K191561 · Exactech, Inc. · PHX · Aug 20, 2019 · Orthopedic
Device Facts
Record ID
K191561
Device Name
Exactech Equinoxe Humeral Augmented Trays
Applicant
Exactech, Inc.
Product Code
PHX · Orthopedic
Decision Date
Aug 20, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Equinoxe Humeral Augmented Tray is indicated for proximal humeral replacement in conjunction with reverse shoulder arthroplasty in skeletally mature individuals with an intact medial calcar for whom the rotator cuff is irreparable or grossly deficient, a functional deltoid muscle is present, and proximal humeral bone loss is present in cases of: - Rheumatoid arthritis, osteoarthritis, osteonecrosis or post-traumatic degenerative problems - Congenital abnormalities in the skeletally mature - Primary and secondary necrosis of the humeral head - Humeral head fracture with displacement of the tuberosities - Pathologies where arthrodesis or resectional arthroplasty of the humeral head are not acceptable - Revisions of humeral prostheses when other treatments or devices have failed (where adequate humeral stem fixation can be achieved) - Displaced three- and four-part upper humeral fractures - Failed ORIF requiring rTSA with an absent or deficient greater tuberosity - Revision of failed previous reconstructions when distal anchorage is required - Treatment of unplanned proximal bone loss during a revision - To restore mobility from previous procedures (e.g. previous fusion) The Equinoxe Humeral Augmented Tray can be used in either primary or revision arthroplasty procedures.
Device Story
Modular orthopedic implant system for proximal humeral bone loss; used in reverse shoulder arthroplasty. Components include humeral tray, modular extension (augment), and universal locking screw. Surgeon selects components based on patient anatomy; modular extension attaches to tray via locking screw to address severe bone loss. Implanted by orthopedic surgeons in clinical/OR settings. Device restores joint mechanics, improves deltoid wrapping, and provides soft tissue tensioning. Titanium alloy construction. Benefits include improved stability and mobility in revision or primary cases with significant bone deficiency.
Clinical Evidence
Bench testing only. Included fatigue strength testing under worst-case lateral and medial loading, range of motion analysis, and cadaveric evaluation. Pyrogen testing performed per USP <161>, USP <85>, and ANSI/AAMI ST72.
Technological Characteristics
Materials: Titanium Alloy (Ti-6Al-4V). Components: Humeral tray, modular extension, universal locking screw. Design: Anatomical, modular, multiple offsets/thicknesses/lateralizations. Sterilization: Compliant with FDA guidance for sterile devices.
Indications for Use
Indicated for skeletally mature patients requiring proximal humeral replacement with reverse shoulder arthroplasty. Requires intact medial calcar, functional deltoid muscle, and irreparable/deficient rotator cuff. Used for proximal humeral bone loss due to arthritis, necrosis, fractures, congenital abnormalities, or failed prior reconstructions.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
Equinoxe Reverse Shoulder System (Humeral Adapter Trays) (K063569)
Equinoxe Reverse Shoulder System + 15mm Humeral Adapter Tray (K082702)
{0}------------------------------------------------
August 20, 2019
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
Exactech, Inc Shing Jen Tai Principal Regulatory Affairs Specialist 2320 NW 66th Court Gainesville, Florida 32653
Re: K191561
Trade/Device Name: Exactech® Equinoxe® Humeral Augmented Trays Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: PHX, KWT, KWS Dated: June 11, 2019 Received: June 13, 2019
Dear Shing Jen Tai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
FOR Vesa Vuniqi Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
#### 510(k) Number (if known) K191561
Device Name
Exactech® Equinoxe® Humeral Augmented Trays
#### Indications for Use (Describe)
The Equinoxe Humeral Augmented Tray is indicated for proximal humeral replacement in conjunction with reverse shoulder arthroplasty in skeletally mature individuals with an intact medial calcar for whom the rotator cuff is irreparable or grossly deficient, a functional deltoid muscle is present, and proximal humeral bone loss is present in cases of:
- · Rheumatoid arthritis, osteoarthritis, osteonecrosis or post-traumatic degenerative problems
- · Congenital abnormalities in the skeletally mature
- · Primary and secondary necrosis of the humeral head
- · Humeral head fracture with displacement of the tuberosities
- · Pathologies where arthrodesis or resectional arthroplasty of the humeral head are not acceptable
- Revisions of humeral prostheses when other treatments or devices have failed (where adequate humeral stem fixation can be achieved)
- · Displaced three- and four-part upper humeral fractures
- · Failed ORIF requiring rTSA with an absent or deficient greater tuberosity
- Revision of failed previous reconstructions when distal anchorage is required
- · Treatment of unplanned proximal bone loss during a revision
- · To restore mobility from previous procedures (e.g. previous fusion)
The Equinoxe Humeral Augmented Tray can be used in either primary or revision arthroplasty procedures.
Type of Use (Select one or both, as applicable)
| <div> <span> <svg class="bi bi-check-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> <path d="M10.97 4.97a.75.75 0 0 1 1.071 1.05l-3.992 4.99a.75.75 0 0 1-1.08.02L4.324 8.384a.75.75 0 1 1 1.06-1.06l2.094 2.093 3.473-4.425a.236.236 0 0 1 .02-.022z" fill-rule="evenodd"></path> </svg> </span>Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg> </span>Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Exactech®Equinoxe® Humeral Augmented Trays Traditional 510(k) - 510(k) Summary
| Sponsor: | Exactech®, Inc<br>2320 N.W. 66th Court<br>Gainesville, FL 32653 |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| | Phone: (352) 377-1140<br>Fax: (352) 378-2617 |
| | FDA Establishment Number 1038671 |
| Contact Person: | Shing Jen Tai, PhD<br>Principal Regulatory Affairs Specialist |
| Date: | August 8, 2019 |
| Proprietary Name: | Exactech® Equinoxe® Humeral Augmented Trays |
| Common Name: | Reverse shoulder prosthesis |
| Regulation Number: | 21 CFR 888.3650<br>21 CFR 888.3660 |
| Classification Name: | Shoulder joint metal/polymer non-constrained cemented<br>prosthesis<br>Shoulder joint metal/polymer semi-constrained cemented<br>prosthesis |
| Device Class: | Class II |
| Product Code: | PHX, KWT, KWS |
| Classification Panel: | Orthopedic |
# Information on devices to which substantial equivalence is claimed:
| 510(k) Number | Trade or Proprietary Model Name | Manufacturer |
|---------------|-----------------------------------------------------------------|---------------|
| K063569 | Equinoxe Reverse Shoulder System (Humeral<br>Adapter Trays) | Exactech, Inc |
| K082702 | Equinoxe Reverse Shoulder System + 15mm<br>Humeral Adapter Tray | Exactech, Inc |
{4}------------------------------------------------
## Exactech® Equinoxe® Humeral Augmented Trays Traditional 510(k) - 510(k) Summary
#### Information on reference device:
| 510(k) Number | Trade or Proprietary Model Name | Manufacturer |
|---------------|--------------------------------------------|---------------|
| K143659 | Equinoxe Humeral Reconstruction Prosthesis | Exactech, Inc |
### Indications for Use
The Equinoxe Humeral Augmented Tray is indicated for proximal humeral replacement in conjunction with reverse shoulder arthroplasty in skeletally mature individuals with an intact medial calcar for whom the rotator cuff is irreparable or grossly deficient, a functional deltoid muscle is present, and proximal humeral bone loss is present in cases of:
- Rheumatoid arthritis, osteoarthritis, osteonecrosis or post-traumatic degenerative ● problems
- Congenital abnormalities in the skeletally mature ●
- Primary and secondary necrosis of the humeral head ●
- Humeral head fracture with displacement of the tuberosities ●
- Pathologies where arthrodesis or resectional arthroplasty of the humeral head are ● not acceptable
- Revisions of humeral prostheses when other treatments or devices have failed (where adequate humeral stem fixation can be achieved)
- Displaced three- and four-part upper humeral fractures
- Failed ORIF requiring TTSA with an absent or deficient greater tuberosity ●
- Revision of failed previous reconstructions when distal anchorage is required ●
- Treatment of unplanned proximal bone loss during a revision
- To restore mobility from previous procedures (e.g. previous fusion) ●
The Equinoxe Humeral Augmented Tray can be used in either primary or revision arthroplasty procedures.
## De vice De scription
The Exactech Equinoxe Humeral Augmented Trays are intended for use in total shoulder arthroplasty in skeletally mature patients with proximal humeral bone loss. Specifically, Equinoxe Humeral Augmented Trays are used in combination with Equinoxe Reverse Shoulder System components to compensate for proximal humeral bone loss in presence of an intact medial calcar, irreparable or grossly deficient rotator cuff and a functional deltoid muscle. One of the overall goals of the anatomically designed Equinoxe Humeral Augmented Trays is to improve joint mechanics and stability by providing increased deltoid muscle wrapping and adequate soft tissue tensioning in patients with proximal humeral bone loss.
The Equinoxe Humeral Augmented Trays are consisted of a humeral tray component, a modular extension (augment) and a universal locking screw for use in primary and revision cases. To accommodate patients' various anatomical needs. Equinoxe Humeral Augmented Trays are anatomically designed with two offsets/thicknesses and multiple lateralizations of left and right humeral trays and left and right modular extensions. Per
{5}------------------------------------------------
## Exactech® Equinoxe® Humeral Augmented Trays Traditional 510(k) - 510(k) Summary
surgeon evaluation, the modular extension secured with a locking screw is used to replace greater and more severe proximal humeral bone loss than can be adequately addressed with use of humeral tray component alone. When the modular extension is used, the locking screw must be used to secure it to the humeral tray component. All three components of the Equinoxe Humeral Augmented Trays are manufactured from Titanium Alloy (Ti-6Al-4V).
## Non-Clinical Testing
The following mechanical testing, engineering analysis, and cadaveric evaluation were performed to demonstrate that the Exactech Equinoxe Humeral Augmented Trays perform as intended and are substantially equivalent to the identified predicate Equinoxe Humeral Adapter Trays previously cleared in K063569 and K082702:
- Fatigue strength testing under worst-case lateral loading
- Fatigue strength testing under worst-case medial loading ●
- Range of Motion Analysis ●
- Cadaveric evaluation
Pyrogen testing was conducted in accordance with USP <161>, USP <85>, and ANSI/AAMI ST72 to ensure the proposed Equinoxe Humeral Augmented Trays meet recommended limits per FDA's Guidance Document Submission and Review of Sterility Information in Premarket (510(k)) Submission for Devices Labeled as Sterile.
## Substantial Equivalence Conclusion
Results of the non-clinical testing including mechanical testing, engineering analysis and cadaveric evaluation demonstrate the proposed Exactech Equinoxe Humeral Augmented Trays are substantially equivalent to the previously cleared Humeral Adapter Trays in the Exactech Equinoxe Reverse Shoulder System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.