The Equinoxe Mega Prosthesis System is intended for use in hemi or total shoulder arthroplasty where proximal humeral resection is deemed necessary in cases of: - Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis - Rheumatoid arthritis - Revision where other devices or treatments have failed - Correction of functional deformity - Treatment of acute or chronic fracture with humeral head involvement, which are unmanageable using other treatment - Traumatic arthritis - Oncology applications including bone loss due to tumor resection. - Significant humeral resection which are unmanageable using other treatment methods The Equinoxe Mega Prosthesis System can be used in either primary or revision arthroplasty procedures. The Equinoxe Mega Prosthesis System is indicated for proximal humeral replacement in conjunction with reverse shoulder arthroplasty in which significant resection of the proximal humerus is necessary, the rotator cuff is irreparable and grossly deficient, and a functional deltoid muscle is present. The Equinoxe Mega Prosthesis System is not indicated for use with the Equinoxe Reverse Shoulder System components in oncology applications. The Equinoxe Mega Prosthesis Distal Stems are for cemented use only, while the HA coated Equinoxe Mega Prosthesis Distal Fixation Rings are only for uncemented, press-fit use.
Device Story
Modular shoulder prosthesis system replacing articulating humeral head and resected humerus; used in hemi, total, or reverse shoulder arthroplasty. Surgeon-assembled modular segments (proximal bodies, middle segments, distal stems) match resected bone length; distal fixation rings provide press-fit option. Used in OR by orthopedic surgeons. Provides structural replacement for bone loss; restores arm length; facilitates soft tissue attachment via suture holes. Benefits patients with significant humeral bone loss by restoring joint function and stability.
Clinical Evidence
No clinical trials performed. Evidence consists of clinical literature review, cadaveric study, and extensive bench testing including fatigue, torsion, taper analysis, axial pull-off, suture abrasion, and plasma spray coating characterization.
Indicated for patients requiring proximal humeral resection due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, revision arthroplasty, functional deformity, acute/chronic fracture, traumatic arthritis, or oncology-related bone loss. Also indicated for proximal humeral replacement with reverse shoulder arthroplasty in patients with irreparable/deficient rotator cuff and functional deltoid. Contraindicated for use with Equinoxe Reverse Shoulder System components in oncology applications.
Regulatory Classification
Identification
A shoulder joint metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-Aluminum 4-Vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-Chromium-Molybdenum Casting Alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-Chromium-Molybdenum Alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal Coatings,”
(vii) F 1378-97 “Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
Biomet Comprehensive Segmental Revision System (K111746)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three faces in profile, overlapping each other to create a sense of depth and unity. The faces are simple and abstract, with no distinct features other than the outline of the head and neck.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 9, 2015
Exactech, Incorporated Mr. Thomas McNamara Regulatory Affairs Specialist 2320 Northwest 66th Court Gainesville, Florida 32653
Re: K143659
Trade/Device Name: Exactech® Equinoxe® Mega Prosthesis Regulation Number: 21 CFR 888.3650 Regulation Name: Shoulder joint metal/polymer non-constrained cemented prosthesis Class: Class II Product Code: KWT, KWS, PHX, HSD Dated: January 8, 2015 Received: January 12, 2015
Dear Mr. McNamara:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K143659
#### Device Name
Exactech® Equinoxe® Mega Prosthesis
#### Indications for Use (Describe)
The Equinoxe Mega Prosthesis System is intended for use in hemi or total shoulder arthroplasty where proximal humeral resection is deemed necessary in cases of:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis
- · Rheumatoid arthritis
- · Revision where other devices or treatments have failed
- · Correction of functional deformity
- Treatment of acute or chronic fracture with humeral head involvement, which are unmanageable using other treatment
- · Traumatic arthritis
- · Oncology applications including bone loss due to tumor resection.
- · Significant humeral resection which are unmanageable using other treatment methods
The Equinoxe Mega Prosthesis System can be used in either primary or revision arthroplasty procedures.
The Equinoxe Mega Prosthesis System is indicated for proximal humeral replacement in conjunction with reverse shoulder arthroplasty in which significant resection of the proximal humerus is necessary, the rotator cuff is irreparable and grossly deficient, and a functional deltoid muscle is present.
The Equinoxe Mega Prosthesis System is not indicated for use with the Equinoxe Reverse Shoulder System components in oncology applications.
The Equinoxe Mega Prosthesis Distal Stems are for cemented use only, while the HA coated Equinoxe Mega Prosthesis Distal Fixation Rings are only for uncemented, press-fit use.
Type of Use (Select one or both, as applicable)
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Exactech® Equinoxe® Mega Prosthesis Traditional 510(k) - 510(k) Summary
| Company: | Exactech®, Inc<br>2320 NW 66th Court<br>Gainesville, FL 32653 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | April 8, 2015 |
| Contact Person: | Thomas McNamara<br>Regulatory Affairs Specialist<br>Phone: (352) 377-1140<br>Fax: (352) 378-2617 |
| Proprietary Name: | Exactech® Equinoxe® Mega Prosthesis |
| Common Name: | Humeral Stem |
| Classification Name: | Prosthesis, Shoulder, Non-Constrained, Meta/Polymer<br>Cemented, (21 CFR Section 888.3650, Class II, Product<br>Code KWT), Prosthesis, Shoulder, Semi-Constrained<br>cemented prosthesis (21 CFR Section 888.3660, Class II<br>Product Code KWS), Shoulder Prosthesis, Reverse<br>Configuration (21 CFR Section 888.3660, Class II Product<br>Code PHX) Prosthesis, Shoulder, Hemi-, Humeral Metallic<br>Uncemented, (21 CFR Section 888.3690, Class II, Product<br>Code HSD) |
### Legally Marketed Device to Which Substantial Equivalence Is Claimed:
- Biomet Comprehensive Segmental Revision System (K111746) ●
#### Device Description
The Equinoxe Mega Prosthesis differs from conventional Total Shoulder Arthroplasty implants in that it is intended to replace both the articulating surface (humeral head) as well as the resected humerus. The modular design allows for the surgeon to construct the middle and proximal segments of the prosthesis in various lengths in order to match the length of resected bone, and to maintain the patient's natural arm length.
The prosthesis is manufactured from Ti-6A1-4V and has both Titanium Plasma Spray and HA coated regions. The proximal bodies are available in four sizes (Small, Medium, Large, Extra-Large) and two lengths (+0mm, +12.5mm). The middle segments are available in 3 lengths (25mm, 50mm, 75mm) with one diameter (20mm). The distal stems are available in 3 lengths (80mm, 120mm, 200mm) and 6 diameters (6mm, 7mm, 8mm, 9mm, 11mm, 13mm). The distal fixation rings are available in 17 diameters (17.5mm-33.5mm, 1mm increments). The distal fixation rings are intended for press-fit fixation. The proximal bodies and middle segments have suture through holes to facilitate soft tissue attachment.
{4}------------------------------------------------
#### Indications for Use
The Equinoxe Mega Prosthesis System is intended for use in hemi or total shoulder arthroplasty where proximal humeral resection is deemed necessary in cases of:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis
- . Rheumatoid arthritis
- . Revision where other devices or treatments have failed
- Correction of functional deformity
- . Treatment of acute or chronic fracture with humeral head involvement, which are unmanageable using other treatment
- . Traumatic arthritis
- . Oncology applications including bone loss due to tumor resection.
- . Significant humeral resection which are unmanageable using other treatment methods
The Equinoxe Mega Prosthesis System can be used in either primary or revision arthroplasty procedures.
The Equinoxe Mega Prosthesis System is indicated for proximal humeral replacement in conjunction with reverse shoulder arthroplasty in which significant resection of the proximal humerus is necessary, the rotator cuff is irreparable and grossly deficient, and a functional deltoid muscle is present.
The Equinoxe Mega Prosthesis System is not indicated for use with the Equinoxe Reverse Shoulder System components in oncology applications.
The Equinoxe Mega Prosthesis Distal Stems are for cemented use only, while the HA coated Equinoxe Mega Prosthesis Distal Fixation Rings are only for uncemented, pressfit use.
#### Summary of Technological Characteristics
The rationale for substantial equivalence is based on consideration of the following device use and characteristics:
- Indications for Use. The proposed Exactech Equinoxe Mega Prosthesis and the predicate device have similar indications for use. The proposed Exactech Equinoxe Mega Prosthesis is also indicated for use in hemi-shoulder arthroplasty and reverse shoulder arthoplasty. In cases with significant proximal humeral resections, joint stability is the primary concern. The stability of the joint is comparable for hemi and total shoulder indications, because stability is achieved by dynamic muscle balancing. The stability of the joint in a reverse shoulder is achieved via the inverted articular concavities without any requirement for dynamic stability provided by the rotator cuff. The additions of the proposed hemi
{5}------------------------------------------------
# Exactech® Equinoxe® Mega Prosthesis Traditional 510(k) - 510(k) Summary
and reverse shoulder arthroplasty indications do not create a new intended use for the proposed device.
- Materials/Surface Finish/Coatings. The proposed Exactech Equinoxe Mega . Prosthesis and the predicate device are composed of similar, biocompatible substrate materials, and the same or similar surface finish/coatings for permanent implants.
- . Design Features. The proposed Exactech Equinoxe Mega Prosthesis and the predicate device share similar design features. Design differences between the cited predicate device and the propose Exactech Equinoxe Mega Prosthesis include the use of a distal fixation ring and different sizes of distal stems and proximal bodies, as well as the hemi- and reverse shoulder indications.
- Dimensions. The proposed Exactech Equinoxe Mega Prosthesis and the predicate device are dimensionally comparable.
- Sterilization. The proposed Exactech Equinoxe Mega Prosthesis and the . predicate device are provided sterile for single use only.
- Performance Requirements. The proposed Exactech Equinoxe Mega Prosthesis and the predicate device conform to recognized performance standards for shoulder replacement devices.
### Non-Clinical Testing
The following clinical literature review, template studies, mechanical testing, and cadaveric evaluation were performed to demonstrate that the Exactech Mega Prosthesis performs as intended and is substantially equivalent to the identified predicate device:
- Clinical Literature Review ●
- Cadaveric Study ●
- Plasma Spray Coating Characterization ●
- Fatigue Testing
- . Torsion Testing
- Taper Analysis
- Axial Pull-Off Testing ●
- . Suture Abrasion Testing
Bench testing was conducted to address the design differences between the proposed Exactech Equinoxe Mega Prosthesis and the identified predicate device, including the use of a distal fixation ring and different sizes of distal stems and proximal bodies, as well as the addition of the indication for reverse shoulder arthroplasty.
### Substantial Equivalence Conclusion
Based on consideration of indications for use, technological characteristics, and results of combined mechanical testing, cadaveric validation study, and clinical literature review described above, it was concluded that Exactech Equinoxe Mega Prosthesis demonstrates substantial equivalence to the referenced predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.