The Levitronix 2nd Generation CentriMag Primary Console and Levitronix Monitor are indicated for use with the Levitronix CentriMag Extracorporeal Blood Pumping System. The Levitronix CentriMag Extracorporeal Blood Pumping System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc).
Device Story
The 2nd Generation CentriMag Primary Console is a microprocessor-based controller for the CentriMag Extracorporeal Blood Pumping System. It monitors system sensors, manages motor control signals, and provides alarm functions. The system includes a graphical user interface with touch pads for status monitoring and command entry. The Levitronix Monitor serves as a redundant, non-sterile, reusable interface, receiving power and data from the Primary Console via a dedicated cable. Used in hospital settings (OR, bedside) by clinicians, the system operates on AC power or internal/external batteries. The console manages the primary motor control, while the monitor displays alpha-numerical and graphical data and stores user commands. The device provides circulatory support by controlling pump speed, assisting clinicians in managing extracorporeal bypass circuits during surgical procedures.
Clinical Evidence
No clinical data was performed. Substantial equivalence is supported by functional testing, risk analysis, and Failure Modes and Effects Analysis (FMEA). The device has received CE Mark approval and has been used clinically in Europe.
Technological Characteristics
Microprocessor-based control system; graphical LCD interface with touch pads; AC power with internal rechargeable battery and optional external UPS; 12V DC-powered monitor; redundant user interface; non-sterile, reusable components.
Indications for Use
Indicated for patients requiring extracorporeal circulatory support via the CentriMag Blood Pumping System for up to six hours, including cardiopulmonary bypass and procedures such as valvuloplasty, mitral valve reoperation, vena cava/aorta surgery, and liver transplants.
Regulatory Classification
Identification
A cardiopulmonary bypass pump speed control is a device used that incorporates an electrical system or a mechanical system, or both, and is used to control the speed of blood pumps used in cardiopulmonary bypass surgery.
{0}------------------------------------------------
# K102129
AUG 2 7 2010
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in This Summary of Federal Regulations, Part 807, and in particular §807.92.
#### Application Information A.
Date Prepared:
July 28, 2010
Levitronix LLC
Submitter's Name & Address:
45 First Avenue Waltham, MA 02451
Contact Person:
Farzad Parsaie V.P., RA/QA (781) 622-5075 Ph: (781) 622-5090 Fax: e-mail: fparsaie@levitronix-us.com
#### Device Information в.
Trade or Proprietary Name:
Common or Usual Name:
Classification Name:
Performance Standard:
1. Levitronix 2nd Generation CentriMag® Primary Console
- 2. Levitronix Monitor
Cardiopulmonary Bypass Pump Console
Class II, DWA, 21 CFR – 870.4380 Control, Pump Speed, Cardiopulmonary Bypass
Performance standards do not currently exist . for these devices. None established under section 514 of the Food, Drug and Cosmetic Act.
#### Legally Marketed Predicate Devices C.
- Levitronix CentriMag System (K020271) .
- Levitronix CentriMag Primary Console (K083340) ●
{1}------------------------------------------------
#### .
D. Device Description
## 2nd Generation CentriMag Primary Console
The 2nd Generation CentriMag Primary Console is a microprocessor based device. The microprocessor generates the primary motor control signal, monitors system sensors, generates display outputs, interprets the inputs through the front keypad, provides alarm functions and handles the Levitronix Monitor interface. The Console is intended to be operated on single phase AC power; however, it also has a built-in rechargeable battery with a battery charger. The rechargeable battery is field replaceable. In addition, an optional external modular battery (UPS) can be used to power the CentriMag Primary Console. The human interface of the CentriMag Primary Console consists of a graphical screen to display data and system options and touch pads to change the system status. The 2nd Generation CentriMag Primary Console is intended to be used together with the Levitronix Monitor; however it can be also operated as a stand-alone unit.
#### Levitronix Monitor .
The Levitronix Monitor provides a redundant user interface containing a display and touch pads. The Primary and Back-Up Consoles also provide a fullyfunctional user interface containing a display and touch pads. The Levitronix Monitor is a non-sterile, reusable device that is designed to work only with the 200 Generation CentriMag Primary Console; therefore, it does not support the 1st Generation CentriMag Primary Console or the CentriMag Back-Up Console. The Levitronix Monitor is a 12V DC-powered device and receives its power directly from 2016 Generation CentriMag Primary Console via a power connector that mates with a connector on the back-panel of 2nd Generation CentriMag Primary Console. Based on the design of its power connector, the Levitronix Monitor cannot be plugged into a hospital AC power outlet. The Levitronix Monitor's core function is to provide multi-color alpha-numerical and graphical displays of information it receives from the 2nd Generation CentriMag Primary Console. The Levitronix Monitor is intended for use as a hospital-based unit (OR and bed-side) and is not intended for use during patient transport from one hospital to another.
The Levitronix Monitor is a processor based device, with a flat color screen (e.g. LCD). In addition, it is equipped with touch pad to allow the user to enter commands. In the hospital-setting configuration, the user is able to operate and monitor the performance of the CentriMag System through the Levitronix Monitor. The Levitronix Monitor can control the functions of the 2nd Generation CentriMag Primary Console, and therefore, of the CentriMag System. The Levitronix Monitor only displays data and stores user commands without interacting directly with the orimary motor control which is managed by the 200 Generation CentriMag Primary Console. The Levitronix Monitor is connected to the 2ª Generation r milling Primary Console through one cable, which includes data and power lines. Power is provided by the 2nd Generation CentriMag Primary Console.
Levitronix LLC
{2}------------------------------------------------
#### E. Intended Use
The Levitronix 2ª Generation CentriMag Primary Console and Levitronix Monitor are indicated for use with the Levitronix CentriMag Extracorporeal Blood Pumping System. The Levitronix CentriMag Extracorporeal Blood Pumping System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc).
#### Technological Characteristics に
The technological characteristics of the Levitronix 2nd Generation CentriMag Primary Console and the Levitronix Monitor are the same as the predicate devices.
#### G. Comparison to Predicate Device
The Levitronix 2nd Generation CentriMag Primary Console and the Levitronix Monitor have an indication for use, design features, and functional characteristics which are substantially equivalent to the predicate devices. Due to the equivalency of indications for use, design features, and functional characteristics, these devices raise no new safety or effectiveness issues.
#### Summary of Performance Data H.
The Levitronix 2014 Generation CentriMag Primary Console and the Levitronix Monitor have successfully undergone functional testing demonstrating substantial equivalence to the predicate devices.
The risk management methods used to assess the operational integrity of the 200 Generation CentriMag Primary Console and the Levitronix Monitor were a Risk Analysis and a Failure Modes and Effects Analysis (FMEA).
#### l. Clinical Performance
- Clinical testing was not performed. The 2nd Generation Console and Levitronix Monitor were CE Mark approved on February 5, 2010 and have been successfully used clinically in Europe in the commercial market.
#### Conclusion J.
The Levitronix 2nd Generation CentriMag Primary Console and the Levitronix Monitor are substantially equivalent to the Levitronix CentriMag Primary Console (K083340) and the Levitronix CentriMag System (K020271).
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement. The text is in all capital letters.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Levitronix LLC c/o Mr. Farzad Parsaie VP Regulatory Affairs/Quality Assurance 45 First Avenue Waltham, MA 02451
AUG 2 7 2010
K102129 Levitronix 2nd Generation CentriMag Primary Console, Levitronix Monitor Regulation Number: 21 CFR 870.4380 Regulation Name: Control, Pump Speed, Cardiopulmonary Bypass Regulatory Class: Class II (two) Product Code: DWA Dated: July 28, 2010 Received: July 29, 2010
Dear Mr. Parsaie:
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Farzad Parsaie
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours
Bra D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## K102129
## Indications for Use Statement
Applicant: Levitronix LLC K102129 510(k) Number (if known): CentriMag® Suq Generation Primary Device Name: Console દ્વ Levitronix® Monitor
### Indications for Use:
The Levitronix 200 Generation CentriMag Primary Console and Levitronix Monitor are indicated for use with the Levitronix CentriMag Extracorporeal Blood Pumping System. The Levitronix CentriMag Extracorporeal Blood Pumping System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc).
OR Prescription Use (per 21 CFR 801.109)
Over-the Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
Division Sign Off Division of Obrdiovascular Devices 510(k) Number
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.