>does not<br>adversely<br>impact safety<br>and<br>effectiveness of<br>subject device | |
| Weight (g) | 48.3 | 52.6 | | |
| Power source | 3 V battery × 1 | 3 V battery × 1 | Identical | |
| Method of Line Current Isolation | Battery supply | Battery supply | Identical | |
| Patient Leakage Current<br>Normal Condition (μΑ) | 2.0 | 2.0 | Identical | |
| Patient Leakage Current<br>Single Fault Condition (μΑ) | 2.1 | 2.1 | Identical | |
| Method of channel isolation | Single channel | Single channel | Identical | |
| Average DC current through | 0 | 0 | Identical | |
| electrodes when device is on but<br>no pulses are being applied (μΑ) | | | | |
| Number of output modes | 1 | 1 | Identical | |
| Regulated current or regulated<br>voltage? | Voltage | Voltage | Identical | |
| Software / Firmware /<br>Microprocessor control? | Yes | Yes | Identical | |
| Automatic overload trip? | No | No | Identical | |
| Automatic no-load trip? | No | No | Identical | |
| Automatic shut-off? | Yes | Yes | Identical | |
| User overrides control? | Yes | Yes | Identical | |
| Indicator display –<br>On / Off Status | No | No | Identical | |
| Indicator display –<br>Low battery? | No | No | Identical | |
| Indicator display –<br>Voltage / Current level | No | No | Identical | |
| Timer range (minutes) | 20 | 20 | Identical | |
| Compliance with voluntary<br>standards? | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-10 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-10 | Identical | |
| Compliance with 21 CFR 898? | Yes | Yes | Identical | |
| Housing material and construction | Silicone | Silicone | Identical |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.