Single Use Endoscope Valves Set

K192048 · Anrei Medical (Hangzhou) Co., Ltd. · ODC · May 14, 2020 · Gastroenterology, Urology

Device Facts

Record IDK192048
Device NameSingle Use Endoscope Valves Set
ApplicantAnrei Medical (Hangzhou) Co., Ltd.
Product CodeODC · Gastroenterology, Urology
Decision DateMay 14, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Single Use Endoscope Valves Set are used to fit to multiple endoscope working channels/ports to enable an endoscope operator control the function of the working channels/ports and prevent retrograde flow of the fluids, gases, and other materials. The Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI endoscopic procedure. The Suction Valve is intended to be used to control the suction on an endoscope during a GI endoscopic procedure. The Biopsy Valve is intended to accommodate various sizes of endoscopic accessory instruments while still providing a seal around the instrument channel inlet.

Device Story

Single-use, sterile endoscope valve set (Air/Water, Suction, Biopsy valves) for GI endoscopy. Valves fit into endoscope working channels/ports; operated by physician/clinician during procedures. Air/Water and Suction valves control respective flows; Biopsy valve provides seal around accessory instruments. Prevents retrograde flow of fluids/gases. Mechanical operation; no electronic or software components. Benefits include reduced cross-contamination risk via single-use design.

Clinical Evidence

No clinical data. Bench testing only. Performance verified via tensile strength, flow rates, pressure maintenance, leakage, backflow, insertion/removal force, and endoscope compatibility. Biocompatibility (cytotoxicity, sensitization, irritation) per ISO 10993. Sterilization validation per ISO 11135.

Technological Characteristics

Materials: ABS, Silica, SUS304. Mechanical valve design. Supplied sterile via Ethylene Oxide (EO). Shelf life: 3 years. Compatible with various endoscope series (e.g., Olympus 140-290, Fujinon 500/600, Pentax i10/90).

Indications for Use

Indicated for use by endoscope operators during GI endoscopic procedures to control working channel functions (air/water, suction) and provide a seal for accessory instrument passage while preventing retrograde fluid/gas flow.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services logo. The logo on the right is the FDA logo, which stands for U.S. Food & Drug Administration. The FDA logo is in blue and white. May 14, 2020 Anrei Medical (Hangzhou) Co., Ltd. % Ray Wang Official Correspondent Beijing Believe-Med Technology Service Co., Ltd. R912, B#15, XiYueHui, No.5, YiHe North Road FangShan District Beijing, 102401 CHINA Re: K192048 > Trade/Device Name: Single Use Endoscope Valves Set Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: ODC Dated: April 13, 2020 Received: April 15, 2020 Dear Ray Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) ### K192048 Device Name Single Use Endoscope Valves Set ### Indications for Use (Describe) The Single Use Endoscope Valves Set are used to fit to multiple endoscope working channels/ports to enable an endoscope operator control the function of the working channels/ports and prevent retrograde flow of the fluids, gases, and other materials. The Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI endoscopic procedure. The Suction Valve is intended to be used to control the suction on an endoscope during a GI endoscopic procedure. The Biopsy Valve is intended to accommodate various sizes of endoscopic accessory instruments while still providing a seal around the instrument channel inlet. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. {3}------------------------------------------------ # Tab #3 510(k) Summary This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92. The assigned 510(k) Number: K192048 - 1. Date of Preparation: 04/13/2020 - 2. Sponsor Identification Anrei Medical (HangZhou) Co., Ltd. No.3 Ave 8, HEDA, HangZhou City, ZheJiang, China 310018 Contact Person: Yang HuiBing Position: Regulations & Quality System Director Tel: +86-571-8691333 ext. 8658 Fax: +86-571-87603502 Email: huibing.yang@anrei.com.cn - Designated Submission Correspondent 3. Mr. Ray Wang Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, BeiJing, China 102401 Tel: +86-18910677558 Fax: +86-10- 56335780 Email: ray.wang@believe-med.com {4}------------------------------------------------ ### 4. Identification of Proposed Device Trade Name: Single Use Endoscope Valves Set Common Name: Endoscope and Accessories Model(s): AMH-EV-01; AMH-EV-02; AMH-EV-03 Regulatory Information Classification Name: Endoscope and Accessories Classification: 2 Product Code: ODC Regulation Number: 21 CFR 876.1500 Review Panel: Gastroenterology/Urology; ### Indications for Use Statement: The Single Use Endoscope Valves Set are used to fit to multiple endoscope working channels/ports to enable an endoscope operator control the function of the working channels/ports and prevent retrograde flow of the fluids, gases, and other materials. The Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI endoscopic procedure. The Suction Valve is intended to be used to control the suction on an endoscope during a GI endoscopic procedure. The Biopsy Valve is intended to accommodate various sizes of endoscopic accessory instruments while still providing a seal around the instrument channel inlet. ### Device Description The Single Use Endoscope Valves Set collects three types of valve products into one package unit, and which are used to fit to multiple endoscope working channels/ports to enable an endoscope operator control the function of the working channels/ports and prevent retrograde flow of the fluids, gases, and other materials. This is single-use device and supplied sterile. There are three main components included, which are Air/Water Valve, Suction Valve and Biopsy Valve. The Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI endoscopic procedure. The Suction Valve is intended to be used to control the suction on an endoscope during a GI endoscopic procedure. The Biopsy Valve is intended to accommodate various sizes of endoscopic accessory instruments while still providing a seal around the instrument channel inlet. There are three models for fitting to varied endoscopes as AMH-EV-01; AMH-EV-02; AMH-EV-03. The main differences between models are physical size of components for varied endoscopes interface, but they share same indication for use, materials and components form. {5}------------------------------------------------ ### ર. Identification of Predicate Device(s) 510(k) Number: K102409 Product Name: DEFENDO Disposable Air/Water Valve for GI Endoscopes 510(k) Number: K102581 Product Name: DEFENDO Disposable Suction Valve for GI Endoscopes 510(k) Number: K090851 Product Name: DEFENDO Biopsy Valve (Model #100301 and I00302) ### 6. Non-Clinical Test Discussion Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. Biological Testing was conducted as following standards, the results shown that the proposed devices has same biocompatibility with predicate devices. ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for Vitro cytotoxicity. ISO 10993-10:2010, Biological Evaluation of Medical Device, Part 10-Test for irritation and delay-type hypersensitivity; Sterile Testing was conducted as following standards, the results shown that the proposed devices has same sterilization performance with predicate devices. ISO 11135:2014 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices ISO 10993-7:2008 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals USP40 NF35<85> Bacterial Endotoxins Test. Package Testing was conducted as following standards, the results shown that the proposed devices has same package performance with predicate devices. ASTM D 3078-02(2013) Standard Test Method for Determination of Leaks in Flexible Packaging by Bubble Emission ASTM F 1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration DIN 58953-6:2016 Sterilization-Sterile supply - Part 6: Microbial barrier testing of packaging materials for medical devices which are to be sterilizaed. ASTM F88/F88M-15 Standard Test Method for Seal Strength of Flexible Barrier Materials ASTM F1140/F1140M-13 Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Packages The shelf Life validation was conducted as following standard, the results shown that the proposed {6}------------------------------------------------ devices meet the requirements of claimed shelf life. ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices The performance testing was conducted for the performance of tensile strength, water and gas flow rates, pressure maintenance, water and air flow leakage, backflow, insertion and removal, endoscope compatibility and insertion force etc., this testing was a test that compares the proposed device with the predicate devices. The results shown that the proposed device has same performance with predicate device. - Clinical Test Conclusion 7. No clinical study is included in this submission. {7}------------------------------------------------ Substantially Equivalent (SE) Compariso ం # ble 1 Comparison of Technology Characteristic | Item | Device name | Classification Name | Regulation Number | Predicate Device (K102409) | Predicate Device(K102581) | Predicate Device (K090851) | Remark | |------|----------------------------------------------------------|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------| | | Proposed Device(s)<br>Single Use Endoscope Valves<br>Set | Endoscope and accessories | 876.1500 | DEFENDO Disposable Air/Water<br>Valve for GI Endoscopes | DEFENDO Disposable Suction<br>Valve for GI Endoscopes | DEFENDO Biopsy Valve<br>(Model #100301 and 100302) | / | | | | Endoscope and accessories | 876.1500 | Endoscope and accessories | Endoscope and accessories | | SAME | | | | | 876.1500 | | | | SAME | | | | Indications for Use | The Single Use Endoscope<br>Valves Set are used to fit to<br>multiple endoscope working<br>channels/ports to enable an<br>endoscope operator control the<br>function of the working<br>channels/ports and prevent<br>retrograde flow of the fluids,<br>gases, and other materials. | The DEFENDO Disposable<br>Air/Water Valve is intended to be<br>used to control the air/water<br>function on an endoscope during a<br>GI endoscopic procedure. | The DEFENDO Disposable<br>Suction Valve is intended to be<br>used to control the suction<br>function on an endoscope<br>during a GI endoscopic<br>procedure. | DEFENDO Disposable Biopsy<br>Valve is indicated for covering<br>the endoscope biopsy port<br>during an endoscopy procedure.<br>The valve provides access for<br>endoscopic device passage and<br>exchange, helps maintain<br>sufflation, and minimizes<br>leakage of biomaterial from the<br>biopsy port throughout the<br>endoscopic procedure. | SAME | K192048 Page 5 of 9 {8}------------------------------------------------ | | endoscopic procedure.<br>The Biopsy Valve is intended<br>to accommodate various sizes<br>of endoscopic accessory<br>instruments while still<br>providing a seal around the<br>instrument channel inlet. | | | | | |------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|------------| | Prescription/OTC | Prescription Use | Prescription Use | Prescription Use | Prescription Use | SAME | | Components | Air/Water Valve, Suction<br>Valve, Biopsy Valve | Air/Water Valve | Suction Valve | Biopsy Valve | SAME | | Supplied Sterile | Yes | Yes | Yes | Yes | SAME | | Single use | Yes | Yes | Yes | Yes | SAME | | Feature | Sterile and single-use;<br>Compatible with multiple<br>endoscope working<br>channels/ports;<br>Enable an endoscope operator<br>control the function of the<br>working channels/ports and<br>prevent retrograde flow of the<br>fluids, gases, and other<br>materials; | Sterile and single-use;<br>Compatible with multiple<br>endoscope working channels/ports;<br>Enable an endoscope operator<br>control the function of the working<br>channels/ports and prevent<br>retrograde flow of the fluids, gases,<br>and other materials; | Sterile and single-use;<br>Compatible with multiple<br>endoscope channels/ports;<br>Enable an endoscope operator<br>control the function of the<br>working channels/ports and<br>prevent retrograde flow of the<br>fluids, gases, and other<br>materials; | Compatible with multiple<br>endoscope biopsy port during an<br>endoscopy procedure | SAME | | Materials | Air/Water Valve:<br>ABS+Silica+SUS304<br>Suction Valve:<br>ABS+Silica+SUS304 | Not Available | Not Available | Not Available | Difference | {9}------------------------------------------------ K192048 Page 7 of 9 {10}------------------------------------------------ | Summary | | |----------|--| | ( | | | 0(k<br>l | | | S | | | | 140/160/180/240/260/290<br>series;<br>Fujinon 500 and 600 series;<br>Pentax i10 and 90 series. | series endoscopes | series endoscopes | Endoscopes | | |----------------------|--------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|-------------------|---------------|------------| | | The air/water valves, suction<br>valve and biopsy valve are<br>housed in a single tray and<br>packaged in sealed initial<br>package. | Suction and air/water valves are<br>housed in a single tray and<br>packaged in a sealed Tyvek<br>pouch | | | | | Packaging | Not Available | Not Available | | | SAME | | Sterilization Method | EO | EO | | Not Available | SAME | | Shelf Life | 3 year | 1 year | | Not Available | Difference | | Biocompatibility | Cytotoxicity, Sensitization and<br>Irritation | Cytotoxicity, Sensitization and<br>Irritation | | Not Available | SAME | | Difference Analysis | | | | | | atibility The proposed device has different on the for this differs, the performance estable to teacher and closes devices and fect the safety and effectiveness of proposed device this different d fe: The proposed device has different shelf life than predicate device, the shelf life validation of 3 years has been conduced, which included perfo after accelerated aging. The testing results shown that 3 years shelf life would not affect the performance of proposed device. So, et the safety and effectiveness of proposed device Materials: Beause the detail materials are not wailable, or could be onsidered as different the tiffers, the Bioompitility testing has betactives and oroposed device {11}------------------------------------------------ ## 9. Substantially Equivalent (SE) Conclusion The conclusions drawn from the nonclinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device predicate
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...