K141921 · Hivox Biotek, Inc. · NGX · Jan 16, 2015 · Physical Medicine
Device Facts
Record ID
K141921
Device Name
HIVOX SPOPAD EMS
Applicant
Hivox Biotek, Inc.
Product Code
NGX · Physical Medicine
Decision Date
Jan 16, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Indications for Use
These Electrical Muscle Stimulation units are indicated for the improvement of muscle tone and firmness, for strengthening muscles in arms, abdomen, thighs, and buttocks areas. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.
Device Story
HIVOX Spopad EMS (SP-910, SP-920, SP-620) are 1-channel, battery-operated powered muscle stimulators; designed for OTC self-use to exercise muscles. Device comprises electronic stimulator module generating stimulation signals; 2 or 4 electrodes connect signals to skin. Power supplied by CR2032 battery. User-friendly system; improves muscle tone, firmness, and strength; relaxes stiff muscles. Device operates via microprocessor control; delivers symmetrical biphasic rectangular waveforms. User applies device to target areas; output affects muscle contraction/relaxation cycles. Benefits include non-medication muscle training and self-training capability.
Clinical Evidence
Bench testing only. No clinical data provided. Evidence consists of comparative electrical output analysis, compliance with IEC 60601-1 (general safety), IEC 60601-2-10 (nerve/muscle stimulators), and IEC 60601-1-2 (EMC).
Technological Characteristics
Battery-powered (3V CR2032); silicone housing. Symmetrical biphasic rectangular waveform. 1-channel output. Microprocessor-controlled. Compliance: IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2. Max current density <2 mA/cm²; max power density <0.0691-0.117 W/cm². Timer: 20 minutes.
Indications for Use
Indicated for healthy individuals seeking improvement of muscle tone, firmness, and strengthening in arms, abdomen, thighs, and buttocks. Not for use in therapy or treatment of medical conditions/diseases.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of three human profiles facing right, connected by flowing lines that resemble ribbons or streams. The profiles are arranged in a staggered formation, with the first profile being the largest and the subsequent profiles gradually decreasing in size. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 16, 2015
HIVOX BIOTEK, INC. Ke-Min Jen. Official Correspondent 5 F., No. 123 Shingde Road San-chong District New Taipei City. TW 24158
Re: K141921
Trade/Device Name: HIVOX Spopad EMS SP-910, SP-920, SP-620 Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: December 7, 2014 Received: December 11, 2014
Dear Dr. Jen. Ke-Min:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Felipe Aquel -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K141921
Device Name HIVOX Spopad EMS SP-910, SP-920, SP-620
#### Indications for Use (Describe)
These Electrical Muscle Stimulation units are indicated for the improvement of muscle tone and firmness, for strengthening muscles in arms, abdomen, thighs, and buttocks areas. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.
Type of Use (Select one or both, as applicable)
- Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "HIVOX" in a sans-serif font. The letters are black, and there is a registered trademark symbol after the "X". To the right of the word is a plus sign made up of four squares. The top left square is green, and the bottom right square is blue.
# 5. 510(k) SUMMARY (According to 21 CFR 807.92)
| ● | 510(K) OWNER'S<br>HIVOX BIOTEK INC. | |
|---|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | NAME | 5F, No.123, Shingde Road, Sanchong Dist.,<br>New Taipei City, 24158, TAIWAN, R.O.C. |
| | | TEL: +886-2-85112668 FAX:+886-2-85112669 |
| ● | Name Of Contact<br>Person | Dr. JEN, KE-MIN<br>TEL: 886-2-85112668 FAX:886-3-5209783<br>Email: ceirs.jen@msa.hinet.net |
| ● | Date Of Submission | July 5, 2014 |
| ● | Trade Name | HIVOX Spopad EMS SP-910, SP-920, SP-620 |
| ● | Common Name | Powered Muscle Stimulator |
| ● | Classification Name | Powered Muscle Stimulator<br>(21 CFR 890.5850, Product Code NGX) |
| ● | Panel | Physical Medicine |
| ● | Intended Use | These Electrical Muscle Stimulation units are indicated for the<br>improvement of muscle tone and firmness, for strengthening<br>muscles in arms, abdomen, thighs, and buttocks areas.<br>Not intended for use in any therapy or for the treatment of any<br>medical conditions or diseases. |
| ● | Device Design | EMS, Electrical Muscle Stimulation, which improves, tones,<br>firms & strengthens muscle and relaxes stiff muscle through the<br>skin. It is recognized as a clinically proven, effective,<br>non-medication method of training muscle from certain causes.<br>It manages muscle strengthen, toning and firming. It is also free<br>from side effects when used properly, and can also be used as a<br>simple means of self-training. |
HIVOX Spopad EMS SP series, SP-910 / SP-920 / SP-620 are the proposed subject devices for this 510(k) submission.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "HIVOX" in bold, black letters. To the right of the word is a plus sign made up of four squares. The top left and bottom right squares are green, and the top right and bottom left squares are blue.
5F., No. 123, Shingde Road, Sanchong District, New Taipei City, Taiwan Phone: +886 2 8511 2668 Fax: +886 2 8511 2669
These Electrical Muscle Stimulation units are indicated for the improvement of muscle tone and firmness, for strengthening muscles in arms, abdomen, thighs, and buttocks areas. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.
SP-910 / SP-920 / SP-620 are 1-channel battery-operated user-friendly muscle stimulation systems specifically designed to exercise the muscles. Each device comprises namely an electronic stimulator module which generates the required stimulation signals. SP-910/620 comprises 2 electrodes, which connects the signals from the stimulator to the skin. SP-920 comprises 4 electrodes, which connects the signals from the stimulator to the skin. Power is supplied from one battery, CR2032, located in a compartment protected by a removable battery cover. The user cannot access the wiring or connectors.
{5}------------------------------------------------
5F., No. 123, Shingde Road, Sanchong District, New Taipei City, Taiwan Phone: +886 2 8511 2668 Fax: +886 2 8511 2669
| Comparison items | Predicate device 1<br>(PD) | Subject device<br>(1) | Subject device<br>(2) | Subject device<br>(3) |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|-----------------------|
| Manufacturer<br>Submitter | SPORT-ELEC S.A. | HIVOX-BIOTEK | | |
| Device name | Body Control<br>System | Spopad EMS, SP series | | |
| Model number | 4M | SP-910 | SP-920 | SP-620 |
| 510(k) number | K092476 | TBA | TBA | TBA |
| Product code | NGX | | NGX | |
| Classification name | Powered Muscle<br>Stimulator | Powered Muscle Stimulator | | |
| Regulation number | 21 CFR 890.5850 | 21 CFR 890.5850 | | |
| Indications for use | Indicated for the<br>improvement of<br>muscle tone and<br>firmness, for<br>strengthening<br>muscles in arms,<br>abdomen, thighs<br>and buttocks areas.<br>Contraindicated<br>use on injured or<br>otherwise impaired<br>muscles<br>Not intended for<br>use in any therapy<br>or for the treatment<br>of any medical<br>conditions or<br>diseases. | Indicated for the improvement of muscle tone<br>and firmness, for strengthening muscles in arms,<br>abdomen, thighs, and buttocks areas.<br>Not intended for use in any therapy or for the<br>treatment of any medical conditions or diseases. | | |
| Technology | Electrical Muscle<br>Stimulation | Electrical Muscle Stimulation | | |
| Power Source | 1.5V battery *3 | 3V Battery *1 | | |
| - Method of Line<br>Current Isolation | Battery supply | Battery Supply | | |
| - Patient Leakage<br>Current<br>Normal Condition (µA) | < 3 | | 2.0 | |
| Single Fault<br>Condition(µA) | < 4 | | 2.1 | |
| Method of channel<br>isolation | Software | 1 channel | | |
## Compare to Legally Marketed Predicate Devices _
{6}------------------------------------------------
5F., No. 123, Shingde Road, Sanchong District, New Taipei City, Taiwan Phone: +886 2 8511 2668 Fax: +886 2 8511 2669
| Average DC current<br>through electrodes when<br>device is on but no<br>pulses are being applied | 0 μΑ | 0 μΑ |
|--------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|------------------------------------------------|
| Number of output<br>modes | 1 | 1 |
| Regulated current or<br>regulated voltage? | Voltage | Voltage |
| Software /firmware /<br>Microprocessor control? | Yes | Yes |
| Automatic overload<br>trip? | No | No |
| Automatic no-load trip? | No | No |
| Automatic shut-off? | Yes | Yes |
| User overrides control? | Yes | Yes |
| Indicator display -<br>On/Off Status | Yes | No |
| Indicator display – Low<br>battery? | Yes | No |
| Indicator display –<br>Voltage /Current | No | No |
| Timer Range (minutes) | N/A | 20 |
| Compliance with<br>voluntary standards? | IEC 60601-1<br>IEC 60601-2-10<br>IEC 60601-1-2<br>IEC 60601-1-4 | IEC 60601-1<br>IEC 60601-2-10<br>IEC 60601-1-2 |
| Compliance with 21<br>CFR 898? | Yes | Yes |
| Housing material and<br>construction | ABS | Silicone |
| Output waveform | Monophasic | Symmetrical biphasic |
| Shape | Rectangular | Rectangular |
| Duration of primary<br>(depolarizing) phase | 0 | 0 |
| Pulse duration (μSec) | N/A | 400 |
| Maximum output<br>voltage (Voltage,<br>+/-10%) at 500 ohms | N/A | 52<br>58.4<br>60 |
| Maximum output<br>voltage (Voltage,<br>+/-10%) at 2k ohms | N/A | 102<br>106<br>109 |
| Maximum output<br>voltage (Voltage,<br>+/-10%) at 10k ohms | N/A | 150<br>146<br>140 |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for HIVOX. The logo consists of the word "HIVOX" in black, sans-serif font. To the right of the word is a plus sign made up of four squares. The top left square is green, the top right square is white, the bottom left square is white, and the bottom right square is blue.
5F., No. 123, Shingde Road, Sanchong District, New Taipei City, Taiwan Phone: +886 2 8511 2668 Fax: +886 2 8511 2669
| Maximum output<br>current (mA +/-10%) at<br>500 ohms | N/A | 104 | 117 | 120 | |
|---------------------------------------------------------|-------------------------|--------|--------|--------|----|
| Maximum output<br>current (mA +/-10%) at<br>2k ohms | N/A | 51 | 53 | 54.5 | |
| Maximum output<br>current (mA +/-10%) at<br>10k ohms | N/A | 15 | 14.6 | 14 | |
| Frequency (Hz) | N/A | 3/4/5 | 2/4/25 | 2/4/25 | |
| Net charge per pulse at<br>500 ohms (µC) | N/A | 0.416 | 0.468 | 0.960 | |
| Maximum charge at 500<br>ohms (µC) | N/A | 41.6 | 46.8 | 48 | |
| Maximum current<br>density at 500 ohms<br>(mA/cm²) | <2 | 1.328 | 1.057 | 1.952 | |
| Maximum average<br>power density at 500<br>ohms (W/cm²) | < 0.25 | 0.0691 | 0.0617 | 0.117 | |
| Burst mode | A. Pulse per burst | N/A | N/A | 25 | 25 |
| | B. Burst per second | N/A | N/A | 1 | 1 |
| | C. Burst duration (sec) | N/A | N/A | 20 | 20 |
| | D. Duty cycle | N/A | N/A | 20 | 20 |
#### Summary of comparison
Basically, the predicate device and Subject Devices are all Over-The-Counter muscle stimulators. Thus, the indications for use for the devices do not differ much.
PD and Subject Devices are all battery-powered, thus electric hazards or safety do not raise much concern, since maximum current / power densities for the devices are all less than 2 (mA/cm2) / 0.25 (W/cm2), which are the FDA recommended ratings. Since the electric output data for PD and Subject Devices only exist minor differences, the minor differences of the electric outputs do not raise any safety and effectiveness aspect. PD and Subject Devices all have the same safety and effectiveness, especially they all pass medical device electric safety standard, IEC 60601-1 and standard for Nerve and Muscle Stimulator, IEC 60601-2-10 and electromagnetic compatibility standard IEC 60601-1-2.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for HIVOX. The logo consists of the word "HIVOX" in a bold, sans-serif font, followed by a registered trademark symbol. To the right of the wordmark is a stylized plus sign composed of four squares, two in green and two in blue. The overall design is clean and modern.
5F., No. 123, Shingde Road, Sanchong District, New Taipei City, Taiwan Phone: +886 2 8511 2668 Fax: +886 2 8511 2669
We know that for effectiveness in achieving repeated muscle contractions, powered muscle stimulators typically are capable of stimulating muscle for at least one second per burst, and are capable of providing at least one second of muscle relaxation between successive pulse bursts. PD should meet these requirements. Subject Devices meet these requirements too. PD and Subject Devices are all validating-software processing, thus keeping regular processing parameters and safety functions normal. The Subject Devices have the same effectiveness as the predicate device.
In conclusion, the Subject Devices do not raise any new safety and effectiveness aspect with respect to the PD. Thus the Subject Devices are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.