APC 3 Argon Plasma Coagulation Unit, VIO 3 Electrosurgical Unit, FiAPC probes, Hybrid APC, APCapplicators, APC Pressure Reducer with Sensor, VIO Cart, Fastening sets
K191234 · Erbe Elektromedizin GmbH · GEI · Dec 20, 2019 · General, Plastic Surgery
Device Facts
Record ID
K191234
Device Name
APC 3 Argon Plasma Coagulation Unit, VIO 3 Electrosurgical Unit, FiAPC probes, Hybrid APC, APCapplicators, APC Pressure Reducer with Sensor, VIO Cart, Fastening sets
Applicant
Erbe Elektromedizin GmbH
Product Code
GEI · General, Plastic Surgery
Decision Date
Dec 20, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Erbe APC 3 with accessories is intended to deliver argon gas for argon plasma coagulation and ablation of tissue as well as argon-assisted cutting of tissue when used in conjunction with a compatible Erbe VIO Electrosurgical Generator (ESU) and applicators or probes.
Device Story
APC 3 unit delivers argon gas to surgical site; used with Erbe VIO 3 electrosurgical generator (ESU) and compatible applicators/probes. Argon gas ionized by high-frequency (HF) electrical field between electrode and tissue, creating argon plasma for non-contact coagulation/ablation. For argon-assisted cutting, HF current applied with contact while argon gas circulates active electrode. System includes pressure reducer, fastening sets, and ESU-integrated touch-screen interface for settings, tutorials, and safety alerts. Operated by physicians in clinical settings. Footswitches or applicator finger switches control activation. Output provides thermal tissue effects; benefits include precise coagulation and cutting. Device is non-sterile, reusable; accessories are sterile, single-use.
Clinical Evidence
Bench testing only. No clinical data provided. Verification and validation included output/mode comparison, tissue testing, capacitive coupling testing per IEC 60601-2-2, and accessory testing. Compliance with consensus standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, IEC 62304, IEC 62366) demonstrated safety and performance.
Technological Characteristics
Unit: metal sheet, plastics, wiring. Applicators/Probes: stainless steel, ceramic, plastics, tungsten, silicone. Sensing: accessory/instrument type detection via VIO ESU. Energy: HF current from VIO 3 generator. Gas: Argon 4.8 (99.998% purity). Connectivity: cable-connected to VIO 3 ESU. Standards: ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-2, IEC 62304, IEC 62366.
Indications for Use
Indicated for argon plasma coagulation, ablation, and argon-assisted cutting of tissue in patients requiring electrosurgical intervention, when used with compatible Erbe VIO ESU and applicators/probes.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Erbe VIO APC Model APC 2 with Accessories (K024047)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 20, 2019
Erbe Elektromedizin GmbH Julia Weller Specialist Regulatory Affairs Waldhoernlestrasse 17 Tuebingen, 72072 De
Re: K191234
Trade/Device Name: Erbe Argon Plasma Coagulation Unit APC 3 with Accessories Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: May 3, 2019 Received: May 8, 2019
Dear Julia Weller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191234
#### Device Name
Erbe Argon Plasma Coagulation Unit APC 3 with Accessories
Indications for Use (Describe)
The Erbe APC 3 with accessories is intended to deliver argon gas for argon plasma coagulation and ablation of tissue as well as argon-assisted cutting of tissue when used in conjunction with a compatible Erbe VIO Electrosurgical Generator (ESU) and applicators or probes.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY [AS REQUIRED BY 21 CFR 807.92(C)]
| Submitted By: | Erbe Elektromedizin GmbH<br>Waldhoernlestrasse 17<br>72072 Tuebingen<br>Germany<br>Tel: 0049-7071-755-0<br>Fax: 0049-7071-755-179 |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Mr. Henry Kutz<br>Specialist Regulatory Affairs |
| Date Prepared: | May 3, 2019 |
| Trade/Proprietary Name: | Erbe APC Unit Model APC 3 with Accessories |
| Common Name: | Argon Plasma Coagulation Unit |
| Classification Name and Code: | Electrosurgical Cutting and Coagulation Device and<br>Accessories (21 CFR Part 878.4400) |
| Regulatory Class: | II |
| Product Code: | GEI |
| Legally Marketed<br>Predicate Device: | Erbe VIO APC Model APC 2 with Accessories,<br>510(k) Number K024047 |
## Device Description:
The Erbe Argon Plasma Coagulation Unit Model APC 3 is an APC unit for use in conjunction with an Erbe Electrosurgical Unit (ESU) compatible Model (currently Model VIO® 3). The APC 3 (as well as VIO 3) can be mounted/secured to a cart or on a ceiling mount using fastening sets, i.e. connecting cables and hardware. The APC 3 is connected to an argon gas source via a pressure reducer and connected to the VIO 3 which powers and controls the APC unit. The ESU further provides a touch-screen monitor (i.e., interactive display) with an onscreen tutorial, settings and operational information as well as safety features (i.e., audio and visual error system). When using the APC 3 together with the VIO 3, the user is offered various APC Modes as well as argon-supported Modes which can be adjusted by the physician via effect levels in order to achieve the desired tissue effect. The ESU VIO 3 provides the high-frequency (HF) current required for thermal coagulation. For argon plasma coagulation and ablation, the HF current is applied non-contact to the tissue to be coagulated through ionized and hence electrically conductive argon gas (argon plasma). The argon gas is ionized in the high-frequency electrical field between the electrode of the applicator and the tissue. For argon-assisted cutting, the ESU provides the regular cutting/coagulation Mode while the argon gas provided by the APC 3 circulates the active electrode. This application is with contact to tissue.
Applicators or probes connected to the APC 3 for coagulating tissue via argon plasma and/or for cutting tissue supported by argon gas may be activated via footswitches connected to the ESU or by use of applicators that have an activation finger switch.
The APC unit and its accessories (i.e. pressure reducer and fastening sets) are supplied non-sterile and are reusable. Cleaning and disinfections are provided in the APC
{4}------------------------------------------------
user manual or notes on use of the involved accessory. The compatible applicators and probes are provided sterile as well as are single-use.
#### Indications for Use Statement:
The Erbe APC 3 with accessories is intended to deliver argon gas for argon plasma coagulation and ablation of tissue as well as argon-assisted cutting of tissue when used in conjunction with a compatible Erbe VIO Electrosurgical Generator (ESU) and applicators or probes.
Similarities and Differences of the Modified Device to the Current Device (Predicate Comparison/Substantial Equivalence):
| Characteristics | Predicate:<br>Erbe VIO APC Model APC 2 with<br>Accessories | Proposed:<br>Erbe APC Unit Model APC 3 with<br>Accessories |
|-----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Erbe Elektromedizin GmbH<br>(Germany) | Same |
| 510(k) | K024047 | N/A |
| Trade/Proprietary<br>Name | Erbe VIO APC Model APC 2 with<br>Accessories | Erbe APC Unit Model APC 3 with<br>Accessories |
| Indications for Use | The Erbe VIO APC with Accessories<br>is intended to deliver argon gas for<br>argon plasma coagulation of tissue<br>when used in conjunction with a<br>compatible Erbe VIO Electrosurgical<br>Generator (ESU) and applicators or<br>probes. | The Erbe APC 3 with accessories is<br>intended to deliver argon gas for<br>argon plasma coagulation and<br>ablation of tissue as well as argon-<br>assisted cutting of tissue when used<br>in conjunction with a compatible<br>Erbe VIO Electrosurgical Generator<br>(ESU) and applicators or probes. |
| Prescription or<br>ОТС | Prescription (Rx) | Same |
| Materials | | |
| Unit | Metal sheet, plastics, wiring, metal<br>fittings | Same |
| Applicators | Stainless steel, ceramic, plastics,<br>tungsten, printing color | Same |
| Probes | Stainless steel, ceramic, plastics,<br>tungsten, glue, silicone | Same |
| Pressure Reducer | Metal, plastic, tubing, metal fittings,<br>pressure gauge | Same |
| Fastening Sets | N/A | metal, wiring, plastics |
| Technical / Performance Data | | |
| Unit gas flow rate<br>(adjustable) | 0.1-8 l/min limited by the<br>respective instrument connected,<br>adjustable in 0.1 I steps | Same |
| Gas Type Used | Argon 4.8 (99.998% purity) or higher | Same |
| Input/Line | No direct line input / powered<br>through compatible VIO ESU | Same |
| Protective Circuits | Detection of specific type of<br>accessory/instruments (e.g.<br>applicator or probe) through the VIO<br>ESU | Same |
| Characteristics | Predicate:<br>Erbe VIO APC Model APC 2 with<br>Accessories | Proposed:<br>Erbe APC Unit Model APC 3 with<br>Accessories |
| Receptacle<br>Configuration | Up to two sockets.<br>(one APC/FiAPC socket or two<br>APC/FiAPC sockets or one<br>APC/FiAPC socket and one<br>monopolar/bipolar/<br>multifunctional socket) | Up to two sockets.<br>(one FiAPC socket or one FiAPC<br>socket and one monopolar/ bipolar/<br>multifunctional socket) |
| Units compatible<br>with the APC Unit | Erbe ESU Model VIO D and VIO S<br>series [K060484,K083452 and<br>K080715] | Erbe ESU Model VIO 3 [K190823 -<br>pending] |
| Applicators and<br>Probes compatible<br>with the APC Unit | APC raspatory handle and tip<br>[K992764] | N/A |
| | APC applicator [K992769] | N/A |
| | APC handle and applicators<br>[K003462] | N/A |
| | APCprobe [K013348] | N/A |
| | FiAPC probes [K060183]<br>HybridAPC probe [K143306] | Same<br>Same |
| | APCapplicators [K183445] | Same |
| Other compatible<br>equipment | VIO Cart [K023886] | Same |
| Labeling | APC 2 Unit<br>- packaging label<br>- user manual<br>- service manual | APC 3 Unit<br>- packaging label<br>- user manual<br>- service manual |
| | Applicators and probes<br>- packaging label<br>- notes on use | Applicators and probes<br>- packaging label<br>- notes on use<br>Note: Compatibility to APC 3 added. |
| | Pressure reducer<br>- packaging label<br>- notes on use | Pressure reducer<br>- packaging label<br>- notes on use<br>Note: Compatibility to APC 3 added. |
| | N/A | Fastening sets<br>- packaging label |
| | VIO Cart<br>- packaging label<br>- user manual | VIO Cart<br>- packaging label<br>- user manual<br>Note: Compatibility to APC 3 added. |
| | APC Unit: N/A - no patient contact | Same |
| Biocompatibility | Applicators and probes: direct patient<br>contact | Same |
| | Pressure reducer and VIO Cart:<br>N/A - no patient contact | Pressure reducer, VIO Cart and<br>fastening sets:<br>N/A - no patient contact |
| Provided condition | APC Unit : non-sterile, reusable | Same |
| | Applicators and probes: sterile,<br>disposable (single-use) | Same |
| | Pressure reducer and VIO Cart: non-<br>sterile, reusable | Pressure reducer, VIO Cart and<br>fastening sets: non-sterile, reusable |
| Shelf Life | Not Applicable for reusable Unit and<br>Accessories (i.e. pressure reducer<br>and VIO Cart) | Not Applicable for reusable Unit and<br>Accessories (i.e. pressure reducer,<br>VIO Cart and fastening set) |
| Characteristics | Predicate:<br>Erbe VIO APC Model APC 2 with<br>Accessories | Proposed:<br>Erbe APC Unit Model APC 3 with<br>Accessories |
| | Three (3) years for single use<br>HybridAPC probe and<br>APCapplicators and five (5) years for<br>FiAPC probes. | Same |
{5}------------------------------------------------
{6}------------------------------------------------
#### Verification/Validation Activities (Including Applied/Certified Standards and Guidance Documents):
FDA Recognized Consensus Standards
- ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance* *Other versions of this standard were also tested to and certified.
- IEC 60601-1-2 Edition 4.0:2014-02 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests* *Other versions of this standard were also tested to and certified.
- · IEC 60601-2-2 Edition 6.0 2017-03 Medical electrical equipment Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surqical accessories* *Other versions of this standard were also tested to and certified
- . IEC 62304:2006 + A1:2015 Medical device software - Software life cycle processes
- · IEC 62366:2007 + A1:2014 Medical Devices Application Of Usability Engineering To Medical Devices
- IEC 60601-1-6:2010 (Third Edition) + A1:2013 Medical electrical equipment Part 1 -. 6: General requirements for basic safety and essential performance - Collateral standard: Usability
- · ISO 15223-1:2016 Medical devices Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements
- EN 1041:2008 + A1:2013 Information supplied by the manufacturer of medical . devices
- ISTA 2A:2011 Partial Simulation Performance Test Packaged-Products 150 Ib (68 . kg) or Less
Performance Evaluations/Testing
- · Output/Mode Comparison
- · Tissue Testing
- · Capacitive Coupling (per Standard Number 60601-2-2)
- · Accessory Testing
#### Conclusion:
The intended use of the proposed Erbe APC 3 was clarified identifying all functions that may be performed with the Unit when used in combination with a compatible Erbe VIO ESU and applicators or probes. The proposed Unit has essentially the same principles of operation, technological characteristics, as well as performance characteristics as the predicate APC unit. Validation and verification activities in design control that includes testing/certification to designated standards (including electrical safety standards) and performance testing of the proposed devices has demonstrated substantial equivalence to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.