Thunderbeat Open Fine Jaw Type X Hand Instrument

K192103 · Olympus Medical Systems Corp. · GEI · May 11, 2020 · General, Plastic Surgery

Device Facts

Record IDK192103
Device NameThunderbeat Open Fine Jaw Type X Hand Instrument
ApplicantOlympus Medical Systems Corp.
Product CodeGEI · General, Plastic Surgery
Decision DateMay 11, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The THUNDERBEAT Open Fine Jaw Type X hand instrument is intended to be used for open surgery to cut, seal, coagulate, grasp, and dissect. Seal & Cut mode: The THUNDERBEAT Open Fine Jaw Type X hand instrument when used in combination with the Seal & Cut mode is indicated for open, general surgery (including plastic and reconstructive, etc.) or in any procedure in which cutting, vessel ligation (sealing and cutting), coagulation, grasping, and dissection is performed. The device has been designed to seal and cut vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics. This mode is also indicated for open ENT procedure in adults (thyroidectomy, parathyroidectomy, parotidectomy, radical neck dissection, and tonsillectomy) for only ligation (sealing and cutting) of vessels, lymphatics and tissue bundles greater than 3 mm away from unintended thermally sensitive structures such as nerves and parathyroid glands. Seal mode: The THUNDERBEAT Open Fine Jaw Type X hand instrument when used in combination with the Seal mode is indicated for open, general surgery (including plastic and reconstructive, etc.) or in any procedure in which vessel sealing, coagulation, grasping is performed. The device has been designed to seal vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics. This mode is also indicated for open ENT procedure in adults (thyroidectomy, parathyroidectomy, parotidectomy, radical neck dissection, and tonsillectomy) for sealing of vessels, lymphatics and tissue bundles greater than 3 mm away from unintended thermally sensitive structures such as nerves and parathyroid glands. The THUNDERBEAT Open Fine Jaw Type X hand instrument has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.

Device Story

Sterile, single-use electrosurgical/ultrasonic hand instrument; used in open surgery for cutting, sealing, coagulating, grasping, and dissecting. Operates via connection to ESG-400 Electrosurgical Generator and USG-400 Ultrasonic Generator. Employs combined HF bipolar and ultrasonic energy (Seal & Cut mode) or HF bipolar energy only (Seal mode). Physician-operated; used in OR. Provides mechanical grasping and energy delivery to tissue bundles/vessels up to 7mm. Output allows for vessel ligation and tissue division; aids in hemostasis. Benefits include efficient tissue management and reduced surgical time.

Clinical Evidence

No clinical data. Evidence consists of bench testing (ex-vivo porcine vessel burst pressure, temperature-time history) and chronic/acute animal studies (porcine/Beagle dogs) evaluating seal maintenance, thermal spread, and healing progression.

Technological Characteristics

Materials: Polycarbonate, ADC12, PEEK, PTFE, SUS630, SUS304, SUS303, Ti-6Al-4V. Sensing/Actuation: Ultrasonic transducer and HF bipolar electrodes. Energy: Ultrasonic and HF bipolar. Form factor: Hemostat-style hand instrument. Connectivity: Connects to ESG-400/USG-400 generators. Sterilization: Sterile, single-use. Standards: ANSI AAMI ES60601-1, IEC60601-1-2, IEC60601-2-2, ISO 10993, ISO 11135, ISO 11607, ASTM F1980.

Indications for Use

Indicated for open general surgery (including plastic/reconstructive) and open ENT procedures in adults (thyroidectomy, parathyroidectomy, parotidectomy, radical neck dissection, tonsillectomy) for cutting, sealing, coagulating, grasping, and dissecting vessels up to 7mm, tissue bundles, and lymphatics. Contraindicated for tubal sterilization/coagulation. ENT procedures require >3mm distance from thermally sensitive structures (nerves, parathyroid glands).

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 11, 2020 Olympus Medical Systems Corp. Christina Flores Manager, Regulatory Affairs Gyrus ACMI inc. 136 Turnpike Road Southborough, Massachusetts 01772 ### Re: K192103 Trade/Device Name: Thunderbeat Open Fine Jaw Type X Hand Instrument Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI, LFL Dated: May 6, 2020 Received: May 7, 2020 ### Dear Christina Flores: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Michael J. Ryan Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K192103 #### Device Name THUNDERBEAT Open Fine Jaw Type X hand instrument #### Indications for Use (Describe) The THUNDERBEAT Open Fine Jaw Type X hand instrument is intended to be used for open surgery to cut, seal, coagulate, grasp, and dissect. #### Seal & Cut mode: The THUNDERBEAT Open Fine Jaw Type X hand instrument when used in combination with the Seal & Cut mode is indicated for open, general surgery (including plastic and reconstructive, etc.) or in any procedure in which cutting, vessel ligation (sealing and cutting), coagulation, grasping, and dissection is performed. The device has been designed to seal and cut vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics. This mode is also indicated for open ENT procedure in adults (thyroidectomy, parotidectomy, partical neck dissection, and tonsillectory) for only ligation (sealing and cutting) of vessels, lymphatics and tissue bundles greater than 3 mm away from unintended thermally sensitive structures such as nerves and parathyroid glands. Seal mode: The THUNDERBEAT Open Fine Jaw Type X hand instrument when used in combination with the Seal mode is indicated for open, general surgery (including plastic and reconstructive, etc.) or in any procedure in which vessel sealing, coagulation, grasping is performed. The device has been designed to seal vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics. This mode is also indicated for open ENT procedure in adults (thyroidectomy, parotidectomy, radical neck dissection, and tonsillectomy) for sealing of vessels, lymphatics and tissue bundles greater than 3 mm away from unintended thermally sensitive structures such as nerves and parathyroid glands. The THUNDERBEAT Open Fine Jaw Type X hand instrument has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures. | Type of Use (Select one or both, as applicable) | | | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------| | <table><tr><td><div style="display:inline-block;"><b></b></div>Prescription Use (Part 21 CFR 801 Subpart D)</td></tr><tr><td><div style="display:inline-block;"></div>Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table> | <div style="display:inline-block;"><b></b></div> Prescription Use (Part 21 CFR 801 Subpart D) | <div style="display:inline-block;"></div> Over-The-Counter Use (21 CFR 801 Subpart C) | | <div style="display:inline-block;"><b></b></div> Prescription Use (Part 21 CFR 801 Subpart D) | | | | <div style="display:inline-block;"></div> Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff *PRAStaff@fda.hhs.gov* "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are slightly rounded and have a glossy appearance. There is a thin, horizontal, yellow line underneath the word. The registered trademark symbol is located to the right of the letter "S". ## 510(k) Summary Gyrus ACMI, Inc. THUNDERBEAT Open Fine Jaw Type X hand instrument ### General Information | Applicant: | OLYMPUS MEDICAL SYSTEMS<br>CORP.<br>2951 Ishikawa-cho, Hachioji-shi, Tokyo,<br>Japan 192-8507<br>Phone: (+81) 42-642-2694<br>Fax: (+81) 42-642-2307 | |------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------| | Establishment Registration Number: | 8010047 | | Manufacturer: | Aomori Olympus<br>2-248-1 okkonoki<br>kuroishi-shi aomori, Japan 036-0357<br>Phone: (+81) 172-52-8543<br>Fax: (+81) 172-52-8515 | | Establishment Registration Number: | 9614641 | | 510(k) Submitter: | Gyrus ACMI, Inc.<br>136 Turnpike Rd.<br>Southborough, MA 01772-2104 | | Establishment Registration Number: | 3003790304 | | Contact Person: | Christina Flores, RAC<br>Manager, Regulatory Affairs | | Date Prepared: | May 10, 2020 | ### Device Description Classification Name: Electrosurgical cutting and coagulation device and accessories Regulation Number: 21 CFR 878.4400 Regulatory Class: Class II, Product Codes: GEI, LFL Review Panel: General & Plastic Surgery {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. A registered trademark symbol is located to the right of the letter "S". Trade Name: THUNDERBEAT Open Fine Jaw Type | | | X hand instrument | |--|----------------------|--------------------------------------------------| | | Model Name | Device Name | | | TB-0009OFX | THUNDERBEAT Open Fine Jaw Type X hand instrument | | | Generic/Common Name: | Ultrasonic and electrosurgical devices | Ultrasonic and electrosurgical devices ### Primary Predicate Device | Device name | 510(k) Submitter | 510(k) No. | |------------------------------------------|----------------------------------|------------| | THUNDERBEAT Open Fine Jaw<br>(TB-0009OF) | OLYMPUS MEDICAL<br>SYSTEMS CORP. | K151743 | ### Secondary Predicate Device | Device name | 510(k) Submitter | 510(k) No. | |----------------------------------------------------|------------------|------------| | LigaSure Curved, Small Jaw, Open<br>Sealer/Divider | COVIDIEN | K113572 | # Product Description THUNDERBEAT Open Fine Jaw Type X hand instrument is intended to be used for open surgery to cut, seal, coagulate, grasp, and dissect. The THUNDERBEAT Open Fine Jaw Type X hand instrument is provided as a sterile, single use device. The subject device is functional device capable of vessel sealing & cutting, tissue coagulating & cutting, grasping, and dissecting. This device has been designed to seal and cut vessels up to and including 7 mm in diameter, tissue bundles, and lymphatics. This device also has been designed for open ENT procedures in adults. # Comparison of Technological Characteristics The basic principle and operational characteristics of the subject device are identical to the predicate device. The proposed THUNDERBEAT Open Fine Jaw Type X Hand Instrument is equivalent to the predicate device in that each device is connected to Surgical Tissue Management System, specifically the ESG-400 Electrosurgical Generator, USG-400 Ultrasonic Generator and THUNDERBEAT Transducer (TD-TB400). Like the predicate device, the subject device activates combined HF Bipolar (FineCoag) output and Ultrasonic output [Seal & Cut mode] simultaneously or only the HF Bipolar output (Seal mode) while grasping vessels, tissue bundles and lymphatics between the Probe and the H electrode in the Grasping section. {5}------------------------------------------------ In terms of the design concept of the subject device, the differences from the predicate device are as follows: - Addition of the Radical Neck Dissection indication - Removal of generator specific information from indications for use - Adding resin cover (PEEK) on the H electrode - Changing a material of H electrode in the grasping section and Sheath cover A side by side comparison of the subject device and the predicate device is provided below. | Item | Subject Device | Primary Predicate<br>Device (PD) K151743 | Secondary<br>Predicate Device<br>(PD2)<br>K113572 | Discussion | |--------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | TB-0009OFX | TB-0009OF | LigaSure Curved,<br>Small Jaw, Open<br>Sealer/Divider | | | Indicatio<br>ns for<br>Use | The THUNDERBEAT<br>Open Fine Jaw Type X<br>hand instrument is<br>intended to be used for<br>open surgery to cut,<br>seal, coagulate, grasp,<br>and dissect.<br><br>Seal & Cut mode:<br>The THUNDERBEAT<br>Open Fine Jaw Type X<br>hand instrument when<br>used in combination<br>with the Seal & Cut<br>mode is indicated for<br>open, general surgery<br>(including plastic and<br>reconstructive, etc.) or<br>in any procedure in<br>which cutting, vessel<br>ligation (sealing and<br>cutting), coagulation,<br>grasping, and<br>dissection is<br>performed. The device<br>has been designed to<br>seal and cut vessels (up<br>to and including 7 mm<br>in diameter), tissue<br>bundles, and<br>lymphatics.<br><br>This mode is also<br>indicated for open<br>ENT procedure in<br>adults (thyroidectomy | The THUNDERBEAT<br>Open Fine Jaw hand<br>instrument is intended<br>to be used with the<br>Ultrasonic Generator<br>(USG-400), the<br>Electrosurgical<br>Generator (ESG-400),<br>and the<br>THUNDERBEAT<br>Transducer, (TD-<br>TB400).<br><br>Seal & Cut mode:<br>The THUNDERBEAT<br>Open Fine Jaw hand<br>instrument when used<br>in combination with<br>the Seal & Cut mode is<br>indicated for open,<br>general surgery<br>(including plastic and<br>reconstructive, etc) or<br>in any procedure in<br>which cutting, vessel<br>ligation (sealing and<br>cutting), coagulation,<br>grasping, and<br>dissection is<br>performed. The device<br>has been designed to<br>seal and cut vessels (up<br>to and including 7 mm<br>in diameter), tissue<br>bundles, and<br>lymphatics. | The LigaSure™<br>Curved, Small Jaw,<br>Open Sealer /<br>Divider is a bipolar<br>electrosurgical<br>instrument<br>intended to be used<br>with the<br>ForceTriad™<br>energy platform.<br>The instrument is<br>indicated for use in<br>open general<br>surgical procedures<br>where ligation and<br>division of vessels<br>(up to 7 mm in<br>diameter), tissue<br>bundles, and<br>lymphatics is<br>performed, such as<br>urologic, thoracic,<br>plastic, and<br>reconstructive, and<br>including such<br>procedures as<br>bowel resections,<br>gall bladder<br>procedures, Nissen<br>fundoplication,<br>adhesiolysis, etc.<br><br>The device is also<br>indicated for open<br>ENT procedures in<br>adults | The<br>differences<br>between SD<br>and PD are the<br>removal of the<br>specific<br>generator<br>information<br>and the<br>addition of the<br>Radical Neck<br>Dissection<br>indication.<br>The Radical<br>Neck<br>Dissection<br>indication is<br>carried by the<br>PD2. | | Item | Subject Device | Primary Predicate<br>Device (PD) K151743 | Secondary<br>Predicate Device<br>(PD2)<br>K113572 | Discussion | | | TB-0009OFX<br>parathyroidectomy,<br>parotidectomy, radical<br>neck dissection, and<br>tonsillectomy) for only<br>ligation (sealing and<br>cutting) of vessels,<br>lymphatics and tissue<br>bundles greater than<br>3mm away from<br>unintended thermally<br>sensitive structures<br>such as nerves and<br>parathyroid glands.<br>Seal mode:<br>The THUNDERBEAT<br>Open Fine Jaw Type X<br>hand instrument when<br>used in combination<br>with the Seal mode is<br>indicated for open,<br>general surgery<br>(including plastic and<br>reconstructive, etc.) or<br>in any procedure in<br>which vessel sealing,<br>coagulation, grasping<br>is performed. The<br>device has been<br>designed to seal<br>vessels (up to and<br>including 7 mm in<br>diameter), tissue<br>bundles, and<br>lymphatics.<br>This mode is also<br>indicated for open<br>ENT procedure in<br>adults (thyroidectomy,<br>parathyroidectomy,<br>parotidectomy, radical<br>neck dissection, and<br>tonsillectomy) for<br>sealing of vessels,<br>lymphatics and tissue<br>bundles greater than<br>3mm away from | TB-0009OF<br>This mode is also<br>indicated for open<br>ENT procedure in<br>adults (thyroidectomy,<br>parathyroidectomy,<br>parotidectomy, and<br>tonsillectomy) for only<br>ligation (sealing and<br>cutting) of vessels,<br>lymphatics and tissue<br>bundles 2-3 mm*1<br>away from unintended<br>thermally sensitive<br>structures such as<br>nerves and parathyroid<br>glands.<br>Seal mode:<br>The THUNDERBEAT<br>Open Fine Jaw hand<br>instrument when used<br>in combination with<br>the Seal mode is<br>indicated for open,<br>general surgery<br>(including plastic and<br>reconstructive, etc.) or<br>in any procedure in<br>which vessel sealing,<br>coagulation, grasping<br>is performed. The<br>device has been<br>designed to seal<br>vessels (up to and<br>including 7 mm in<br>diameter), tissue<br>bundles, and<br>lymphatics.<br>This mode is also<br>indicated for open<br>ENT procedure in<br>adults (thyroidectomy,<br>parathyroidectomy,<br>parotidectomy, and<br>tonsillectomy) for<br>sealing of vessels, | LigaSure Curved,<br>Small Jaw, Open<br>Sealer/Divider<br>(thyroidectomy,<br>radical neck<br>dissection,<br>parotidectomy, and<br>tonsillectomy) for<br>ligation and<br>division of vessels,<br>lymphatics and<br>tissue bundles 2-3<br>mm away from<br>unintended<br>thermally sensitive<br>structures such as<br>nerves and<br>parathyroid glands.<br>The LigaSure<br>system has not<br>been shown to be<br>effective for tubal<br>sterilization or<br>tubal<br>coagulation for<br>sterilization<br>procedures. Do not<br>use the LigaSure<br>system for these<br>procedures. | | | Item | Subject Device | Primary Predicate Device (PD) K151743 | Secondary Predicate Device (PD2) K113572 | Discussion | | | TB-00090FX<br>unintended thermally<br>sensitive structures<br>such as nerves and<br>parathyroid glands.<br>The THUNDERBEAT<br>Open Fine Jaw Type X<br>hand instrument has<br>not been shown to be<br>effective for tubal<br>sterilization or tubal<br>coagulation for<br>sterilization<br>procedures, and should<br>not be used for these<br>procedures. | TB-0009OF<br>lymphatics and tissue<br>bundles 2-3mm*1<br>away from unintended<br>thermally sensitive<br>structures such as<br>nerves and parathyroid<br>glands.<br>The THUNDERBEAT<br>Open Fine Jaw hand<br>instrument has not<br>been shown to be<br>effective for tubal<br>sterilization or tubal<br>coagulation for<br>sterilization<br>procedures, and should<br>not be used for these<br>procedures.<br>*1 It should be<br>extended appropriately<br>depending on the<br>operation situation. | LigaSure Curved,<br>Small Jaw, Open<br>Sealer/Divider | | | Regulati<br>on<br>Number | 878.4400 | 878.4400 | 878.4400 | Same as PD | | Regulati<br>on Name | Electrosurgical, cutting<br>& coagulation &<br>accessories | Electrosurgical, cutting<br>& coagulation &<br>accessories | Electrosurgical,<br>cutting &<br>coagulation &<br>accessories | | | Regulato<br>ry Class | II | II | II | | | Product<br>Code | GEI, LFL | GEI, LFL | GEI | | | Classific<br>ation<br>Panel | General & Plastic<br>Surgery | General & Plastic<br>Surgery | General & Plastic<br>Surgery | | | Basic<br>principle | Seal and Cut mode:<br>Activating the<br>Ultrasonic output<br>(generated by the<br>USG-400, Ultrasonic<br>Generator) and the HF<br>Bipolar output<br>(generated by the<br>ESG-400) | Seal and Cut mode:<br>Activating the<br>Ultrasonic output<br>(generated by the<br>USG-400, Ultrasonic<br>Generator) and the HF<br>Bipolar output<br>(generated by the<br>ESG-400) | High-frequency<br>current generated<br>by Covidien's<br>Electrosurgical<br>Generators (such<br>as Force Triad) is<br>delivered to the<br>jaw (electrodes),<br>and coagulate<br>(seal) vessels, | Same as PD | | Item | Subject Device<br>TB-0009OFX | Primary Predicate<br>Device (PD) K151743<br>TB-0009OF | Secondary<br>Predicate Device<br>(PD2)<br>K113572<br>LigaSure Curved, Small Jaw, Open<br>Sealer/Divider | Discussion | | | sealing and cutting of<br>vessels, tissue bundles,<br>and lymphatics and<br>cutting and coagulation<br>soft tissue.<br>Seal mode: Uses only<br>the HF Bipolar (ESG-<br>400) energy output<br>which enables vessel,<br>tissue bundle and<br>lymphatic sealing and | sealing and cutting of<br>vessels, tissue bundles,<br>and lymphatics and<br>cutting and coagulation<br>soft tissue.<br>Seal mode: Uses only<br>the HF Bipolar (ESG-<br>400) energy output<br>which enables vessel,<br>tissue bundle and<br>lymphatic sealing and | tissue bundles,<br>lymphatics<br>clamped by the<br>jaw. The internal<br>mechanical blade<br>in the jaw makes<br>the soft tissues<br>divided | | | Handle<br>shape | hemostasis.<br>hemostat-style body | hemostasis.<br>hemostat-style body | hemostat-style<br>body | Same as PD | | Tip<br>shape | Probe and H electrode | Probe and H electrode | Jaw | Same as PD | | Resin<br>cover on<br>the tip of<br>grasping<br>section | Equipped | N/A | Unknown | The resin<br>cover is added<br>on the H<br>electrode. | | Grasping<br>section | PTFE/SUS630 | PTFE/Aluminum alloy | SUS | The material<br>of the H<br>electrode was<br>changed from<br>aluminum<br>alloy to<br>SUS63. | | Device<br>performa<br>nce | Performance testing including Bench and Animal testing was conducted in support of the<br>substantial equivalence determination. | | | | | Reproce<br>ssing | Sterile, for single use…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...