K192260 · GC America, Inc. · EBF · Nov 22, 2019 · Dental
Device Facts
Record ID
K192260
Device Name
everX Flow
Applicant
GC America, Inc.
Product Code
EBF · Dental
Decision Date
Nov 22, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
everX Flow is a reinforcing dentin replacement material suitable for: 1. All direct composite restorations including large posterior cavities. 2. Deep cavities and endo-treated teeth. 3. Cavities with missing cusps or after amalgam removal and cavities where inlays would also be recommended. 4. Core build-up
Device Story
everX Flow is a light-cured, short-fiber-reinforced composite resin used as a dentin replacement material in dental clinics. It is supplied in a syringe and applied by dental professionals to restore posterior and anterior teeth. The material functions as a reinforcing base for direct composite restorations, deep cavities, endo-treated teeth, and core build-ups. It polymerizes via a photo-initiator system upon exposure to a dental curing light. By providing structural reinforcement, it benefits patients by supporting restorations in compromised tooth structures, such as those with missing cusps or significant decay. The clinician uses the material to fill the cavity preparation before placing a final composite layer, improving the durability and integrity of the restoration.
Clinical Evidence
No clinical testing performed. Substantial equivalence is supported by bench testing confirming compliance with ISO 4049:2009 standards for sensitivity to ambient light, depth of cure, flexural strength, water sorption, solubility, color stability, and radiopacity. Biocompatibility was assessed per ISO 10993-1:2009 and ISO 10993-5, confirming the material is non-cytotoxic.
Technological Characteristics
Light-cured, short-fiber-reinforced composite resin. Materials: methacrylate ester monomers. Curing: photo-initiator system. Form factor: syringe-dispensed. Standards: ISO 4049:2009 (polymer-based restorative materials), ISO 10993-1:2009 (biocompatibility).
Indications for Use
Indicated for patients requiring direct composite restorations, including large posterior cavities, deep cavities, endo-treated teeth, cavities with missing cusps, post-amalgam removal, or core build-up.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.
November 22, 2019
GC America Inc. Dr. Mark Heiss Director, Regulatory Affairs 3737 W. 127th Street Alsip, Illinois 60803
Re: K192260
Trade/Device Name: everX Flow Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: August 20, 2019 Received: August 21, 2019
Dear Mark Heiss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration |
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
| 510(k) Number (if known) |
|--------------------------|
|--------------------------|
| Device Name | everX Flow |
|--------------------------------|------------|
| Indications for Use (Describe) | |
everX Flow is a reinforcing dentin replacement material suitable for:
- 1. All direct composite restorations including large posterior cavities.
- 2. Deep cavities and endo-treated teeth.
- 3. Cavities with missing cusps or after amalgam removal and cavities where inlays would also be recommended.
- 4. Core build-up
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the letters 'GC' in a stylized, sans-serif font. The letters are a teal color, and they appear to be slightly italicized or slanted to the right. There are single quotation marks on either side of the letters.
Image /page/3/Picture/1 description: The image shows a simple, abstract design. On the left side, there's a partial outline of a triangle, depicted in a light teal color. To the right of the triangle, there's a cluster of small, rectangular shapes, also in teal, arranged in a pattern that suggests a scattered or slightly organized grouping. The background is plain white, which makes the teal elements stand out.
### 510(k) Summary - K192260
| 1. Submitter Information: | GC America Inc. |
|---------------------------|----------------------------|
| | 3737 W. 127th Street |
| | Alsip, IL 60803 |
| Contact Person: | Mark Heiss, D.D.S. |
| Phone: | (708) 926-3090 |
| Alternate Contact: | Lori Rietman |
| Phone: | (708) 926-3092 |
| Fax: | (708) 925-0373 |
| Date Prepared: | October 28, 2019 |
| 2. Device Name: | |
| Proprietary Name: | everX Flow |
| Classification Name: | Tooth shade resin material |
| Device Classification: | Class II, 872.3690 |
- 3. Predicate Devices:
Product Code:
| Product | Applicant | 510(k) No. | Code No | Predicate | Decision Date |
|-----------------|-----------------|------------|---------|-----------|---------------|
| everX Posterior | GC America Inc. | K153127 | EBF | Primary | 05/20/2016 |
| Gradia Core | GC America Inc. | K082171 | EBF | Reference | 10/30/2008 |
| G-aenial Flo X | GC America Inc. | K133182 | EBF | Reference | 4/9/2014 |
#### 4. Description of Device:
everX Flow is a light cured, short-fiber-reinforced composite for dentin replacement for the restoration of posterior and anterior teeth. EverX Flow is filled in a syringe. The material is available in two shades, Bulk shade and Dentin shade.
### 5. Indications for Use:
- everX Flow is a reinforcing dentin replacement material suitable for: ● All direct composite restorations including large posterior cavities.
EBF
- . Deep cavities and endo-treated teeth.
- . Cavities with missing cusps or after amalgam removal and cavities where inlays would also be recommended.
- . Core build-up
#### 6. Packaging
everX Flo Package:
- i Syringe (3.7 g / 2 mL) QTY: 1
- Dispensing Tip III QTY: 20 -
- Light protective cap QTY: 1 i
- 7. Shades
Shades available: Bulk Shade, Dentin Shade
{4}------------------------------------------------
- 8. Shelf Life and Storage Conditions:
- Shelf Life 3 years ।
- Recommended for optimal performance, store in a cool and dark place (4-25°C / 39.2-77.0°F) away from high temperatures i or direct sunlight.
#### Performance Bench Tests 9.
It is confirmed that the device conforms to the required specifications and is suitable for its intended use. Performance testing includes:
- · Sensitivity to ambient light
- · Depth of cure
- · Flexural strength
- · Water sorption
- · Solubility
- · Color stability after irradiation and water sorption
- · Radiopacity
The applicant device complies with all the requirements of ISO 4049: 2009 (Dentistry - Polymer-based restorative materials) (see table below).
| | Property | Requirements |
|---|------------------------------------------------------|-------------------------------------------------|
| 1 | Sensitivity to ambient light | Remain physically homogeneous |
| 2 | Depth of cure | Opaque shade: > 1.0 mm<br>Other shade: > 1.5 mm |
| 3 | Flexural strength | > 80 MPa |
| 4 | Water sorption | < 40 µg/mm³ |
| 5 | Solubility | < 7.5 µg/mm³ |
| 6 | Color stability after irradiation and water sorption | No more than slight change in color |
| 7 | Radiopacity | Greater than the same thickness of aluminum |
### 10. Non-Clinical Performance Testing
A biocompatibility assessment was completed according to ISO 10993-1:2009, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.
Based on the criteria of the protocol and the ISO 10993-5 guidelines, the test article meets the requirements of the test and is not considered to have a cytotoxic effect.
- 11. Clinical Performance Testing
No clinical testing has been performed on this device.
# 12. Comparison of Technology
The main difference between the applicant device and primary predicate device is the flow. The indications for use were similar as well. The formula was modified without issue associated with safety and efficacy. In addition, biocompatibility studies were successfully completed. Mechanism for curing and polymerization did not change. Also, depth of cure was similar.
- 2
{5}------------------------------------------------
| | Applicant device | Primary Predicate Device | Reference Predicate Device | Reference Device |
|-----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product<br>category | Tooth shade resin material, Class II | Tooth shade resin material, Class II | Tooth shade resin material, Class II | Tooth shade resin material, Class II |
| Trade name | everX Flow | everX Posterior | Gradia Core | G-AENIAL FLO X (HTFX-222) |
| 510(k) | Pending | K153127 | K082171 | K133182 |
| Manufacturer | GC Corporation | GC Corporation | GC Corporation | GC Corporation |
| Indications for<br>use | everX Flow is a reinforcing dentin<br>replacement material suitable for:<br>1. All direct composite restorations<br>including large posterior cavities.<br>2. Deep cavities and endo-treated teeth.<br>3. Cavities with missing cusps or after<br>amalgam removal and cavities where<br>inlays and onlays would also be<br>recommended.<br>4. Core build-up | everX Posterior is suitable for use as the<br>reinforcing material for direct composite<br>restorations, especially in large posterior<br>cavities, for example:<br>1. Cavities including 3 surfaces or more.<br>2. Cavities with missing cusps.<br>3. Deep cavities (including class I, II and<br>endodontically treated teeth).<br>4. Cavities after amalgam replacement.<br>5. Cavities where onlays & inlays would also<br>be indicated. | GRADIA CORE is intended to be used for<br>restoration, core build-up and post<br>cementation. | 1. Liner or base<br>2. Blocking out undercuts<br>3. Repair of (in) direct aesthetic<br>restorations, temporary crown &<br>bridge, defect margins when margins<br>are in enamel<br>4. Sealing hypersensitive areas<br>5. Fissure sealant<br>6. Direct restorative for small Class II,<br>III, IV, I and V cavities |
| Product<br>description | everX Flow is a light cured, short-fiber-<br>reinforced composite for dentin<br>replacement for the restoration of<br>posterior and anterior teeth. EverX Flow<br>is filled in a syringe. The material is<br>available in two shades, Bulk shade and<br>Dentin shade. | everX Posterior is a fiber-reinforced<br>composite resin filled in a unitip. The device<br>is a universal type. The material is available<br>in one universal shade. The device is<br>suitable for use as the reinforcing material<br>for direct composite restorations, especially<br>in large posterior cavities. | GRADIA CORE consists of the two<br>articles, GRADIA CORE cartridge and<br>GRADIA CORE self-etching bond.<br>GRADIA CORE cartridge is a two-paste<br>(base and catalyst) type, dual-cured<br>composite resin filled in double syringe to<br>provide auto-mixing system. GRADIA<br>CORE cartridge paste comprises four<br>shades, Universal, Dentin, White and Blue.<br>GRADIA CORE self-etching bond is a two-<br>liquid (liquid A and liquid B), dual-cured<br>self-etching bonding agent. | G-AENIAL FLO X (HTFX-222) is a light<br>cured nano-filled radiopaque composite<br>resin filled in a syringe. The device is<br>used for the restorations of both<br>anterior and posterior teeth |
| Technological<br>Characteristics<br>and Mode of<br>action | The curing mechanism is polymerization<br>of uncured methacrylate ester<br>monomers. This reaction is caused by<br>photo initiator system. | The curing mechanism of the predicate is<br>polymerization of uncured methacrylate<br>ester monomers. This reaction is caused by<br>photo initiator system. | The curing mechanism of the predicate is<br>polymerization of uncured methacrylate<br>ester monomers. This reaction is caused<br>by photo initiator system and is assisted by<br>chemical curing. | The curing mechanism of the predicate<br>devices is polymerization of uncured<br>methacrylate ester monomers. This<br>reaction is caused by photo initiator<br>system. |
# 13. Conclusion
Based upon similarities in technology and indical performance testing, we believe that ever X Flow is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.