K191557 · Thoratec Corporation (Now Part of Abbott) · DWA · Jul 10, 2019 · Cardiovascular
Device Facts
Record ID
K191557
Device Name
CentriMag Acute Circulatory Support System
Applicant
Thoratec Corporation (Now Part of Abbott)
Product Code
DWA · Cardiovascular
Decision Date
Jul 10, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4380
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CentriMag Extracorporeal Blood Pumping System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc).
Device Story
CentriMag Acute Circulatory Support System provides temporary mechanical circulatory support; used during cardiopulmonary bypass or other extracorporeal circulatory support procedures. System components: primary console, motor, blood pump, flow probe. Motor uses Full MagLev technology to drive pump impeller; rotation creates pressure gradient, moving blood from inflow to outflow port. User sets pump speed via console. Device used in clinical settings (e.g., OR); operated by healthcare professionals. Output (blood flow) assists hemodynamics in patients with acute compromise. Redesign focuses on motor cable with enhanced bend protection to improve durability, flexibility, and reduce risk of cable breaks/short circuits.
Clinical Evidence
No clinical data. Bench testing only. Mechanical performance verification included motor cable endurance testing and kink testing under simulated severe handling to compare the redesigned cable against the predicate design. Compliance with IEC 60601-1 confirmed.
Technological Characteristics
CentriMag Motor uses Full MagLev technology for impeller rotation. System includes primary console, motor, blood pump, and flow probe. Redesign involves motor cable assembly with extended bend protection. Complies with IEC 60601-1. No changes to materials or fundamental sensing/actuation principles.
Indications for Use
Indicated for patients requiring extracorporeal circulatory support during cardiopulmonary bypass or other procedures not requiring complete bypass (e.g., valvuloplasty, mitral valve reoperation, vena cava/aorta surgery, liver transplants) for up to six hours.
Regulatory Classification
Identification
A cardiopulmonary bypass pump speed control is a device used that incorporates an electrical system or a mechanical system, or both, and is used to control the speed of blood pumps used in cardiopulmonary bypass surgery.
Predicate Devices
CentriMag Extracorporeal Blood Pumping System (K020271)
CentriMag 2nd Generation Primary Console and Mag Monitor (K102129)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services-USA seal. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
Thoratec Corporation (now part of Abbott) Lori Dondiego Associate Director, Regulatory Affairs 6035 Stoneridge Drive Pleasanton, California 94588
Re: K191557
Trade/Device Name: CentriMag Acute Circulatory Support System Regulation Number: 21 CFR 870.4380 Regulation Name: Cardiopulmonary Bypass Pump Speed Control Regulatory Class: Class II Product Code: DWA Dated: June 10, 2019 Received: June 12, 2019
Dear Lori Dondiego:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Fernando Aguel Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### 3.0 Indications for Use Statement (Form FDA 3881)
| Applicant: | Thoratec (now part of Abbott) |
|---------------------------|----------------------------------------------------------|
| 510(k) Number (if known): | |
| Device Name: | CentriMag <sup>TM</sup> Acute Circulatory Support System |
### Indications for Use:
The CentriMag Extracorporeal Blood Pumping System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc).
Prescription Use _ X (Part 21 CFR 801 Subpart D) Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
AND/OR
{3}------------------------------------------------
#### 4.0 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title 21 of the Code of Federal Regulations, Part 807, and in particular §807.92.
## A. Application Information
| | Date Prepared: | June 10, 2019 |
|----|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Submitter's Name & Address: | Thoratec (now part of Abbott)<br>6035 Stoneridge Drive<br>Pleasanton, CA 94588<br>Establishment Registration<br>No.: 2916596 |
| | Owner / Operator No.: | 1415939 |
| | Contact Person: | Kim Bondarenko<br>Manager, Regulatory Affairs<br>Ph: 925-734-4097<br>Fax: (925) 847-8628<br>Thoratec (now part of Abbott)<br>6035 Stoneridge Drive<br>Pleasanton, CA 94588 |
| B. | Device Information | |
| | Trade or Proprietary Name: | CentriMag™ Acute Circulatory Support<br>System |
| | Common or Usual Name: | CentriMag Extracorporeal Blood Pumping<br>System |
| | Classification Name: | Class II, DWA, 21 CFR - 870.4380<br>Control, Pump Speed, Cardiopulmonary<br>Bypass |
| | Performance Standard: | Performance standards do not currently<br>exist for these devices. None established<br>under section 514 of the Food, Drug and<br>Cosmetic Act. |
{4}------------------------------------------------
## C. Legally Marketed Predicate Device
CentriMag Extracorporeal Blood Pumping System (K020271) — Original blood pumping system with 1st generation Primary console and Back-up console.
CentriMag 2nd Generation Primary Console and Mag Monitor (K102129) -Redesigned (2nd generation) console and new peripheral monitor introduced.
# D. Device Description
The CentriMag™ Circulatory Support System, hereafter referred to as the CentriMag System, was designed to provide temporary mechanical circulatory support. To date, the CentriMag system in the United States (US) is indicated as a component of an extracorporeal bypass circulatory support circuit for use during cardiopulmonary bypass (CPB) surgery (up to 6 hours). The CentriMag System provides circulatory assistance for patients in acute hemodynamic compromise, a population whose treatment options are limited.
The CentriMag System is composed of:
- CentriMag Primary Console
- CentriMag Motor
- CentriMag Blood Pump
- CentriMag Flow Probe
- Mag Monitor (optional, not shown)
Image /page/4/Picture/14 description: The image shows a clear, plastic, circular medical device. The device has a cylindrical protrusion on top and a tube extending from the side. Inside the circular part, there is a visible turbine-like structure with multiple blades.
CentriMag Blood Pump
Image /page/4/Picture/16 description: The image shows a close-up of a device with a black cable attached to it. The device has a cylindrical shape with a silver-colored, ribbed exterior. The cable is thick and black, and it is connected to the device via a gray connector. The device appears to be some sort of electronic component or sensor.
CentriMag Motor
Image /page/4/Picture/18 description: The image shows a CentriMag device. The device is silver and black, and it has a screen on the front. There are also several buttons on the front of the device. The device has two handles on the right side.
## CentriMag Console
Image /page/4/Picture/20 description: The image shows a gray and black flow sensor with a gray cable attached to it. The sensor has a gray top with a blue arrow and some text on it. The sensor is designed to clamp onto a pipe or tube to measure the flow of liquid or gas.
Flow Probe (fits standard 3/8" tubing)
{5}------------------------------------------------
The CentriMag Motor is a reusable, non-sterile component of the CentriMag Acute Circulatory Support System. The CentriMag Motor holds the blood pump and drives the impeller inside the blood pump. The motor turns the magnet (and impeller) within the blood pump (Full MagLev™ technology) at a speed that is set on the console by the user. When the impeller is rotated, a pressure gradient develops between the center and outside edge of the pump, causing blood to flow from the inflow to the outflow port of the pump. The amount of flow through the pump depends on the speed of the impeller, and the difference between the inlet and outlet pressures.
#### نن Intended Use
The CentriMag Motor is intended for use only with the CentriMag Acute Circulatory Support System. The CentriMag System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc).
## F. Technological Characteristics
The technological characteristics of the CentriMag Acute Circulatory Support System have not changed. The redesigned CentriMag Motor cable with enhanced (longer) bend protection is the focus of this special 510(k) notification. The fundamental technological and functional characteristics of the motor and its cable are the same as the predicate device motor cable.
## G. Comparison to Predicate Device
The redesigned CentriMag Motor cable assembly with longer bend protection has indication for use, technological characteristics, and performance an characteristics which are substantially equivalent to the predicate CentriMag Motor cable. Due to this, the redesigned CentriMag Motor cable structure with enhanced bend protection raises no new safety or effectiveness concerns. These design modifications enhance cable flexibility and durability, and reduce the impact of rough handling and the occurrence of cable breaks and short circuiting.
No other components of the CentriMag Acute Circulatory Support System are affected by the design changes proposed in this Special 510(k) premarket notification.
{6}------------------------------------------------
### H. Summary of Performance Data
Comparison bench testing was performed to evaluate the mechanical performance characteristics of the redesigned CentriMag Motor cable with new bend protection with those of the current motor cable design.
The following mechanical performance verification tests were conducted on both designs:
### Motor Cable Endurance Test:
Evaluated the durability and functionality of the Motor cable and new bend protection assembly under simulated severe handling to demonstrate the new design meets the expected service life of the CentriMag Motor under worst-case conditions. In addition, visual inspection of the test bend area showed the integrity of the outer surface of the motor cable or bend protection in the area of bending.
### Motor Cable Kink Test:
Verified the Motor cable with new bend protection can tolerate excessive kinking or bending with small radius (e.g. when the cable is wound around the housing for storage) for the targeted expected service life. Visual inspection showed no defects on the surface of the motor cable or bend protection in the area of bending.
The redesigned CentriMag Motor cable with new custom-made bend protection successfully passed the cable endurance (bend) and kink test protocols, both designed to represent extreme / severe handling scenarios. The redesigned robust motor cable with extended bend relief successfully passed both tests (no failures), whereas the original cable design was less robust under high stress. Therefore, the redesigned CentriMag Motor cable with new bend protection performance test results demonstrate it to be substantially equivalence to the performance of the predicate CentriMag Motor cable design. Therefore, it can be concluded that the redesigned Motor cable offers enhanced cable flexibility and durability and improved protection from bending and kinking at the bend protection over the current commercial cable design.
Compliance to IEC 60601-1, Medical electrical equipment, Part 1: General requirements for basic safety and essential performance for the redesigned CentriMag Motor cable and new bend protection assembly has been confirmed, and it is therefore, substantially equivalent.
Given the nature of the proposed change, pre-clinical in-vivo testing, biocompatibility, software or sterility testing was not necessary.
No other device components of the CentriMag Motor assembly, or of the CentriMag Circulatory Support System, are affected by the design changes proposed in this Special 510(k) premarket notification.
{7}------------------------------------------------
# I. Clinical Performance
Clinical testing was not performed on the redesigned motor cable.
# J. Conclusion
The redesigned CentriMag Motor cable is substantially equivalent to the predicate CentriMag System motor cable.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.