DEFENDO DISPOSABLE AIR/WATER VALVE MODEL 100304

K102409 · Byrne Medical, Inc. · ODC · Nov 4, 2010 · Gastroenterology, Urology

Device Facts

Record IDK102409
Device NameDEFENDO DISPOSABLE AIR/WATER VALVE MODEL 100304
ApplicantByrne Medical, Inc.
Product CodeODC · Gastroenterology, Urology
Decision DateNov 4, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesPediatric

Indications for Use

The DEFENDO™ Disposable Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI endoscopic procedure.

Device Story

The DEFENDO™ Disposable Air/Water Valve is a single-use accessory for GI endoscopes; it replaces reusable valves to control air and water flow during endoscopic procedures. The device is manually operated by the clinician during the procedure. It functions by regulating the air/water channel of compatible Olympus 140/160/180/240/260 series endoscopes. By providing a sterile, disposable alternative to reusable valves, it eliminates the need for reprocessing/sterilization of the valve between uses, thereby reducing potential cross-contamination risks for patients.

Clinical Evidence

Bench testing only. Comparative water flow rate testing was performed between the DEFENDO™ valve and the Olympus MH-438 predicate. Testing measured flow rates over time through the endoscope. Results showed high correlation (r=0.989 for full duration; r=0.96 for 0-7 second cleaning interval), demonstrating equivalent performance.

Technological Characteristics

Disposable air/water valve for GI endoscopes. Compatible with Olympus 140/160/180/240/260 series. Sterile, single-use design. Mechanical operation via manual control of air/water channel. No software or electronic components.

Indications for Use

Indicated for control of air/water function on endoscopes during GI endoscopic procedures in pediatric to adult male and female patients.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K107409. Pg 1 of 3 ## Summary of Safety and Effectiveness Data NOV - 4 2010 The Byrne Medical DEFENDO™ Disposable Air/Water Valve is intended to be used The Bythe Medical DEFEND - Disposable Mining a GI endoscopic procedure. The Byrne Medical DEFENDO™ Disposable Air/Water Valve is substantialiy The Dyme McGlour DEF ENDO - Sports Water Valve. The following table is a comparison between the two: Table 14-1: Comparison of features and principles of operation between the DEFENDO™ Disposable Air/Water Valve and Predicate Device (Olympus®MH-438 Air/Water Valve, K001241) | Characteristic | Byrne Medical | Olympus® | Same? | |------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------| | Part number | 100304 | MH-438 | N/A | | Trade Name | 100304: DEFENDO™<br>Disposable Air/Water Valve | Olympus® EVIS EXERA<br>Colovideoscopes | N/A | | 510(k) Document<br>Number | This submission | K001241 | N/A | | Product Code | KOG | FDF | No | | Regulation<br>Number | 876.1500 | 876.1500 | Yes | | Class | II | II | Yes | | Review Advisory<br>Committee | Gastroenterology/Urology | Gastroenterology/Urology | Yes | | Indications for use | The DEFENDO™<br>Disposable Air/Water<br>Valve is intended to be<br>used to control the<br>air/water function on an<br>endoscope during a GI<br>endoscopic procedure. | The Olympus® Air/Water<br>valve is intended to be<br>used to control the<br>air/water function on an<br>Olympus ®endoscope<br>during a GI endoscopic<br>procedure. | Yes | | Sterile | Yes | No, user must sterilize | No | | Single Use | Yes | No, re-sterilize, re-use | No | | Compatible<br>Endoscope(s) | Olympus®<br>140/160/180/240/260<br>series endoscopes | Olympus®<br>140/160/180/240/260<br>series endoscopes | Yes | | Patient Population | Male/Female, Pediatric to<br>Adult | Male/Female Pediatric to<br>Adult | Yes | | Reusable or<br>disposable | Disposable | Reuseable | No | {1}------------------------------------------------ K 102 409 pg r g3 # Summary of Safety and Effectiveness Data #### Performance Testing Comparitive Testing Water Flow Test Ten BMI 100304 Air/Water valves and ten Olympus MH-438 Air/Water valves Ten DMF 100004 Mirrolet Varros varros side-by-side comparison of flow rates were compared to ouon other in a same as a surements were made over a time through the endoscope, whereby 100 most of flow of water through the scope. penod of 07.4 8000nds measures and intervals for comparison and statistical analysis yielded the following: Image /page/1/Figure/7 description: The image is a graph comparing the flow rate of the Olympus and Byrne air water valves. The x-axis represents time in seconds, ranging from 0 to 40, while the y-axis represents flow rate in ml/min, ranging from -20 to 100. Both valves show a rapid increase in flow rate initially, reaching a plateau around 80-90 ml/min before sharply decreasing around 25 seconds. n = 188 Correlation Coefficient r = 0.989 Image /page/1/Figure/9 description: The image is a black and white photograph that appears to be a scan of a document or a piece of paper. The majority of the image is white, with a scattering of small black specks or dots across the surface, giving it a slightly textured or aged appearance. A dark, horizontal line is present near the bottom of the image, stretching across the majority of the width. {2}------------------------------------------------ K 102409 pg 3 of 3 Additionally, measurements from 0 – 7 seconds were analyzed for comparison, the 0-7 second period is significant as this is the typical amount of oompanoon, the o i so i so r sens cleaning. The following is a graphical depiction of the results: > n = 36 Correlation coefficient r = 0.96 Image /page/2/Figure/4 description: This image is a line graph comparing two different data sets, labeled as 'Olympus' and 'Byrne'. The x-axis ranges from 0 to 8, while the y-axis ranges from -10 to 90. Both lines show an increasing trend, with 'Byrne' starting at a slightly higher value and generally remaining above 'Olympus'. The 'Olympus' line has a more pronounced dip around x=2 before continuing its upward trajectory. Olympus vs. Byrne Based on the above Comparative Testing results and the Bench Testing results (see Bench Testing results, Section 20 of this submission)from the comparison between the Olympus® and the Byrne Medical Air/Water Valves, we have concluded that the Byrne Medical DEFENDO™Disposable Air/Water Valve is substantially equivalent to the Olympus® MH-438 Air/Water Valve. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an image of an eagle with its wings spread. Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002 Mr. John Willis Director of Regulatory Affairs Byrne Medical Inc. 3150 Pollok Dr. CONROE TX 77303 - 4 2010 NOV Re: K102409 Trade/Device Name: DEFENDO™ Disposable Air/Water Valve for GI Endoscopes Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscopes and accessories Regulatory Class: II Product Code: ODC, FDF Dated: August 23, 2010 Received: August 24, 2010 Dear Mr. Willis: -----------------------------------------------------------------------------------------------------------------------------------------------------------------------------referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related {4}------------------------------------------------ #### Page 2 adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Hukut Lemur MD Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive. Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health #### Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known) Device Name: DEFENDO™ Disposable Air/Water Valve for GI Endoscopes 2102409 Indications for Use: The DEFENDO™ Disposable Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI endoscopic procedure. Prescription Use _____________________________________________________________________________________________________________________________________________________________ X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) ### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Ohlut hum (Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices 510(k) Number K102409 Page 1 of 1 ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...