K113572 · Covidien, Formerly Valleylab, A Division of Tyco H · GEI · Sep 5, 2012 · General, Plastic Surgery
Device Facts
Record ID
K113572
Device Name
LIGASURE CURVED, SMALL JAW, OPEN SEALER/DIVIDER
Applicant
Covidien, Formerly Valleylab, A Division of Tyco H
Product Code
GEI · General, Plastic Surgery
Decision Date
Sep 5, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K113572 · Sep 5, 2012
LIGASURE CURVED, SMALL JAW, OPEN SEALER/DIVIDER
Covidien, Formerly Valleylab, A Division of Tyco H
Published clinical literature
The sponsor utilized a review of published clinical literature to demonstrate that the use of LigaSure instruments in ENT procedures is safe and effective, supporting the expansion of the indications for use.
Clinical literature; ENT procedures; Safety and effectiveness
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Assessment of published clinical literature; Literature review
Adult patients undergoing ENT procedures (thyroidectomy, radical neck dissection, parotidectomy, and tonsillectomy)
Not applicable for this study
Safety and effectiveness of LigaSure instruments in ENT procedures
Indications for Use
The LigaSure™ Curved, Small Jaw, Open Sealer / Divider is a bipolar electrosurgical instrument intended to be used with the ForceTriad™ energy platform. The instrument is indicated for use in open general surgical procedures where ligation and division of vessels (up to 7 mm in diameter), tissue bundles, and iymphatics is performed, such as urologic, thoracic, plastic, and reconstructive; and including such procedures as bowel resections, gall bladder procedures, Nissen fundoplication, adhesiolysis, etc. The device is also indicated for open ENT procedures in adults (thyroidectorny, radical neck dissection, parotidectomy, and tonsillectomy) for ligation of vessels, lymphatics and tissue bundles 2-3 mm away from unintended thermally sensitive structures such as nerves and parathyroid glands. The LigaSure system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.
Device Story
Sterile, single-use, hand-held bipolar electrosurgical instrument; used with ForceTriad energy platform. Inputs: user-applied pressure to grasp tissue via hemostat-style ring handles; RF energy activation via handle button or footswitch. Operation: generator delivers precise energy for tissue fusion; mechanical stainless steel blade cuts tissue. Used in open surgery by surgeons. Output: fused and divided tissue. Benefits: minimal sticking, charring, or thermal spread; improved visibility in confined spaces via small, curved jaws. Smart connector enables generator recognition.
Clinical Evidence
No new clinical trials conducted. Evidence includes preclinical comparative bench testing (burst pressure, thermal profile) and animal testing (thermal spread). Usability evaluated via formative and summative human factors studies in simulated operating environments. Clinical literature assessment supports safety and effectiveness in ENT procedures.
Technological Characteristics
Bipolar electrosurgical instrument; RF energy source; stainless steel cutting blade; hemostat-style handle with ring grips; smart connector for generator recognition. Sterile, single-use. Dimensions/materials unchanged from K102470.
Indications for Use
Indicated for open general surgical procedures (urologic, thoracic, plastic, reconstructive, bowel resection, gall bladder, Nissen fundoplication, adhesiolysis) and open ENT procedures in adults (thyroidectomy, radical neck dissection, parotidectomy, tonsillectomy) for ligation/division of vessels up to 7mm, tissue bundles, and lymphatics. Contraindicated for tubal sterilization or tubal coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
LigaSure™ Curved, Small Jaw, Open Sealer/Divider (K102470)
LigaSure™ Precise Instrument Vessel Sealing System (K010010)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary
Date summary prepared: 8/29/2012
### 510(k) Submitter/Holder
Covidien IIc 5920 Lonabow Drive Boulder, CO 80301
#### Contact
John Van Hoven Sr. Associate, Regulatory Affairs Telephone: 303-476-7502 Fax: 303-516-8307 Email: john vanhoven@covidien.com
#### Name of Device
LigaSure™ Curved, Small Jaw, Open Sealer / Divider Trade Name: Catalog Number: LF1212 Common Name: Bipolar electrosurgical instrument Classification Name: Electrosurgical cutting and coagulation device and accessories (21 CFR § 878.4400, class II, GEI).
#### Purpose of Submission
The purpose of this submission is to modify the indications for use statement to include the specific indication to use the device in ENT procedures and to remove the warning statement surrounding specific surgical procedures around which precision is required. There were no physical changes made to the device associated with the proposed expanded indication; the device design is unchanged from that cleared under K102470.
This submission followed the "Guidance for Industry and FDA Staff. Format for Traditional and Abbreviated 510(k)s, 2005," the "Guidance for Industry on General / Specific Intended Use, 1998," and the "Draft Guidance for Industry and FDA Staff: Factors to Consider when Making Benefit-Risk Determinations in Medical Device Premarket Review, 2011."
#### Predicate Devices
The LigaSure™ Curved, Small Jaw, Open Sealer/Divider was compared and found to be substantially equivalent to the following products of comparable type in commercial distribution:
| Trade Name: | LigaSure™™ Curved, Small Jaw, Open Sealer/Divider |
|---------------------|-------------------------------------------------------------|
| Device Common Name: | Bipolar electrosurgical instrument |
| Catalog Number: | LF1212 |
| 510(k) Number: | K102470 (cleared 2/7/2011) |
| Manufacturer: | Covidien, formerly Valleylab, a division of Tyco Healthcare |
| Trade Name: | LigaSure™™ Precise Instrument Vessel Sealing System |
| Device Common Name: | Bipolar electrosurgical instrument |
| Catalog Number: | LS1200 |
| 510(k) Number: | K010010 (cleared 4/2/2001) |
| Manufacturer: | Covidien, formerly Valleylab, a division of Tyco Healthcare |
#### Device Description
The LigaSure™ Curved, Small Jaw, Open Sealer/Divider is a sterile, single-use, hand-held bipolar electrosurgical instrument designed exclusively for use with the ForceTriad™ energy platform (generator) to ligate (seal) and divide (cut) vessels, tissue bundles, and lymphatics during open general surgical
510(k): LigaSure ™ Curved, Small Jaw, Open Sealer/Divider - LF1212
SFP 5 2012
{1}------------------------------------------------
procedures (as indicated). The ForceTriad's tissue-fusion (LigaSure) mode delivers precise energy to tissue for a controlled time period to achieve complete and permanent tissue fusion, and has been designed to produce minimal sticking, charring, or thermal spread to adjacent tissue.
LigaSure instruments attach to the generator with a ten-foot containing a "smart" connector. This connector functions as a unique product identifier for device-specific recognition by the generator. The Curved, Small Jaw, Open Sealer/Divider instrument is designed to be both ergonomic and intuitive for the user. Its hemostat-style body and symmetrically placed controls facilitate handling by both left and righthanded users, and its small, curved jaws maximize visibility and access when the instrument is used in confined surgical spaces.
Ring handles function to allow the user to grasp tissue by opening and closing the jaws of the instrument. The interior surfaces of the jaws contain the electrodes, which serve to ligate by delivering energy to the grasped tissue. RF energy can be activated by the user in two ways: (1) through the use of a single button incorporated into the handle body or (2) through the use of a footswitch attached to the generator. A cutting mechanism functions to mechanically divide tissue following tissue fusion. It consists of a stainless steel blade and is controlled by the user through a trigger located on the handle body.
#### Intended Use
The LigaSure™ Curved, Small Jaw, Open Sealer/Divider is a bipolar electrosurgical instrument intended to be used with the ForceTriad™ energy platform. The instrument is indicated for use in open general surgical procedures where ligation and division of vessels (up to 7 mm in diameter), tissue bundles, and iymphatics is performed, such as urologic, thoracic, plastic, and reconstructive; and including such procedures as bowel resections, gall bladder procedures, Nissen fundoplication, adhesiolysis, etc.
The device is also indicated for open ENT procedures in adults (thyroidectorny, radical neck dissection, parotidectomy, and tonsillectomy) for ligation of vessels, lymphatics and tissue bundles 2-3 mm away from unintended thermally sensitive structures such as nerves and parathyroid glands.
The LigaSure system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.
#### Technological and Performance Characteristics
The LigaSure Curved, Small Jaw, Open Sealer / Divider is the same device as the predicate K102470. The function of the device has not changed. The LigaSure Curved, Small Jaw, Open Sealer/Divider is similar in design and utilizes radio frequency (RF) technology to achieve its intended use similarly to the predicate devices.
#### Performance Data
The design of the LigaSure Curved, Small Jaw, Open Sealer/Divider is unchanged and the information provided in K102470 is applicable to the device subject of this submission.
Preclinical Bench Testing: Preclinical comparative bench testing for burst pressure and thermal profile was conducted with the LigaSure Curved, Small Jaw, Open Sealer/Divider and its predicates to demonstrate equivalent performance. The subject device was shown to yield higher mean burst pressures on isolated vessels than the predicates; in addition, thermal profile studies showed the subject device has consistently lower mean temperatures than the predicates after single and multiple activations.
Animal Testing: The mean thermal spread of the subject device was shown to be comparable to the predicates for all vessel sizes evaluated.
#### Usability / Human Factors
Usability was evaluated with users in simulated operating environments, including in the ENT space. These studies consisted of several formative and one summative study, which demonstrate the instrument provides adequate assurance.of safety and performance (in regards to human factors/usability aspects) for the patient and operator.
#### Clinical Literature
An assessment of published clinical literature demonstrates that there is a considerable body of knowledge and experience that shows the use of LigaSure instruments in ENT procedures is safe and effective.
{2}------------------------------------------------
K1/3572
pg 3 of 3
## Substantial Equivalence
In establishing substantial equivalence of the LigaSure Curved, Small Jaw, Open Sealer / Divider to the predicate devices, Covidien EbD evaluated the intended use, indications for use, technological characteristics, reported adverse events, published clinical literature, and instrument risk profiles. The use of the LigaSure Curved, Small Jaw, Open Sealer / Divider in ENT procedures does not raise any new types of questions of safety and effectiveness compared with the predicate devices currently used for these procedures.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter of the circle. Inside the circle is a stylized image of an eagle, with its wings spread. The eagle is facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Covidien % Mr. John Van Hoven Surgical Solutions 5920 Longbow Drive Boulder, CO 80301
# SEP 0 5 2012
Re: K113572
Trade/Device Name: LigaSure™ Curved, Small Jaw, Open Sealer / Divider (LF1212) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: August 29, 2012 Received: August 30, 2012
Dear Mr. Van Hoven:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
# Page 2 - Mr. John Van Hoven
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Erine Keith
- Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use Statement
510(k) Number (if known): K113572
Device Name: LigaSure™ Curved, Small Jaw, Open Sealer / Divider (LF1212)
Indications for Use:
The LigaSure™ Curved, Small Jaw, Open Sealer / Divider is a bipolar electrosurgical instrument intended to be used with the ForceTriad™ energy platform. The instrument is indicated for use in open general surgical procedures where ligation and division of vessels (up to 7 mm in diameter), tissue bundles, and lymphatics is performed, such as urologic, thoracic, plastic, and reconstructive; and including such procedures as bowel resections, gall bladder procedures, Nissen fundoplication, adhesiolysis, etc.
The device is also indicated for open ENT procedures in adults (thyroidectorny, radical neck dissection, parotidectorny, and tonsillectorny) for ligation and division of vessels, lymphatics and tissue bundles 2-3 mm away from unintended thermally sensitive structures such as nerves and parathyroid glands.
The LigaSure system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.
Prescription Use (Part 21 CFR 801 Subpart D) and/or
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
E.D. Keith
(Division Sign-Off) (Division Sign-On)
Division of Surgical, Orthopedic, Division of Surg.
and Restorative Devices
510(k) Number K113572
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.