THINQ

K192051 · Corticometrics, LLC · LLZ · Sep 30, 2020 · Radiology

Device Facts

Record IDK192051
Device NameTHINQ
ApplicantCorticometrics, LLC
Product CodeLLZ · Radiology
Decision DateSep 30, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Volumetric quantification of brain structuresProbabilistic neuroanatomical atlas-based segmentationWhole Brain Dice 0.94; Total Gray Matter Dice 0.82; Total White Matter Dice 0.87; Left Hippocampus Dice 0.78; Right Hippocampus Dice 0.79645 unique MR images

Indications for Use

THINQ™ is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures from a set of MR images. It is intended to automate the manual process of identifying, labeling and quantifying the volume of segmentable brain structures identified on MR images.

Device Story

THINQ is a software-only medical device for quantitative brain imaging; accepts 3D T1-weighted MRI scans as input. Processing pipeline performs atlas-based segmentation of brain structures; includes automated QA checks for scan sequence, atlas alignment, and cortical surface integrity; uses classifier to filter corrupted inputs. Produces PDF neuromorphometry report containing volume measurements, visualizations, and comparisons to age/gender-matched reference data; outputs segmentation overlays in DICOM-JPEG format. Deployed as a container in high-performance computing environments within clinical workflows. Physician reviews output to confirm accuracy; assists in clinical decision-making by providing objective morphometric data. Benefits include automated, reproducible quantification of brain structures compared to manual methods.

Clinical Evidence

Bench testing only. Validation used 645 unique MR images. Accuracy evaluated against expert-labeled ground truth using Dice similarity coefficient (range 0.64-0.94) and Absolute Volume Error. Intracranial Volume accuracy measured by Absolute Percent Difference (3.42%). Reproducibility assessed via test-retest on repeated scans using Absolute Percent Difference (range 0.30%-2.00%).

Technological Characteristics

Software-only; operates on off-the-shelf hardware. Processing uses atlas-based segmentation and probabilistic neuroanatomical models. Supports DICOM input/output. Containerized deployment. Developed per ISO 13485, ISO 14971, ISO 62304, and ISO 62366-1 standards.

Indications for Use

Indicated for automatic labeling, visualization, and volumetric quantification of segmentable brain structures from MR images in patients, with volumetric measurements compared to reference percentile data.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ September 30, 2020 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size. CorticoMetrics LLC % Mr. Nick Schmansky Co-Founder, CEO 128 Granite Street ROCKPORT MA 01966 Re: K192051 Trade/Device Name: THINO Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: August 26, 2020 Received: August 31, 2020 Dear Mr. Schmansky: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see {1}------------------------------------------------ https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K192051 Device Name THINQ Indications for Use (Describe) THINQ is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures from a set of MR images. Volumetric measurements may be compared to reference percentile data. Type of Use (Select one or both, as applicable) | <div> <span style="text-decoration: line-through;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | |--------------------------------------------------------------------------------------------------------------------| | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image contains the logo for CorticoMetrics. The logo consists of a stylized brain graphic on the left, followed by the company name "CorticoMetrics" in text. The word "Cortico" is in a teal color, while "Metrics" is in a dark brown color. # THINQ™ 510(k) Summary CorticoMetrics, LLC August 22, 2020 {4}------------------------------------------------ This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92: #### 1 Submitter | Name | CorticoMetrics, LLC | |------------------|----------------------------------------| | Address | 128 Granite St., Rockport MA 01966 USA | | Contact Person | Nick Schmansky | | Telephone Number | 617-329-5042 | | Fax Number | none | | Email | nicks@corticometrics.com | | Date Prepared | August 22, 2020 | #### 2 Device | Device Trade Name | THINQ™ | |-----------------------------|---------------------------------------------| | Common Name | Medical Imaging Processing Software | | Classification Name | System, Image Processing, Radiological | | Regulation Number | 21 CFR §892.2050 | | Regulation Description | Picture archiving and communications system | | Regulatory Class | Class II | | Product Classification Code | LLZ | | Classification Panel | Radiology | #### Predicate Device 3 | Device | NeuroQuant | |-----------------------------|---------------------------------------------| | 510(k) Number | K170981 | | Manufacturer | CorTechs Labs, Inc. | | Common Name | Medical Imaging Processing Software | | Classification Name | System, Image Processing, Radiological | | Regulation Number | 21 CFR §892.2050 | | Regulation Description | Picture archiving and communications system | | Regulatory Class | Class II | | Product Classification Code | LLZ | | Classification Panel | Radiology | Copyright ©2019-2020 CorticoMetrics LLC. All Rights Reserved. {5}------------------------------------------------ # Device Description 4 THINQ™ is a software-only, non-interactive, medical device for quantitative imaging, accepting as input 3D T1-weighted MRI scan data of the human head. THINQ™ produces as output a quantitative neuromorphometry report in PDF format. The report contains morphometric (volume) measurements and visualizations of various structures in the brain, and compares these measures to age and gender-matched reference percentile data. The report includes images of the brain with color-coded segmentations, as well as plots showing how measurements compare to reference data. Additionally, in order to visually confirm the accuracy of the results, three segmentation overlays are created in DICOM-JPEG format; one in each anatomical plane: sagittal, coronal and axial. The THINQ™ processing pipeline performs an atlas-based segmentation of brain structures followed by measurement of those structures and a comparison to a reference dataset. The pipeline includes automated QA checks on the input DICOM 3D T1 MRI series to ensure adherence to imaging sequence requirements, checks on the data elements generated during the processing pipeline, and usage of a classifier to filter potentially incorrect reports due to corrupted image input. THINQ™ is packaged as a container, for deployment and operation in a high-performance computing environment within a clinical workflow. #### Intended Use 5 THINQ™ is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures from a set of MR images. It is intended to automate the manual process of identifying, labeling and quantifying the volume of segmentable brain structures identified on MR images. # Comparison of Technological Characteristics with the б Predicate Device | | Subject Device | Predicate Device | |---------------------------|------------------------------------------------|------------------------------------------------| | Device | THINQ™ | NeuroQuant v2.2 | | 510(k) Number | K192051 | K170981 | | Regulation<br>Number | 21 CFR §892.2050 | 21 CFR §892.2050 | | Regulation<br>Description | Picture archiving and<br>communications system | Picture archiving and<br>communications system | {6}------------------------------------------------ | | Subject Device | Predicate Device | |------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device: | THINQ™ | NeuroQuant v2.2 | | Classification<br>Name | System, Image Processing,<br>Radiological | System, Image Processing,<br>Radiological | | Classification: | Class II | Class II | | Product Code: | LLZ | LLZ | | Indications<br>for Use | THINQ™ is intended for automatic<br>labeling, visualization and<br>volumetric quantification of<br>segmentable brain structures from<br>a set of MR images. Volumetric<br>measurements may be compared<br>to reference percentile data. | NeuroQuant is intended for<br>automatic labeling, visualization<br>and volumetric quantification of<br>segmentable brain structures and<br>lesions from a set of MR images.<br>Volumetric measurements may be<br>compared to reference percentile<br>data. | | Design and<br>Incorporated<br>Technology | • Automated measurement of brain<br>tissue volumes and structures<br>• Automatic segmentation and<br>quantification of brain structures<br>using a probabilistic<br>neuroanatomical atlas based on<br>the MR image intensity | • Automated measurement of brain<br>tissue volumes and structures and<br>lesions<br>• Automatic segmentation and<br>quantification of brain structures<br>using a dynamic probabilistic<br>neuroanatomical atlas, with age<br>and gender specificity, based on<br>the MR image intensity | | Physical<br>Characteristics | • Software package<br>• Operates on off-the-shelf<br>hardware (multiple vendors) | • Software package<br>• Operates on off-the-shelf<br>hardware (multiple vendors) | | Operating<br>System | Supports Linux | Supports Linux, Mac OS X and<br>Windows | | Deployment: | Container installation | Cloud based or installed | | Processing<br>Architecture | Automated internal pipeline that<br>performs:<br>• Artifact correction<br>• Segmentation<br>• Volume calculation<br>• Report generation | Automated internal pipeline that<br>performs:<br>• Artifact correction<br>• Segmentation<br>• Lesion quantification<br>• Volume calculation<br>• Report generation | | Data Source | • MRI scanner: 3D T1 MRI scans<br>acquired with specified protocols<br>• THINQ™ supports DICOM format as<br>input | • MRI scanner: 3D T1 MRI scans<br>acquired with specified protocols<br>• NeuroQuant supports DICOM<br>format as input | | | Subject Device | Predicate Device | | Device: | THINQ™ | NeuroQuant v2.2 | | Output: | • Provides volumetric<br>measurements of brain structures<br>• Includes segmented color<br>overlays and morphometric reports<br>• Automatically compares results to<br>reference percentile data<br>• Report output is PDF file format<br>• Outputs segmentation results as<br>overlay in DICOM Encapsulated<br>JPEG format allowing display on<br>DICOM workstations and Picture<br>Archive and Communications<br>System | • Provides volumetric<br>measurements of brain structures<br>and lesions<br>• Includes segmented color<br>overlays and morphometric reports<br>• Automatically compares results to<br>reference percentile data and to<br>prior scans when available<br>• Supports DICOM format as output<br>of results that can be displayed on<br>DICOM workstations and Picture<br>Archive and Communications<br>Systems | | Safety: | • Automated quality control<br>functions:<br>• Scan sequence (protocol) checks<br>• Atlas alignment checks<br>• Cortical surface checks<br>• Result validity checks<br>• Results must be reviewed by a<br>trained physician | • Automated quality control<br>functions:<br>• Tissue contrast check<br>• Scan protocol verification<br>• Atlas alignment check<br>• Results must be reviewed by a<br>trained physician | {7}------------------------------------------------ # Performance Data 7 THINQ™ was designed, developed and tested in accordance with the following process standards: - 21 CFR part 820 Quality System Regulation / Medical Device Good Manufacturing Practices - ISO 13485:2016 Medical devices Quality management systems - ISO 14971:2007 Medical devices Application of risk management to medical devices - ISO 62304:2006 Medical device software Software life-cycle processes - ISO 62366-1:2015 Medical devices Part 1: Application of usability engineering to medical devices Validation of THINQ included performance testing for accuracy, where comparisons were made to expert-labeled brain images, and reproducibility, where test-retest image data was used. The accuracy of the reference population model was validated against statistical tests of normality using both subject test data expected to align within the reference ranges, as well as subjects with neurological disorders where affected brain structures are known to lie outside {8}------------------------------------------------ the reference ranges. A literature review of neuroimaging publications informed the selection of acceptance criteria. THINQ's segmentation accuracy of brain structures compared to ground-truth segmentations of 3D T1 MRI scans was evaluated using the Dice similarity coefficient metric. Accuracy of Intracranial Volume (ICV) was evaluated using the Absolute Percent Difference (APD) metric. Table 1 lists the results, including the Absolute Volume Error (AVE) and Relative Volume Error (RVE) metrics. | Structure | Accuracy Metric | Mean (StDev) | | |-----------------------------|-----------------|-----------------|---------------| | Whole Brain | Dice | 0.94 (0.01) | | | | AVE (cm³) | 327.00 (111.48) | | | | RVE | 0.30 (0.13) | | | Total Gray Matter | Dice | 0.82 (0.02) | | | | AVE (cm³) | 174.63 (46.42) | | | | RVE | 0.24 (0.05) | | | Total White Matter | Dice | 0.87 (0.04) | | | | AVE (cm³) | 65.67 (37.38) | | | | RVE | 0.18 (0.12) | | | Left Cortical Gray Matter | Dice | 0.92 (0.06) | | | | AVE (cm³) | 10.59 (6.51) | | | | RVE | 0.05 (0.03) | | | Right Cortical Gray Matter | Dice | 0.92 (0.07) | | | | AVE (cm³) | 10.43 (6.67) | | | | RVE | 0.05 (0.03) | | | Left Frontal Lobe | Dice | 0.90 (0.06) | | | | AVE (cm³) | 5.44 (3.83) | | | | RVE | 0.07 (0.05) | | | Right Frontal Lobe | Dice | 0.90 (0.06) | | | | AVE (cm³) | 5.07 (3.86) | | | | RVE | 0.06 (0.05) | | | Left Parietal Lobe | Dice | 0.88 (0.08) | | | | AVE (cm³) | 4.06 (3.04) | | | | RVE | 0.08 (0.06) | | | Right Parietal Lobe | Dice | 0.88 (0.08) | | | | AVE (cm³) | 3.85 (2.86) | | | | RVE | 0.07 (0.06) | | | Left Occipital Lobe | Dice | 0.82 (0.07) | | | | AVE (cm³) | 1.57 (1.22) | | | | RVE | 0.07 (0.05) | | | Right Occipital Lobe | Dice | 0.82 (0.08) | | | | AVE (cm³) | 1.92 (1.97) | | | | RVE | 0.08 (0.09) | | | Left Temporal Lobe | Dice | 0.89 (0.06) | | | | AVE (cm³) | 2.08 (1.89) | | | | RVE | 0.04 (0.04) | | | Right Temporal Lobe | Dice | 0.89 (0.06) | | | | | AVE (cm³) | 2.11 (1.81) | | | | RVE | 0.04 (0.04) | | Left Cerebral White Matter | Dice | | 0.86 (0.04) | | | | AVE (cm³) | 32.60 (18.80) | | | | RVE | 0.18 (0.12) | | Right Cerebral White Matter | Dice | | 0.86 (0.04) | | | | AVE (cm³) | 33.07 (18.88) | | | | RVE | 0.18 (0.12) | | Left Lateral Ventricle | Dice | | 0.86 (0.07) | | | | AVE (cm³) | 2.32 (1.69) | | | | RVE | 0.17 (0.15) | | Right Lateral Ventricle | Dice | | 0.85 (0.07) | | | | AVE (cm³) | 2.19 (1.59) | | | | RVE | 0.18 (0.14) | | Left Hippocampus | Dice | | 0.78 (0.03) | | | | AVE (cm³) | 0.45 (0.29) | | | | RVE | 0.14 (0.09) | | Right Hippocampus | Dice | | 0.79 (0.03) | | | | AVE (cm³) | 0.39 (0.28) | | | | RVE | 0.12 (0.10) | | Left Amygdala | Dice | | 0.66 (0.05) | | | | AVE (cm³) | 0.60 (0.17) | | | | RVE | 0.68 (0.24) | | Right Amygdala | Dice | | 0.64 (0.06) | | | | AVE (cm³) | 0.69 (0.19) | | | | RVE | 0.74 (0.27) | | Left Caudate | Dice | | 0.78 (0.07) | | | | AVE (cm³) | 0.50 (0.35) | | | | RVE | 0.17 (0.14) | | Right Caudate | Dice | | 0.78 (0.07) | | | | AVE (cm³) | 0.53 (0.34) | | | | RVE | 0.18 (0.13) | | Left Putamen | Dice | | 0.82 (0.04) | | | | AVE (cm³) | 0.83 (0.35) | | | | RVE | 0.20 (0.10) | | Right Putamen | Dice | | 0.82 (0.03) | | | | AVE (cm³) | 0.89 (0.35) | | | | RVE | 0.21 (0.08) | | Left Thalamus | Dice | | 0.82 (0.03) | | | | AVE (cm³) | 1.51 (0.53) | | | | RVE | 0.19 (0.05) | | Right Thalamus | Dice | | 0.83 (0.03) | | | | AVE (cm³) | 1.38 (0.45) | | | | RVE | 0.18 (0.04) | | Left Cerebellum | Dice | | 0.91 (0.02) | | | | AVE (cm³) | 2.10 (1.43) | | Right Cerebellum | Dice | 0.92 (0.02) | | | | AVE (cm³) | 2.07 (1.49) | | | | RVE | 0.03 (0.02) | | | Intracranial Volume ICV | APD | 3.42 (2.05) | | | | AVE (cm³) | 50.18 (31.92) | | | | RVE | 0.03 (0.02) | | {9}------------------------------------------------ {10}------------------------------------------------ Table 1: Accuracy testing results Reproducibility of brain structure segmentation of repeated 3D T1 MRI scans for same-subjects was evaluated by using the Absolute Percent Difference (APD) metric, shown in Table 2. | Structure | Reproducibility APD<br>Mean (StDev) | |-----------------------------|-------------------------------------| | Whole Brain | 0.34 (0.29) | | Total Gray Matter | 0.83 (0.80) | | Total White Matter | 1.04 (1.12) | | Left Cortical Gray Matter | 1.08 (0.89) | | Right Cortical Gray Matter | 1.04 (0.88) | | Left Frontal Lobe | 1.31 (1.30) | | Right Frontal Lobe | 1.57 (2.41) | | Left Parietal Lobe | 1.50 (1.31) | | Right Parietal Lobe | 1.67 (2.73) | | Left Occipital Lobe | 1.49 (1.16) | | Right Occipital Lobe | 2.00 (1.41) | | Left Temporal Lobe | 1.24 (1.37) | | Right Temporal Lobe | 1.39 (1.15) | | Left Cerebral White Matter | 1.15 (1.09) | | Right Cerebral White Matter | 1.10 (1.20) | | Left Lateral Ventricle | 1.44 (1.21) | | Right Lateral Ventricle | 1.55 (1.12) | | Left Hippocampus | 1.56 (1.76) | | Right Hippocampus | 1.49 (1.57) | | Left Amygdala | 1.25 (1.23) | | Right Amygdala | 1.72 (1.36) | | Left Caudate | 1.14 (1.22) | | Right Caudate | 1.24 (1.10) | | Left Putamen | 1.41 (1.15) | | Right Putamen | 1.29 (0.90) | | Left Thalamus | 0.86 (0.59) | | Right Thalamus | 0.75 (0.63) | | Left Cerebellum | 0.62 (0.58) | | Right Cerebellum | 0.60 (0.54) | | Intracranial Volume ICV | 0.30 (0.29) | Table 2: Reproducibility testing results {11}------------------------------------------------ THINQ performs a quantitative imaging function. The validation of THINQ addressed typical sources of error in quantitative imaging values by including in its image testing dataset a wide range of patient characteristics (e.g. age, gender, disease case) and image acquisition varieties (e.g. scanner manufacturer, image acquisition protocols, data noise and artifacts). The validation dataset was composed of 645 unique MR images. # Conclusion 8 The performance testing presented above shows that the device is as safe, as effective, and performs as well as the predicate device, and as well as gold-standard computer-aided expert manual segmentation. By virtue of the physical characteristics and intended use, THINQ™ is substantially equivalent to its predicate device, and its technological differences do not raise questions of safety and effectiveness.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...