K153127 · GC America, Inc. · EBF · May 20, 2016 · Dental
Device Facts
Record ID
K153127
Device Name
everX Posterior
Applicant
GC America, Inc.
Product Code
EBF · Dental
Decision Date
May 20, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
everX Posterior is indicated for use as a fiber reinforced composite for the restoration of prepared carious teeth, including large posterior cavities with loss of dentin.
Device Story
everX Posterior is a light-cured, fiber-reinforced, universal-shade composite resin paste. It is used by dental professionals in a clinical setting to restore prepared carious teeth, particularly large posterior cavities with dentin loss. The device consists of randomly oriented short E-glass fibers and inorganic fillers in a semi-interpenetrating polymer network matrix. The material is applied to the tooth cavity and polymerized via light curing to form a solid dental restoration. The fibers provide structural reinforcement to the composite mix. The device is supplied in unitips.
Clinical Evidence
Bench testing only. Testing performed in accordance with ISO 4049:2009 for flexural strength, depth of cure, water sorption, solubility, color stability, and sensitivity to ambient light. Biocompatibility testing conducted per ISO 10993-1, including cytotoxicity (ISO 10993-5), sensitization, and irritation (ISO 10993-10). Results confirm the device meets specifications for its intended use.
Technological Characteristics
Light-cured composite resin; contains randomly oriented short E-glass fibers (0.8 mm avg length) and inorganic particulate fillers. Matrix: bisphenol A glysidyl methacrylate, triethyleneglycol dimethacrylate, and polymethylmethacrylate. Form factor: unitip paste. Standards: ISO 4049:2009 (physical properties), ISO 10993-1/5/10 (biocompatibility).
Indications for Use
Indicated for restoration of prepared carious teeth, specifically large posterior cavities with loss of dentin.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized design featuring a series of overlapping human profiles, creating a sense of unity and connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 20, 2016
GC America Inc. Mark Heiss, D.D.S. Director, Academic & Regulatory Affairs 3737 W. 127th Street Alsip, Illinois 60803
Re: K153127
Trade/Device Name: everX Posterior Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: II Product Code: EBF Dated: April 21, 2016 Received: April 22, 2016
Dear Dr. Heiss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mark Heiss, D.D.S.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang
-s
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Section 4 – Indications for Use Statement
# Indications for Use
510(k) Number: K153127
Device Name: everX Posterior
Indications for Use:
everX Posterior is indicated for use as a fiber reinforced composite for the restoration of prepared carious teeth, including large posterior cavities with loss of dentin.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 4.1 of 4.1
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the text "'GC'" in a stylized, sans-serif font. The text is a teal color, and the letters are slightly italicized. The apostrophes are also teal and are placed at the beginning and end of the text.
GC AMERICA INC.
3737 W 127th STREET
ALSIP, ILLINOIS 60803
TEL (708) 597-0900
FAX (708) 926-9100
www.gcamerica.com
- 1. Submitter Information:
GC America Inc. 3737 W. 127th Street Alsip, IL 60803
| Contact Person: | Mark Heiss, D.D.S. |
|--------------------|--------------------|
| Phone: | (708) 926-3090 |
| Alternate Contact: | Lori Rietman |
| Phone: | (708) 926-3092 |
| Fax: | (708) 926-9100 |
Date Prepared: April 21, 2016
- 2. Device Name:
| 510(k) Number: | K153127 |
|------------------------|----------------------------|
| Proprietary Name: | everX Posterior |
| Classification Name: | Tooth shade resin material |
| Device Classification: | Class II, 872.3690 |
| Product Code: | EBF |
- 3. Predicate Devices:
| | Product | Applicant | Submission | Procode |
|-----------|------------------------------|------------------------------------|------------|---------|
| Primary | Build-It Total Core | Sybron Dental Specialties,<br>Inc. | K102703 | EBF |
| Reference | G-aenial Sculpt<br>(MFP-051) | GC America Inc. | K123631 | EBF |
#### 4. Description of Device:
everX Posterior is a light-cured, fiber-reinforced, universal type composite resin packaged in a unitip and available in one universal shade.
- న్. Indications for Use:
everX Posterior is indicated for use as a fiber reinforced composite for the restoration of prepared carious teeth, including large posterior cavities with loss of dentin.
### Technological characteristics: 6.
The applicant device is composed of randomly oriented short E-glass and inorganic particulate fillers in combination with a semi-interpenetrating polymer network matrix that consists of bisphenol A glysidyl methacrylate, triethyleneglycol dimethacrylate and polymethylmethacrylate. The glass fibers are pre-impregnated into the composite resin and are 0.8 mm average length.
{4}------------------------------------------------
| | Applicant device | Predicate Device | Reference Device |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name | everX Posterior | Build-It Total Core | G-ænial Sculpt (MFP-051) |
| Manufacturer | GC Corporation | Sybron Dental Specialties | GC Corporation |
| Indications<br>for Use | everX Posterior is indicated for<br>use as a fiber reinforced<br>composite for the restoration of<br>prepared carious teeth, including<br>large posterior cavities with loss<br>of dentin. | Build-it Total Core is indicated for<br>use as<br>1. Core build up material on vital<br>and non-vital teeth<br>2. Sealing root canal openings,<br>3. Cavity floor liner,<br>4. Post cementation,<br>5. Restorative material<br>6. Dentin replacement material. | 1. Direct restorative for class I,<br>II, III, IV, V cavities.<br>2. Direct restorative for<br>wedge-shaped defects and root<br>surface cavities.<br>3. Direct restorative for veneers<br>and diastema closure. |
| Product<br>description | everX Posterior is a<br>fiber-reinforced, universal shade<br>composite resin paste filled in a<br>unitip suitable as a reinforcing<br>material for direct composite<br>restorations, especially in large<br>posterior cavities.<br>Physical properties are within the<br>acceptable range for flexural<br>strength, depth of cure, water<br>sorption, solubility in accordance<br>with ISO 4049:2009 Dentistry -<br>Polymer based restorative materials | Build-it Total Core is a two<br>component, catalyst and base, fiber<br>reinforced, self-adhesive composite<br>that contains fluoride, and is<br>available in esthetic and contrast<br>shades<br>Physical properties tested include<br>flexural strength, compressive<br>strength, water solubility, water<br>absorption | G-aenial Sculpt is a light cured<br>nano-filled radiopaque<br>composite resin filled in a<br>syringe and unitip and<br>available in several shades.<br>The device is used for the<br>restorations of both anterior<br>and posterior teeth.<br>Physical properties tested<br>In accordance with ISO<br>4049:2009 Dentistry - Polymer<br>based restorative materials |
| Mode of action | The paste material with glass<br>fibers contains a<br>photo initiator to allow for<br>polymerization through light<br>curing resulting in a solid<br>material for dental<br>restorations. | The two component paste material<br>uses dual cure to become a<br>solid material for dental<br>restorations | G-aenial Sculpt is a paste<br>composite reinforced by a<br>dispersion of filler particles<br>bonded to the resin matrix with a<br>photo initiator to allow for<br>curing/conversion. The end result<br>is a conversion from a paste to a<br>solid material for dental<br>restorations. |
### 7. Non clinical performance testing
Bench testing was performed in accordance with ISO 4049:2009 Dentistry - Polymer based restorative materials, and the testing results indicate that the device conforms to the required specifications for sensitivity to ambient light, depth of cure, flexural strength, water sorption, solubility, and color stability after irradiation and water sorption and is therefore suitable for the intended use.
## 8. Biocompatibility
Biocompatibility assessment was based upon ISO 10993-1 and testing included cytotoxicity in accordance with ISO 10993-5, and sensitization and irritation in accordance with ISO 10993-10.
### Substantial equivalence: 9.
everX Posterior is similar to the primary predicate (K102703) Build-It Total Core in that both are fiber
{5}------------------------------------------------
reinforced composite material used for dental restorations; both devices make use of pre-impregnated glass fibers to provide support to the composite mix for their intended use: both devices are indicated for use in large cavities with loss of dentin including posterior teeth, tooth core structure replacement or build up. The predicate device is not indicated for use as a root canal sealer and cavity liner because it is not self-adhesive meaning like the predicate K102703. A technologic difference between the applicant device and primary predicate device is that the applicant device is a light cured paste with a photo initiator, while the predicate device K102703 is a dual cure, two paste system of catalyst and base, and that contains fluoride. In principle, the curing mechanism of the applicant and predicate devices is substantially equivalent in principle as they both undergo polymerization to become a solid material adapted to a cavitation in a tooth to become a dental restoration.
everX Posterior and the reference predicate G-aenial Sculpt (K123631) are both light cured composite resins intended to be used as dental restorative materials: the device formulations are identical for all components, except that the applicant device also includes glass fibers and polymethylmethacrylate. everX Posterior is available only in universal shade whereas both of the predicates are available in a range of shades.
All three devices have been tested for physical properties including flexural strength, water sorption, depth of cure. The results of testing the physical properties of the applicant device in accordance with the FDA recognized standard for polymer based restorative materials used in dentistry, demonstrate that the applicant device is similar to the predicate devices in function and intended use.
# 10. Conclusion
Given the similarities in intended use, mode of action, and chemical composition, and in consideration of the results of non-clinical performance testing and biocompatibility testing, everX Posterior is substantially equivalent to the predicate and reference devices.
{6}------------------------------------------------
{7}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.