Salem Sump Dual Lumen Stomach Tube with ENFit Connection
Applicant
Cardinalhealth
Product Code
PIF · Gastroenterology, Urology
Decision Date
Nov 7, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Salem Sump™ Dual Lumen Stomach Tube with ENFit™ Connection is intended for gastric decompression and administration of nutrition, fluids and medication. The device is intended for patients with age of two and older.
Device Story
Double-lumen PVC tube used for naso/orogastric drainage and feeding; larger lumen facilitates drainage and nutrition/medication delivery; smaller lumen draws in ambient air to moderate suction pressure at drainage eyes. Device incorporates ISO 80369-3 compliant ENFit connector for secure enteral access. Single-use, EO-sterilized device; intended for clinical use by healthcare professionals. Provides gastric decompression and enteral support; ENFit connection prevents misconnection with non-enteral systems, enhancing patient safety.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including functional verification, occlusion, leakage, patency, tensile strength, stress cracking, flow rate, simulated gastric indwell, and biocompatibility evaluation.
Technological Characteristics
Double-lumen PVC tube; sizes 6Fr to 18Fr. Features ISO 80369-3 compliant ENFit connector. Sterilized via EO. Complies with EN 1615:2000, EN 1618:1997, ISO 80369-1:2010, ISO 80369-3:2016, ISO 80369-20:2015, and ISO 10993-1:2009.
Indications for Use
Indicated for gastric decompression and administration of nutrition, fluids, and medication in patients aged two years and older.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
Salem Sump Dual Lumen Stomach Tube with Multi-Functional Port (K150711)
Reference Devices
Argyle Salem Sump Tube Anti-Reflex Valve Insertion Tray (K935781)
Argyle™ Polyvinyl Chloride (PVC) and Kangaroo™ Polyurethane (PU) Neonatal and Pediatric Feeding Tubes with ENFit connector (K150084)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 7, 2019
Cardinal Health Alexis Erazo Principal Regulatory Affairs Specialist 777 West Street Mansfield, MA 02048
Re: K190923
> Trade/Device Name: Salem Sump Dual Lumen Stomach Tube with ENFit Connection Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: PIF, FEG Dated: October 7, 2019 Received: October 8, 2019
Dear Alexis Erazo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Martha W. Betz, Ph.D. Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190923
Device Name
Salem Sump™ Dual Lumen Stomach Tube with ENFit™ Connection
Indications for Use (Describe)
The Salem Sump™ Dual Lumen Stomach Tube with ENFit™ Connection is intended for gastric decompression and administration of nutrition, fluids and medication. The device is intended for patients with age of two and older.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K190923 Page 1 of 3
#### 510(k) Summary Salem Sump™ Dual Lumen Stomach Tube with ENFit™ Connection
### Preparation date:
November 4, 2019
## Submitter Information:
Alexis Erazo Principal Regulatory Affairs Specialist Cardinal Health 777 West Street, Mansfield, MA 02048 Phone: 508.618.3727 Email: alexis.erazo@cardinalhealth.com
### Name of Medical Device:
| Trade Name: | Salem Sump™ Dual Lumen Stomach Tube with ENFit™ Connection |
|----------------------|------------------------------------------------------------|
| Common Name: | Tubes, gastrointestinal |
| Classification Name: | Gastrointestinal tube and accessories |
| Regulation Number: | 21 CFR 876.5980 |
| Product Code: | PIF; FEG |
| Class: | II |
## Identification of Predicate Device:
| 510(k) Number | K150711 |
|--------------------|-------------------------------------------------------------------|
| Device Description | Salem Sump Dual Lumen Stomach Tube with Multi-<br>Functional Port |
| Submitter | Covidien |
## Identification of Reference Device:
| 510(k) Number | K935781 |
|--------------------|---------------------------------------------------------|
| Device Description | Argyle Salem Sump Tube Anti-Reflex Valve Insertion Tray |
| Submitter | Sherwood Medical Co. |
#### Identification of Reference Device:
| 510(k) Number | K150084 |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Device Description | Argyle™ Polyvinyl Chloride (PVC) and Kangaroo™<br>Polyurethane (PU) Neonatal and Pediatric Feeding Tubes<br>with ENFit connector |
| Submitter | Covidien |
#### Device Description:
The Salem Sump™ Dual Lumen Stomach Tube with ENFit™ Connection is a double-lumen tube made of PVC used for naso/orogastric drainage and feeding. The larger (main) lumen is
{4}------------------------------------------------
for feeding and drainage, while the smaller lumen draws in outside air to moderate the amount of suction at the drainage eyes. The device contains an ISO 80369-3 compliant ENFit connector which is inserted into the main lumen after decompression is no longer required to administer enteral fluids, including enteral nutrition, hydration and medication. It is a single use device which is provided sterilized via EO sterilization.
The Salem Sump™ Dual Lumen Stomach Tube with ENFit™ Connection is available in the following configurations:
- 6Fr x 24in ●
- . 8Fr x 24in
- 10Fr x 36in
- . 12Fr x 48in
- . 14Fr x 48in
- 16Fr x 48in ●
- 18Fr x 48in
## Intended Use/Indications for Use:
The Salem Sump™ Dual Lumen Stomach Tube with ENFit™ Connection is intended for gastric decompression and administration of nutrition, fluids and medication. The device is intended for patients with age of two and older.
## Product Comparison Summary:
The proposed Salem Sump™ Dual Lumen Stomach Tube with ENFit™ Connection has the same intended use, indications for use and similar fundamental technological characteristics and as compared to the predicate Salem Sump Dual Lumen Stomach Tube with Multi-Functional Port (K150711). Verification testing provided demonstrates that the technological characteristics differences to the Salem Sump™ Dual Lumen Stomach Tube with ENFit™ Connection do not raise new questions of safety and efficacy than that of the predicate K150711. The test results demonstrate that the proposed device is as safe and effective as the legally marketed predicate device.
## Non-Clinical Performance Data:
Laboratory testing was completed to support substantial equivalence between the modified device and the current device. The modified device was evaluated to show compliance to the standards requirements (listed below) as well as performance characteristics related to the modification of the device.
- EN 1615:2000 Enteral feeding catheters and enteral giving sets for single use and ● their connectors - Design and testing
- EN 1618:1997 Catheters other than intravascular catheters. Test methods for . common properties
- . ISO 80369-1:2010 Small-bore connectors for liquids and gases in healthcare applications - Part 1: General requirements
- ISO 80369-3:2016 Small-bore connectors for liquids and gases in healthcare . applications - Part 3: Connectors for enteral applications
{5}------------------------------------------------
Salem Sump™ Dual Lumen Stomach Tube with ENFit™ Connection Traditional 510(k) Pre-Market Notification
- . ISO 80369-20:2015 Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods
- . ISO 10993-1:2009 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
The following testing was conducted to demonstrate that the device continues to meet the requirements of the product specifications and supports the determination of substantial equivalence.
- . Functional Verification
- Occlusion Verification ●
- Liquid/Fluid Leakage .
- Patency Verification .
- Tensile Strength
- Resistance to separation from axial load ●
- . Resistance to separation from unscrewing
- . Resistance to overriding
- . Disconnection by unscrewing
- . Stress Cracking
- Dimension verification ●
- Flow Rate ●
- Simulated Gastric Indwell ●
- Biocompatibility Evaluation ●
- Shelf Life (2 years) ●
The results of the testing show that the modified device continues to meet the requirements of the product specifications and supports the determination of substantial equivalence.
## Clinical Data:
Clinical evaluations were not relied upon for evidence of safety of effectiveness, or for a determination of substantial equivalence.
## Conclusion:
The proposed Salem Sump™ Dual Lumen Stomach Tube with ENFit™ Connection has the same indications for use, intended use and similar fundamental technological characteristics and as compared to the predicate Salem Sump Dual Lumen Stomach Tube with Multi-Functional Port (K150711). Verification testing demonstrates that the difference in technological characteristics between the K157011 predicate and the proposed Salem Sump™ Dual Lumen Stomach Tube with ENFit™ Connection do not raise new questions of safety and efficacy. In addition, test results demonstrate that the proposed device is as safe and effective as the legally marketed predicate device. Based on the above evaluation, the proposed Salem Sump™ Dual Lumen Stomach Tube with ENFit™ Connection is substantially equivalent to the predicate Salem Sump Dual Lumen Stomach Tube with Multi-Functional Port (K150711).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.