The TK Insulin Syringe with Safety Retractable Device is a sterile, single-use, disposable and non-reusable, manual retractable safety insulin syringe intended for injection of U-100 insulin into the body, while reducing the risk of sharps injuries and the potential for insulin syringe reuse. The TK Insulin Syringe is a sterile, single-use, disposable, insulin syringe intended for injection of U-100 insulin into the body.
Device Story
Device consists of sterile, single-use, disposable insulin syringes, available with or without a manual retractable safety mechanism. Input is U-100 insulin; device delivers insulin via integrated needle and piston syringe. Safety version features anti-needle-stick mechanism; after injection, plunger is withdrawn and snapped off to contain needle within barrel, rendering it unusable. Used in clinical or home settings by patients or healthcare providers. Output is insulin delivery; safety feature provides sharps injury protection. Benefits include accurate insulin dosing and reduced risk of accidental needle sticks and syringe reuse.
Clinical Evidence
Bench testing only. No clinical data presented. Performance verified via testing to ISO 8537, ISO 7864, ISO 9626, ISO 23908, and USP <788>. Biocompatibility testing performed per ISO 10993 series (cytotoxicity, sensitization, irritation, systemic toxicity, hemolysis, pyrogenicity).
Technological Characteristics
Materials: Polypropylene (barrel, plunger, hub, sheath), rubber (piston), stainless steel (needle). Lubricant: Silicone oil. Energy source: Manual. Form factor: Integrated needle/syringe, various sizes (0.3cc-1cc, 27G-31G). Sterilization: Ethylene oxide. Connectivity: None. Standards: ISO 8537, ISO 7864, ISO 9626, ISO 23908, ISO 10993, USP <788>.
Indications for Use
Indicated for injection of U-100 insulin into the body for patients requiring insulin therapy. The safety version is indicated for use while reducing the risk of sharps injuries and syringe reuse.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 11, 2020
Anhui Tiankang Medical Technology Co.,Ltd. Bai Baodong RA Manager No.228 Weiyi Road, Economic Development Zone Tianchang, 239300 Cn
Re: K191639
Trade/Device Name: TK Insulin Syringe with/without Safety Retractable Device Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: MEG, FMF Dated: August 13, 2020 Received: August 13, 2020
#### Dear Bai Baodong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K191639
Device Name
TK Insulin Syringe with/without Safety Retractable Device
Indications for Use (Describe)
The TK Insulin Syringe with Safety Retractable Device is a sterile, single-use, disposable and non-reusable, manual retractable safety insulin syringe intended for injection of U-100 insulin into the body, while reducing the risk of sharps injuries and the potential for insulin syringe reuse.
The TK Insulin Syringe is a sterile, single-use, disposable, insulin syringe intended for injection of U-100 insulin into the body.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-size:16px"> ☑ </span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------------|
| <span style="font-size:16px"> ☐ </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(K) Summary: K191639
### Date prepared: 09.11.2020
#### 1. Submitter Name and Address:
-
| Owner Name: | Anhui Tiankang Medical Technology Co.,Ltd. |
|-----------------|---------------------------------------------------------------------|
| Address: | No.228 Weiyi Road, Development Zone,<br>Tianchang City, Anhui,China |
| Contactor Name: | Bai Baodong |
| TEL: | +86-550-7309187 |
| E-mail: | zy@tkmedical.com |
#### Manufacturer
| Name: | Anhui Tiankang Medical Technology Co.,Ltd |
|----------|------------------------------------------------------------------------|
| Address: | No.228 Weiyi Road, Economic Development<br>Tianchang City, Anhui,China |
#### US Agent:
| US Agent: | James H . Liao |
|-----------|---------------------------------------------------|
| Address: | 6775 Verde Ridge Rd Rancho Palos Verdes, CA 90275 |
| TEL: | 310 3758169 Ext |
| Email: | James@Sino2us.Com |
#### 2.Submission Devices :
Product Name: TK Insulin Syringe with/without Safety Retractable Device Classification Name: Syringe, Antistick Piston Syringe Classification : II. Product code: MEG and FMF Submission Type: 510(K) Traditional Regulation Number: 21 CFR 880.5860
#### 3.Predicate Device :
K152808
Product Name: U&U Insulin Syringe with/without Safety Retractable Device Classification name: Syringe, Antistick Piston Syringe Classification: II Product Code: MEG and FMF Regulation No. 21 CFR 880.5860 Review Panel :80
{4}------------------------------------------------
### 4. Indication for Use
The TK Insulin Syringe with Safety Retractable Device is a sterile, single-use, disposable and non-reusable, manual retractable safety insulin syringe intended for injection of U-100 insulin into the body, while reducing the risk of sharps injuries and the potential for insulin syringe reuse.
The TK Insulin Syringe is a sterile, single-use, disposable and non-reusable, insulin syringe intended for injection of U-100 insulin into the body.
# 5. Device Description:
The TK Insulin Syringe with/without Safety Retractable Device are similar devices except one has a safety feature and the other is a standard insulin syringe.
The TK Insulin Syringe with/without Safety Retractable Device is an integrated needle and piston syringe with an anti-needle-stick mechanism. The mechanism allows clear visualization of the injection site at all times. The mechanism shows the needle is contained within the syringe barrel. After standard techniques for injection, the plunger is withdrawn and snapped off which renders the needle unusable and prevents accidental needle sticks. The used syringe is then discarded into a sharp's container.
The subject devices are sterile, single-use, disposable and non-reusable, insulin syringe intended for injection of U-100 insulin into the body.
The TK Insulin Syringe with/without Safety Retractable Device will be available in numerous sizes and combinations in the following table.
| Ref Number | Model Number | Description | Size |
|------------|--------------|-----------------|----------------------------|
| TKIS001 | TKIS | Insulin Syringe | 0.3cc/ml Needle 27G × 1/2" |
| | | | 0.3cc/ml Needle 28G × 1/2" |
| | | | 0.3cc/ml Needle 29G × 1/2" |
| | | | 0.3cc/ml Needle 30G × 1/2" |
| | | | 0.3cc/ml Needle 31G × 3/8" |
| TKIS002 | TKIS | Insulin Syringe | 0.5cc/ml Needle 27G × 1/2" |
| | | | 0.5cc/ml Needle 28G × 1/2" |
| | | | 0.5cc/ml Needle 29G × 1/2" |
| | | | 0.5cc/ml Needle 30G × 1/2" |
{5}------------------------------------------------
| | | | 0.5cc/ml Needle 31G ×3/8" |
|----------|-------|---------------------------------------------------|----------------------------|
| TKIS003 | TKIS | Insulin Syringe | 1cc/ml Needle 27G × 1/2" |
| | | | 1cc/ml Needle 28G × 1/2" |
| | | | 1cc/ml Needle 29G × 1/2" |
| | | | 1cc/ml Needle 30G × 1/2" |
| | | | 1cc/ml Needle 31G × 3/8" |
| TKSIS001 | TKSIS | Insulin Syringe with Safety<br>Retractable Device | 0.5cc/ml Needle 27G × 1/2" |
| | | | 0.5cc/ml Needle 28G × 1/2" |
| | | | 0.5cc/ml Needle 29G × 1/2" |
| | | | 0.5cc/ml Needle 30G × 1/2" |
| | | | 0.5cc/ml Needle 31G × 3/8" |
| TKSIS002 | TKSIS | Insulin Syringe with Safety<br>Retractable Device | 1cc/ml Needle 27G × 1/2" |
| | | | 1cc/ml Needle 28G × 1/2" |
| | | | 1cc/ml Needle 29G × 1/2" |
| | | | 1cc/ml Needle 30G × 1/2" |
| | | | 1cc/ml Needle 31G × 3/8" |
### 6. Comparison of technological characteristics with the predicate:
Through comparisons between the submitted devices with the predicate devices as follows Tables. We believe the applicant devices are substantially equivalent with the predicate devices.
The main components of a TK insulin syringe with a safety retractable device are manufactured to the same or similar specifications as the said assertion device. The intended use, basic design, function, and materials of the predicate device are identical to or similar to the predicate device. The difference between the subject device and the predicate does not affect the safety and effectiveness of the subject device because the evaluation is conducted using industry consensus standards and based on the performance requirements of the device. Therefore, different technical features do not introduce new safety and effectiveness issues and are essentially equivalent to predicate devices.
{6}------------------------------------------------
### Comparison Table
| Element of<br>Comparison | Submission Device | Predicate Device K152808 | Comment |
|-----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The TK Insulin Syringe with<br>Safety Retractable Device is a<br>sterile, single-use, disposable<br>and non-reusable, manual<br>retractable safety insulin<br>syringe intended for injection<br>of U-100 insulin into the body,<br>while reducing the risk of<br>sharps injuries and the<br>potential for insulin syringe<br>reuse.<br><br>The TK Insulin Syringe is a<br>sterile, single-use, disposable<br>and non-reusable, insulin<br>syringe intended for injection<br>of U-100 insulin into the body. | The U&U Insulin Syringe with<br>Safety Retractable Device is a<br>sterile, single-use, disposable and<br>non-reusable, manual<br>retractable safety insulin syringe<br>intended for injection of U-100<br>insulin into the body, while<br>reducing the risk of sharps injuries<br>and the potential for insulin<br>syringe reuse.<br><br>The U&U Insulin Syringe is a<br>sterile, single-use, disposable and<br>non-reusable, insulin syringe<br>intended for injection of U-100<br>insulin into the body. | Same |
| Principle of Operation | Normal | Normal | Same |
| Syringe Capacity | TK Insulin Syringe:<br>0.3cc/ml Needle 27G × 1/2"<br>0.3cc/ml Needle 28G × 1/2"<br>0.3cc/ml Needle 29G × 1/2"<br>0.3cc/ml Needle 30G × 1/2"<br>0.3cc/ml Needle 31G × 3/8"<br>0.5cc/ml Needle 27G × 1/2"<br>0.5cc/ml Needle 28G × 1/2"<br>0.5cc/ml Needle 29G × 1/2"<br>0.5cc/ml Needle 30G × 1/2" | U&U Insulin Syringe<br>0.3cc/ml Needle 27G to 31G<br>0.5cc/ml Needle 27G to 31G<br>1cc/ml Needle 27G to 31G<br><br>U&U Insulin Syringe with Safety<br>Retractable Device:<br>1cc/ml only Needle 27G to 31G | Different<br>Because there<br>are many<br>customers in<br>our market who<br>need 0.5cc/ml,<br>which indicates<br>that customers<br>have demand<br>for this<br>specification,<br>we have added<br>0.5cc/ml in the<br>syringe |
| | 0.5cc/ml Needle 31G × 3/8" | | capacity.<br>Adding 0.5mL<br>does not affect<br>the safety and<br>effectiveness of<br>the subject<br>device and is<br>verified by<br>non-clinical<br>testing listed in<br>Section 7. |
| | 1cc/ml Needle 27G × 1/2" | | |
| | 1cc/ml Needle 28G × 1/2" | | |
| | 1cc/ml Needle 29G × 1/2" | | |
| | 1cc/ml Needle 30G × 1/2" | | |
| | 1cc/ml Needle 31G × 3/8" | | |
| | TK Insulin Syringe with<br>Safety Retractable Device: | | |
| | 0.5cc/ml Needle 27G × 1/2" | | |
| | 0.5cc/ml Needle 28G × 1/2" | | |
| | 0.5cc/ml Needle 29G × 1/2" | | |
| | 0.5cc/ml Needle 30G × 1/2" | | |
| | 0.5cc/ml Needle 31G × 3/8" | | |
| | 1cc/ml Needle 27G × 1/2" | | |
| | 1cc/ml Needle 28G × 1/2" | | |
| | 1cc/ml Needle 29G × 1/2" | | |
| | 1cc/ml Needle 30G × 1/2" | | |
| | 1cc/ml Needle 31G × 3/8" | | |
| Nozzle Type | N.A | N.A | Same |
| Lubricant for Barrel | Silicone Oil | Silicone Oil | Same |
| Barrel Transparency | Transparent and Clear | Transparent and Clear | Same |
| Gradations Legibility | Legible | Legible | Same |
| Materials | | | |
| Barrel | PP | PP | Same |
| Plunger | PP | PP | |
| | | | Same |
| Piston | Rubber | Rubber | |
| Lubricant for Barrel | HC-SS36 | | |
| Needle Hub | PP | | |
| Needle | Stainless | | |
| Needle Sheath | PP | | |
| Needle Gauge and<br>Length | Various Sizes | Same | |
| Lubricant for Needle | Silicone Oil | Same | |
| TK Insulin Syringe<br>with Safety Retractable<br>Device: Sharps Injury<br>Prevention Features | Manual Retractable Conforms<br>to ISO 23908 | Same | |
| Performances | Conforms to ISO7864<br>ISO8537 | Same | |
| Biocompatibility | Conforms to ISO10993(Part 1: Evaluation and testing, Part 4: Selection of tests for interactions with blood, Part 5: Tests for in vitro cytotoxicity, Part 7: Ethylene oxide sterilization residuals, Part 10: Tests for irritation and delayed-type hypersensitivity, Part 11: Tests for systemic toxicity, Tests for Bacterial endotoxins, Tests for Pyrogenicity)<br>Conforms to USP <788>:<br>Particulate Matter for injection | Conforms to ISO10993(Part 1: Evaluation and testing, Part 4: Selection of tests for interactions with blood, Part 5: Tests for in vitro cytotoxicity, Part 7: Ethylene oxide sterilization residuals, Part 10: Tests for irritation and delayed-type hypersensitivity, Part 11: Tests for systemic toxicity)<br>Conforms to USP <788>:<br>Particulate Matter for injection | Same |
| Labeling | Meet the requirements of<br>21 CFR Part 801 | Same | |
{7}------------------------------------------------
{8}------------------------------------------------
# 7.Non-Clinical Tests performed on the subject device:
The proposed devices were tested per the following standards, to evaluate its performance.
[PMN-510(K) SUBMISSION]
{9}------------------------------------------------
- ISO 8537: 2016 Third edition, Sterile single-use syringes, with or without ● needle, for insulin.
- ISO 7864:2016 Fourth edition, Sterile hypodermic needles for single use-● Requirements and test method.
## For the TK Insulin Syringe with Safety Retractable Device Only:
- ISO 9626:2016 Second edition 2016-08-01 Stainless steel needle tubing for manufacture of medical device-Requirements and test method.
- . ISO 23908:2011 First edition 2011-06-11 Sharps Injury Protection-Requirements and test methods-Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling.
- Guidance for Industry and FDA Staff: Medical Devices with Sharps Injury . Prevention Features:
## Biocompatibility
Conforms to ISO10993. The TK Insulin Syringe with/without Safety Retractable Device is classified as prolonged, external communicating device, indirect blood contact with the following endpoints:
- . Cytotoxicity
- . Sensitization
- Irritation or Intracutaneous reactivity ●
- . Acute systemic toxicity
- . Material mediated pyrogenicity
- Subacute/Subchronic toxicity
- Indirect hemolysis
- . USP 788
- . ISO 10993-1:2018 Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process
- · ISO 10993-4:2017 Biological evaluation of medical devices Part 4: Selection of tests for interactions with blood
- ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity
- · ISO 10993-7:2008 Biological evaluation of medical devices Part 7: Ethylene oxide sterilization residuals
{10}------------------------------------------------
- ISO 10993-10: 2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization
- ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic toxicity
- USP <788>: Particulate Matter for injection
### 8. Conclusions:
The TK Insulin Syringe with/without Safety Retractable Device is substantially equivalent to the U&U Insulin Syringe with Safety Retractable Device. The materials, performance, and operational features of both the subject device and the predicate devices are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.